← Product Code [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI) · K182355

# Secret RF Smartcure Applicator with electrodes (K182355)

_Ilooda Co,., Ltd. · GEI · Sep 6, 2019 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K182355

## Device Facts

- **Applicant:** Ilooda Co,., Ltd.
- **Product Code:** [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI.md)
- **Decision Date:** Sep 6, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

Secret RF Smartcure Applicator is intended for use in dermatologic and general surgical procedures for electro-coagulation and hemostasis

## Device Story

Secret RF Smartcure Applicator is an accessory for the Secret RF system; enables monopolar RF energy delivery for electro-coagulation and hemostasis. System comprises main unit, handpiece, and disposable monopolar micro-needle electrodes. Physician operates handpiece, placing electrodes in light contact with epidermis at right angles to target tissue. RF energy passes through skin, generating electro-thermal reaction to coagulate tissue. Device used in clinical settings. Output allows precise thermal coagulation of dermal tissue. Benefits include controlled hemostasis and coagulation in dermatologic/surgical procedures.

## Clinical Evidence

Bench testing included biocompatibility (ISO 10993-1), electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), and high-frequency surgical equipment standards (IEC 60601-2-2). In vivo animal study (micropig model) evaluated histological depth and zone of ablation/thermal damage at low, mid, and high power intensities at immediate, 7-day, and 14-day post-treatment intervals.

## Technological Characteristics

High-frequency (RF) electrosurgical accessory. Operates at 2MHz or 4MHz (±10%). Monopolar micro-needle electrodes constructed of stainless steel 304 with Teflon PTFE insulating coating. Single-use, EO gas sterilized. Max power 18W-45W depending on mode. Connects to Secret RF main unit via handpiece.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- SecretRF ([K170325](/device/K170325.md))

## Reference Devices

- AGNES ([K160469](/device/K160469.md))
- EVRF ([K112334](/device/K112334.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services - USA. To the right of the symbol, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".

September 6, 2019

ILOODA Co., Ltd. % Dave Kim Medical Device Regulatory Affairs Mtech Group 8310 Buffalo Speedway Houston, Texas 77025

Re: K182355

Trade/Device Name: Secret RF Smartcure Applicator Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI, OUH Dated: July 29, 2019 Received: August 8, 2019

Dear Dave Kim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K182355

Device Name Secret RF Smartcure Applicator

Indications for Use (Describe)

Secret RF Smartcure applicator is intended for use in dermatologic and general surgical procedures for electro-coagulation and hemostasis

Type of Use (Select one or both, as applicable)

| <span style="font-size:100%;"><b> <div style="display:inline-block; vertical-align:middle; margin: 0px; padding: 0px;"> <svg fill="none" height="16px" viewbox="0 0 16 16" width="16px" xmlns="http://www.w3.org/2000/svg"> <path d="M13.3333 2.66663H2.66663C1.9307 2.66663 1.3333 3.26396 1.3333 3.99989V12C1.3333 12.7359 1.9307 13.3333 2.66663 13.3333H13.3333C14.0693 13.3333 14.6666 12.7359 14.6666 12V3.99989C14.6666 3.26396 14.0693 2.66663 13.3333 2.66663Z" stroke="#000000" stroke-linecap="round" stroke-linejoin="round" stroke-width="1.5"></path> <path d="M5.33337 8L7.33337 10L10.6667 6" stroke="#000000" stroke-linecap="round" stroke-linejoin="round" stroke-width="1.5"></path> </svg> </div> </b></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:100%;"><b> <div style="display:inline-block; vertical-align:middle; margin: 0px; padding: 0px;"> <svg fill="none" height="16px" viewbox="0 0 16 16" width="16px" xmlns="http://www.w3.org/2000/svg"> <path d="M13.3333 2.66663H2.66663C1.9307 2.66663 1.3333 3.26396 1.3333 3.99989V12C1.3333 12.7359 1.9307 13.3333 2.66663 13.3333H13.3333C14.0693 13.3333 14.6666 12.7359 14.6666 12V3.99989C14.6666 3.26396 14.0693 2.66663 13.3333 2.66663Z" stroke="#000000" stroke-linecap="round" stroke-linejoin="round" stroke-width="1.5"></path> </svg> </div> </b></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# K182355

