← Product Code [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI) · K211019

# Disposable Monopolar Laparoscopic Tips and Reusable Handles (K211019)

_Dannik, LLC · GEI · Oct 28, 2021 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K211019

## Device Facts

- **Applicant:** Dannik, LLC
- **Product Code:** [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI.md)
- **Decision Date:** Oct 28, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

The Disposable Monopolar Laparoscopic Tips and Reusable Handles are designed to cut, dissect, manipulate and/or cauterize various tissues during endoscopic/laparoscopic, general surgical procedures.

## Device Story

Device consists of sterile, single-use monopolar tips (graspers, dissectors, forceps, scissors) and a non-sterile, reusable stainless steel handle. Used in endoscopic/laparoscopic surgery by surgeons. Handle features 360-degree rotation knob and RF post for connection to an external electrosurgical generator. During operation, surgeon manipulates tissue via handle; RF energy is delivered through the tip to the tissue to achieve coagulation. Device requires an external neutral electrode. Output is mechanical tissue manipulation and thermal coagulation. Benefits include precise tissue handling and hemostasis in minimally invasive settings.

## Clinical Evidence

No clinical data. Bench testing only. Performance verified via biocompatibility (ISO 10993), electrical safety (IEC 60601-1, 60601-2-2, 60601-2-18), and sterilization validation. Thermal spread efficacy compared to predicate on kidney and muscle tissue at 30W, 50W, and 90W power settings, demonstrating equivalent thermal damage profiles.

## Technological Characteristics

Materials: Biocompatible plastic and stainless steel. Energy: Monopolar RF (3800 Vp, 300-2000 kHz). Form factor: 5mm diameter, 34cm length. Connectivity: Connects to external electrosurgical generator. Sterilization: Ethylene Oxide (tips), Autoclave (handle). Standards: ISO 10993, ISO 11135, IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2, IEC 60601-2-18.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- Hunan Handlike Minimally Invasive Surgery Co. Endoscopic Surgical Instruments ([K171825](/device/K171825.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

October 28, 2021

DANNIK Olga Haberland Regulatory Compliance 941 W. Morse Blvd. Suite #100 Winter Park, Florida 32789

Re: K211019

Trade/Device Name: Disposable Monopolar Laparoscopic Tips and Reusable Handles Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: October 27, 2021 Received: October 1, 2021

Dear Olga Haberland:

We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801): medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531 -542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K211019

#### Device Name

Disposable Monopolar Laparoscopic Tips and Reusable Handles

Indications for Use (Describe)

The Disposable Monopolar Laparoscopic Tips and Reusable Handles are designed to cut, dissect, manipulate and/or cauterize various tissues during endoscopic/laparoscopic, general surgical procedures.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for DANNIK. The logo has a blue swoosh above the name, with a purple line underneath the swoosh. The name DANNIK is written in large, blue, sans-serif letters.

# K211019 510(k) Summary

## 1. SUBMITTER'S CONTACT INFORMATION

Company: DANNIK Address: 941 West Morse Blvd. Suite #100 Winter Park, FL 32789 Contact Person: Mrs. Olga Haberland, President & Requlatory Compliance Phone: (407) 927-1743

### 2. DEVICE NAME

Trade Name – Disposable Monopolar Laparoscopic Tips and Reusable Handles Common Name - Monopolar Laparoscopic Instruments Regulation Number - 21 CFR 878.4400 Classification Name - Electrosurgical, Cutting & Coaqulation Device and Accessories Product Code - GEI Device Classification - Class II Classification Panel - General and Plastic Surgery

## 3. PREDICATE DEVICE

The Disposable Monopolar Laparoscopic Tips and Reusable Handles claim Substantial Equivalence to Hunan Handlike Minimally Invasive Surgery Co. Endoscopic Surgical Instruments cleared under 510(k) K171825.

### 4. DEVICE DESCRIPTION

The Disposable Monopolar Laparoscopic Tips and Reusable Handles are sterile packaged single use monopolar attachments intended for use in combination with the Reusable Handle. It is designed to include graspers, dissectors, forceps, and scissors, intended to grasp, manipulate, cut, and cauterize soft tissue.

