BiPad Hand Activated, Disposable Bipolar Electrocautery Cords
K201833 · Bipad Surgical, Inc. · GEI · Jul 31, 2020 · General, Plastic Surgery
Device Facts
Record ID
K201833
Device Name
BiPad Hand Activated, Disposable Bipolar Electrocautery Cords
Applicant
Bipad Surgical, Inc.
Product Code
GEI · General, Plastic Surgery
Decision Date
Jul 31, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The BiPAD® Hand Activated, Disposable Bipolar Electrocautery Cords are intended to connect an electrosurgical device to an electrosurgical generator. They are indicated for use with bipolar forceps during general surgical procedures.
Device Story
Sterile, disposable bipolar cord connecting electrosurgical generator to bipolar forceps; enables hand-activated coagulation. Features integrated hand switch (actuator assembly) near forceps socket; removable to allow traditional foot-pedal-only operation. Model BP105-M uses 3-prong connector for Medtronic-style generators; Model BP105-C uses Codman-style connector with Y-connector accessory (BP105-CY) to parallel hand switch with existing foot pedal. Used in surgical settings by clinicians. Benefits include hand-controlled activation of RF current for hemostasis, reducing reliance on foot pedals. Device is single-use; EO sterilized.
Clinical Evidence
No clinical data; substantial equivalence established via bench testing, including biocompatibility, electrical safety (AAMI/ANSI ES 60601-1, IEC 60601-2-2), EMC (IEC 60601-1-2), usability (IEC 60601-1-6), sterilization validation (ISO 11135), and packaging performance (ISO 11607, ASTM D4169).
Technological Characteristics
Disposable bipolar cord with integrated hand switch. Materials evaluated per ISO 10993-1. Connectivity: standard US-spaced pins, Medtronic 3-prong, or Codman-style connectors. Energy: RF current from external electrosurgical generator. Sterilization: Ethylene Oxide (EO) per ISO 11135. Standards: AAMI/ANSI ES 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-2-2.
Indications for Use
Indicated for use with bipolar forceps during general surgical procedures to control bleeding. Contraindicated for laparoscopic procedures.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of the square.
July 31, 2020
BiPad Surgical, Inc. % Rafael Aguila Responsible Third-Part Official Accelerated Device Approval Services, LLC 6800 S.W. 40th Street, Ste. 403 Ludlum, Florida 33155
Re: K201833
Trade/Device Name: BiPad Hand Activated. Disposable Bipolar Electrocautery Cords Model BP105-C (Codman) and Model BP105-M (Medtronic) Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: July 1, 2020 Received: July 2, 2020
Dear Rafael Aguila:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Long Chen. Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### K201833
# Section 4 - Indications for Use Statement (FDA 3881)
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K201833
Device Name
BIPAD Hand Activated. Disposable Bipolar Electrocautery Cords. Model BP105-C (Codman) and Model BP105-M (Medtronic)
Indications for Use (Describe)
BIPAD Hand Activated, Disposable Bipolar Electrocautery Cords are intended to connect an electrosurgical device to and electrosurgical generator. They are indicated for use with bipolar forceps during general surgical procedures.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## K201833
## Section 5 - 510(k) Summary
In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CFR 807.92, this information serves as a Summary of Safety and Effectiveness for the BIPAD® Hand Activated, Disposable Electrocautery Cords, Model BP105-C (Codman) and Model BP105-M (Medtronic).
| Submitted by: | BiPAD Surgical, Inc. |
|------------------------------------|-------------------------------------------------------------------------------------------------------------------|
| | 110 Ocean Blvd. |
| | Point Lookout, NY 11569 |
| Date: | April 30, 2020 |
| Contact Person: | Louis Cornacchia, MD |
| | Chief Executive Officer |
| | lcornacchia@bipadsurgical.com |
| Proprietary Name: | BIPAD® Hand Activated, Disposable Bipolar Electrocautery<br>Cords, Model BP105-C (Codman) and BP105-M (Medtronic) |
| Common Name: | Disposable Bipolar Electrocautery Cord |
| Classification Name and Reference: | 21 CFR 878.4400, Electrosurgical cutting and coagulation device<br>and accessories, Device Class II |
| Device Product Code, Device Panel: | GEI, General & Plastic Surgery |
| Predicate Device: | Kepler MedTec Disposable Bipolar Cable (K110462) |
#### 5.1 Device Description
The BiPAD® Hand Activated, Disposable Bipolar Electrocautery Cords, Model BP105-C (Codman) and BP105-M (Medtronic) have been designed as an accessory to electrosurgical instruments where bipolar electrosurgical coagulation is desired during surgery. The device is a sterile, disposable, bipolar cord that connects to the electrosurgical generator on one end and the active instrument (bipolar forceps) on the other end. The cord has an integrated hand switch to activate the coagulating current of the RF generator. The device is intended for use with currently marketed Electrosurgical generators and bipolar surgical forceps to control bleeding.