# 510(K) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.

# Date 510k summary prepared: 7/29/2019

#### I. SUBMITTER

| Submitter's Name :      | ILOODA CO LTD.                             |
|-------------------------|--------------------------------------------|
| Submitter's HQ Address: | 120, Jangan-ro 458 Beon-gil, Jangan-Gu,    |
|                         | Suwon-Si Gyeonggido, KOREA, 16200          |
| Submitter's Telephone:  | +82-31- 210-1622                           |
| Contact person:         | Yun-Jung HA (yjha@ilooda.com) / RD Manager |
| Official Correspondent: | Dave Kim (davekim@mtech-inc.net)           |
| Address:                | 8310 Buffalo Speedway, Houston, TX 77025   |
| Telephone:              | +713-467-2607                              |
| Fax:                    | +713-583-8988                              |

#### II. DEVICE NAME

| Trade/proprietary name: | Secret RF Smartcure Applicator                              |
|-------------------------|-------------------------------------------------------------|
| Common or Usual Name:   | Micro-needle Fractional RF                                  |
| Regulation Name:        | Electrosurgical, cutting & coagulation device & accessories |
| Regulation Number:      | 21 CFR 878.4400 (Product Code: GEI, OUH)                    |
| Regulatory Class:       | Class II                                                    |
| Prescription Use.       |                                                             |

#### III. PREDICATE DEVICE

| Trade/proprietary name: | SecretRF                                                    |
|-------------------------|-------------------------------------------------------------|
| Common or Usual Name:   | Micro-needle Fractional RF                                  |
| 510K number:            | K170325                                                     |
| Regulation Name:        | Electrosurgical, cutting & coagulation device & accessories |
| Regulation Number:      | 21 CFR 878.4400 (Product Code: GEI, OUH)                    |

{4}------------------------------------------------

### K182355

| Regulatory Class: | Class II |
|-------------------|----------|
| Prescription Use. |          |

#### IV. REFERENCE DEVICE 1

| Trade/proprietary name: | AGNES                                                                     |
|-------------------------|---------------------------------------------------------------------------|
| Common or Usual Name:   | RF Electrosurgical device                                                 |
| 510K number:            | K160469                                                                   |
| Regulation Name:        | Electrosurgical, cutting & coagulation device & accessories               |
| Regulation Number:      | 21 CFR 878.4400 (Product Code: GEI)<br>21 CFR 878.5350 (Produce Code:KCW) |
| Regulatory Class:       | Class II                                                                  |

#### v. REFERENCE DEVICE 2

| Trade/proprietary name: | EVRF                                                        |
|-------------------------|-------------------------------------------------------------|
| 510K number:            | K112334                                                     |
| Common or Usual Name:   | Electrosurgical, cutting & coagulation device & accessories |
| Regulation Number:      | 21 CFR 878.4400 (Product Code: ONQ)                         |
| Regulatory Class:       | Class II                                                    |

#### VI. DEVICE DESCRIPTION

Secret RF's High Frequency(=Radio Frequency) includes the system main body, a handpices with singleuse micro-needle type electrodes, footswitch and an LCD touch screen control panel.

The HF energy is delivered to the target tissue using a handpiece and disposable tip(micro needle electrode tip), the tip being placed in light contact with the epidermis and the handpiece being held at right angles to the target tissue. As the HF energy passes through the skin, it generates an electro thermal reaction, which is capable of coagulating the tissue. Using the micro needle tip, the Secret RF system creates heat within the target dermal tissue via micro-needles inserted from the tip.

The Secret RF is consists of ;

- 1) Secret RF main unit (FDA cleared K170325)
- 2) Bipolar type handpieces with Bipolar type micro-needle electrodes (FDA cleared K170325)
- 3) Secret RF Smartcure applicator

The Secret RF Smartcure Applicator consists of :

- Smartcure handpiece
- Monopolar type micro-needle electrodes

(MTR-AC-01 , MTR-AC-04, MTR-AC-27G, K3i)

{5}------------------------------------------------

#### VII. INDICATIONS FOR USE

Secret RF Smartcure Applicator is intended for use in dermatologic and general surgical procedures for electro-coagulation and hemostasis

### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE VIII. DEVICE

Secret RF's Smartcure applicator changes the treatment(operation) mode depending on the electrode to be connected.

