K211000 · Ilooda Co,., Ltd. · GEI · Jun 9, 2022 · General, Plastic Surgery
Device Facts
Record ID
K211000
Device Name
ACUTRON
Applicant
Ilooda Co,., Ltd.
Product Code
GEI · General, Plastic Surgery
Decision Date
Jun 9, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
ACUTRON is intended for use in dermatologic and general surgical procedures for electro-coagulation and hemostasis
Device Story
ACUTRON is a high-frequency (RF) electrosurgical system for dermatologic and general surgical procedures. The system comprises a main unit, a footswitch, an LCD touch screen control panel, and a handpiece with single-use monopolar micro-needle electrodes. The physician operates the device by placing the micro-needle tip in light contact with the epidermis at a right angle to the target tissue. RF energy (2MHz) is delivered through the needles to generate an electro-thermal reaction, coagulating dermal tissue. The device is used in clinical settings to achieve hemostasis and tissue coagulation. The output is controlled via the touch screen and activated by the footswitch. The device benefits patients by providing a controlled method for surgical coagulation.
Clinical Evidence
Bench testing only. No clinical data provided. Safety and performance verified via IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-2 standards. Biocompatibility and ex-vivo histology testing were leveraged from the predicate device (K182355).
Technological Characteristics
RF electrosurgical system; 2MHz frequency; 0.1W-45W output power at 500Ω. Materials: Polycarbonate (body), Stainless Steel 304 with Perylene coating (needles). Monopolar operation. Single-use micro-needle electrodes sterilized via E.O. gas. Connectivity: LCD touch screen interface, footswitch activation. Dimensions: 274(W)x187(D)x291(H) mm.
Indications for Use
Indicated for dermatologic and general surgical procedures requiring electro-coagulation and hemostasis.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 9, 2022
ILOODA Co., Ltd % Dave Kim Medical Device Regulatory Affairs Mtech Group 7505 Fannin St. Ste 610 Houston, Texas 77054
Re: K211000
Trade/Device Name: Acutron Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: April 4, 2022 Received: April 11, 2022
Dear Dave Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory top ic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K211000
Device Name ACUTRON
Indications for Use (Describe)
ACUTRON is intended for use in dermatologic and general surgical procedures for electro-coagulation and hemostasis
Type of Use (Select one or both, as applicable)
| <span> <svg class="bi bi-check-square" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M14 1H2a1 1 0 0 0-1 1v12a1 1 0 0 0 1 1h12a1 1 0 0 0 1-1V2a1 1 0 0 0-1-1zM2 0a2 2 0 0 0-2 2v12a2 2 0 0 0 2 2h12a2 2 0 0 0 2-2V2a2 2 0 0 0-2-2H2z" fill-rule="evenodd"></path> <path d="M10.97 4.97a.75.75 0 0 1 1.071 1.05l-3.992 4.99a.75.75 0 0 1-1.08.02L4.324 8.384a.75.75 0 1 1 1.06-1.06l2.094 2.093 3.473-4.425a.236.236 0 0 1 .02-.022z" fill-rule="evenodd"></path> </svg> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span> <svg class="bi bi-square" fill="currentColor" height="1em" viewbox="0 0 16 16" width="1em" xmlns="http://www.w3.org/2000/svg"> <path d="M14 1H2a1 1 0 0 0-1 1v12a1 1 0 0 0 1 1h12a1 1 0 0 0 1-1V2a1 1 0 0 0-1-1zM2 0a2 2 0 0 0-2 2v12a2 2 0 0 0 2 2h12a2 2 0 0 0 2-2V2a2 2 0 0 0-2-2H2z" fill-rule="evenodd"></path> </svg> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary K211000
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR Part 807.92.
