← Product Code [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI) · K203522

# Surn5 Electrosurgical Argon Beam Generator and Accessories (K203522)

_Surnic Corporation · GEI · Feb 22, 2021 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K203522

## Device Facts

- **Applicant:** Surnic Corporation
- **Product Code:** [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI.md)
- **Decision Date:** Feb 22, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Surn5 Electrosurgical Argon Beam Generator and Accessories are indicated for monopolar or bipolar surgery to achieve argon beam and cut or coagulation for the tissue. It is intended to be used with monopolar handpiece and dispersive electrode or bipolar handpiece and footswitches.

## Device Story

Electrosurgical generator system; provides high-frequency (HF) power (400-600 kHz) for surgical cutting and coagulation. Inputs: user-selected modes and power settings via front panel; activation via handpiece or footswitch. Outputs: HF electrical current delivered to tissue via monopolar pencil or bipolar handpiece; argon beam delivery for coagulation. Used in clinical surgical settings by physicians. Device transforms electrical input into specific waveforms (sinusoidal, constant/modulated) to achieve desired tissue effects. Benefits include precise tissue management during surgery. Output status indicated on front panel.

## Clinical Evidence

Bench testing only. Compliance demonstrated with IEC 60601-1 (general safety), IEC 60601-1-2 (EMC), IEC 60601-2-2 (HF surgical equipment safety), and IEC 60601-1-6 (usability). Performance testing conducted per FDA guidance 'Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery'.

## Technological Characteristics

HF generator; 400-600 kHz output; 100-240VAC 50/60Hz. Modes: 4 Monopolar Cut, 8 Monopolar Coagulation, Argon Beam, 4 Bipolar. Max power: 320W (Monopolar Cut), 120W (Coag), 150W (Argon), 100W (Bipolar). Waveforms: Sinusoidal constant/modulated, Cut/Coag pulse phases. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2, IEC 60601-1-6.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- HelixAR ABC Systems ([K172671](/device/K172671.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

February 22, 2021

Surnic Corporation Helen Xie RA Assistant 480 Apollo Street, Suite D. Brea, California 92821

Re: K203522

Trade/Device Name: Surn5 Electrosurgical Argon Beam Generator and Accessories Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: December 29, 2020 Received: December 31, 2020

Dear Helen Xie:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Actinclude requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Long Chen. Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K203522

#### Device Name

Surn5 Electrosurgical Argon Beam Generator and Accessories

Indications for Use (Describe)

The Surns Electrosurgical Argon Beam Generator and Accessories are indicated for monopolar or bipolar surgery to achieve argon beam and cut or coagulation for the tissue. It is intended to be used with monopolar handpiece and dispersive electrode or bipolar handpiece and footswitches.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the title "510(k) Summary" in a large font. Below the title is a horizontal line, and below the line is the text "Traditional 510(k) Submission". To the left of the title is a black graphic that looks like a check mark.

| K203522<br>Section 5 |             |
|----------------------|-------------|
| Version              | 1           |
| Pages                | 3           |
| Date                 | Feb-19-2021 |

### 1. Submitter Information

| Company: | Surnic Corporation                        |
|----------|-------------------------------------------|
| Address: | Suite D, 480 Apollo Street, Brea CA 92821 |
| Contact: | Helen Xie                                 |
| Phone:   | 1(909) 859 - 9055                         |
| Email:   | hx9055@gmail.com                          |

# 2. Device Identification

| Common name:             | Electrosurgical Generator and Accessories                  |
|--------------------------|------------------------------------------------------------|
| Classification name:     | Electrosurgical, Cutting & Coagulation & Accessories       |
| Trade name:              | Surn5 Electrosurgical Argon Beam Generator and Accessories |
| Model:                   | Surn5                                                      |
| Device Class             | Class II                                                   |
| Product Code:            | GEI                                                        |
| 21CFR Regulation Number: | 878.4400                                                   |

### 3. Predicate Device

Product HelixAR ABC Systems 510(K) K172671

## 4. Device Description

Surn5 electrosurgical argon beam Generator and accessories have Monopolar Pure cut, Monopolar Bland Cut, Argon Beam, Bipolar functions. Surn5 electrosurgical argon beam Generator and accessories consists of HF generator and accessories. The device generates HF power, which is transmitted to the tissue through the output Pencil. Operator selects the output function and power with the front panel. The indicator is used to indicate the setup status and output status of the device. The output is controlled by foot or hand switch.