Since the specifications for each operation mode are different, we provide an equivalent comparison table for each mode.

|                                 | Proposed device                                                                                                                                    | Predicate Device                                                                                                              |                        |
|---------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|------------------------|
| Model name                      | SecretRF<br>Smartcure Applicator (Model:<br>MTR-AC-01, MTR-AC-04, MTR-<br>AC-27G)                                                                  | SecretRF                                                                                                                      |                        |
| Manufacturer                    | ILOODA CO.,LTD                                                                                                                                     | ILOODA CO.,LTD                                                                                                                |                        |
| 510(k)number                    | K182355                                                                                                                                            | K170325                                                                                                                       |                        |
| Intended use                    | SecretRF Smartcure Applicator is<br>intended for use in dermatologic<br>and general surgical procedures for<br>electro-coagulation and hemostasis. | SecretRF is intended for use in<br>dermatologic and general surgical<br>procedures for electro-coagulation<br>and hemostasis. | Same                   |
| Main unit                       | SecretRF                                                                                                                                           | SecretRF                                                                                                                      |                        |
| Connected handpiece             | Monopolar                                                                                                                                          | Bipolar handpiece                                                                                                             | Reference<br>device #1 |
| Output energy type              | High frequency                                                                                                                                     | High frequency                                                                                                                | Same                   |
| Electrode type                  | Monopolar micro needle                                                                                                                             | Bipolar micro needle                                                                                                          | Reference<br>device #2 |
| Frequency                       | 2MHz ± 10%                                                                                                                                         | 2MHz ± 10%                                                                                                                    | Same                   |
| Max power                       | Max 45W at 500Ω                                                                                                                                    | Max 25W at 500Ω                                                                                                               | Different              |
| RF duration<br>(ON/OFF RF TIME) | Continuous,<br>100 ms ~ 3000 ms                                                                                                                    | 50 ms ~ 950 ms                                                                                                                | Different              |
| Treatment time                  | 10~15min<br>(recommended)                                                                                                                          | 10~15min<br>(recommended)                                                                                                     | Same                   |

# Table 1. Technical Comparison for the SC mode

|              | Proposed device                                                             | Reference device 1                  | Remark |
|--------------|-----------------------------------------------------------------------------|-------------------------------------|--------|
| Model name   | Secret RF<br>Smartcure Applicator (Model: MTR-AC-01 , MTR-AC-04, MTR-AC-27G | AGENS                               |        |
| Manufacturer | ILOODA CO.,LTD                                                              | Gowoonsesang<br>Cosmetics Co., Ltd. |        |
| 510(k)number | K182355                                                                     | K160469                             |        |

{6}------------------------------------------------

|                                 | Proposed device                                                                                                                                       | Reference device 1                                                                                                           | Remark                                                     |      |
|---------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------|------|
| Intended use                    | Secret RF SmartcureApplicator is<br>intended for use in dermatologic<br>and general surgical procedures<br>for electro-coagulation and<br>hemostasis. | AGNES is indicated for use in<br>dermatological and general surgical<br>procedures for electrocoagulation<br>and hemostasis. | Same                                                       |      |
| Main unit                       | Secret RF                                                                                                                                             | AGENS                                                                                                                        | Same                                                       |      |
| Connected handpiece             | Monopolar handpiece                                                                                                                                   | Monopolar handpiece                                                                                                          | Same                                                       |      |
| Output energy type              | High frequency                                                                                                                                        | High frequency                                                                                                               | Same                                                       |      |
| Electrode type                  | Monopolar micro needle type                                                                                                                           | Monopolar micro needle type                                                                                                  | Same                                                       |      |
| Frequency                       | 2MHz ± 10%                                                                                                                                            | 1MHz                                                                                                                         |                                                            |      |
| Max power                       | Max 45W at 500Ω                                                                                                                                       | Max 46 W at 200 Ω                                                                                                            |                                                            |      |
| RF duration<br>(ON/OFF RF TIME) | Continuous,<br>100 ms ~ 3000 ms                                                                                                                       | 50ms ~ 2000ms                                                                                                                |                                                            |      |
| Treatment time                  | 10~15min<br>(recommended)                                                                                                                             | Unknown                                                                                                                      |                                                            |      |
| Connected<br>electrodes         | Needle<br>Length                                                                                                                                      | 0.8/1.25/1.5/ 2.0 mm                                                                                                         | 0.8/1.25/1.5/ 2.0 mm                                       | Same |
|                                 | Thickness                                                                                                                                             | 0.2mm                                                                                                                        | 0.2mm                                                      | Same |
|                                 | Materials                                                                                                                                             | Stainless steel 304,<br>Insulating Coating:<br>Teflon PTFE                                                                   | Stainless steel 304,<br>Insulating Coating:<br>Teflon PTFE | Same |
| Single Use                      | Single Use                                                                                                                                            | Single Use                                                                                                                   | Same                                                       |      |
| Sterilization                   | EO gas                                                                                                                                                | EO gas                                                                                                                       | Same                                                       |      |

