← Product Code [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI) · K211107

# BTL-899 FP (K211107)

_BTL Industries, Inc. · GEI · Jul 9, 2021 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K211107

## Device Facts

- **Applicant:** BTL Industries, Inc.
- **Product Code:** [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI.md)
- **Decision Date:** Jul 9, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

BTL-899 FP is indicated to be used for: - Non-invasive lipolysis (breakdown of fat) of the abdomen. - Reduction in circumference of the abdomen. - Non-invasive lipolysis (breakdown of fat) of the thighs. - Reduction in circumference of the thighs. - BTL-899 FP is intended for use with skin type I VI.

## Device Story

BTL-899 FP is a non-invasive therapeutic device for fat reduction; utilizes combined bipolar radiofrequency (RF) and electromagnetic stimulation; system consists of main console with touch-screen interface and two applicators for hands-free treatment; operator (clinician) sets parameters via touch screen; device provides real-time therapy monitoring; emergency button included for safety; intended for use in clinical settings; benefits patients through non-invasive fat breakdown and circumference reduction; clinical efficacy supported by studies in Fitzpatrick Skin Types IV-VI.

## Clinical Evidence

Two prospective clinical studies (n=34) evaluated safety and efficacy in Fitzpatrick Skin Types IV-VI. Subjects received three 30-minute treatments spaced 1 week apart. Primary endpoint (fat thickness reduction) met; 75% of subjects showed circumference reduction. 88% of subjects reported satisfaction. No adverse events reported; mild erythema and muscle soreness resolved quickly. Bench testing confirmed electrical safety (IEC 60601-1, IEC 60601-1-2) and software life cycle (IEC 62304).

## Technological Characteristics

System console with bipolar RF (27.12 MHz, 60W max) and electromagnetic stimulation (0.5-1.8 T). Biphasic, sinusoidal waveform. Pulse repetition 1-150 Hz; pulse duration 280 µs. Hands-free applicator with fixation belt. Class II, BF electrical protection. Software-controlled via touch screen. Complies with IEC 60601-1, IEC 60601-1-2, IEC 62304, ISO 14971, and ISO 10993 (biocompatibility).

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- BTL-899ST ([K202199](/device/K202199.md))

## Reference Devices

- [K192224](/device/K192224.md)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

July 9, 2021

BTL Industries Inc. David Chmel VP of Operations 362 Elm Street Marlborough, Massachusetts 01752

Re: K211107

Trade/Device Name: BTL-899 FP Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: May 12, 2021 Received: May 17, 2021

Dear David Chmel:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531 -542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### Indications for Use

510(k) Number (if known) K211107

Device Name BTL-899 FP

Indications for Use (Describe)

BTL-899 FP is indicated to be used for:

- · Non-invasive lipolysis (breakdown of fat) of the abdomen.
- · Reduction in circumference of the abdomen.
- · Non-invasive lipolysis (breakdown of fat) of the thighs.
- · Reduction in circumference of the thighs.
- · BTL-899 FP is intended for use with skin type I VI.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120

Expiration Date: 06/30/2023 See PRA Statement below.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains the logo for BTL Industries. The logo consists of three blue squares nested inside each other, with the letters "BTL" in white inside the innermost square. To the right of the logo is the text "BTL Industries" in a gray, sans-serif font. The text is aligned horizontally with the logo.

# 510(k) Summary

# General Information

| Sponsor:                     | BTL Industries, Inc.<br>362 Elm Street<br>Marlborough, MA 01752<br>Tel: +1-866-285-1656<br>Fax: +1-888-499-2502 |
|------------------------------|-----------------------------------------------------------------------------------------------------------------|
| Applicant:                   | BTL Industries, Inc.<br>362 Elm Street<br>Marlborough, MA 01752<br>Tel: +1-866-285-1656<br>Fax: +1-888-499-2502 |
| Contact Person:              | David Chmel<br>BTL Industries, Inc.<br>chmel@btlnet.com                                                         |
| Summary Preparation<br>Date: | 07 April 2021                                                                                                   |

### Device Name

| Trade/Proprietary Name:      | BTL-899 FP                                                  |
|------------------------------|-------------------------------------------------------------|
| Primary Classification Name: | Electrosurgical, Cutting & Coagulation Device & Accessories |
| Classification Regulation:   | 878.4400, Class II                                          |
| Classification Product Code: | GEI                                                         |

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for BTL Industries. The logo consists of three blue squares nested inside each other, with the letters "BTL" inside the innermost square. To the right of the square logo is the text "BTL Industries" in a dark gray sans-serif font.