### 5. Indication for use

The Surn5 Electrosurgical Argon Beam Generator and Accessories are indicated for monopolar or bipolar surgery to achieve argon beam and cut or coagulation for the tissue. It is intended to be used with monopolar handpiece and dispersive electrode or bipolar handpiece and footswitches.

{4}------------------------------------------------

# 6. Substantial Equivalence Discussion

# Technological characteristics

| Device                 | Subject device                                                | Predicate device                                              |
|------------------------|---------------------------------------------------------------|---------------------------------------------------------------|
| Manufacturer           | Surnic Corporation                                            | Conmed Corporation                                            |
| Model                  | Surn5 Electrosurgical Argon Beam<br>Generator and Accessories | HelixAR ABC System                                            |
| 510(k) Number          | K203522                                                       | K172671                                                       |
| Class                  | II                                                            | II                                                            |
| Product Code           | GEI                                                           | GEI                                                           |
| Regulation Number      | 21 CFR 878.4400                                               | 21 CFR 878.4400                                               |
| Mechanism of<br>Action | The device generates high<br>Frequency current<br>400-600 kHz | The device generates high<br>Frequency current<br>385-600 kHz |
| System Voltage         | 100-240VAC- 50/60Hz                                           | 100-240VAC- 50/60Hz                                           |
|                        | 4 Monopolar Cut Modes                                         |                                                               |
| Max. Power             | 320 W (at 300Ω)                                               | 300 W (at 300Ω)                                               |
| Wave Forms             | Sinusoidal<br>constant / modulated<br>Cut / Coag Pulse Phases | Sinusoidal<br>constant / modulated<br>Cut / Coag Pulse Phases |
|                        | 8 Monopolar Coagulation Modes                                 |                                                               |
| Max. Power             | 120 W (at 500Ω)                                               | 120 W (at 500Ω)                                               |
| Wave Forms             | Sinusoidal<br>constant / modulated<br>Cut / Coag Pulse Phases | Sinusoidal<br>constant / modulated<br>Cut / Coag Pulse Phases |
|                        | Argon Beam                                                    |                                                               |
| Max. Power             | 150W (at 500Ω)                                                | 150W (at 500Ω)                                                |
|                        | 4 Bipolar Modes                                               |                                                               |
| Max. Power             | 100 W (at 100Ω)                                               | 90 W (at 300Ω)                                                |
| Wave Forms             | Sinusoidal constant / modulated<br>Pause Phases               | Sinusoidal constant / modulated<br>Pause Phases               |

{5}------------------------------------------------

## K203522

### 7. Non-Clinical Performance Data

To demonstrate the safety and effectiveness of Surn5 Electrosurgical Argon Beam Generator and Accessories show substantial equivalence to the predicate device (Surn5 Electrosurgical Argon Beam Generator and Accessories), Medrange completed successfully the following non-clinical performance tests:

| General Safety Testing | IEC 60601-1:2005+C1+C2+A1:2012 Medical electrical equipment, Part 1<br>General requirements for basic safety and essential performance                                                                                   |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| EMC Safety Testing     | IEC 60601-1-2:2014 (Edition 4.0) Medical electrical equipment, Part 1-2<br>General requirements for basic safety and essential performance Collateral<br>Standard: Electromagnetic disturbances - Requirements and tests |
| HF Safety Testing      | IEC 60601-2-2:2017 Medical electrical equipment, Part 2-2<br>Particular requirements for the basic safety and essential performance of high<br>frequency - Surgical equipment and high frequency surgical accessories    |
| Performance Testing    | Applicable parts of FDA Guidance Document “Premarket Notification<br>(510(k)) Submissions for Electrosurgical Devices for General Surgery”, May<br>2016                                                                  |
| Usability Testing      | IEC 60601-1-6:2010, AMD1:2013 Medical electrical equipment Part 1-6<br>General requirements for safety - Collateral Standard: Usability                                                                                  |

## 8. Statement of Substantial Equivalence

The Surn5 Electrosurgical Argon Beam Generator and Accessories have the same intended use as the HelixAR, and the same technological characteristics. The non-clinical test results, such as performance data, software data, electrical safety and electromagnetic compatibility data have demonstrated that Surn5 Electrosurgical Argon Beam Generator and Accessories are as safe and effective as the predicate device. Therefore it is concluded, that the Surn5 Electrosurgical Argon Beam Generator and Accessories fulfill the requirements of a substantially equivalent device and that no new questions of safety and effectiveness were raised.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K203522](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K203522)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