The BiPAD® Hand Activated, Disposable Bipolar Electrocautery Cords are compatible with FDA cleared and marketed Codman and Medtronic Electrosurgical Generators and FDA cleared bipolar forceps using standard US connection spaced pins. We have provided minimum specifications for use with the subject device with other electrosurgical generators using Medtronic style connectors with a fixed pin spacing and/or Codman generators with the identified bipolar connection pin spacing.
## Device Components
BiPAD® Hand Activated, Disposable Bipolar Electrocautery Cords consist of:
- 1) A connector specific to the intended electrosurgical generator (Codman type or Medtronic type)
- 2) A cable capable of carrying the high frequency coagulation current from the electrosurgical generator to the surgical forceps
- 3) BiPad Actuator Assembly which consists of the integrated hand switch, an actuator arm, and a molded housing.
The actuator assembly is designed to be removable from the housing to allow the surgeon to use the cord as a traditional disposable electrocautery cord (no hand switch) if desired. When used in this configuration, a foot switch (foot pedal) is required to control the generator output.
{5}------------------------------------------------
### BiPAD Surgical BIPAD® Hand Activated, Disposable Bipolar Electrocavery Cords, Model BP105-C (Codman) and Model BP105-M (Medronic) Traditional 510(k)
The BiPAD® Hand Activated, Disposable Bipolar Electrocautery Cords will be available in two configurations that are essentially identical except for the proximal connection to the intended electrosurgical generator. The Model BP105-M (Medtronic) is compatible with any electrosurgical generator that accepts the standard 3-prong, molded bipolar plug with fixed pin spacing such as Medtronic generators.
The BP105-C (Codman) is compatible with the Codman Malis generators. Both models of the subject device have the same cord, cord length and integrated hand switch.
The Codman version, Model BP105-C, requires an additional Y-connector accessory (BP105-CY). The Y-connector (BP105-CY) allows the hand switch on the Codman version of the cord to share control of the flow of energy from the generator to the forceps with the existing foot pedal. It is intended that the Y-connector will be attached to the Codman Malis generator using the supplied adhesive and remain with the generator. This accessory is provided non-sterile and is intended to be reusable.
The BiPAD® Hand Activated, Disposable Bipolar Electrocautery Cords are meant to be used in a surgical setting. These are single use devices that should not be reprocessed or reused. The cords will be EO sterilized and have a shelf life of 12 months.
#### 5.2 Intended Use / Indication for Use
The BiPAD® Hand Activated, Disposable Bipolar Electrocautery Cords are intended to connect an electrosurgical device to an electrosurgical generator. They are indicated for use with bipolar forceps during general surgical procedures.
#### 5.3 Contraindications
The BiPAD® Hand Activated, Disposable Bipolar Electrocautery Cords should not be used for laparoscopic procedures.
#### Technology and Fundamental Scientific Technology 5.4
BiPAD® Hand Activated, Disposable Bipolar Electrocautery Cords work together with FDA cleared and marketed bipolar electrosurgical generators and FDA cleared and marketed 2-prong standard bipolar surgical forceps.
BiPAD® Hand Activated, Disposable Bipolar Electrocautery Cords have an integrated hand switch in the cord near the forceps socket, which allows activation of the electrocautery current from the attached generator.
The Model BP105-M (Medtronic) cord has a molded three prong connector which allows the hand switch to share the foot pedal circuit to activate the generator.
The Model BP105-C (Codman) cord connects the hand switch in parallel with the foot pedal switch using the BP105-CY Y-connector accessory as a splitter. This allows both the hand switch and the foot pedal switch to close the pedal circuit to activate the generator's electrocautery current.
#### 5.5 Determination of Substantial Equivalence
BiPad Surgical, Inc. claims that the BiPAD® Hand Activated, Disposable Bipolar Electrocautery Cords are substantially equivalent to the predicate device, (Kepler MedTec Disposable Bipolar Cable, K110462). BiPAD Surgical claims this equivalence because the proposed device has an equivalent intended use, manufacturing materials, operating principles and specifications, target population, clinical setting and environments, sterilization method, physical and operational specifications, and electrical safety characteristics. The comparison of these specifications demonstrates the functional equivalence of the products
{6}------------------------------------------------
BiPAD Surgical believes that the differences that exist in working mechanism and structure between the BiPAD® Hand Activated, Disposable Bipolar Electrocautery Cords and the predicate (i.e. finger switch and cable length) are addressed in the Intertek Electrical Safety reports and have been addressed in the Risk Management File for the device. The differences do not raise new issues of safety and effectiveness. Verification and Usability testing demonstrated that no adverse effects have been introduced by these differences.
