K203640 · Bolder Surgical, LLC · GEI · May 6, 2021 · General, Plastic Surgery
Device Facts
Record ID
K203640
Device Name
CoolSeal Reveal
Applicant
Bolder Surgical, LLC
Product Code
GEI · General, Plastic Surgery
Decision Date
May 6, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The CoolSeal™ Reveal is a bipolar electrosurgical instrument intended for use in open surgical procedures in adults and pediatrics where ligation and division of vessels, tissue bundles, and lymphatics is desired. The CoolSeal™ Reveal can be used on vessels (arteries and veins) up to and including 6 mm in diameter. It is indicated for use in general surgery and in such surgical specialties as urologic, thoracic, plastic, and reconstructive. Procedures may include but are not limited to, bowel resections, gall bladder procedures, Nissen fundoplication, and adhesiolysis. The CoolSeal™ Reveal is also indicated for open ENT procedures in adults (thyroidectomy, radical neck dissection, and parotidectomy) for ligation and division of vessels, lymphatics and tissue bundles 2-3 mm away from unintended thermally-sensitive structures such as nerves and parathyroid glands. The CoolSeal™ Reveal has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the CoolSeal™ Reveal for these procedures.
Device Story
CoolSeal Reveal is a sterile, single-use, hand-held bipolar electrosurgical instrument for open surgery. Used by surgeons in ORs. Device features 5 mm shaft, 10 cm length, and curved Maryland-style rotating jaw. Input: RF energy from CoolSeal Generator applied to tissue/vessels between jaws. Operation: surgeon grasps tissue, activates RF via finger button for sealing, then uses manual trigger to actuate internal blade for division. Output: sealed and divided vessels/tissue bundles up to 6 mm. Benefits: provides hemostasis and tissue dissection; enables ligation near thermally-sensitive structures (nerves/parathyroid glands) in ENT procedures. Device includes automatic sealing cycle.
Clinical Evidence
Bench testing included ex-vivo/in-vivo vessel burst pressure testing on porcine models to verify sealing performance. In-vivo thermal spread comparison via histological assessment (length, width, depth) demonstrated safety comparable to predicate. Chronic animal study (23-day duration) confirmed maintained hemostasis and anticipated healing without complications.
Technological Characteristics
Materials: stainless steel, titanium, polymers, silicone, polymer adhesives, insulating coatings. Energy: Bipolar RF (190 Vpeak, 30W max). Form factor: 5 mm shaft, 10 cm length, 335° jaw rotation. Sterilization: Ethylene Oxide (SAL 10^-6). Connectivity: Connects to CoolSeal Generator.
Indications for Use
Indicated for ligation and division of vessels, tissue bundles, and lymphatics in open surgical procedures. Patient population: adults and pediatrics for general surgery; adults for ENT procedures (thyroidectomy, radical neck dissection, parotidectomy). Contraindicated for tubal sterilization/coagulation.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a sans-serif font.
May 6, 2021
Bolder Surgical, LLC Nicholas Wong Regulatory Affairs Manager 331 S. 104th Street, Suite 200 Louisville, Colorado 80027
Re: K203640
Trade/Device Name: CoolSeal Reveal Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: April 9, 2021 Received: April 12, 2021
Dear Nicholas Wong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrl-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K203640
Device Name CoolSeal Reveal
#### Indications for Use (Describe)
The CoolSeal™ Reveal is a bipolar electrosurgical instrument intended for use in open surgical procedures in adults and pediatrics where ligation and division of vessels, tissue bundles, and lymphatics is desired. The CoolSeal™ Reveal can be used on vessels (arteries and veins) up to and including 6 mm in diameter. It is indicated for use in general surgery and in such surgical specialties as urologic, thoracic, plastic, and reconstructive. Procedures may include but are not limited to, bowel resections, gall bladder procedures, Nissen fundoplication, and adhesiolysis.
The CoolSeaT™ Reveal is also indicated for open ENT procedures in adults (thyroidectorny, radical neck dissection, and parotidectomy) for ligation and division of vessels, lymphatics and tissue bundles 2-3 mm away from unintended thermallysensitive structures such as nerves and parathyroid glands.
The CoolSeal™ Reveal has not been shown to be effective for tubal coagulation for sterilization procedures. Do not use the CoolSeal™ Reveal for these procedures.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Bolder Surgical. The word "BOLDER" is in large, bold, blue letters. The word "SURGICAL" is in smaller, thinner, blue letters below the word "BOLDER". The letter "O" in "BOLDER" has a black diamond shape in the upper left quadrant.