### Legally Marketed Predicate Device

The BTL-899 FP is a state-of-the-art high-frequency energy device with accessories and is substantially equivalent to the current product that is already cleared for distribution in the USA under the following 510(k) Premarket Notification numbers:

- BTL-899ST (K202199) ●
#### Product Description

The BTL-899 FP is a non-invasive therapeutic device.

BTL-899 FP consists of a main unit and applicator(s). The main unit is equipped with a color touch screen that makes the device easy to use. The on-screen information guides the Operator through the entire therapy procedure. The therapeutic parameters are easily set using the touch screen. During therapy, the screen displays information about the remaining therapy time and other therapy parameters. The device is equipped with an emergency button to terminate the therapy. The two outputs (applicators) of the device enable hands-free simultaneous treatment by two applicators.

#### Indications for Use

BTL-899 FP is indicated to be used for:

- Non-invasive lipolysis (breakdown of fat) of the abdomen. ●
- . Reduction in circumference of the abdomen.
- Non-invasive lipolysis (breakdown of fat) of the thighs.
- . Reduction in circumference of the thighs.
- . BTL-899 FP is intended for use with skin types I - VI.

#### Non-clinical Testing (Performance, Bench Testing)

The BTL-899 FP device has been thoroughly evaluated for electrical safety. The device has been found to comply with the following applicable medical device safety standards:

| IEC 60601-1   | Medical electrical equipment – Part 1: General requirements for basic safety<br>and essential performance                                                                                   |
|---------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-1-2 | Medical electrical equipment – Part 1-2: General requirements for basic safety<br>and essential performance - Collateral standard: Electromagnetic disturbances<br>– Requirements and tests |
| IEC 62304     | Medical device software - Software life cycle processes                                                                                                                                     |

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for BTL Industries. The logo consists of three nested diamond shapes in blue, with the letters "BTL" inside the innermost diamond. To the right of the diamond shapes is the text "BTL Industries" in a gray sans-serif font.

- Medical devices Application of risk management to medical devices ISO 14971
- ISO 10993-1 Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process
- Biological evaluation of medical devices Part 5: Tests for in vitro cytotoxicity ISO 10993-5
- ISO 10993-10 Biological evaluation of medical devices – Part 10: Tests for irritation and skin sensitization

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for BTL Industries. The logo consists of three blue squares nested inside each other, with the letters "BTL" in white inside the innermost square. To the right of the squares is the text "BTL Industries" in a gray, sans-serif font. The logo is simple and modern, and the use of blue and gray gives it a professional look.

### Clinical Testing

BTL Industries conducted two clinical investigations in the United States of America.

In total 34 subjects with Fitzpatrick Skin Types (FST) IV-VI were recruited to participate in the two IRBapproved clinical studies investigating clinical efficacy and safety of the BTL-899 FP device for toning of abdomen and reduction of subcutaneous fat.

In summary 9 subjects with skin type IV, 8 subjects with skin type V and 8 subjects with skin type VI finished three active 30-minute treatments with BTL-899 FP device spaced 1 week apart. Average BMI of all subjects was 27.7±3.3 kg·m2. They were scheduled to 1-month, 3-month and 6-month follow-up (optional) visits and examined for adverse reactions to the treatment, fat tissue reduction (by diagnostic ultrasound or magnetic resonance), abdominal circumference reduction, therapy comfort (using 5-point Likert scale questionnaire and 10-point visual analogue scale - VAS) and subjective satisfaction (using 5-point Likert scale questionnaire).

The primary endpoint in subjects with FST IV-VI, who underwent active treatments with the BTL-899 FP device has been met. All subjects exhibited reduction in the fat thickness in their abdominal area, matching the efficacy previously evidenced in the 510(k) K192224. Circumference reduction was seen in 75% of subjects in total.

The secondary endpoint of the study, subjects' satisfaction with the therapy, has also been met with 88% of the subjects satisfied with the achieved results. Subjects felt none (92%) to minimal (8%) discomfort during the study treatment.

No adverse events were reported in the course of the clinical investigations and in particular in the subjects with FST IV-VI. There was no inflammation or post-inflammatory hyperpigmentation in the treated areas immediately after the treatments or at the follow-up visits. The only observed side effects related to the therapy were mild erythema and muscle soreness that occur in FST I-III subjects as well. These resolved very quickly for all subjects.