BiPad Surgical, Inc. believes that the BiPAD® Hand Activated, Disposable Bipolar Electrocautery Cords are as safe and effective and perform in a substantially equivalent manner to the predicate device.
#### 5.6 Summary of Non-Clinical Tests
The BiPAD® Hand Activated, Disposable Bipolar Electrocautery Cords, Model BP105-C (Codman) and BP105-M (Medtronic) have been evaluated for biocompatibility, sterilization, usability, as well as electrical, electromagnetic, and mechanical safety and has been found to conform with applicable medical device safety standards. The BiPAD® Hand Activated, Disposable Bipolar Electrocautery Cords comply with voluntary standards and FDA Guidance documents. The following performance data has been provided in support of the substantial equivalence determination. All testing was performed on final devices.
#### 5.6.1 Bench Testing
Performance bench testing was conducted in alignment with FDA's quidance Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery issued (August 15, 2016) and internal requirements.
#### 5.6.2 Electrical Safety and EMC
Electrical safety and EMC testing were conducted in accordance with:
- AAMI/ANSI ES 60601-1 (2012), Medical Electrical Equipment -Medical electrical equipment-Part1: General requirements for basic safety and essential performance
- . IEC 60601-1-2 (2014), Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Disturbances -Requirements And Tests
- . IEC 60601-1-6 (2013), Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
- IEC 60601-2-2 (2017), Medical Electrical Equipment -Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories.
#### 5.6.3 Biocompatibility
The subject device passed all required biocompatibility tests per:
- ISO 10993-1 (2018) Biological Evaluation Of Medical Devices Part 1: Evaluation And Testing ● Within A Risk Management Process
- . ISO10993-5 (2009) Biological Evaluation Of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity
- ISO10993-10 (2010). Biological Evaluation Of Medical Devices Part 10: Tests For Irritation ● And Skin Sensitization
- Guidance for Industry and FDA Staff – Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process" (issued June 16, 2016)
#### 5.6.4 Sterilization
The BIPAD® Hand Activated, Disposable Bipolar Electrocautery Cord is provided sterile using a validated ethylene oxide (EO) sterilization cycle. The sterilization cycle has been validated to ensure a sterility assurance level (SAL) of 106 in accordance with ISO 11135:2014, Sterilization of health care products - Ethylene oxide: Requirements for the development, validation, and routine control of a
{7}------------------------------------------------
sterilization process for medical devices and ISO 11138-1 Sterilization of health care products, Biological indicators – Part 1: General requirements.
#### Packaging & Distribution 5.6.5
Packaging Performance testing was successfully completed per the standards listed below prior to, and immediately following, accelerated age conditioning to support the proposed shelf life. Real time aging is in process.
- ISO 11607-1:2019 Packaging for Terminally Sterilized Medical Devices, Part 1
- ISO 11607-2:2019 Packaging for Terminally Sterilized Medical Devices, Part 2 ●
- ASTM D4169-16 (2016) Standard Practice for Performance Testing of Shipping Containers ●
#### 5.6.6 Other
- . ISO 15223-1 (2016), Medical devices - Symbols to be used with medical device labels, labelling and information to be supplied -Part 1: General requirements
- ISO14971 (2007), Medical device - Application of risk management to medical devices
- . Guidance for Industry and Staff, Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery (issued March 24, 2014)
- Guidance for Industry and Staff, Applying Human Factors and Usability Engineering to ● Medical Devices (issued February 3, 2016)
- . FDA Guidance, Design Control Guidance for Medical Device Manufacturers (issued March 11, 1997)
- FDA Guidance for Industry, Pyrogen and Endotoxins Testing: Questions and Answers (issued June 2012)
- Guidance for Industry and Staff, Submission and Review of Sterility Information in Premarket Notification (510(k)) Submissions for Devices Labeled as Sterile (issued January 21, 2016)
- Guidance for Industry and Staff, Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling (issued March 17, 2015)
#### 5.7 Summary of Clinical Tests
The proposed BIPAD® Hand Activated, Disposable Bipolar Electrocautery Cord did not require formal clinical study since substantial equivalence to the legally marketed predicate device was proven with verification/validation testing.
#### 5.8 Conclusion
It was determined through comparison with the predicate device (Kepler MedTec Disposable Bipolar Cable - K110462) and device testing that the BIPAD® Hand Activated, Disposable Bipolar Electrocautery Cord is safe and effective and does not raise any new questions of safety and effectiveness and is therefore considered substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.