# Date 510k summary prepared: July 6, 2022
#### I. SUBMITTER
| Submitter's Name : | Ilooda Co.,Ltd |
|-------------------------|-------------------------------------------------------------------------------|
| Submitter's HQ Address: | 120, Jangan-ro 458 Beon-gil, Jangan-Gu,<br>Suwon-Si Gyeonggi do, KOREA, 16200 |
| Submitter's Telephone: | +82-31- 210-1622 |
| Contact person: | Yun-Jung HA (yjha@ilooda.com) / RD Manager |
| Official Correspondent: | Dave Kim (davekim@mtech-inc.net) |
| Address: | 7505 Fannin St. Ste 610, Houston, TX 77054 |
| Telephone: | +713-467-2607 |
| DEVICE | |
| Trade/proprietary name: | ACUTRON |
| Common or Usual Name: | Micro-needle RF system |
| Regulation Name: | Electrosurgical, cutting & coagulation device & accessories |
| Regulation Number: | 21 CFR 878.4400 (Product Code: GEI) |
| Regulatory Class: | Class II |
| Prescription Use. | |
{4}------------------------------------------------
### PREDICATE DEVICE
| Device Manufacturer: | ILOODA CO.,LTD |
|----------------------|-------------------------------------------------------------|
| Device Name: | Secret RF Smartcure Applicator |
| 510(k) Number: | K182355 |
| Regulation Name: | Electrosurgical, cutting & coagulation device & accessories |
| Regulation Number: | 21 CFR 878.4400 (Product Code: GEI, OUH) |
| Regulatory Class: | Class II |
### REFERENCE DEVICE
| Device Manufacturer: | ILOODA CO., LTD |
|-------------------------|-------------------------------------------------------------|
| Trade/proprietary name: | Secret RF |
| Common or Usual Name: | Micro-needle Fractional RF |
| 510K number: | K170325 |
| Regulation Name: | Electrosurgical, cutting & coagulation device & accessories |
| Regulation Number: | 21 CFR 878.4400 (Product Code: GEI, OUH) |
| Regulatory Class: | Class II |
These devices have not been subject to a design-related recall.
#### II. DEVICE DESCRIPTION
ACUTRON is High Frequency(=Radio Frequency, RF) includes the system main body, a handpieces with single-use micro-needle type electrodes, footswitch and an LCD touch screen control panel.
The RF energy is delivered to the target tissue using a handpiece and disposable tip(micro needle electrode tip), the tip being placed in light contact with the epidermis and the handpiece being held at right angles to the target tissue. As the RF energy passes through the skin, it generates an electro thermal reaction, which is capable of coagulating the tissue.
Using the micro needle tip, the ACUTRON creates heat within the target dermal tissue via micro-needles inserted from the tip.
The ACUTRON is consists of ;
- 1) Main unit
- 2) Smartcure handpiece(FDA cleared K182355)
- 3) Monopolar type micro-needle electrodes(FDA cleared K182355)
{5}------------------------------------------------
#### INDICATIONS FOR USE: IV.
ACUTRON is intended for use in dermatologic and general surgical procedures for electrocoagulation and hemostasis
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE V.
| | | Proposed Device | Predicate Device | Remark |
|--|-------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| | | ACUTRON<br>(K211000) | Secret RF Smartcure<br>applicator (K182355) | |
| | Indications for use | ACUTRON is intended<br>for use in dermatologic and<br>general surgical procedures<br>for electro-coagulation and<br>hemostasis. | Secret RF Smartcure<br>applicator is intended for<br>use in dermatologic and<br>general surgical procedures<br>for electro-coagulation and<br>hemostasis | |
| | Handpieces | Image: Handpiece | Image: Handpiece | Same |
| | Handpiece<br>material | Aluminum | Aluminum | Same |
| | Patient return<br>electrode | All-in-one connection with<br>Patient plate connector.<br>The handpiece and the counter<br>plate connector are connected to<br>one connector of the main body<br>and do not affect the safety or<br>effectiveness performance. | The electrode handpiece<br>and the patient plate are<br>separate. | Same |
| | Frequency | $2MHz \pm 10%$ | $2MHz \pm 10%$ | Same |
| | Output Power | $0.1W ~ 45W at 500\Omega$ | $0.1W ~ 45W at 500\Omega$ | Same |
| | Treatment time | 10~15min<br>(recommended) | 10~15min<br>(recommended) | Same |
| | With Connected<br>Needle electrodes | MTR-AC series | | Same |
| | | MTR-AC-01, MTR-AC-04, MTR-AC-27G | | Same |
| | Body | PC(Polycarbonate) | PC(Polycarbonate) | Same |
| | Needle<br>material needle | Stainless Steel 304 +<br>Perylene coating | Stainless Steel 304 +<br>Perylene coating | Same |
{6}------------------------------------------------
| Electrode Needle Specification | | | | | | | | |
|------------------------------------|------------------------|-------------------------------------------|--------------|--------------|-------------------------------------------|--------------|--------------|------|
| Model name | | ACUTRON (K211000) | | | Secret RF Smartcure applicator (K182355) | | | |
| 510(k)number | | K211000 | | | K182355 | | | |
| Model name of electrode | | MTR<br>AC-27G | MTR<br>AC-01 | MTR<br>AC-04 | MTR<br>AC-27G | MTR<br>AC-01 | MTR<br>AC-04 | Same |
| Configuration | | Body + Needle | | | Body + Needle | | | Same |
