K191732 · Sutter Medizintechnik GmbH · GEI · Aug 30, 2019 · General, Plastic Surgery
Device Facts
Record ID
K191732
Device Name
Sutter RaVoR Bipolar Electrodes
Applicant
Sutter Medizintechnik GmbH
Product Code
GEI · General, Plastic Surgery
Decision Date
Aug 30, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Sutter RaVoR™ Bipolar Electrodes are to be connected to the Sutter CURIS® RF Generator. Sutter Ra VoR™ Bipolar Electrodes are indicated for ablation of soft tissue in otorhinolaryngology (ENT) surgery. The Ra VoR™ Bipolar Electrode for the soft palate is indicated for submucosal tissue coagulation in the uvula/soft palate for the treatment of snoring. The RaVoR™ Bipolar Electrode for the turbinates is indicated for submucosal tissue shrinkage for reduction of hypertrophic nasal turbinates in patients with nasal turbinate airway obstruction.
Device Story
Sutter RaVoR™ Bipolar Electrodes are electrosurgical instruments designed for connection to the Sutter CURIS® RF Generator. The device delivers radiofrequency energy to soft tissue in ENT procedures; specifically for submucosal shrinkage or coagulation. The electrodes feature various angled sheath configurations and are single-use, sterile devices. Operation is performed by qualified medical personnel in a clinical setting. The device functions by applying bipolar RF energy to target tissues, resulting in controlled thermal effects (lesions) for tissue reduction. The healthcare provider uses the device to treat snoring or nasal airway obstruction; the resulting tissue shrinkage/coagulation provides therapeutic benefit by reducing tissue volume. The device is substantially equivalent to the predicate Celon ENT System.
Clinical Evidence
No clinical or animal studies were performed. Substantial equivalence is supported by bench testing, including electromagnetic compatibility, electrical safety (IEC 60601-1, 60601-2-2, 60601-1-2), mechanical strength, thermal damage/spread experiments, and biocompatibility (ISO 10993-1).
Technological Characteristics
Materials: Stainless steel, Fluoropolymer, Polyester. Energy: Bipolar RF. Design: Parallel electrode configuration, various angled sheaths. Connectivity: Connects to Sutter CURIS® RF Generator. Sterilization: Ethylene Oxide (SAL 10^-6). Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2, ISO 11137-1/2, ISO 10993-1.
Indications for Use
Indicated for ENT soft tissue ablation/coagulation. Specifically: submucosal tissue coagulation in uvula/soft palate for snoring treatment; submucosal tissue shrinkage for reduction of hypertrophic nasal turbinates in patients with nasal turbinate airway obstruction.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below.
August 30, 2019
Sutter Medizintechnik GmbH Ulrike Zeissler Manager Regulatory Affairs Tullastrasse 87 79108 Freiburg, Germany
Re: K191732
Trade/Device Name: Sutter RaVoR Bipolar Electrodes Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: June 25, 2019 Received: June 28, 2019
Dear Ulrike Zeissler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Srinivas Nandkumar, Ph.D. Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K191732
Device Name Sutter RaVoR™ Bipolar Electrodes
#### Indications for Use (Describe)
Sutter RaVoR™ Bipolar Electrodes are to be connected to the Sutter CURIS® RF Generator. Sutter Ra VoR™ Bipolar Electrodes are indicated for ablation of soft tissue in otorhinolaryngology (ENT) surgery.
The Ra VoR™ Bipolar Electrode for the soft palate is indicated for submucosal tissue coagulation in the uvula/soft palate for the treatment of snoring.
