← Product Code [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI) · K182354

# SafeAir Telescopic Smoke Pencil (K182354)

_Lina Medical Aps · GEI · Dec 14, 2018 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K182354

## Device Facts

- **Applicant:** Lina Medical Aps
- **Product Code:** [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI.md)
- **Decision Date:** Dec 14, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

SafeAir Telescopic Smoke Pencil is designed for general electrosurgical applications, including cutting and coagulation, and for removing smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system. The pencil enables the operator to remotely conduct an electrosurgical current from the output connector of an electrosurgical unit to the operative site for the desired surgical effect.

## Device Story

SafeAir Telescopic Smoke Pencil is a sterile, single-use electrosurgical handpiece integrating smoke evacuation. It connects to an electrosurgical unit (generator) and an external smoke evacuation system. The device features a telescopic shaft extendable by up to 120 mm via a locking ring, allowing operators to adjust reach during surgery. It conducts electrosurgical current to the operative site while simultaneously evacuating surgical smoke. Used in clinical surgical environments by physicians/surgeons. The device benefits patients by reducing exposure to surgical smoke and providing adjustable reach for precise electrosurgical cutting and coagulation.

## Clinical Evidence

No clinical data. Substantial equivalence established through non-clinical bench testing, including biocompatibility (ISO 10993), electrical safety (IEC 60601-1/2-2), sterilization (ISO 11135), packaging/aging (ASTM F88, F1929, F1980, D4169), thermal spread, flammability, and smoke evacuation performance testing.

## Technological Characteristics

Monopolar electrosurgical pencil with integrated smoke evacuation. Materials: 303 SST (electrode), Trianmx731 (suction sleeve), ABS/TPE (housing). Dimensions: 18mm x 200mm. Telescopic mechanism extends shaft up to 120mm. Electrical: 5.5 kV peak voltage, US-3-Pin connector. Sterilization: EtO (SAL 10^-6). Standards: ISO 10993-1, AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-2-2, ISO 11135.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- Stryker Neptune E-SEP Smoke Evacuation Pencil ([K160693](/device/K160693.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 14, 2018

LiNA Medical ApS % Kevin MacDonald Clinical/Regulatory Consultant MacDonald Regulatory Consulting 4297 D Street Sacramento, California 95819

Re: K182354

Trade/Device Name: SafeAir Telescopic Smoke Pencil Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: November 16, 2018 Received: November 19, 2018

Dear Kevin MacDonald:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Image /page/1/Picture/5 description: The image shows a digital signature. The signature is from Long H. Chen -S, and the date of the signature is 2018.12.14. The time of the signature is 14:56:52 -05'00'. The signature is for something.

Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K182354

Device Name SafeAir Telescopic Smoke Pencil

#### Indications for Use (Describe)

SafeAir Telescopic Smoke Pencil is designed for general electrosurgical applications, including cutting and coagulation, and for removing smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system. The pencil enables the operator to remotely conduct an electrosurgical current from the output connector of an electrosurgical unit to the operative site for the desired surgical effect.

| Type of Use (Select one or both, as applicable)                                               |                                                                                              |
|-----------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|
| <div><span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div><span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)</div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# K182354 510(k) Summary SafeAir Telescopic Smoke Pencil

#### 1. Submission Sponsor:

 LiNA Medical ApS Formervangen 5 DK-2600 Glostrup, Denmark Natalia Szychulska Regulatory Affairs Officer Email: nhe@lina-medical.com Office number: +48 61 222 21 43

#### 2. Submission Correspondent:

Kevin MacDonald U.S. Regulatory Consultant Email: kma@lina-medical.com Office number: 1-415-609-9875

#### 3. Date prepared:

December 11, 2018

#### 4. Device Identification

| Type of 510(k) Submission: | Traditional                                                         |
|----------------------------|---------------------------------------------------------------------|
| Trade or Proprietary Name: | SafeAir Telescopic Smoke Pencil                                     |
| Regulation Number:         | 21 CFR 878.4400                                                     |
| Product Code:              | GEI, Electrosurgical Cutting And Coagulation Device And Accessories |
| Class of Device: Class     | II                                                                  |
| Panel:                     | General and Plastic Surgery                                         |
| Reason for Submission:     | New device                                                          |
| Prior Related Submissions: | No prior submissions for the device                                 |
| Multiple Devices:          | n/a                                                                 |

{4}------------------------------------------------

#### K182354

### 5. Legally Marketed Predicate Device(s)

Stryker Neptune E-SEP Smoke Evacuation Pencil (K160693)

#### 6. Indication for Use Statement

SafeAir Telescopic Smoke Pencil is designed for general electrosurgical applications, including cutting and coagulation, and for removing smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system. The pencil enables the operator to remotely conduct an electrosurgical current from the output connector of an electrosurgical unit to the operative site for the desired surgical effect

#### 7. Device Description

SafeAir Telescopic Smoke Pencil is designed for general electrosurgical applications and for removing smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system. The pencil enables the operator to remotely conduct an electrosurgical current from the output connector of an electrosurgical unit to the operative site for the desired surgical effect.