In conclusion, the treatments with BTL-899 FP have proven to be safe for patients with Fitzpatrick skin types I to VI.

### Technological Characteristics

The BTL-899 FP device has the same intended use and identical technological characteristics and principles of operation to its predicate device. The BTL-899 FP device and its predicate are comprised of a system console and applicator(s). The system console consists of the generators, computer, and the touch-screen control panel.

The mechanism of action and technological similarities and differences between the BTL-899 FP device and the predicate device are described below in the comparison table. The differences do not raise any new types of safety or effectiveness questions.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for BTL Industries. The logo consists of three nested squares, with the letters "BTL" inside the innermost square. To the right of the square logo is the text "BTL Industries" in a simple, sans-serif font. The logo and text are both in a dark gray color.

# Comparison with the Predicate Device

| 510(k) number<br>Device name<br>Company name          | Not assigned<br>BTL-899 FP<br>BTL Industries, Inc.                                                                                                                                                                                                                                                                                                  | K202199<br>BTL-899ST<br>BTL Industries, Inc.                                                                                                                                                                                                                                                                                                                   | Significant<br>Difference         |
|-------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------|
| Product Code<br>and Regulation                        | General & Plastic Surgery<br>21 CFR 878.4400<br>GEI - Electrosurgical, Cutting &<br>Coagulation & Accessories                                                                                                                                                                                                                                       | General & Plastic Surgery<br>21 CFR 878.4400<br>GEI - Electrosurgical, Cutting &<br>Coagulation & Accessories                                                                                                                                                                                                                                                  | None                              |
| Indications for Use                                   | BTL-899 FP is indicated to be used<br>for:<br>· Non-invasive lipolysis<br>(breakdown of fat) of the<br>abdomen.<br>• Reduction in circumference of<br>the abdomen<br>· Non-invasive lipolysis<br>(breakdown of fat) of the thighs.<br>• Reduction in circumference of<br>the thighs.<br>• BTL-899 FP is intended for use<br>with skin types I - VI. | BTL-899ST is indicated to be<br>used for:<br>· Non-invasive lipolysis<br>(breakdown of fat) of the<br>abdomen.<br>• Reduction in circumference of<br>the abdomen<br>·  Non-invasive lipolysis<br>(breakdown of fat) of the<br>thighs.<br>· Reduction in circumference of<br>the thighs.<br>• BTL-899ST is intended for<br>use with Skin Type I, II and<br>III. | Not<br>significantly<br>different |
| Basic Technology                                      | The system combines bipolar<br>radiofrequency with<br>electromagnetic stimulation.                                                                                                                                                                                                                                                                  | The system combines bipolar<br>radiofrequency with<br>electromagnetic stimulation.                                                                                                                                                                                                                                                                             | None                              |
| Clinical Use                                          | Prescription use                                                                                                                                                                                                                                                                                                                                    | Prescription use                                                                                                                                                                                                                                                                                                                                               | None                              |
| Electrical Protection                                 | Class II, BF                                                                                                                                                                                                                                                                                                                                        | Class II, BF                                                                                                                                                                                                                                                                                                                                                   | None                              |
| User Interface                                        | Touch screen                                                                                                                                                                                                                                                                                                                                        | Touch screen                                                                                                                                                                                                                                                                                                                                                   | None                              |
| Firmware Controlled                                   | Yes                                                                                                                                                                                                                                                                                                                                                 | Yes                                                                                                                                                                                                                                                                                                                                                            | None                              |
| Number of output<br>channels                          | 2                                                                                                                                                                                                                                                                                                                                                   | 2                                                                                                                                                                                                                                                                                                                                                              | None                              |
| RF Type                                               | bipolar                                                                                                                                                                                                                                                                                                                                             | bipolar                                                                                                                                                                                                                                                                                                                                                        | None                              |
| Max. RF Power                                         | 60 W (2x30 W)                                                                                                                                                                                                                                                                                                                                       | 60 W (2x30 W)                                                                                                                                                                                                                                                                                                                                                  | None                              |
| RF Frequency                                          | 27.12 Mhz                                                                                                                                                                                                                                                                                                                                           | 27.12 Mhz                                                                                                                                                                                                                                                                                                                                                      | None                              |
| Number of Magnetic<br>coils in the<br>Applicator      | 1                                                                                                                                                                                                                                                                                                                                                   | 1                                                                                                                                                                                                                                                                                                                                                              | None                              |
| Magnetic Field<br>Intensity (on the coil<br>surface)  | 0.5 to 1.8 T                                                                                                                                                                                                                                                                                                                                        | 0.5 to 1.8 T                                                                                                                                                                                                                                                                                                                                                   | None                              |