The RaVoR™ Bipolar Electrode for the turbinates is indicated for submucosal tissue shrinkage for reduction of hypertrophic nasal turbinates in patients with nasal turbinate airway obstruction.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
Section 807.92
Section 807.92
| Date: | August 29, 2019 | |
|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Name:<br>Address: | SUTTER MEDIZINTECHNIK GmbH<br>Tullastrasse 87<br>79108 Freiburg<br>Germany |
| | Contact person:<br>Titel:<br>Telephone:<br>Fax: | Ulrike Zeissler<br>Manager Regulatory Affairs<br>+49 (0) 761 51551-14<br>+49 (0) 761 51551-30 |
| Product: | Trade Name:<br>Common Name:<br>Classification<br>name:<br>Product Code: | Sutter RaVoR™ Bipolar Electrodes<br>Bipolar Electrodes<br>Electrosurgical, cutting & coagulation &<br>accessories<br>GEI |
| | Regulation<br>Number:<br>Classification:<br>Classification<br>Panel: | CFR 21 § 878.4400<br>Class II<br>General and Plastic Surgery |
| Predicate<br>Device: | Predicate device to which Sutter RaVoR™ Bipolar Electrodes are<br>claimed to be substantially equivalent is manufactured by Celon AG medical instruments, Celon ENT System (K032838) | |
| Device<br>Description: | Sutter RaVoR™ Bipolar Electrodes are electrosurgical instruments<br>available in different angled sheath configurations. They are to be<br>connected to the bipolar output of the Sutter CURIS® RF Generator. The<br>electrodes are provided sterile and are single-use devices. | |
| Indications<br>for Use: | Sutter RaVoR™ Bipolar Electrodes are to be connected to the bipolar<br>output of the Sutter CURIS® RF Generator. Sutter RaVoR™ Bipolar<br>Electrodes are indicated for ablation and coagulation of soft tissue in<br>otorhinolaryngology (ENT) surgery.<br>The RaVoR™ Bipolar Electrode for the soft palate is indicated for<br>submucosal tissue shrinkage and tissue coagulation in the uvula/soft<br>palate for the treatment of snoring.<br>The RaVoR™ Bipolar Electrode for the turbinates is indicated for<br>submucosal tissue shrinkage for reduction of hypertrophic nasal<br>turbinates in patients with nasal turbinate airway obstruction. | |
{4}------------------------------------------------
### APPLICANT: SUTTER MEDIZINTECHNIK GMBH DEVICES: SUTTER RaVoR™ BIPOLAR ELECTRODES
| Techno-<br>logical<br>Characteris-<br>tics: | The table below provides a comparison between the subject device and<br>the predicate device | | | | |
|---------------------------------------------|----------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Sutter RaVOR™<br>Bipolar Electrodes | Predicate device<br>K032838 | Discussion | | |
| | Indication for<br>Use | Sutter RaVOR™<br>Bipolar Electrodes are<br>to be connected to the<br>bipolar output of the<br>Sutter CURIS® RF<br>Generator. Sutter<br>RaVoR™ Bipolar<br>Electrodes are<br>indicated for ablation<br>and coagulation of soft<br>tissue in<br>otorhinolaryngology<br>(ENT) surgery.<br><br>The RaVoR™ Bipolar<br>Electrode for the soft<br>palate is indicated for<br>submucosal tissue<br>shrinkage and tissue<br>coagulation in the<br>uvula/soft palate for<br>the treatment of<br>snoring.<br><br>The RaVoR™ Bipolar<br>Electrode for the<br>turbinates is indicated<br>for submucosal tissue<br>shrinkage for reduction<br>of hypertrophic nasal<br>turbinates in patients<br>with nasal turbinate<br>airway obstruction. | This application<br>system, comprising the<br>CelonLab ENT power<br>control unit and the<br>CelonProBreafh,<br>CelonProSleep, and<br>CelonProSleep PLUS<br>bipolar coagulation<br>electrodes, is indicated<br>for ablation and<br>coagulation of soft<br>tissue in<br>otorhinolaryngology<br>(ENT) surgery<br>including:<br>CelonProBreath:<br>Submucosal tissue<br>shrinkage for nasal<br>airway obstruction by<br>reduction of<br>hypertrophic nasal<br>turbinates.<br><br>CelonProSleep:<br>Submucosal tissue<br>shrinkage and tissue<br>coagulation in the<br>uvula/soft palate for<br>the treatment of<br>snoring.<br><br>CelonPro Sleep PLUS:<br>Submucosal tissue<br>shrinkage and tissue<br>coagulation in the<br>uvula/soft palate for<br>the treatment of<br>snoring.<br><br>The system is intended<br>for use by qualified<br>medical personnel<br>trained in the use of<br>electrosurgical<br>equipment. | Same - for the<br>respective electrode<br>designs. | |
| | Maximum<br>peak voltage | 600 Vp | 200 Vp | Similar- the difference<br>does not affect safety<br>and effectiveness. | |
| | Electrode<br>Design | parallel | coaxial | Similar- the differences<br>do not affect safety<br>and effectiveness.<br>Thermal damage and<br>thermal spread<br>performance tests<br>show equivalence of<br>obtained lesions. | |
| | | | | | |
| | Dimensions<br>Tip size [mm]: | $Ø$ 0.8 | $Ø$ 1.1-1.3 | | do not affect safety<br>and effectiveness<br>Similar- the differences<br>do not affect safety<br>and effectiveness.<br>Thermal damage and<br>thermal spread<br>performance tests<br>show equivalence of<br>obtained lesions. |
