K202938 · Aesculap, Inc. · GEI · Oct 30, 2020 · General, Plastic Surgery
Device Facts
Record ID
K202938
Device Name
Aesculap Caiman 5 Seal and Cut Technology System
Applicant
Aesculap, Inc.
Product Code
GEI · General, Plastic Surgery
Decision Date
Oct 30, 2020
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Caiman Seal and Cut Technology System instruments are intended for use in general surgery and gynecologic surgical procedures where ligation and division of vessels is desired.
Device Story
Sterile, single-use bipolar electrosurgical instrument; connects to dedicated bipolar RF generator. Device grasps, seals, and mechanically divides tissue/vessels. Operation: bipolar RF energy (coagulation) applied to vascular structures between jaws; integrated cutting blade manually actuated for tissue division. Used in OR by surgeons. Output: physical tissue seal and division. Benefits: efficient vessel ligation and tissue dissection in laparoscopic/open surgery. Modifications include updated jaw materials and articulation features.
Clinical Evidence
Bench testing only. No clinical data. Verification testing included life cycle (100 cycles), burst pressure (vessels up to 7mm), and cutting blade activation force (≤ 31 N). Thermal spread metrics (max ≤ 5mm, average ≤ 1mm) and sealing times (≤ 6.5s standard, ≤ 7.5s plus) met acceptance criteria.
Indicated for general surgery (urologic, vascular, thoracic, thoracoscopic) and gynecologic procedures (e.g., vaginal hysterectomies, Nissen fundoplication, colectomy, adhesiolysis, bowel resection, oophorectomy) requiring vessel ligation, tissue grasping, and dissection. Suitable for vessels up to 7mm and tissue bundles. 36cm/44cm lengths for laparoscopic; 24cm for open procedures. Contraindicated for tubal sterilization or tubal coagulation for sterilization.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
K110824 — ARAGON SURGICAL RF SYSTEM - 5MM LAPAROSCOPIC INSTRUMENT · Aragon Surgical, Inc. · Jul 11, 2011
K031011 — LIGASURE 5MM LAPAROSCOPIC SEALER-DIVIDER, MODEL LS1500 · Valleylab, Inc. · May 29, 2003
K182588 — Medline ReNewal Reprocessed LigaSure Maryland Jaw Sealer/Divider · Surgical Instrument Service and Savings Inc.(Dba Medline Ren · Oct 31, 2018
K172624 — Voyant 5mm Fusion Device · Applied Medical Resources Corporation · Feb 13, 2018
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
October 30, 2020
Aesculap Inc. Omunique (Nikki) Luke Regulatory Affairs Specialist 3773 Corporate Parkway Center Valley, Pennsylvania 18034
Re: K202938
Trade/Device Name: Aesculap Caiman 5 Seal and Cut Technology System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: September 23, 2020 Received: September 30, 2020
Dear Omunique (Nikki) Luke:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K202938
#### Device Name
Aesculap Caiman 5 Seal and Cut Technology System
#### Indications for Use (Describe)
Caiman Seal and Cut Technology System consists of dedicated bipolar electrosurgical instruments intended for use in general surgery and gynecologic surgical procedures where ligation of vessels is desired. The instruments create a seal by the application of bipolar electrosurgical RF energy (coagulation) to vascular structure (vessels) interposed between the jaws of the device. A cutting blade is actuated for the division of tissue.
The Caiman 12 Plus (44cm) and the Caiman 5 (36cm and 44cm) are indicated for laparoscopic procedures and the Caiman 12 Plus (24cm) and the Caiman 5 (24cm) are indicated for open procedures. The indications for use include general surgical procedures, (including urologic, vascular, thoracoscopic), and gynecological procedures where ligation and division of vessels is performed. These procedures include: vaginal hysterectomies, Nissen fundoplication, colectory, adhesiolysis, bowel resection, and oophorectorny etc., or any procedure where vessel ligation (seal and cut), tissue grasping, and dissection is performed. The devices can be used on vessels up to and including 7mm and bundles as large as will fit in the jaws of the instrument.
