K193004 · New Deantronics Taiwan , Ltd. · GEI · Dec 19, 2019 · General, Plastic Surgery
Device Facts
Record ID
K193004
Device Name
Monopolar Cord
Applicant
New Deantronics Taiwan , Ltd.
Product Code
GEI · General, Plastic Surgery
Decision Date
Dec 19, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
A Monopolar Cord is intended to be used with a compatible electrosurgical generator and an RF electrode to cut tissue and control bleeding by use of high-frequency electrical current.
Device Story
Monopolar Cord acts as electrical conduit between electrosurgical generator and RF electrode; facilitates delivery of high-frequency electrical current for tissue cutting and coagulation. Used in hospital settings by trained professionals. Device connects to generator via adapter and to monopolar instruments via shrouded 2.5 mm female electrical connections. Activation controlled by foot switch. Output is high-frequency energy delivered to target tissue. Benefits include reliable energy transmission for surgical hemostasis and tissue dissection.
Clinical Evidence
Bench testing only. No clinical data. Testing included electrical safety (HF leakage, dielectric strength, hi-pot, mains frequency), mechanical integrity (strain relief, connector pull force, anchorage), and transit testing per ASTM D4169. All tests passed.
Technological Characteristics
Materials: PE insulation, copper wire, brass generator connector, nickel-plated brass electrode connector. Energy: Radiofrequency electrical current. Form factor: 10 ft cable with shrouded 2.5 mm female connector. Connectivity: Wired connection to ESU via adapter. Sterilization: Gamma (VDmax25). Standards: IEC 60601-1:2005+AM1:2012, IEC 60601-1-2:2014, IEC 60601-2-2:2017.
Indications for Use
Indicated for use with compatible electrosurgical generators and RF electrodes to cut tissue and control bleeding in surgical procedures. For use by trained professionals in hospital settings.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 19, 2019
New Deantronics Taiwan Ltd % Mr. Craig J. Coombs President Coombs Medical Device Consulting, Inc 1100 Pacific Marina, Suite 806 Alameda. California 94501
Re: K193004
Trade/Device Name: Monopolar Cord Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: October 24, 2019 Received: October 28, 2019
Dear Mr. Craig Coombs:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803) for devices or post marketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K193004
Device Name Monopolar Cord
#### Indications for Use (Describe)
A Monopolar Cord is intended to be used with a compatible electrosurgical generator and an RF electrode to cut tissue and control bleeding by use of high-frequency electrical current.
Type of Use (Select one or both, as applicable)
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|--------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for New Deantronics. The word "NEW" is in blue, and the word "DEANTRONICS" is in a larger, bolded blue font. There is a blue circle with a red dot in the middle between the two words.
# Section 5: 510(k) Summary
### A. Device Information:
| Category | Comments |
|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor: | New Deantronics Taiwan Ltd.<br>12F., No.51, Sec. 4, Chong Yang Rd.,<br>Tu Cheng District,<br>New Taipei City 23675, Taiwan R.O.C.<br>Tel: (886) 2-2268-1726<br>Fax: (886) 2-2268-3800<br>Sponsor Contact: Ms. Jane Liu, President<br>Email: jane@newdean.com.tw |
| Correspondent Contact<br>Information: | Mr. Craig Coombs<br>President<br>Coombs Medical Device Consulting<br>1100 Pacific Marina, Suite 806<br>Alameda, CA 94501<br>Tel: 510-995-8499<br>Email: CraigJCoombs@gmail.com |
| Device Common Name: | Electrosurgical accessory<br>Monopolar Cord |
| Device Classification Number: | 21 CFR 878.4400 |
| Device Classification &<br>Product Code: | Class 2,<br>GEI |
| Device Proprietary Name: | Monopolar Cord |
### Predicate Device Information:
| Predicate Device: | New Deantronics Foot Control Pencil |
|----------------------------------------------------|-------------------------------------------------------------------------------------|
| Predicate Device Manufacturer: | New Deantronics Taiwan Ltd. |
| Predicate Device Common Name: | Electrosurgical accessory |
| Predicate Device Premarket Notification # | K982749 |
| Predicate Device Classification: | 21 CFR 878.4400<br>Electrosurgical, Cutting & Coagulation<br>Device and Accessories |
| Predicate Device Classification &<br>Product Code: | Class 2,<br>GEI |
## B. Date Summary Prepared
13 December 2019
## C. Description of Device
The Monopolar Cord, sold as sterile packaged, and ready for use devices, is intended to be used with compatible accessories to cut tissue and control bleeding by use of high-frequency electrical current.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for New Deantronics. The word "NEW" is in blue, and the word "DEANTRONICS" is in a darker blue. There is a blue circle with a red dot in the middle between the two words. The logo is simple and modern.
