K193601 · Micro-Tech (Nanjing) Co., Ltd. · GEI · Jul 24, 2020 · General, Plastic Surgery
Device Facts
Record ID
K193601
Device Name
Single Use Electrosurgical Knife
Applicant
Micro-Tech (Nanjing) Co., Ltd.
Product Code
GEI · General, Plastic Surgery
Decision Date
Jul 24, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
These instruments have been designed to be used with endoscopes and electrosurgical for dissection, elevation, irrigation and preparation of tissue layers in combination with monopolar cutting and coagulation within the digestive tract.
Device Story
Single-use endoscopic electrosurgical knife; used with endoscope (working channel ≥2.8mm) and electrosurgical unit (ESU). Device consists of handle, tube assembly, and retractable electrode. Physician inserts device through endoscope; extends electrode; activates via ESU footswitch. Models with injection function allow submucosal fluid delivery via syringe or pump. Provides monopolar high-frequency current for tissue cutting and coagulation. Used in digestive tract procedures. Benefits include tissue dissection, elevation, and hemostasis. Sterile, EO-sterilized, single-use.
Clinical Evidence
No clinical data or animal studies included. Substantial equivalence supported by bench testing, including dimension, actuation, sealing, flow, mucosa lift, cutting, and connection force performance. Biocompatibility per ISO 10993-1; electrical safety/EMC per IEC 60601 series; shelf-life per ASTM F1980-16; sterilization per ISO 11135:2014.
Technological Characteristics
Materials: ABS, Stainless Steel (SUS304), PTFE, Zirconia, TPU, Gold. Energy: Monopolar RF. Dimensions: 1.5mm-4mm knife length; 1950mm-2350mm working length. Sterilization: EO (SAL 10^-6). Standards: ISO 10993-1, IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2, IEC 60601-2-18, ASTM F1980-16, ISO 11135:2014.
Indications for Use
Indicated for use with endoscopes and electrosurgical units for dissection, elevation, irrigation, and preparation of tissue layers in combination with monopolar cutting and coagulation within the digestive tract.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 24, 2020
Micro-Tech (Nanjing) Co., Ltd. Sally He, RA Engineer No.10 Gaoke Third Road Nanjing National Hi-Tech, Industrial Development Zone Nanjing, Jiangsu Province 210032 China
Re: K193601
Trade/Device Name: Single Use Electrosurgical Knife Regulation Number: 21 CFR 876.4300 Regulation Name: Endoscopic Electrosurgical Unit and Accessories Regulatory Class: Class II Product Code: KNS Dated: December 19, 2019 Received: December 23, 2019
Dear Sally He:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
#### K193601
Device Name Single Use Electrosurgical Knife
Indications for Use (Describe)
These instruments have been designed to be used with endoscopes and electrosurgical for dissection, elevation, irrigation and preparation of tissue layers in combination with monopolar cutting and coagulation within the digestive tract.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows a blue logo with the letters 'MT' in a stylized, blocky font. The letters are interconnected and appear to be in a three-dimensional perspective. To the upper right of the letters, there is a small 'R' in a circle, indicating a registered trademark symbol.
## 510K Summary
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92. The assigned 510(k) Number: K193601
# 1. Date of Preparation: 2020-05-28
- 2. Sponsor Identification
# Micro-Tech (Nanjing) Co., Ltd.
No.10 Gaoke Third Road, Nanjing National Hi-Tech, Industrial Development Zone,
Nanjing, Jiangsu Province, PRC
## Establishment Registration Number: 3004837686
Contact Person: Sally He
Position: RA Engineer
Tel: +86-25-58646395
Fax: +86-25-58350006
Email: RA.Micro-Tech@outlook.com
# 3. Identification of Proposed Device
Trade Name: Single Use Electrosurgical Knife
Common Name: Electrosurgical Knife
# Regulatory Information
Classification Name: Endoscopic electrosurgical unit and accessories
Classification: 2
Product Code: KNS
Regulation Number: 876.4300
Review Panel: Gastroenterology/Urology
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows a blue logo with the letters "MT" in a stylized, three-dimensional font. The letters are bold and appear to be interconnected, with the "M" on the left and the "T" on the right. A small "R" in a circle, indicating a registered trademark, is positioned in the upper right corner of the logo.
### 4. Identification of Predicate Device
Predicate Device 510(k) Number: K171158 Product Name: Single Use Electrosurgical Knife Manufacturer: Olympus Medical Systems Corporation
Reference device 510(k) Number: K083608 Product Name: ERBE Hybrid Knife™ Manufacturer: ERBE USA, Inc
## 5. Indications for Use
These instruments have been designed to be used with endoscopes and electrosurgical units for dissection, elevation, irrigation and preparation of tissue layers in combination with monopolar cutting and coagulation within the digestive tract.