The Disposable Monopolar Laparoscopic Tips are offered in a working diameter of 5 mm with various jaw patterns including scissors, graspers, forceps and dissectors.

The Reusable Handle is supplied non-sterile and in intended to be sterilized by the user using an autoclave process. The shaft of the handle is 34 cm in length and is made from stainless steel covered with an external insulation. Handles are offered in ratcheting configurations and are intended to be connected to the sterile disposable tips. Once connected the rotation knob provides 360 degrees of rotation for the instrument shaft and jaws. The handle includes an RF Post for electrosurgery when attached to an approved electrosurgical generator to provide coagulation of tissue when used with an appropriate ground (neutral) electrode.

### 5. INDICATIONS FOR USE

The Disposable Monopolar Laparoscopic Tips and Reusable Handles are designed to cut. dissect, manipulate and/or cauterize various tissues during endoscopic, general surgical procedures.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "DANNIK" in large, bold, blue letters. Above the word is a curved line that is blue on the outside and purple on the inside. The word "DANNIK" is the main focus of the image and is likely the name of a company or organization.

### 6. SUBSTANTIAL EQUIVALENCE TABLE

| Device            | Disposable Monopolar Laparoscopic Tips and<br>Reusable Handles (510(k) TBD)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Hunan Handlike Minimally Invasive Surgery Co.<br>Endoscopic Surgical Instruments (510(k) K171825)                                                                                                                                                                                                                                                                                                                                                                        |
|-------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use      | The Disposable Monopolar Laparoscopic Tips and Reusable Handles<br>are intended to cut, dissect, manipulate and/or cauterize various<br>tissues during endoscopic/laparoscopic, general surgical procedures.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Product Picture   | Image: Disposable tip and reusable handle & shaft                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Image: Disposable tip and reusable handle & shaft                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Design            | The Disposable Monopolar Laparoscopic Tips and Reusable Handles<br>are sterile single use monopolar attachments intended for use in<br>combination with the Reusable Handle. These devices are made from<br>biocompatible plastic and stainless steel. The handle activates the<br>instrument jaws and scissor blades. The rotation knob provides 360<br>degrees of rotation for the instrument shaft and jaws. The shaft<br>includes an external insulation that runs from the rotation knob to the<br>instrument jaws and is provided in working diameters of 5 mm and<br>length of 34 cm. The handle includes an RF Post for electrosurgery<br>when attached to an approved electrosurgical generator to provide<br>coagulation of tissue when used with an appropriate ground (neutral)<br>electrode. | Endoscopic Surgical Instruments Tip-Detachable Type: Disposable<br>tip and reusable shaft/handle can be disassembled but shaft cannot<br>be disassembled from handle. Handles have ratchet and non-ratchet.<br>Tips have 14 different models available. This product is monopolar<br>high frequency electrode as the applied part to be used with a<br>matching HF surgical generator with peak output voltage 3800V and<br>frequency 0.3MHz-2MHz and neutral electrode. |
| Diameter          | 5 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Length            | 34 cm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Power Settings    | Maximum Peak Voltage = 3800 Vp                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Maximum Peak Voltage = 3800 Vp                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|                   | Operating Frequency = 300 to 2000 kHz                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Operating Frequency = 300 to 2000 kHz                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|                   | Operating Current = 0.8 Arms                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Operating Current = 0.8 Arms                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Biocompatibility  | Conforms to ISO 10993                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Sterilization     | Sterilized using Ethylene Oxide for single patient use<br>in accordance with ISO 11135 to an SAL of 10^-6.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Ethylene Oxide                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Electrical Safety | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-18<br>IEC 60601-2-2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Prescription Use  | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |

## 7. NONCLINICAL TESTS

Nonclinical testing has been conducted to verify that the Disposable Monopolar Laparoscopic Tips and Reusable Handles met all design specifications and are substantially equivalent to the predicate device. Testing included the following:

- . Biocompatibility Testing performed in accordance with the following:
	- ISO 10993-1: 2018 Biological evaluation of medical devices -Part 1: Evaluation and O testing within a risk management process (Recognition No. 2-258)
	- ISO 10993-5:2009 Biological evaluation of medical devices Part 5: Tests for in vitro O cytotoxicity (Recognition No. 2-245)
	- ISO 10993-7:2008 Biological evaluation of medical devices Part 7: Ethylene oxide o sterilization residuals (Recognition No. 14-408)
	- ISO 10993-10:2010 Biological evaluation of medical devices Part 10: Tests for о irritation and skin sensitization (Recognition No. 2-174)

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for "DANNIK". The logo is in blue, with a purple swoosh above the text. The text is in a bold, sans-serif font. The logo is simple and modern.

- ISO 10993-11:2017 Biological evaluation of medical devices Part 11: Tests for o systemic toxicity (Recognition No. 2-255)
- Medical Electrical Equipment Safety Testing performed in accordance with the following:
	- o ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 (Consolidated Text) Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD) (Recognition No. 19-4)
	- ANSI AAMI IEC 60601-1-2:2014 Medical electrical equipment Part 1-2: General o requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests (Recognition No. 19-8)
	- IEC 60601-2-2:2017 Medical electrical equipment Part 2-2: Particular requirements о for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories (Recognition No. 6-389)
	- IEC 60601-2-18:2009 Medical electrical equipment Part 2-18: Particular o requirements for the basic safety and essential performance of endoscopic equipment (Recognition No. 9-114)
- Aging Study ●
- Autoclave Sterilization Validation ●
- Ethylene Oxide Sterilization per ISO 11135:2014 Sterilization of health-care . products - Ethylene Oxide (Recognition No. 14-529)

In addition, The Disposable Monopolar Laparoscopic Tips and Reusable Handles have been compared to the predicate device through various performance studies designed to test appearance, dimensions, corrosion resistance, operational forces, cutting efficacy and thermal effects on tissue.

Electrical performance of the device was completed following FDA guidance "Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery" issued August 15, 2016. This requires testing on three different tissue types at minimum, default, and maximum generator power in order to simulate thermal spread across different tissue types. The spread is then measured under magnification, and recorded to be compared with the predicate product. Results showed an equivalent thermal spread under the same conditions across the different tissue types and power settings. Summary results are shown below.

| Power Level | Thermal Damage Width<br>( $$\overline{X}$$ ±SD, mm) | Thermal Damage Depth<br>( $$\overline{X}$$ ±SD, mm) |
|-------------|-----------------------------------------------------|-----------------------------------------------------|
| 30W         | 0.42 ±0.26                                          | 3.05 ±2.10                                          |
| 50W         | 0.28 ±0.30                                          | 2.69 ±0.47                                          |
| 90W         | 0.47 ±0.36                                          | 4.43 ±1.42                                          |

### Table 1 – Thermal Damage on Kidney Tissue

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the word "DANNIK" in large, blue, sans-serif font. Above the word is a blue arc, with a smaller purple arc inside of it. The arcs appear to be a design element associated with the brand name.

| Power Level | Thermal Damage Width<br>(X ±SD, mm) | Thermal Damage Depth<br>(X ±SD, mm) |
|-------------|-------------------------------------|-------------------------------------|
| 30W         | 1.28 ±0.58                          | 2.62 ±0.20                          |
| 50W         | 1.51 ±0.65                          | 3.26 ±0.32                          |
| 90W         | 1.20 ±0.85                          | 2.82 ±0.37                          |

#### Table 3 – Thermal Damage on Muscle Tissue

## 8. CLINICAL TESTS

There were no clinical trials performed on these devices.

# 9. CONCLUSIONS

The subject device has equivalent indications for use as the predicate device. There are no new technologies being added to this device from the predicate, in terms of finished device functions. The device has the same intended use and application as the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K211019](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K211019)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