The Monopolar Cord is compatible with the monopolar instruments with shrouded 2.5 mm female electrical connections, electrosurgical generator, and patient return electrode.
During the operation, the sterile cord should connect to the sterile accessory, and then generator connector end is inserted into the adapter and generator which it receives the high frequency current and delivers the current onto a target tissue for cutting and coagulation in a procedure. These devices can be used in hospitals and are used by trained professionally only.
## D. Indications for Use
A Monopolar Cord is intended to be used with a compatible electrosurgical generator and an RF electrode to cut tissue and control bleeding by use of high-frequency electrical current.
## E. Comparison to Predicate Device
As described below, the application Monopolar Cord is substantially equivalent in intended use, technology, design and physician use to the predicate New Deantronics Foot Control Pencil (K982749).
| Feature | Application Device:<br>Monopolar Cord<br>(K193004) | Predicate Device:<br>New Deantronics Foot Control<br>Pencil (K982749) | Pertinence of Feature<br>to Consideration of<br>Substantial<br>Equivalence. |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | A Monopolar Cord is intended<br>to be used with a compatible<br>electrosurgical generator and an<br>RF electrode to cut tissue and<br>control bleeding by use of high-<br>frequency electrical current. | Electrosurgical accessory. A<br>foot activated electrosurgical<br>pencil is intended to remove<br>tissue and control bleeding by<br>use of high-frequency electrical<br>current. | The Indications for Use<br>of the application and<br>predicate devices are<br>the same; both devices<br>are for use in standard<br>cutting or coagulation.<br>The application device<br>does not include an<br>attached electrode, so<br>the specific action of<br>the electrode is<br>omitted. |
| Product Code | GEI | GEI | Identical |
| Technology | | | |
| Mechanism of<br>Standard<br>Electrosurgery | The Monopolar Cord is<br>designed to connect a<br>electrosurgical generator with<br>an RF electrode. When<br>electrosurgery is activated by<br>foot pedal, the electrosurgical<br>energy will be delivery from<br>electrosurgical generator<br>through the Cord and to the<br>electrode, to cut or coagulate in<br>surgical procedures | The Foot Control Pencil is<br>designed to connect with<br>electrosurgical generator. When<br>electrosurgery is activated by<br>foot pedal, the electrosurgical<br>energy will be delivery from<br>electrosurgical generator<br>through the cord and to the<br>electrode, to cut or coagulate in<br>surgical procedures | Identical as far as the<br>connecting cord<br>portion is concerned |
| Energy Used | Radiofrequency Electrical<br>Current | Radiofrequency Electrical<br>Current | Identical |
| Operation<br>Principle | Monopolar Electrosurgery | Monopolar Electrosurgery | Identical |
| Equipment<br>Mated | Electrosurgical Generator:<br>Force FX or Force Triad | Electrosurgical Generator:<br>Force FX or Force Triad | Identical |
| Feature | Application Device:<br>Monopolar Cord<br>(K193004) | Predicate Device:<br>New Deantronics Foot Control<br>Pencil (K982749) | Pertinence of Feature<br>to Consideration of<br>Substantial<br>Equivalence. |
| Required<br>Adapter | New Deantronics AG001<br>adapter or Covidien E0502<br>Series or E0017 adapter. | New Deantronics AG001<br>adapter or Covidien E0502<br>Series or E0017 adapter. | Identical |
| Design -Mechanism | | | |
| Control Type | Foot Control | Foot Control | Identical |
| Activation<br>Mechanism | Foot Switch | Foot Switch | Identical |
| Cable Set Length | 10 ft | 10 ft | Identical |
| | ESU: banana lead with adapter | ESU: banana lead with adapter | Identical |
