K203299 · Pulse Biosciences, Inc. · GEI · Feb 2, 2021 · General, Plastic Surgery
Device Facts
Record ID
K203299
Device Name
CellFX System
Applicant
Pulse Biosciences, Inc.
Product Code
GEI · General, Plastic Surgery
Decision Date
Feb 2, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The CellFX® System is intended for dermatological procedures requiring ablation and resurfacing of the skin.
Device Story
CellFX System is a platform technology for dermatological ablation and resurfacing. Device consists of console, handpiece, and sterile treatment tips with stainless steel microneedles. Operates via Nano-Pulse Stimulation (NPS); delivers series of timed, nanosecond duration electrical pulses (100-700 ns) at amplitudes up to 15 kV. Pulses disrupt cellular function leading to cell death while sparing non-cellular tissue. Used in clinical settings by healthcare providers. User selects treatment parameters via console touchscreen. Output is targeted electrical energy applied to skin. Benefits include precise tissue ablation/resurfacing for dermatologic conditions.
Clinical Evidence
No human clinical data provided. Evidence consists of bench testing (biocompatibility per ISO 10993-1, electrical safety/EMC per IEC 60601 series) and a GLP animal safety study. Animal study used 15 Yucatan pigs comparing CellFX to predicate across 5 cohorts (0-30 days). Results showed successful degeneration/necrosis of target tissue with intact epidermis, no procedure-related complications, and equivalent safety/performance profiles.
Technological Characteristics
Console-based system delivering nanosecond electrical pulses (100-700 ns, 300V-15kV). Treatment tips feature stainless steel microneedles. Power input 100-240 VAC. Compliance with IEC 60601-1, 60601-1-2, 60601-2-2, 60601-1-6, IEC 62366, and IEC 62304. Biocompatibility per ISO 10993-1. Software level of concern: moderate.
Indications for Use
Indicated for dermatological procedures requiring ablation and resurfacing of the skin in patients requiring such dermatological treatment.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".
February 2, 2021
Pulse Biosciences, Inc. William Knape VP Regulatory, Clinical, & Quality Affairs 3957 Point Eden Way Hayward. California 94545
Re: K203299
Trade/Device Name: CellFX® System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: November 6, 2020 Received: November 9, 2020
Dear William Knape:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrb/ofdocs/cfpmp/pmn.cfm_identifies_combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Long Chen. Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K203299
Device Name CellFX® System
Indications for Use (Describe)
The CellFX® System is intended for dermatological procedures requiring ablation and resurfacing of the skin.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(K) SUMMARY
# 510(k) Summary
#### I. Submitter
| Applicant: | Pulse Biosciences, Inc.<br>3957 Point Eden Way<br>Hayward, CA 94545<br>Phone: (510) 906-4649 |
|-----------------|--------------------------------------------------------------------------------------------------------------------|
| Contact Person: | William A. Knape<br>VP Regulatory, Clinical, & Quality Affairs<br>Pulse Biosciences, Inc.<br>Phone: (919) 757-2033 |
| Date Prepared: | February 2, 2021 |
#### DEVICE INFORMATION II.
| Trade Name: | CellFX® System |
|-----------------------|-------------------------------------------------------------------|
| Regulation Number | 21 CFR § 878.4400 |
| Regulation Name: | Electrosurgical Cutting and Coagulation Device and<br>Accessories |
| Regulation Class: | Class II |
| Product Code: | GEI |
| Classification Panel: | General and Plastic Surgery |
#### III. PREDICATE DEVICE
K102461 - Fractora by Invasix Ltd. No reference devices were used in this submission.
#### IV. DEVICE DESCRIPTION
The Pulse Biosciences® CellFX® System is a proprietary platform technology. The CellFX System consists of the CellFX Console, CellFX Handpiece, CellFX Treatment Tips, and CellFX Software. The CellFX System delivers nanosecond duration electrical pulses that disrupt the function of cells leading to cell death, while sparing non-cellular tissue. The CellFX System delivers a series of timed, nanosecond electrical pulses (referred to as a "Cycle") to ablate and resurface tissue areas in dermatologic conditions.
The CellFX Console is capable of delivering short electric pulses at amplitudes up to 15 kV and pulse widths up to 700 ns. The electrical energy pulses are applied directly to targeted tissue using sterile Treatment Tips with stainless steel microneedles. The
{4}------------------------------------------------
treatment parameters are selected by the user through a user interface on the Touchscreen Display of the CellFX Console.
## V. INDICATIONS FOR USE STATEMENT
The CellFX® System is intended for dermatological procedures requiring ablation and resurfacing of the skin.
### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
To support a determination of substantial equivalence, Pulse Biosciences performed verification and validation testing demonstrating the subject device performs as intended; thus, even though the CellFX System has different technological characteristics than the predicate device, the differences do not raise new or different questions of safety and effectiveness.
| | Subject Device | Predicate Device (K102461) |
|------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | CellFX® System | Fractora |
| Prescription Only | Yes | Yes |
| Regulation Number | 878.4400 | 878.4400 |
| Product Code | GEI | GEI |
| Indications for Use | Indications for Use | |
| Indications for Use Statement | The CellFX System is intended for dermatological procedures requiring ablation and resurfacing of the skin. | Fractora is intended for dermatological procedures requiring ablation and resurfacing of the skin. |
| Technological Characteristics | | |
| Mechanism of Action | Nano-Pulse Stimulation: ultrafast nanosecond electrical pulses to the targeted tissue via treatment tips with an array of microneedles | Non-homogeneous (fractional) pulsed RF energy delivered to skin via an array of multi-electrode pins resulting in heating of skin directly below the electrodes |
| Pulse Frequency | 1-10 Pulses per second | Up to 2 Pulses per second |
| Pulse Amplitude/Voltage | 300V to 15kV | ~300 Volts |
| Pulse Width | 100 ns – 700 ns | Not Applicable |
| Power Input | 100-240 VAC, 50/60 Hz, 2A | 100-240 VAC, 50-60 Hz, 1.4 A |
| Power Output | 30 Watts | 75 Watts |
| Energy Range | Minimum: 0.5 Joules @ 0.06 Watts<br>Maximum: 4.3 Joules @ 0.6 Watts | Minimum: 0.2 Joules @ 75 Watts<br>Maximum: 3.7 Joules @ 75 Watts |
| Integrated Skin Temperature Sensor | Not Included | Included |
| Use of Local Anesthesia | Yes | Yes |
| Treatment Tip | Yes | Yes |
| Handpiece | Yes | Yes |
{5}------------------------------------------------
CELLFX® SYSTEM 510(k) Premarket Notification
| System Physical Characteristics | | |
|---------------------------------|-------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------|
| Height | 132 cm | 100 cm |
| Length | 46 cm | 36 cm |
| Width | 53 cm | 36 cm |
| Weight | 54 kg | 15 kg |
| Power Input | 100-240 VAC, 50/60 Hz, 2A | 100-240 VAC, 50-60 Hz, 1.4 A |
| Tip Characteristics | | |
| Microneedle<br>Configuration | 1.5x1.5mm: 2 rows of 3 pins per row<br>2.5x2.5mm: 2 rows of 3 pins per row<br>5.0x5.0mm: 2 rows of 4 pins per row | 24 pin tip: 6 rows of 4 pins per row<br>60 pin tip: 10 rows of 6 pins per row |
#### PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
#### Biocompatibility Testing
The testing was conducted in accordance with FDA Guidance, issued June 16, 2016: Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process." All biocompatibility tests met their respective acceptance criteria.
The battery of testing included the following:
- Material-Mediated Pyrogenicity ●
- Systemic Toxicity ●
- Sensitization ●
- Irritation ●
- Cytotoxicity
- . Hemolysis
All biocompatibility tests met their respective acceptance criteria.
#### Electrical safety and electromagnetic compatibility (EMC)
The CellFX System complies with IEC 60601-1, IEC 60601-1-2, 60601-2-2. IEC 60601-1-6, IEC 62366, and IEC 62304. All test conditions were performed as outlined within the IEC standards. All electrical safety and EMC tests passed
#### Software Verification and Validation Testing
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern since a failure of the device could directly result in minor injury to the patient or operator. Completed software testing supports the safety and effectiveness of the device.
Cybersecurity controls have been implemented to mitigate the risk of malware being introduced into the CellFX System as recommended by FDA's Draft Guidance for Industry and FDA Staff, "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices."
{6}------------------------------------------------
# Animal Study
A GLP animal safety study was conducted to validate the safety and performance of the CellFX System. The study was conducted on 15 Yucatan pigs that underwent skin treatments with either the CellFX System (test) or the predicate device (control). Animals were divided into 5 cohorts of 3 animals each for the following timepoints: 0, 2, 6, 14, and 30-days.
Safety
- All animals survived to their termination date with no procedure related . complications. Based on the results of intra-procedural treatments, gross evaluation, histopathological analysis, clinical pathology, and in-life physical examinations the CellFX treatments are considered to be safe.
## Performance
- All test articles met the acceptance criteria for device performance. The treatment . zones met the acceptance criteria. The CellFX treatment sites successfully achieved degeneration and necrosis of active treatment sites while the epidermis remained intact. The device is expected to perform as intended.
## Summary
Based on the GLP animal safety study, the CellFX System was found to have a safety and performance profile that is equivalent to the predicate device.
## CONCLUSION
The CellFX® System has the same intended use as the predicate. The differences in technological characteristics between the CellFX System and predicate device do not raise any different questions of safety and effectiveness. The non-clinical performance testing and in vivo animal data provided in this submission demonstrate and support that the CellFX System is as safe and as effective as the predicate device. Therefore, the CellFX System is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.