Caiman Seal and Cut Technology System has not been shown to be effective for tubal coagulation for sterilization procedures. Do not use the system for these procedures.
| Type of Use (Select one or both, as applicable) | <span> <span style="padding-right: 20px;"> <b> [X] Prescription Use (Part 21 CFR 801 Subpart D) </b> </span> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </span> |
|-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Paperwork Reduction Act (PRA) Staff
{3}------------------------------------------------
#### 510(k) SUMMARY (as required by 21 CFR 807.92)
#### Caiman Seal and Cut Technology System October 30, 2020
| COMPANY: | Aesculap, Inc.<br>3773 Corporate Parkway<br>Center Valley, PA 18034<br>Establishment Registration Number: 2916714 |
|----------------------|-------------------------------------------------------------------------------------------------------------------|
| CONTACT: | Nikki Luke<br>484-523-6868 (phone)<br>nikki.luke@aesculapimplants.com<br>610-791-6882 (fax) |
| TRADE NAME: | Aesculap® Caiman Seal and Cut Technology System |
| COMMON NAME: | Electrosurgical, Cutting & Coagulation & Accessories |
| CLASSIFICATION NAME: | General and Plastic Surgery |
| REGULATION NUMBER: | 21 CFR 878.4400 |
| PRODUCT CODE: | GEI |
| DEVICE CLASS: | Class II per 21 CFR 878.4400 |
### PREDICATE DEVICE
Caiman Seal and Cut Technology (K151696)
### DEVICE DESCRIPTION
The Aesculap® Caiman Seal and Cut Technology System consists of sterile, single-use bipolar instruments that connect to a dedicated bipolar RF generator. The Caiman instruments are designed to grasp, seal (ligate), and mechanically divide (cut) varying lengths of tissue (for example mesentery) per device application. This is accomplished by a two electrode (side by side-top and bottom) design. The instruments are capable of vessel sealing, blunt dissection, grasping, and dividing tissue enclosed within its dissection clips. This reason for this submission is to capture modifications to the caiman blunt instruments.
### INDICATIONS FOR USE
Caiman Seal and Cut Technology System consists of dedicated bipolar electrosurgical instruments intended for use in general surgery and gynecologic surgical procedures where ligation and division of vessels is desired. The instruments create a seal by the application of
{4}------------------------------------------------
bipolar electrosurgical RF energy (coagulation) to vascular structure (vessels) interposed between the jaws of the device. A cutting blade is actuated for the division of tissue.
Instruments 36cm and 44cm in length are indicated for laparoscopic procedures and instruments 24 cm in length are indicated for open procedures. The indications for use include general surgical procedures, (including urologic, vascular, thoracic, and thoracoscopic), and gynecological procedures where ligation and division of vessels is performed. These procedures include: vaginal hysterectomies, Nissen fundoplication, colectomy, adhesiolysis, bowel resection, and oophorectomy etc., or any procedure where vessel ligation (seal and cut), tissue grasping, and dissection is performed. The devices can be used on vessels up to and including 7mm and bundles as large as will fit in the jaws of the instrument.
Caiman Seal and Cut Technology System has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the system for these procedures.
# TECHNOLIGICAL CHARACTERISTICS (compared to predicate devices)
The table below provides a summary of the device technological characteristics comparing the subject device and predicate device. The modifications made to these devices do not raise any new issues of safety and effectiveness, as confirmed by the testing and validation activities described in the submission.