| Connection<br>Interface | Active Electrode: None.<br>Designed to connect with<br>monopolar instruments with<br>shrouded 2.5 mm female<br>electrical connections. | Active Electrode: Standard<br>Blade | The application device<br>is a subset of the<br>predicate device. |
| Design- Electrical Feature | | | |
| Electrical Safety | Withstand 3.3kVp for <b>High</b><br><b>Frequency</b> breakdown<br>Withstand 3.3kVp AC at <b>Main</b><br><b>Frequency (60Hz)</b> breakdown | Withstand 3.3kVp for <b>High</b><br><b>Frequency</b> breakdown<br>Withstand 3.3kVp AC at <b>Main</b><br><b>Frequency (60Hz)</b> breakdown | Identical |
| Design- Material | | | |
| Handle | N/A | ABS | Different |
| | Insulation: PE | Insulation: PE | |
| Cable | Colorant: Blue, PMS300 | Colorant: Blue, PMS300 | Identical |
| | Wire: Copper | Wire: Copper | |
| Generator<br>Connector (pin<br>tip) | ESU: Brass | ESU: Brass | Identical |
| Active Electrode<br>Connector | LDPE / Nickel plated brass | N/A | Different |
| Other Attributes | | | |
| Single Use or<br>Reusable | Single use | Single use | Identical |
| Sterilization | Gamma (VDmax25) | Gamma (VDmax25) | Identical |
| Performance/<br>Electrical Safety<br>Testing in | IEC 60601-1: 2005 +<br>AM1:2012<br>IEC 60601-1-2: 2014 | AAMI HF-18:1993 | Application device in<br>conformance with<br>latest version of the<br>listed Standards |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for New Deantronics. The word "NEW" is in blue, and next to it is a blue circle with a red dot in the middle. The word "DEANTRONICS" is in blue and is to the right of the circle.
New Deantronics concludes that the devices are substantially equivalent.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for New Deantronics. The word "NEW" is in a blue outlined font. To the right of the word "NEW" is a blue circle with a red dot in the center. To the right of the circle is the word "DEANTRONICS" in a bold blue font.
### F. Summary of Supporting Data
A summary of the bench testing and electrical safety testing is presented on the next page.
Electrical safety testing and bench testing has demonstrated that the performance of the Monopolar Cord meets the requirements of its pre-defined acceptance criteria and intended uses. The results of the non-clinical testing demonstrate that the Monopolar Cord is as safe and effective as the predicate device.
### Summary Tables of Supporting Data
Performance Data for FDA Guidance for Reviewer of Electrosurgical Device for General Surgery, issued on August 15, 2016 Method of Scope (major Verification: Specification Name or Pass / Item component or Standard (list), Comments Description or Test Purpose Fail? system) Testing Method or Audit Provide graphical display of the This submission is not A ESU --output waveform an ESU. This submission is an Active Accessory and Connector pull Pass has demonstrated the force risks associated with mechanical failure and short circuit is 0.26kg Dynamic Pass Perform mechanical testing of minimized by Strain Relief electrosurgical instruments to mechanical testing Active Component / B minimize the risks associated before and after transit. Accessorv with mechanical failure and The mechanical testing 10 Pound Static short circuit includes anchorage test Pass Strain Relief and connector pull force. Also, this submission covers the devices intended to be Anchorage Test Pass single used rather than re-used. This submission does C Neutral Electrodes Perform the following testing not include a Neutral ---Electrode The Monopolar Cord is Miscellaneous an Active Component Meet the design and D Components / ーー instead of performance specifications Accessories Miscellaneous Component
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for New Deantronics. The word "NEW" is in blue, followed by a blue circle with a red dot in the center. The word "DEANTRONICS" is in blue and to the right of the circle. The logo is simple and professional.