| | Material | Stainless steel,<br>Fluoropolymer,<br>Polyester | Stainless steel,<br>Plastic material | | Similar- the difference<br>does not affect safety<br>and effectiveness. |
| | Meets IEC 60601-1 | yes | yes | | Same |
| | Meets IEC 60601-1-2 | yes | yes | | Same |
| | Meets IEC 60601-2-2 | yes | yes | | Same |
| | Sterility Assurance Level | 10-6 | 10-6 | | Same |
| | Sterilization Method | Ethylene Oxide | Gamma irradiation | | Similar- the differences<br>do not affect safety<br>and effectiveness.<br>Different commonly<br>used sterilization<br>methods, both lead to<br>the same value of<br>sterility assurance<br>level. |
| Non-Clinical<br>Performance Data: | | The following performance data has been obtained for the substantial equivalence determination.<br><br><b>Bench Testing</b><br>Performance testing has been executed in line with the internal R&D process and in compliance with the proposals and recommendations of the FDA guidance: "Premarket Notification [510(k)] Submissions for Electrosurgical Devices for General Surgery" - Guidance for Industry and Food and Drug Administration Staff, August 15, 2016.<br><br>In particular, tests have been carried out with respect to following subject areas: | | | |
| | | Performance Data Bench Tests | | | |
| | | Test | Conclusion | | |
| | | Electromagnetic Compatibility and<br>Electrical Safety | Pass | | |
| | | Mechanical strength | Pass | | |
| | | System Performance /Bench-top<br>validation testing | Pass | | |
| | | Thermal effects on tissue | Pass | | |
| | | Transport and Storage | Pass | | |
{5}------------------------------------------------
## APPLICANT: SUTTER MEDIZINTECHNIK GMBH
DEVICES: SUTTER RaVoR™ BIPOLAR ELECTRODES
{6}------------------------------------------------
| <b>Electromagnetic Compatibility and Electrical Safety Testing</b><br>Electrical and electromagnetic tests have been performed to<br>demonstrate that design specifications and performance requirements<br>are met. Compliance to the voluntary standards IEC 60601-1<br>(AAMI/ANSI ES60601-1:2005), IEC 60601-2-2 and IEC 60601-1-2 has<br>been shown. | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <b>Sterilization</b><br>Sutter RaVoR™ Bipolar Electrodes are sterilized by using a validated<br>ethylene oxide cycle. The sterilization cycle has been validated to<br>ensure a sterility level of (SAL) 10-6 in accordance with ISO 11137-1<br>and ISO 11137-2. | |
| <b>Shelf Life Testing</b><br>Shelf-life testing has been conducted in accordance to of FDA<br>guidance: "Premarket Notification [510(k)] Submissions for<br>Electrosurgical Devices for General Surgery" – Guidance for Industry<br>and Food and Drug Administration Staff, August 15, 2016. Sutter<br>RaVoR™ Bipolar Electrodes have been subjected to accelerated aging.<br>The aging studies established that the device and packaging remain<br>functional and maintain sterility for up to 3 years. | |
| <b>Biocompatibility Testing</b><br>Biological evaluation and Biocompatibility testing has been performed in<br>compliance to ISO 10993-1 Fourth edition 2009-10-15, “Biological<br>evaluation of medical devices - Part 1: Evaluation and testing within a<br>risk management process”. | |
| <b>Thermal damage / thermal spread testing</b><br>Thermal damage / thermal spread experiments were performed for the<br>subject and predicate device. It was possible to examine and determine<br>the tissue cool-down time from RF activation to baseline temperature.<br>Furthermore, results of these experiments showed that subject and<br>predicate device generate comparable thermal effects in tissue with<br>regards to lesion size and cool-down time. | |
| <b>Animal Studies</b><br>No animal studies have been performed as appropriate verification and<br>validation of the subject device has been achieved based on<br>comparison to the predicate device and from results of the bench<br>testing, biocompatibility evaluation, and electrical/safety testing. | |
| <b>Clinical Studies</b><br>No clinical studies have been performed as appropriate verification and<br>validation of the subject device has been achieved based on<br>comparison to the predicate device and from results of the bench<br>testing, biocompatibility evaluation, and electrical/safety testing. | |
| <b>Conclusion:</b> | In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR<br>Part 807 and based on the similarity to the predicate device in terms of<br>technology, performance, and indications for use, Sutter Medizintechnik<br>GmbH concludes that the subject device, Sutter RaVoR™ Bipolar<br>Electrodes, does not raise any issues of safety or effectiveness and are<br>substantially equivalent to the predicate device as described above. |
| | |
{7}------------------------------------------------
## APPLICANT: SUTTER MEDIZINTECHNIK GMBH
DEVICES: SUTTER RaVoR™ BIPOLAR ELECTRODES
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.