{5}------------------------------------------------
| | Subject Device: Caiman Seal and<br>Cut Technology System<br>Product Code: GEI<br>TBD | Predicate Device: Caiman Seal and<br>Cut Technology System<br>Product Code: GEI<br>K151696 | Reference Device: Caiman Seal<br>and Cut Technology System<br>Product Code: GEI | |
|------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|
| Intended Use | Caiman Seal and Cut Technology<br>System instruments are intended for<br>use in general surgery and<br>gynecologic surgical procedures<br>where ligation and division of vessels<br>is desired. | Caiman Seal and Cut Technology<br>consists of dedicated bipolar<br>electrosurgical instruments<br>intended for use in general surgery<br>and gynecologic surgical<br>procedures where ligation and<br>division of vessels is desired. | Caiman Seal and Cut<br>Technology consists of<br>dedicated bipolar<br>electrosurgical instruments<br>intended for use in general<br>surgery and gynecologic<br>surgical procedures where<br>ligation and division of vessels<br>is desired. | Same |
| Indications for Use | Caiman Seal and Cut Technology<br>System consists of dedicated bipolar<br>electrosurgical instruments intended<br>for use in general surgery and<br>gynecologic surgical procedures<br>where ligation and division of vessels<br>is desired. The instruments create a<br>seal by the application of bipolar<br>electrosurgical RF energy<br>(coagulation) to vascular structure<br>(vessels) interposed between the<br>jaws of the device. A cutting blade is<br>actuated for the division of tissue.<br>Instruments 36cm and 44cm in<br>length are indicated for laparoscopic<br>procedures and instruments 24 cm in<br>length are indicated for open<br>procedures. The indications for use<br>include general surgical<br>procedures, (including urologic,<br>vascular, thoracic, and<br>thoracoscopic), and gynecological<br>procedures where ligation and<br>division of vessels is performed.<br>These procedures include: vaginal<br>hysterectomies, Nissen<br>fundoplication, colectomy,<br>adhesiolysis, bowel resection, and<br>oophorectomy etc., or any procedure<br>where vessel ligation (seal and cut),<br>tissue grasping, and dissection is<br>performed. The devices can be used<br>on vessels up to and including 7mm<br>and bundles as large as will fit in the<br>jaws of the instrument. | Caiman Seal and Cut Technology<br>consists of<br>dedicated bipolar electrosurgical<br>instruments intended for use in<br>general surgery and gynecologic<br>surgical procedures where<br>ligation and division of vessels is<br>desired. The instruments create a<br>seal by the application of bipolar<br>electrosurgical RF energy<br>(coagulation) to vascular structure<br>(vessels) interposed between the<br>jaws of the device. A cutting blade is<br>actuated for the division of tissue.<br>Instruments 24 cm in length are<br>indicated for open procedures<br>and instruments 36 cm and 44 cm in<br>length are indicated for laparoscopic<br>procedures. The indications for use<br>include general surgical procedures,<br>(including urologic, vascular,<br>thoracic, and thoracoscopic), and<br>gynecological procedures where<br>ligation and division of vessels is<br>performed. These procedures<br>include: vaginal hysterectomies,<br>Nissen fundoplication, colectomy,<br>adhesiolysis, bowel resection, and<br>oophorectomy etc., or any procedure<br>where vessel ligation (seal and cut),<br>tissue grasping, and dissection is<br>performed. The devices can be used<br>on vessels up to and including 7mm<br>and bundles as large as<br>will fit in the jaws of the instrument.<br>Caiman Seal and Cut Technology | Caiman Seal and Cut Technology<br>consists of dedicated bipolar<br>electrosurgical instruments intended<br>for use in general surgery and<br>gynecologic surgical procedures<br>where ligation and division of vessels<br>is desired. The instruments create a<br>seal by the application of bipolar<br>electrosurgical RF energy<br>(coagulation) to vascular structure<br>(vessels) interposed between the<br>jaws of the device. A cutting blade is<br>actuated for the division of tissue.