Monopolar Cord K193004: Traditional Premarket Notification
| Item | Scope (major<br>component or<br>system) | Specification Name or<br>Description or Test Purpose | Method of<br>Verification:<br>Standard (list),<br>Testing Method<br>or Audit | Pass /<br>Fail? | Comments |
|------|-----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------|-----------------|--------------------------------------------------------------------------------------------------------------------------------------|
| E | | | | | |
| E.1 | Thermal Effects on<br>Tissue | Active component or electrode<br>should demonstrate the<br>measurement of the size of the<br>thermal damage zone | -- | - | This submission does not<br>include an Active<br>Electrode. This accessory<br>is not intended to create<br>thermal damage by itself. |
| E.2 | Temperature<br>Monitoring | The temperature sensing under<br>simulated conditions meets<br>design specification and<br>performance requirements | -- | - | This submission does not<br>intend to include any<br>temperature sensing<br>capability. |
| E.3 | CQM and/or<br>Continuity<br>Monitoring | Test to demonstrate the CQM<br>and/or continuity monitoring<br>features work as intended. | -- | - | This is the requirement for<br>ESU and Neutral<br>Electrodes. This submission<br>does not include both of<br>these components. |
| E.4 | Capacitive Coupling | For minimally invasive surgery<br>application, test for active<br>coupling resistance between the<br>subject device and a conductive<br>cannula/trocar device under<br>simulated normal use condition. | -- | - | This device is not for<br>minimally invasive surgery<br>application. |
The following table summarizes the electrical safety tests performed and test results. Some listed here are repeated from the Guidelines table above. The results demonstrated that the Monopolar Cord is safe for its intended use and supports a finding of substantial equivalence with the predicate device.
| Standard Used for<br>Testing Protocol | Specific Clause in<br>Standard used for<br>Testing Protocol | Test Description | Results |
|---------------------------------------|-------------------------------------------------------------|---------------------------------------------|---------|
| (General Mfr Requirement) | | Visual Inspection | PASS |
| (General Mfr Requirement) | | Continuity test | PASS |
| IEC 60601-1:2005+<br>AM1:2012 | 15.3.4.1 | Drop Test (preconditioning) | PASS |
| IEC 60601-1:2005+<br>AM1:2012 | 8.10.2 | Cable pull force from Terminal Pin | PASS |
| | | 0.26kg Dynamic Strain Relief | PASS |
| | | 10 Pound Static Strain Relief | PASS |
| IEC 60601-2-2:2017 | 201.8.10.4.2 | Anchorage Test | PASS |
| IEC 60601-2-2:2017 | 201.8.8.3.102 | HF Leakage Current Test | PASS |
| IEC 60601-2-2:2017 | 201.8.8.3.103 | High Frequency Dielectric Test | PASS |
| IEC 60601-2-2:2017 | 201.8.8.3.104 | Hi-Pot Test | PASS |
| | | Mains Frequency Dielectric<br>Strength Test | PASS |
| IEC 60601-2-2:2017 | 201.15.4.1 | Connector pull force | PASS |
| IEC 60601-2-2:2017 | 201.17<br>202 | EMC Test | PASS |
# Monopolar Cord Electrical Testing
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for New Deantronics. The word "NEW" is in blue outlined letters. To the right of the word "NEW" is a blue letter "D" with a red circle in the middle. The word "DEANTRONICS" is in blue and to the right of the letter "D".
The transit test was to define the product ship testing requirements and package and product performance after ship testing. The Package system performance testing per ASTM D4169 met the acceptance criteria. The results demonstrate that the Monopolar Cord is functional and remains in the sterile barrier after shipping transportation.
## G. Conclusion
After comparing the Indications for Use, technology and design of the Monopolar Cord, along with all electrical safety (including IEC 60601-1: 2005 + AM1:2012; IEC 60601-1-2: 2014; IEC 60601-2-2: 2017) and performance testing, in accordance with the FDA's guidelines and FDArecognized consensus standards for electrical safety, New Deantronics concludes that the Monopolar Cord is substantially equivalent to the predicate New Deantronics Foot Control Pencil (K982749).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.