| Pulse Repetition<br>Rate - supported by<br>the device | 1 – 150 Hz                                                                                                                                                                                                                                                                                                                                          | 1 – 150 Hz                                                                                                                                                                                                                                                                                                                                                     | None                              |
| Pulse Duration                                        | 280 µs ± 20% µs                                                                                                                                                                                                                                                                                                                                     | 280 µs ± 20% µs                                                                                                                                                                                                                                                                                                                                                | None                              |
| Waveform                                              | Biphasic                                                                                                                                                                                                                                                                                                                                            | Biphasic                                                                                                                                                                                                                                                                                                                                                       | None                              |
| Shape                                                 | Sinusoidal                                                                                                                                                                                                                                                                                                                                          | Sinusoidal                                                                                                                                                                                                                                                                                                                                                     | None                              |
| Temperature Sensor                                    | Yes                                                                                                                                                                                                                                                                                                                                                 | Yes                                                                                                                                                                                                                                                                                                                                                            | None                              |
| Selection of<br>parameters                            | Yes                                                                                                                                                                                                                                                                                                                                                 | Yes                                                                                                                                                                                                                                                                                                                                                            | None                              |
| Application                                           | Hands-free, applicator fixed by<br>fixation belt                                                                                                                                                                                                                                                                                                    | Hands-free, applicator fixed by<br>fixation belt                                                                                                                                                                                                                                                                                                               | None                              |
| Therapy Time                                          | Up to 30 min                                                                                                                                                                                                                                                                                                                                        | Up to 30 min                                                                                                                                                                                                                                                                                                                                                   | None                              |
| Energy Source                                         | 100 V to 230 V AC, 50 Hz<br>100 V to 130 V AC, 60 Hz                                                                                                                                                                                                                                                                                                | 100 V to 230 V AC, 50 Hz<br>100 V to 130 V AC, 60 Hz                                                                                                                                                                                                                                                                                                           | None                              |
| System Dimensions<br>(W×H×D)                          | 23 x 39 x 29 in<br>(592 x 985 x 730 mm)                                                                                                                                                                                                                                                                                                             | 23 x 39 x 29 in<br>(592 x 985 x 730 mm)                                                                                                                                                                                                                                                                                                                        | None                              |
| System Weight                                         | 155 lb (70 kg) / 200 lb (91 kg)<br>including packaging and accessories                                                                                                                                                                                                                                                                              | 155 lb (70 kg) / 200 lb (91 kg)<br>including packaging and<br>accessories                                                                                                                                                                                                                                                                                      | None                              |
| Operating Ambient<br>Temperature                      | 64 °F to 86 °F<br>(+18 °C to +30 °C)                                                                                                                                                                                                                                                                                                                | 64 °F to 86 °F<br>(+18 °C to +30 °C)                                                                                                                                                                                                                                                                                                                           | None                              |
| Operating Relative<br>Humidity                        | 30% to 75%                                                                                                                                                                                                                                                                                                                                          | 30% to 75%                                                                                                                                                                                                                                                                                                                                                     | None                              |
| Environmental<br>Specifications                       | For indoor use only                                                                                                                                                                                                                                                                                                                                 | For indoor use only                                                                                                                                                                                                                                                                                                                                            | None                              |

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for BTL Industries. The logo consists of three blue squares nested inside each other, with the letters "BTL" in white inside the innermost square. To the right of the square logo is the text "BTL Industries" in a gray, sans-serif font. The logo is simple and modern, and the colors are clean and professional.

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for BTL Industries. The logo consists of a blue geometric design on the left and the text "BTL Industries" on the right. The geometric design is made up of three interlocking squares, each with the letters "B", "T", and "L" inside. The text "BTL Industries" is written in a dark gray sans-serif font.

### Substantial Equivalence

The BTL-899 FP device has expanded indications for use to include Fitzpatrick Skin Types I - VI.

There are no technological modifications done to the device and its applicator compared to the predicate device.

The safety of the device has been evaluated and no new risks have been identified.

Therefore these differences do not raise any new questions of safety or effectiveness.

#### Conclusion

Based upon the intended use and the known technical and clinical data provided in this pre-market notification, the BTL-899 FP device has been shown to be substantially equivalent to the currently marketed predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K211107](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K211107)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