### 6. Device Description
The proposed device Single Use Electrosurgical Knife is a sterile, single-use endoscopic device, intended to be used with endoscopes and electrosurgical units for dissection, elevation, irrigation and preparation of tissue layers in combination with monopolar cutting and coagulation within the digestive tract.
The Single Use Electrosurgical Knife is used with the Electrosurgical Unit (ESU), the ESU supplies high frequency (HF) energy through a retractable electrode of the Single Use Electrosurgical Knife for the cutting and coagulation of tissue. The Single Use Electrosurgical Knife consists of Cutting Knife, outer tube assembly, Middle Tube assembly and handle assembly. Physicians would attach it to the ESU. For endoscopic procedures, the Single Use Electrosurgical Knife is placed down the channel of an endoscope that has a working channel equal to or greater than 2.8mm.
2/7
{5}------------------------------------------------
Image /page/5/Picture/2 description: The image shows a blue logo with the letters "MT" in a stylized, three-dimensional font. The letters are interconnected and appear to be made of solid blocks. A small registered trademark symbol is located in the upper right corner of the logo.
Upon the setup of the ESU, the Single Use Electrosurgical Knife is ready for use. To activate cautery, the Cutting Knife is extended out and the ESU's footswitch is depressed. The subject devices with injection models can supply fluid into the submucosa to supply liquid, use a syringe or pump to connect with luer tap on the hand components.
The Single Use Electrosurgical Knife is divided into I type, T type, O type and IT type according to the shape of the Cutting Knife. The main materials include ABS, Stainless Steel, PTFE, Zirconia, TPU and Gold. The physical and chemical performances of the device are stable. The proposed devices are EO sterilized to achieve the Sterility Assurance Level (SAL) of 106 and placed in a sterility maintenance package to ensure a shelf life of 3 years.
The Single Use Electrosurgical Knife has twenty-two (22) specifications; the main differences of these specifications are Cutting Knife Shape, Cutting Knife Length, and Effective Working Length and with injection function or without injection function.
## 7. Comparison of Technological Characteristics
The Single Use Electrosurgical Knife incorporates substantially equivalent device materials, design, configuration, packaging fundamental technology, sterilization process and intended use as those featured in the predicate device Single Use
| Electrosurgical Knife (K171158): | | | |
|----------------------------------|-----------------|--|--|
| | Proposed Device | | |
| ltem | Proposed Device<br>Single Use Electrosurgical<br>Knife | Predicate Device<br>Single Use<br>Electrosurgical Knife<br>(K171158) | Remark |
|--------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------|
| Product Code | KNS | KNS | Same |
| Requlation No. | 878.4300 | 878.4300 | Same |
| Class | 2 | 2 | Same |
| Supplied in Sterile | Yes | Yes | Same |
| Main Material | PTFE, ABS, SUS304 | Polymer materials,<br>Stainless Steel | Similar |
| Configuration | Cutting Knife, Tube, and | Distal end, Tube, and | Same |
| Item | Proposed Device<br>Single Use Electrosurgical<br>Knife | Predicate Device<br>Single Use<br>Electrosurgical Knife<br>(K171158) | Remark |
| Handle | Handle | Handle | |
| Injection | With/Without Injection | With Injection | Similar |
| Cutting Knife<br>Shape | I,T,O,IT | T | Similar |
| Cutting Knife<br>length | 1.5mm,2mm,4mm | 2mm,1.5mm | Similar |
| Working Length | 1950mm,2350mm | 1650mm,1950mm,2300mm | Similar |
| Compatible<br>endoscopy<br>working channel | ≥2.8mm | ≥2.8mm | Same |
| Energy used/<br>Delivered | Monopolar Radio Frequency<br>Current | Monopolar Radio<br>Frequency Current | Same |
| Rated<br>High-Frequency<br>Voltage | CUT/COAG:2400Vp-p | CUT: 3200Vp-p<br>COAG: 5800Vp-p | Similar |
| Indications for Use | These instruments have<br>been designed to be used<br>with endoscopes and<br>electrosurgical for dissection,<br>elevation, irrigation and<br>preparation of tissue layers<br>in combination with<br>monopolar cutting and<br>coagulation within the<br>digestive tract. | These instruments have<br>been designed to be used<br>with Olympus endoscopes,<br>electrolsurgical units to cut<br>tissue and coagulate or to<br>perform hemostasis using<br>high-frequency current and<br>flushing devices for<br>submucosal injection in the<br>digestive tract. | Substantial<br>same |
| Single Use | Yes | Yes | Same |
| Packaging | Single-use EO sterilized<br>pouch with one device per<br>pouch | Single-use EO sterilized<br>pouch with one device per<br>pouch | Same |