| Physical<br>SPEC. | Total length (mm) | 40 | 20 | 20 | 40 | 20 | 20 | Same |
| | Needle(tip) Diameter | $Ø$ 0.4 | $Ø$ 0.25 | $Ø$ 0.25 | $Ø$ 0.4 | $Ø$ 0.25 | $Ø$ 0.25 | Same |
| | Needle(tip) Length(mm) | 3.0 | 2.5 | 1.5 | 3.0 | 2.5 | 1.5 | Same |
| | Needle(tip) Unit | 1 | 1 | 4 | 1 | 1 | 4 | Same |
| Patient<br>Contacting<br>Materials | Body (indirect) | Polycarbonate | | | Polycarbonate | | | Same |
| | Needle | Stainless steel 304 +<br>Perylene coating | | | Stainless steel 304 +<br>Perylene coating | | | |
| Output power | | Max 45W | | | Max 45W | | | Same |
| Type and operation mode | | Monopolar (Normal mode) | | | Monopolar (SC mode) | | | Same |
| Single Use | | Single use | | | Single use | | | Same |
| Sterilization method | | E.O gas | | | E.O gas | | | Same |
| | Proposed device | Reference device | Remark |
|-----------------|---------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|--------|
| Model name | ACUTRON | SecretRF | |
| Manufacturer | ILOODA CO.,LTD | ILOODA CO.,LTD | |
| 510(k)number | K211000 | K170325 | |
| Intended use | ACUTRON is intended<br>for use in dermatologic and<br>general surgical procedures<br>for electro-coagulation and<br>hemostasis. | SECRET RF is intended for<br>use in dermatologic and<br>general surgical procedures for<br>electro-coagulation and<br>hemostasis | Same |
| Rated voltage | 100-240V~, 50/60Hz | 100-240V~, 50/60Hz | |
| User interface | Color touch Panel | Color touch Panel | Same |
| Dimension(mm) | Main unit :<br>274(W)x187(D)x291(H) | Main unit :<br>180(W)x460(D)x1100(H) | |
| Delivery system | Monopolar handpieces +<br>Micro needle electrodes | Monopolar handpiece + Micro<br>needle electrodes | Same |
{7}------------------------------------------------
| Mode of operation | Monopolar mode (Normal mode) | Monopolar (SC mode) | Same monopolar |
|----------------------|------------------------------|---------------------|----------------|
| Method of activation | Footswitch | Footswitch | Same |
| Output energy type | High frequency | High frequency | Same |
| Frequency | 2MHz ± 10% | 2MHz ± 10% | Same |
| Max power | Max 45W at 500Ω | Max 45W at 500Ω | Same |
ACUTRON has the same technical specifications and same indications for use as those of the predicate device and the reference device.
The main differences are:
- Change in the dimension of the device -
- Change in GUI design -
- 1. Main unit(RF console)
ACUTRON has been specifically designed for the Secret RF Smartcure applicator handpieces, the predicate device (K182355) and identical operating specifications compared with the Secret RF system, the reference device (K170325). Electrical and mechanical safety and performance of ACUTRON electrosurgical system have been verified according to international standards.
- 2. Handpiece
ACUTRON handpieces include the identical handpiece needles cleared for the predicate device, Smartcure Applicator (K182355). Therefore, ACUTRON handpieces are subject to the same biocompatibility, sterilization, cleaning, shelf-life performance validation testing for the Smartcure Applicator handpieces, the predicate device, K182355.
- 3. GUI Design
ACUTRON has been specifically designed for the Secret RF Smartcure applicator handpieces, the predicate device(K182355). The normal mode is identical to the SC mode of the Secret RF system , the reference device(K170325).
#### VI. PERFORMANCE DATA
Design verification process were performed following risk assessment to verify that no new questions of safety and effectiveness have been raised due to the modifications
{8}------------------------------------------------
introduced.
### Biocompatibility testing:
The patient contact components and materials are identical to the predicate device. The biocompatibility is being leveraged from the predicate device.
### Non Clinical testing:
IEC 60601-1: 2005, AMD 1: 2012 Medical Electrical Equipment was performed for General Requirements for basic safety and essential performance. The requirements of specified standards were fulfilled.
IEC 60601-1-2: 2014 Medical Electrical Equipment was performed for General Requirements for basic safety and essential performance (collateral standards: electromagnetic compatibility.
IEC 60601-2-2: 2017 Medical electrical equipment Part 2: Particular requirements for the basic safety and essential performance of high-frequency surgical equipment and high-frequency surgical accessories.
### Animal testing:
Ex-vivo histology testing for ACUTRON handpieces is being leveraged from the Smartcure Applicator handpieces, the predicate device.
#### VII. CONCLUSIONS
There are no significant differences between ACUTRON and the predicate device and the reference device.
The proposed device does not raise any additional questions regarding safety and effectiveness. ACUTRON has the same indication of use and shares the same technological characteristics as the predicate devices.
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification, it is the opinion of Ilooda Co, Ltd. that ACUTRON is substantially equivalent in comparison with the predicate device and the reference device as described herein.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.