#### 510(k) Summary
#### SUBMITTER
Bolder Surgical, LLC. 331 S. 104th Street, Suite 200 Louisville, CO 80027
Phone: 720-287-7130 Fax: 720-287-7135
Contact Person: Nick Wong Regulatory Affairs Manager
Date Prepared: May 6th, 2021
## DEVICE
| Trade Name / Model #: | CoolSeal™ Reveal / CSL-RV105-10 |
|-----------------------|----------------------------------------------------------------|
| Common Name: | Bipolar Vessel Sealing System |
| Classification Name: | Electrosurgical cutting and coagulation device and accessories |
| Regulatory Class: | II |
| Produce Code: | GEI |
#### PREDICATE DEVICES
| Trade Name: | LigaSure Exact Dissector, Nano-Coated |
|-----------------------|----------------------------------------------------------------|
| 510(k): | K173281 |
| Common or Usual Name: | Bipolar Vessel Sealing System |
| Classification Name: | Electrosurgical cutting and coagulation device and accessories |
| Regulatory Class: | II |
| Produce Code: | GEI |
| Trade Name: | JustRight™ Sealer |
| 510(k) | K160602 |
| Common or Usual Name: | Bipolar Vessel Sealing System |
| Classification Name: | Electrosurgical cutting and coagulation device and accessories |
| Regulatory Class: | II |
| Produce Code: | GEI |
## DEVICE DESCRIPTION
#### CoolSeal™ Reveal:
The CoolSeal™ Reveal device is a sterile, single-use, hand-held bipolar vessel sealing device designed for use with the CoolSeal™ Generator. The CoolSeal™ Reveal is a hemostat-style
{4}------------------------------------------------
design with a 5 mm shaft diameter and 10 cm shaft length. The jaw is a curved Maryland-style jaw that rotates. The Reveal creates seals by application of radiofrequency (RF) electrosurgical energy to vascular structures (vessels and lymphatics) or tissue bundles interposed between its iaws. A blade within the instrument is surgeon-actuated to divide tissue. The double-action jaws have been designed to dissect tissue, which includes separating tissue planes and widening openings as necessary for the surgical procedure.
# INDICATIONS FOR USE
## CoolSeal™ Reveal:
The CoolSeal™ Reveal is a bipolar electrosurgical instrument intended for use in open surgical procedures in adults and pediatrics where ligation and division of vessels, tissue bundles, and lymphatics is desired. The CoolSeal™ Reveal can be used on vessels (arteries and veins) up to and including 6 mm in diameter. It is indicated for use in general surgery and in such surgical specialties as urologic, thoracic, plastic, and reconstructive. Procedures may include but are not limited to, bowel resections, gall bladder procedures, Nissen fundoplication, and adhesiolysis.
The CoolSeal™ Reveal is also indicated for open ENT procedures in adults (thyroidectomy, radical neck dissection, and parotidectomy) for ligation and division of vessels, lymphatics and tissue bundles 2-3 mm away from unintended thermally-sensitive structures such as nerves and parathyroid glands.
The CoolSeal™ Reveal has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the CoolSeal™ Reveal for these procedures.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Bolder Surgical. The word "BOLDER" is in large, bold, blue font, with a circle in place of the letter "O". A small black diamond is in the upper left quadrant of the circle. Below "BOLDER" is the word "SURGICAL" in a smaller, thinner, black font.
# COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES
Tables 1 and 2 below present the comparison of technological characteristics between the subject device and the two identified predicate devices.
