K203461 · Aesculap, Inc. · GEI · Dec 18, 2020 · General, Plastic Surgery
Device Facts
Record ID
K203461
Device Name
Aesculap Caiman 12 Seal and Cut Technology System
Applicant
Aesculap, Inc.
Product Code
GEI · General, Plastic Surgery
Decision Date
Dec 18, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Caiman Seal and Cut Technology System instruments are intended for use in general surgery and gynecologic surgical procedures where ligation and division of vessels is desired.
Device Story
Sterile, single-use bipolar electrosurgical instrument; connects to dedicated bipolar RF generator. Used in general/gynecologic surgery for vessel ligation and tissue division. Surgeon grasps tissue between jaws; applies bipolar RF energy to seal vessels; actuates mechanical blade to cut tissue. Used in OR for open or laparoscopic procedures. Provides hemostasis and tissue separation; benefits patient by enabling efficient vessel ligation and dissection.
Clinical Evidence
Bench testing only. Verification testing included instrument cycling and jaw clamping force (170 N - 350 N) to confirm design modifications met acceptance criteria.
Indicated for general surgery and gynecologic procedures (urologic, vascular, thoracic, thoracoscopic) requiring vessel ligation and division, tissue grasping, and dissection. Suitable for vessels up to 7mm and tissue bundles fitting in jaws. 36cm/44cm lengths for laparoscopic use; 24cm length for open procedures. Contraindicated for tubal sterilization or tubal coagulation.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 18, 2020
Aesculap Inc. Omunique Luke Regulatory Affairs Specialist 3773 Corporate Parkway Center Valley, Pennsylvania 18034
Re: K203461
Trade/Device Name: Aesculap Caiman 12 Seal and Cut Technology System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: November 23, 2020 Received: November 24, 2020
Dear Omunique Luke:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm_identifies_combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Long Chen. Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K203461
#### Device Name
Aesculap Caiman 12 Seal and Cut Technology System
#### Indications for Use (Describe)
Caiman Seal and Cut Technology System consists of dedicated bipolar electrosurgical instruments intended for use in general surgery and gynecologic surgical procedures where ligation and division of vessels is desired. The instruments create a seal by the application of bipolar electrosurgical RF energy (coagulation) to vascular structure (vessels) interposed between the jaws of the device. A cutting blade is actuated for the division of tissue.
Instruments 36cm and 44cm in length are indicated for laparoscopic procedures and instruments 24 cm in length are indicated for open procedures. The indications for use include general surgical procedures, (including urologic, vascular, thoracic, and thoracoscopic), and gynecological procedures where ligation and division of vessels is performed. These procedures include: vaginal hysterectomies, Nissen fundoplication, colectomy, adhesiolysis, bowel resection, and oophorectomy etc., or any procedure where vessel ligation (seal and cut), tissue grasping, and dissection is performed. The devices can be used on vessels up to and including 7mm and bundles as large as will fit in the jaws of the instrument.
Caiman Seal and Cut Technology System has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the system for these procedures.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------|-----------------------------------------------|
| <span style="text-decoration: underline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
*PRAStaff@fda.hhs.gov*
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) SUMMARY (as required by 21 CFR 807.92)
#### Caiman 12 Seal and Cut Technology System December 17, 2020
| COMPANY: | Aesculap, Inc.<br>3773 Corporate Parkway<br>Center Valley, PA 18034<br>Establishment Registration Number: 2916714 |
|----------------------|-------------------------------------------------------------------------------------------------------------------|
| CONTACT: | Nikki Luke<br>484-523-6868 (phone)<br>nikki.luke@aesculapimplants.com<br>610-791-6882 (fax) |
| TRADE NAME: | Aesculap® Caiman 12 Seal and Cut Technology System |
| COMMON NAME: | Electrosurgical, Cutting & Coagulation & Accessories |
| CLASSIFICATION NAME: | General and Plastic Surgery |
| REGULATION NUMBER: | 21 CFR 878.4400 |
| PRODUCT CODE: | GEI |
| DEVICE CLASS: | Class II per 21 CFR 878.4400 |
### PREDICATE DEVICE
Caiman Seal and Cut Technology (K130596)
### DEVICE DESCRIPTION
The Aesculap® Caiman Seal and Cut Technology System consists of sterile, single-use bipolar instruments that connect to a dedicated bipolar RF generator. The Caiman instruments are designed to grasp, seal (ligate), and mechanically divide (cut) varying lengths of tissue (for example mesentery) per device application. This is accomplished by a two electrode (side by side-top and bottom) design. The instruments are capable of vessel sealing, grasping, and dividing tissue enclosed within its dissection clips.
