← Product Code [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI) · K221989

# The Oligio (K221989)

_Wontech Co., Ltd. · GEI · Oct 13, 2022 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K221989

## Device Facts

- **Applicant:** Wontech Co., Ltd.
- **Product Code:** [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI.md)
- **Decision Date:** Oct 13, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Oligio is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis of soft tissue.

## Device Story

The Oligio is a radio frequency (RF) therapy system used in dermatologic and general surgical procedures. It delivers 6.78 MHz RF energy to soft tissue for selective coagulation and hemostasis while conductively cooling the epidermis. The system comprises a main unit with a touch LCD monitor, an RF handpiece, return pad cable, non-sterile treatment tips, cooling gas, and coupling fluids. The device employs RF tuning to deliver energy across varying tissue impedances. Operated by a clinician, the system is controlled manually or via footswitch. The output provides therapeutic coagulation; the cooling mechanism protects the skin surface. The device is intended for clinical settings.

## Clinical Evidence

No clinical data; bench testing only.

## Technological Characteristics

RF therapy system operating at 6.78 MHz ± 1% with 145 W max output. Components include touch LCD, RF handpiece, return pad, and disposable tips. RF tip material is polyethylene terephthalate (ISO 10993-1 compliant for limited contact). Features conductive epidermal cooling, manual/footswitch operation, and adjustable handpiece vibration (30-50 dB). Complies with IEC 60601-1, IEC 60601-2-2, IEC 60601-1-2, and IEC 61000-3-2.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- Thermage FLX System ([K170758](/device/K170758.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" on top and "ADMINISTRATION" on the bottom.

10/13/2022

WON TECH Co., Ltd. Hyun Yoon General Manager 64 Techno 8-ro Yuseong-gu, Daejeon 34028 Korea, South

### Re: K221989

Trade/Device Name: The Oligio Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI Dated: October 13, 2022 Received: October 13, 2022

## Dear Hyun Yoon:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

for Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K221989

Device Name The Oligio

### Indications for Use (Describe)

'The Oligio' is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis of soft tissue.

Type of Use (Select one or both, as applicable)

| <div> <span></span>Prescription Use (Part 21 CFR 801 Subpart D)    </div> |
|---------------------------------------------------------------------------|
| <div> <span></span>Over-The-Counter Use (21 CFR 801 Subpart C)    </div>  |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for WONTECH. The logo consists of an orange circle with a stylized "W" inside, followed by the text "WONTECH". The "WON" part of the text is orange, while the "TECH" part is gray.

# 510(k) Summary

[As required by 21 CFR 807.92]

K221989

#### 1. Date Prepared [21 CFR 807.92(a)(a)]

September 13, 2022

#### 2. Submitter's Information & Contact Person [21 CFR 807.92(a)(1)]

- Name of Manufacturer: WON TECH Co., Ltd.

- Address: 64 Techno 8-ro, Yuseong-gu, Daejeon, 34028, Republic of Korea - Contact Name: Hyun Sik Yoon - Telephone No.: +82-10-6750-5346

- Fax No.: +82-70-7836-0110
- Email Address: yoonhs21@wtlaser.com

#### 3. Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]

Common name: Radio Frequency Therapy System Trade name: The Oligio

| Classification Description                                        | 21 CFR Section | Product Code |
|-------------------------------------------------------------------|----------------|--------------|
| Electrosurgical Cutting and Coagulation Device and<br>Accessories | 878.4400       | GEI          |

As stated in 21 CFR, parts 878.4400, this generic types of devices has been classified as Class II.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for WONTECH. The logo consists of an orange circle with a stylized "W" inside, followed by the text "WONTECH". The "WON" part of the text is orange, while the "TECH" part is gray. The logo appears to be for a technology company.

#### 4. Identification of Predicate Device(s) [21 CFR 807.92(a)(3)]

The identified predicate devices within this submission are shown as follow:

Predicate device

- 510(k) Number: K170758
- Applicant: Solta Medical Inc. ●
- Classification Name: Electrosurgical Cutting and Coagulation Device and Accessories ●
- Trade Name: ● Thermage FLX System

#### ട്. Description of the Device [21 CFR 807.92(a)(4)]

'The Oligio' is a radio frequency therapy system. The radio frequency output of the device is 6.78 MHz, and the maximum power is 145 W. "The Oligio" consists of a main body including touch LCD monitor, a RF handpiece, return pad cable, non-sterile treatment tips, cooling gas, coupling fluids and a power cable.

'The Oligio' RF System delivers radio frequency energy for selective coagulation of tissue while conductively cooling the epidermis. "The Oligio' delivers energy form the disposable tip to the patient, and employs radio frequency turning to provide radio frequency energy across a range of impedances for delivery to the patient through the tip.

#### Indications for Use [21 CFR 807.92(a)(5)] 6.

'The Oligio' is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis of soft tissue.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for WONTECH. The logo consists of an orange circle with a stylized "W" inside, followed by the word "WONTECH" in orange and gray. The "WON" part of the word is in orange, while the "TECH" part is in gray.