| Shelf Life | Three years | Three years | Same |
| Biocompatibility | Conform to ISO 10993-1 | Conform to ISO 10993-1 | Same |
| Item | Proposed Device<br>Single Use Electrosurgical<br>Knife | Predicate Device<br>Single Use<br>Electrosurgical Knife<br>(K171158) | Remark |
| Sterilization | EO Sterilized, SAL: $10^{-6}$ | EO Sterilized, SAL: $10^{-6}$ | Same |
| Labeling | Conform to 21 CFR part 801 | Conform to 21 CFR part 801 | Same |
| Electrical Safety<br>and<br>Electromagnetic<br>compatibility | Conform to:<br>IEC60601-1<br>IEC 60601-1-2<br>IEC 60601-2-2<br>IEC 60601-2-18 | Conform to:<br>IEC60601-1<br>IEC 60601-1-2<br>IEC 60601-2-2<br>IEC 60601-2-18 | Same |
{6}------------------------------------------------
Image /page/6/Picture/2 description: The image shows a blue logo with the letters "MT" in a stylized, three-dimensional design. The letters are bold and appear to be interconnected, with the "M" slightly overlapping the "T". A registered trademark symbol is visible in the upper right corner of the logo.
{7}------------------------------------------------
Image /page/7/Picture/2 description: The image shows a blue logo with the letters "MT" in a stylized, three-dimensional design. The letters are connected and appear to be angled slightly, giving the logo a sense of depth. A registered trademark symbol is present in the upper right corner of the logo.
## 8. Performance Data
Performance testing was conducted to demonstrate the essential performance of the proposed device and confirmed that the proposed device works as intended with the compatible devices. Additionally, the results of the tests below were evaluated as substantially equivalent to the predicate device.
- > Dimension
- レ Actuation Performance
- > Compatible Performance
- > Sealing Performance
- > Flowing Performance
- > Mucosa Lift Performance
- > Cutting Performance
- ♪ Connected Force Performance
The bench testings performed demonstrated that the proposed device and predicate device are substantially equivalent.
Shelf-life testing was conducted based on an accelerated aging test in accordance with ASTM F1980-16 Standard Guide for Accelerated Aging of Sterile Barrier Systems
{8}------------------------------------------------
Image /page/8/Picture/2 description: The image shows a blue logo with the letters "MT" in a stylized, three-dimensional font. The letters are interconnected, with the "M" appearing to be in front of the "T". A small registered trademark symbol is located in the upper right corner of the logo. The logo has a clean and modern design.
for Medical Devices. Three-year aging test will be performed to demonstrate longer stability and support the results of the accelerated aging test. Sterilization validation was carried out in accordance with ISO 11135:2014 "Sterilization of Health Care products - Ethylene Oxide - Part 1: Requirements for Development, Validation, and Routine Control of Sterilization processes for Medical Devices".
Biocompatibility testing was performed in accordance with the FDA Guidance, Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing within a risk management process" issued on June 16,2016.
Electromagnetic compatibility, electric safety, and thermal safety had been confirmed according to the following standards:
IEC60601-1:2005+A1:2012 Medical Electrical Equipment - Part 1: Medical electrical equipment - general requirements for the basic safety and essential performance IEC 60601-2-2:2017 Medical electrical equipment - Part 2-2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories
IEC 60601-2-18:2009 Medical electrical equipment - Part 2-18: Particular
requirements for the safety of endoscopic equipment
IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
#### 9. Animal Study
No animal study is included in this submission.
#### Clinical Study 10.
No clinical study is included in this submission.
{9}------------------------------------------------
Image /page/9/Picture/2 description: The image shows a blue logo with the letters "MT" in a stylized, three-dimensional font. The letters are interconnected, with the "M" appearing to be in front of the "T". A small registered trademark symbol is located in the upper right corner of the logo. The logo has a clean and modern design.
#### Substantially Equivalent (SE) Conclusion 11.
Based on the indications for use, technological characteristics, and safety and performance testing, the Single Use Electrosurgical Knife has been shown to be appropriate for its intended use and is considered to be substantially equivalent to the currently cleared predicate device Single Use Electrosurgical Knife (K171158).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.