The reason for this submission is to capture modifications to the caiman 12 (24cm and 44cm) instruments.
### INDICATIONS FOR USE
Caiman Seal and Cut Technology System consists of dedicated bipolar electrosurgical instruments intended for use in general surgery and gynecologic surgical procedures where ligation and division of vessels is desired. The instruments create a seal by the application of
{4}------------------------------------------------
bipolar electrosurgical RF energy (coagulation) to vascular structure (vessels) interposed between the jaws of the device. A cutting blade is actuated for the division of tissue.
Instruments 36cm and 44cm in length are indicated for laparoscopic procedures and instruments 24 cm in length are indicated for open procedures. The indications for use include general surgical procedures, (including urologic, vascular, thoracic, and thoracoscopic), and gynecological procedures where ligation and division of vessels is performed. These procedures include: vaginal hysterectomies, Nissen fundoplication, colectomy, adhesiolysis, bowel resection, and oophorectomy etc., or any procedure where vessel ligation (seal and cut), tissue grasping, and dissection is performed. The devices can be used on vessels up to and including 7mm and bundles as large as will fit in the jaws of the instrument.
Caiman Seal and Cut Technology System has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the system for these procedures.
#### TECHNOLIGICAL CHARACTERISTICS (compared to predicate devices)
The table below provides a summary of the device technological characteristics comparing the subject device and predicate device. The modifications made to these devices do not raise any new issues of safety and effectiveness, as confirmed by the testing and validation activities described in the submission.
{5}------------------------------------------------
| | Subject Device: Caiman Seal<br>and Cut Technology System<br>Product Code: GEI<br>K203461 | Predicate Device: Caiman Seal<br>and Cut Technology System<br>Product Code: GEI<br>K130596 | |
|------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|
| Intended Use | Caiman Seal and Cut Technology<br>System instruments are intended<br>for use in general surgery and<br>gynecologic surgical procedures<br>where ligation and division of<br>vessels is desired. | Caiman Seal and Cut Technology<br>consists of dedicated bipolar<br>electrosurgical instruments<br>intended for use in general surgery<br>and gynecologic surgical<br>procedures where ligation and<br>division of vessels is desired. | Same |
| Indications for Use | Caiman Seal and Cut Technology<br>System consists of dedicated<br>bipolar electrosurgical instruments<br>intended for use in general surgery<br>and gynecologic surgical<br>procedures where ligation and<br>division of vessels is desired. The<br>instruments create a seal by the<br>application of bipolar<br>electrosurgical RF energy<br>(coagulation) to vascular structure<br>(vessels) interposed between the<br>jaws of the device. A cutting blade<br>is actuated for the division of<br>tissue.<br>Instruments 36cm and 44cm in<br>length are indicated for<br>laparoscopic procedures and<br>instruments 24 cm in length are<br>indicated for open procedures. The<br>indications for use include general<br>surgical procedures, (including<br>urologic, vascular, thoracic, and<br>thoracoscopic), and gynecological<br>procedures where ligation and<br>division of vessels is performed.<br>These procedures include: vaginal<br>hysterectomies, Nissen<br>fundoplication, colectomy,<br>adhesiolysis, bowel resection, and<br>oophorectomy etc., or any<br>procedure where vessel ligation<br>(seal and cut), tissue grasping,<br>and dissection is performed. The<br>devices can be used on vessels up | Caiman Seal and Cut Technology<br>consists of dedicated bipolar<br>electrosurgical instruments<br>intended for use in general surgery<br>and gynecologic surgical<br>procedures where ligation and<br>division of vessels is desired. The<br>instruments create a seal by the<br>application of bipolar<br>electrosurgical RF energy<br>(coagulation) to vascular structure<br>(vessels) interposed between the<br>jaws of the device. A cutting blade<br>is actuated for the division of<br>tissue.<br>The Caiman 12 Plus (44cm) and<br>the Caiman 5 (36cm) are indicated<br>for laparoscopic procedures and<br>the Caiman 12 Plus (24cm) is<br>indicated for open procedures. The<br>indications for use include general<br>surgical procedures, (including<br>urologic, vascular, thoracic, and<br>thoracoscopic), and gynecological<br>procedures where ligation and<br>division of vessels is performed.<br>These procedures include: vaginal<br>hysterectomies, Nissen<br>fundoplication, colectomy,<br>adhesiolysis, bowel resection, and<br>oophorectomy etc., or any<br>procedure where vessel ligation<br>(seal and cut), tissue grasping,<br>and dissection is performed. The<br>devices can be used on vessels up | Same |