64 Techno 8-ro, Yuseong-gu, Daejeon, Republic of Korea

TEL: +82 (42) 934 6800 FAX: +82 (42) 934 9491

# 7. Determination of Substantial Equivalence [21 CFR 807.92(a)(6) and 21 CFR 807.92(b)]

There are no significant differences between The Oligio and the predicate devices that would adversely affect the use of the product. It is substantially equivalent to this device in design, function, and technical characteristics.

|                                                                                                                                                                                                                                                                                                                                                             | Proposed Device                                                                                                                                                                                                                                                                                                                                                                         | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | SE Decision |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| K Number                                                                                                                                                                                                                                                                                                                                                    | K221989                                                                                                                                                                                                                                                                                                                                                                                 | K170758                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | -           |
| Manufacturer                                                                                                                                                                                                                                                                                                                                                | WON TECH Co., Ltd.                                                                                                                                                                                                                                                                                                                                                                      | Solta Medical Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | -           |
| Model                                                                                                                                                                                                                                                                                                                                                       | The Oligio                                                                                                                                                                                                                                                                                                                                                                              | Thermage FLX System                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | -           |
| Intended Use                                                                                                                                                                                                                                                                                                                                                | Dermatologic and general surgical<br>procedures for electrocoagulation<br>and hemostasis of soft tissue.                                                                                                                                                                                                                                                                                | Dermatologic and general surgical<br>procedures for electrocoagulation<br>and hemostasis.                                                                                                                                                                                                                                                                                                                                                                                                    | Same        |
| Principle/Method of<br>Operation                                                                                                                                                                                                                                                                                                                            | 'The Oligio' RF System delivers<br>radio frequency energy for selective<br>coagulation of tissue while<br>conductively cooling the epidermis.<br>'The Oligio' delivers energy form<br>the disposable tip to the patient, and<br>employs radio frequency turning to<br>provide radio frequency energy<br>across a range of impedances for<br>delivery to the patient through the<br>tip. | The Thermage FLX System<br>delivers RF energy for selective<br>coagulation of tissue while<br>conductively cooling the epidermis.<br>The Thermage FLX System<br>delivers energy from the disposable<br>tip to the patient. The System and its<br>Handpiece monitor skin contact<br>during treatment. The System<br>employs RF tuning to provide RF<br>energy across a range of<br>impedances for delivery to the<br>patient through single and multiple<br>pass stamping motions of the tip. | Same        |
| Output frequency                                                                                                                                                                                                                                                                                                                                            | 6.78MHz ± 1%                                                                                                                                                                                                                                                                                                                                                                            | 6.78MHz                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same        |
| Max output power                                                                                                                                                                                                                                                                                                                                            | 145 W                                                                                                                                                                                                                                                                                                                                                                                   | 400 W                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Different   |
| Different - Max output power<br>The Max output power for proposed device is different from the predicate devices. However, the Max output power<br>of the proposed The Oligio is covered by the range of the Max output power of the predicate device. Therefore, this<br>difference does not affect substantially equivalence on safety and effectiveness. |                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |             |
| Mode of Operation                                                                                                                                                                                                                                                                                                                                           | Manual or footswitch                                                                                                                                                                                                                                                                                                                                                                    | Manual or footswitch                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Same        |

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for WONTECH. The logo consists of an orange circle with a stylized "W" inside, followed by the text "WONTECH" in orange and gray. The "WON" part of the text is in orange, while the "TECH" part is in gray.

### Non-Clinical Test Summary [21 CFR 807.92(b)(1)]

### 1) Electrical Safety, Electromagnetic Compatibility Testing

Bench tests were conducted to verify that the proposed device met all design specifications. The test results demonstrated that the proposed device complies with the following standards:

| Standard<br>(Edition)                                                                                  | Standard Title                                                                                                                                                                                     |
|--------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-1:2005,<br>AMD1:2012                                                                         | Amendment 1 - Medical electrical equipment - Part 1: General<br>requirements for basic safety and essential performance                                                                            |
| IEC 60601-2-2:2017 for use in<br>conjunction with IEC 60601-<br>1:2005, IEC 60601-<br>1:2005/AMD1:2012 | Medical electrical equipment - Part 2-2: Particular requirements for the<br>basic safety and essential performance of high frequency surgical<br>equipment and high frequency surgical accessories |
| IEC 60601-1-2:2014                                                                                     | Medical electrical equipment - Part 1-2: General requirements for basic<br>safety and essential performance - Collateral Standard: Electromagnetic<br>disturbances - Requirements and tests        |
| IEC 61000-3-2:2018                                                                                     | Electromagnetic compatibility (EMC) - Part 3-2: Limits - Limits for<br>harmonic current emissions (equipment input current ≤16 A per phase)                                                        |

### 2) Software Validation

The Oligio contains Basic Documentation Level software. Software was designed and developed according to a software development process and was verified and validated.

The software information is provided in accordance with FDA guidance: Content of premarket submissions for Device Software Functions, on November 04, 2021.

### 3) Biocompatibility

| Part   | Material                      | Patient<br>Contact | Duration of Contact<br>by ISO 10993-1 | Bio-<br>compatibility |
|--------|-------------------------------|--------------------|---------------------------------------|-----------------------|
| RF tip | Polyethylene<br>terephthalate | Intact Skin        | Limited<br>(< 24 hours)               | Yes                   |

### 4) Performance Testing

The performance of The Oligio has been defined as follows.

- RF output frequency: 6.78MHz ± 1% -
- -Max output power: 145 W
- Handpiece vibration: Low (30dB), Medium (40dB), High (50dB) -

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for WONTECH. The logo consists of an orange circle with a stylized white "W" inside, followed by the text "WON" in orange and "TECH" in gray. The "W" in the circle is made up of two curved lines that intersect in the middle.

### Clinical Test Summary [21 CFR 807.92(b)(2)]

No clinical studies were considered necessary and performed.

### Conclusion [21 CFR 807.92(b)(3)]

In according with the Federal Food & drug and cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification WON TECH Co., Ltd. concludes that The Oligio is substantially equivalent to predicate devices as described herein.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K221989](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K221989)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