| Description | CoolSeal™ Reveal (Subject Device) | LigaSure Exact (Primary Predicate) |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The CoolSeal™ Reveal is a bipolar electrosurgical<br>instrument intended for use in open surgical procedures in<br>adults and pediatrics where ligation and division of<br>vessels, tissue bundles, and lymphatics is desired. The<br>CoolSeal™ Reveal can be used on vessels (arteries and<br>veins) up to and including 6 mm in diameter. It is<br>indicated for use in general surgery and in such surgical<br>specialties as urologic, thoracic, plastic, and<br>reconstructive. Procedures may include but are not limited<br>to, bowel resections, gall bladder procedures, Nissen<br>fundoplication, and adhesiolysis.<br>The CoolSeal™ Reveal is also indicated for open ENT<br>procedures in adults (thyroidectomy, radical neck<br>dissection, and parotidectomy) for ligation and division of<br>vessels, lymphatics and tissue bundles 2-3 mm away from<br>unintended thermally-sensitive structures such as nerves<br>and parathyroid glands.<br>The CoolSeal™ Reveal has not been shown to be effective<br>for tubal sterilization or tubal coagulation for sterilization<br>procedures. Do not use the CoolSeal™ Reveal for these<br>procedures. | The LigaSure Sealer/Divider is a bipolar electrosurgical instrument<br>intended for use in open surgical procedures where ligation and<br>division of vessels, tissue bundles, and lymphatics is desired. The<br>LigaSure Sealer/ Divider can be used on vessels (arteries and veins)<br>up to and including 7 mm. It is indicated for use in general surgery<br>and in such surgical<br>specialties as urologic, thoracic, plastic, and reconstructive.<br>Procedures may include, but are not limited to, bowel resections, gall<br>bladder procedures, Nissen fundoplication, and adhesiolysis.<br>The instrument is also indicated for open ENT procedures in adults<br>(thyroidectomy, radical neck dissection, parotidectomy, and<br>tonsillectomy) for ligation and division of vessels, lymphatics and<br>tissue bundles 2-3 mm away from unintended thermally-sensitive<br>structures such as nerves and parathyroid glands.<br>The LigaSure system has not been shown to be effective for tubal<br>sterilization or tubal coagulation for sterilization<br>procedures. Do not use the LigaSure system for these procedures. |
| Where used<br>(environment) | Operating Room | Operating Room |
| Intended User | Surgeon | Surgeon |
| Description | CoolSeal™ Reveal (Subject Device) | LigaSure Exact (Primary Predicate) |
| Anatomical Sites | Vessels, tissue bundles, and lymphatics | Vessels, tissue bundles, and lymphatics |
| Anatomical Size | Up to and including 6 mm diameter | Up to and including 7 mm diameter |
| Patient<br>Population | Adult and Pediatric in General Surgery.<br>Adult in ENT surgery. | Not specified (Adults assumed) |
| Power Source | Bipolar energy platform | Bipolar energy platform |
| Primary<br>Functions | Grasp, Dissect, Seal, Divide | Grasp, Dissect, Seal, Divide |
| Mechanism of<br>Grasping | Hand actuated lever allows user to open or close | Hand actuated lever allows user to open or close |
| Mechanism of<br>Dissection<br>(Separation) | Bilateral jaw allows user to separate planes of tissue | |
| Mechanism of<br>Action (Sealing) | Seal is created by application of RF energy to structures<br>interposed between the jaws of the instrument. | Seal is created by application of RF energy to structures interposed<br>between the jaws of the instrument. |
| Seal Activation | Button on the sealer instrument is activated by user finger<br>depression. | Button on the sealer instrument is hand activated by closing the<br>handles until the button is depressed. |
| Knife Activation | Cutting trigger – non-energized | Cutting trigger – non-energized |
| Automatic<br>sealing cycle | Yes | Yes |
| Rated Voltage<br>(Vpeak) | 190 Vpeak | 288 Vpeak |
| Shaft Diameter | 5 mm | N/A |
| Shaft Length | 10 cm | N/A |
| Jaw Rotation | $≤$ 335° | N/A |
| Jaw Dimensions<br>– Jaw Length<br>and Side Width | Jaw Length: 12 mm<br>Proximal Side Width: 4.3 mm<br>Distal Side Width: 2.0 mm | Jaw Length: 20.6 mm<br>Proximal Side Width: 4.2 mm<br>Distal Side Width: 2.3 mm |
| Jaw Dimensions<br>- Tip Width<br>(proximal and<br>distal) | Proximal: 5.2 mm<br>Distal: 3.2 mm | Proximal: 6.7 mm<br>Distal: 3.6 mm |
| Seal Plate Width | Seal Plate Width: 1.8 mm throughout | Seal Plate Width: |
| Description | CoolSeal™ Reveal (Subject Device) | LigaSure Exact (Primary Predicate) |
| | | Proximal: 4.0 mm<br>Distal: 1.75 mm |
| Maximum size<br>structure (vessel,<br>tissue bundles,<br>lymphatic) for<br>sealing | Up to 6 mm | Up to 7 mm |
| Tissue Contact<br>Materials | Stainless steel, titanium, polymers, silicone, polymer<br>adhesives and insulating coatings | Stainless steel, titanium, polymers, silicone, polymer adhesives and<br>insulating coatings |
| Surgical<br>Approach | Open | Open |
| Sterilization | Ethylene Oxide…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.