| | to and including 7mm and bundles | to and including 7mm and bundles | |
| | as large as will fit in the jaws of the | as large as will fit in the jaws of the | |
| | instrument. | instrument. | |
| | Caiman Seal and Cut Technology<br>System has not been shown to be<br>effective for tubal sterilization or<br>tubal coagulation for sterilization<br>procedures. Do not use the system<br>for these procedures. | Caiman Seal and Cut Technology<br>has not been shown to be effective<br>for tubal sterilization or tubal<br>coagulation for sterilization<br>procedures. Do not use the system<br>for these procedures. | |
| Material Composition | Handle - molded thermoplastic<br>Shaft - stainless steel and molded<br>thermoplastic<br>Jaw - stainless steel<br>Jaw Insulation Components -<br>PEEK | Handle - molded thermoplastic<br>Shaft - stainless steel and molded<br>thermoplastic<br>Jaw - stainless steel<br>Jaw Insulation Components -<br>PEEK | Same |
| Functional Use | Grasping Ligation and Coagulation | Grasping Ligation and Coagulation | Same |
| Surgical Approach | Caiman 12 (44cm)-Laparoscopic<br>Caiman 12 (24cm)- Open | Caiman 12 (44cm)-Laparoscopic<br>Caiman 12 (24cm)- Open | Same |
| Length | 24 cm<br>44 cm | 24 cm<br>44 cm | Same |
| Jaw | Upper Jaw Assembly<br>Stainless-Steel Material: 420 SST<br>(EN ISO 1.4021)<br>Stainless-Steel Material: 304 (EN<br>ISO 1.4301)<br>Electrical Insulation Features:<br>PEEK<br>Lower Jaw Assembly<br>Stainless-Steel Material: 17-4 PH<br>(EN ISO 1.4542)<br>Stainless-Steel Material: 304 (EN<br>ISO 1.4301)<br>Electrical Insulation Features:<br>PEEK<br>Jaw Shape: Straight | Upper Jaw Assembly<br>Stainless-Steel Material: 420 SST<br>(EN ISO 1.4021)<br>Stainless-Steel Material: 304 (EN<br>ISO 1.4301)<br>Electrical Insulation Features:<br>PEEK<br>Lower Jaw Assembly<br>Stainless-Steel Material: 17-4 PH<br>(EN ISO 1.4542)<br>Stainless-Steel Material: 304 (EN<br>ISO 1.4301)<br>Electrical Insulation Features:<br>PEEK<br>Jaw Shape: Straight | Same |
| Jaw Clamp Force | Within Specification 170 N - 350 N | Within Specification 170 N - 350 N | Same |
| Diameter | 12mm | 12mm | Same |
| Number of Electrodes (Pairs) | 2 | 2 | Same |
| Electrode<br>Length | 5 cm | 5 cm | Same |
| Electrode<br>Width | 0.25 cm | 0.25 cm | Same |
| Electrode Texture | Smooth with PEEK stops to<br>maintain consistent gap between<br>electrode surfaces | Smooth with PEEK stops to<br>maintain consistent gap between<br>electrode surfaces | Same |
| Cable | 10ft | 10ft | Same |
| Sterile/Single Use | Yes | Yes | Same |
| Shelf Life | 3 years | 3 years | Same |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
# PERFORMANCE TESTING
Non-clinical testing was conducted as part of demonstrating substantial equivalence to the predicate device and ensure the design changes met the predetermined acceptance criteria. A risk analysis was completed to identify the risks associated with the modifications to the Caiman Seal and Cut Technology System. Verification testing was conducted to evaluate the modification. The following tests associated with the device modifications were performed on the subject device according to the methods and acceptance criteria as required per test protocol.
See Design verification table below:
#### Design verification
| Test | Acceptance Criteria | Results |
|-------------------------|----------------------------------------------------------------------|---------|
| Instrument Cycling Test | No pin walking shall be observed | Pass |
| Jaw Clamping Force Test | Jaw clamping force must remain within<br>specification 170 N - 350 N | Pass |
The testing performed verified and validated that the Caiman 12 Seal and Cut Technology System has met all acceptance criteria.
## Conclusion
The conclusions drawn from the nonclinical tests demonstrate that the modifications described in this submission do not affect the intended use of the device or alter the fundamental scientific technology of the device.