<br>Instruments 12.5 cm, 17 cm, and 24<br>cm in length are indicated for open<br>procedures<br>and instruments 36 cm and 44 cm in<br>length are indicated for laparoscopic<br>procedures. The indications for use<br>include general surgical procedures,<br>(including urologic, vascular,<br>thoracic, and thoracoscopic), and<br>gynecological procedures where<br>ligation and division of vessels is<br>performed. These procedures<br>include: vaginal hysterectomies,<br>Nissen adhesiolysis, bowel resection,<br>and oophorectomy etc., or any<br>procedure where vessel ligation (seal<br>and cut), tissue grasping, and<br>dissection is performed. The devices<br>can be used on vessels up to and<br>including 7mm and bundles as large<br>as<br>will fit in the jaws of the instrument. | Same |
| | Caiman Seal and Cut<br>Technology System has not<br>been shown to be effective for<br>tubal sterilization or tubal<br>coagulation for sterilization<br>procedures. Do not use the<br>system for these procedures. | has not been shown to be effective<br>for tubal sterilization or tubal<br>coagulation for sterilization<br>procedures. Do not use the system<br>for these procedures. | Caiman Seal and Cut Technology<br>has not been shown to be effective<br>for tubal sterilization or tubal<br>coagulation for sterilization<br>procedures. Do not use the system<br>for these procedures. | |
| Material Composition | Handle - molded thermoplastic<br>Shaft - stainless steel<br>Insulation - polyester tube<br>(Caiman 5)<br>EFTE (Caiman 5 Articulating)<br>Dissection clip - PEEK | Handle - molded thermoplastic<br>Shaft - stainless steel<br>Insulation - polyester tube<br>(Caiman 5)<br>EFTE (Caiman 5 Articulating)<br>Dissection clip - Valox HX 420 | Handle - molded thermoplastic<br>Shaft - stainless steel<br>Insulation - polyester tube<br>(Caiman 5)<br>ETFE (Caiman 5 Articulating)<br>Dissection clip - PEEK | Same |
| Functional Use | Grasping Ligation and Coagulation | Grasping Ligation and Coagulation | Grasping Ligation and Coagulation | Same |
| Surgical Approach | 240mm - Open<br>240mm, 360mm and 440mm -<br>Laparoscopic | 240mm - Open<br>360mm and 440mm - Laparoscopic | 125mm and 170mm - Open<br>360mm and 440mm - Laparoscopic | Same |
| Length | 240mm<br>360mm<br>440mm | 240mm<br>360mm<br>440mm | 125mm<br>170mm<br>360mm<br>440mm | Same |
| Jaw | Upper Jaw<br>Stainless-Steel Material: 17-4 PH<br>(EN ISO 1.4542)<br>Isolation Features: None<br>Pivot Jaw<br>Stainless-Steel Material: 17-4 PH<br>(EN ISO 1.4542)<br>Isolation Features: Standing ceramic<br>features<br>Jaw Shape: Straight | Upper Jaw<br>Stainless-Steel Material: 1.4021<br>Isolation Features: Peek Stripes<br>Pivot Jaw<br>Stainless-Steel Material: 17-4 PH<br>(EN ISO 1.4542)<br>Isolation Features: None<br>Jaw Shape: Straight | Jaw Shape: Curved | Same |
| Diameter | 5mm | 5mm | 5mm | Same |
| | | | | Same |
| Articulation | 240, 360, and 440mm - non-<br>articulating<br>240, 360, and 440mm - articulating | 240, 360, and 440mm - non-<br>articulating<br>360mm - articulating | 125, 170, 360, and 440mm -<br>non-articulating<br>360 and 440 - articulating | Same |
| Number of Electrodes (Pairs) | 1 | 1 | 1 | Same |
| Electrode<br>Length | 2.65 cm | 2.65 cm | 21.5 mm | Same |
| Electrode<br>Width | 0.13-0.15 cm | 0.13-0.15 cm | 1.01- 1.63 mm - varies over the<br>length of the jaw due to curve | Same |
| Electrode Texture | Smooth with nonconductive (Alumina<br>Titania) stop members on the Jaws,<br>to maintain consistent gap between<br>electrode surfaces. | Smooth, with PEEK stop members at<br>the distal end of the jaw and Valox<br>stop members at the proximal end of<br>the jaw, to maintain consistent gap<br>between electrode surface. | Smooth with conductive (steel) and<br>non conductive (Alumina Titania)<br>stop members on the Jaws, to<br>maintain consistent gap between<br>electrode surfaces. | Same |
| Cable | 10ft | 10ft…
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.