# Table 2. Technical Comparison for the VC mode

|                                 | Proposed device                                                                                                                                        | Reference device 2                                                                                                                                | Remark |
|---------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| Model name                      | Secret RF<br>Smartcure Applicator (Model K3i)                                                                                                          | EVRF                                                                                                                                              |        |
| Manufacturer                    | ILOODA CO.,LTD                                                                                                                                         | F Care System NV                                                                                                                                  |        |
| 510(k)number                    | K182355                                                                                                                                                | K130283                                                                                                                                           |        |
| Intended use                    | Secret RF Smartcure Applicator is<br>intended for use in dermatologic<br>and<br>general surgical procedures for<br>electro-coagulation and hemostasis. | The EVRF System is intended<br>for the epilation and for the<br>treatment of lower limb spider<br>vein or telangiectasia by<br>thermocoagulation. | Same   |
| Output energy<br>type           | High frequency                                                                                                                                         | High frequency                                                                                                                                    | Same   |
| Main unit                       | Secret RF                                                                                                                                              | EVRF                                                                                                                                              | Same   |
| Electrode type                  | Monopolar                                                                                                                                              | Monopolar                                                                                                                                         | Same   |
| Frequency                       | 4MHz ± 10%                                                                                                                                             | 4MHz ± 10%                                                                                                                                        | Same   |
| Max power                       | Max 18W at 500Ω                                                                                                                                        | Max 25W at 500Ω                                                                                                                                   | Same   |
| RF duration<br>(ON/OFF RF TIME) | 10ms~900ms                                                                                                                                             | 50 ms ~ 950 ms                                                                                                                                    | Same   |
| Treatment time                  | 10~15min                                                                                                                                               | 10~15min                                                                                                                                          | Same   |

{7}------------------------------------------------

|                      | Proposed device       | Reference device 2                | Remark |
|----------------------|-----------------------|-----------------------------------|--------|
|                      | (recommended)         | (recommended)                     |        |
| Connected electrodes | Cleared FDA           | Needles are purchased from Ballet |        |
|                      | (Recommended K883892) | Technologies, Ltd (K883892)       |        |

#### IX. PERFORMANCE DATA

The following performance data was provided in support of the substantial equivalence determination.

### Biocompatibility testing:

The patient contact components and materials are tested and validated according to ISO10993-1;2009.

### Non Clinical testing:

IEC 60601-1 Test for Medical Electrical Equipment was performed for General Requirements for basic safety and essential performance. The requirements of specified standards were fulfilled.

IEC 60601-1-2 Test for Medical Electrical Equipment was performed for General Requirements for basic safety and essential performance (collateral standards: electromagnetic compatibility.

IEC 60601-2-2: 2009 Medical electrical equipment Part 2: Particular requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories. The requirements of specified standards were fulfilled.

### Animal testing :

In vivo animal testing using micropig models was also conducted to obtain histological data of values for depth and zone of ablation and thermal damage immediately post treatment: 7 days post treatment; and 14 days post treatment.

The treatment was performed at the intensity(power) low, mid, high.

#### X. CONCLUSIONS

Secret 87 Smartcure applicator with electrode device has the same intended use and similar indications as its predicate devices. The technology of the predicate devices is also the same.

The envelope of power and frequency of the submitted Secret 88 Smartcure applicator with electrodes are covered by the envelopes of its predicate devices.

Any minor differences in the human interface and accessories design do not raise any new types of safety and effectiveness issues, as verified by performance testing.

{8}------------------------------------------------

# K182355

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification Ilooda Co, Ltd. concludes that SECRET RF Smart Applicator is substantially equivalent in comparison SECRET RF (K170325), the predicate device, while it is similar to AGENS and EVRF, the reference devices, as described herein.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K182355](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K182355)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