SafeAir Telescopic Smoke Pencil has got telescopic function which allows to extend the length of pencil by up to 120 mm. Extension is possible after twisting the green lock-ring.

| Model number |                                 |
|--------------|---------------------------------|
| SHK-TSP-US   | SafeAir Telescopic Smoke Pencil |

#### 8. Substantial Equivalence Discussion

Both the SafeAir Telescopic Smoke Pencil and the Stryker Neptune E-SEP Smoke Evacuation Pencil (K160693) are manufactured by LiNA Medical in Poland. The subject and predicate devices are substantially equivalent in terms of technological characteristics and intended use. The primary difference is the telescopic feature present on the SafeAir Telescopic Smoke Pencil (Subject device).

| Substantial Equivalence Table            |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                                                                                                                               |
|------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Characteristic                           | Predicate Device<br>Stryker Neptune E-SEP Smoke Evacuation<br>Pencil K160693                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Subject Device<br>SafeAir Telescopic Smoke Pencil                                                                                                                                                                                                                                                                                                                                             |
| Product Code                             | GEI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Same                                                                                                                                                                                                                                                                                                                                                                                          |
| Regulatory Class                         | II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same                                                                                                                                                                                                                                                                                                                                                                                          |
| Regulation Name                          | Electrosurgical Cutting And Coagulation Device<br>And Accessories                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Same                                                                                                                                                                                                                                                                                                                                                                                          |
| Regulation Number                        | 21 CFR 878.4400                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Same                                                                                                                                                                                                                                                                                                                                                                                          |
| Prescription                             | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Same                                                                                                                                                                                                                                                                                                                                                                                          |
| Intended Use                             | The Neptune E-SEP is an Integrated Smoke<br>Evacuation Pencil (pencil) designed for general                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | SafeAir Telescopic Smoke Pencil is<br>designed for general electrosurgical                                                                                                                                                                                                                                                                                                                    |
| Substantial Equivalence Table            |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                                                                                                                               |
| Characteristic                           | Predicate Device<br>Stryker Neptune E-SEP Smoke Evacuation<br>Pencil K160693                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Subject Device<br>SafeAir Telescopic Smoke Pencil                                                                                                                                                                                                                                                                                                                                             |
|                                          | electrosurgical applications including cutting<br>and coagulation, and for removing smoke<br>generated by electrosurgery when used in<br>conjunction with an effective smoke<br>evacuation system. The pencil enables the<br>operator to remotely conduct an<br>electrosurgical current from the output<br>connector of an electrosurgical unit (generator)<br>to the operative site for the desired surgical<br>effect                                                                                                                                                                                                                                                 | applications, including cutting and<br>coagulation, and for removing smoke<br>generated by electrosurgery when<br>used in conjunction with an effective<br>smoke evacuation system. The pencil<br>enables the operator to remotely<br>conduct an electrosurgical current<br>from the output connector of an<br>electrosurgical unit to the operative<br>site for the desired surgical effect. |
| Overall Design                           | Designed to integrate smoke evacuation into<br>electrosurgery by combining both features into<br>a single handpiece                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Designed to integrate smoke<br>evacuation into electrosurgery by<br>combining both features into a single<br>handpiece. Incorporates a<br>telescoping feature allowing the<br>length top extended up to 120mm                                                                                                                                                                                 |
| Power Supply                             | Monopolar Generator                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Same                                                                                                                                                                                                                                                                                                                                                                                          |
| Voltage Rating                           | 5.5 kV peak                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                                                                                                                                                                                                                                                          |
| Electrical Connector                     | US-3-Pin                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Same                                                                                                                                                                                                                                                                                                                                                                                          |
| Electrical Safety<br>Testing             | ISO 60601-1<br>ISO 60601-1-2<br>ISO 60601-2-2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Same                                                                                                                                                                                                                                                                                                                                                                                          |
| Sterility                                | Sterile Single Use, EtO, SAL 10-6                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Same                                                                                                                                                                                                                                                                                                                                                                                          |
| Packaging                                | Individually packaged pencil and electrode in<br>Tyvek pouch, sold 10 per box                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Same                                                                                                                                                                                                                                                                                                                                                                                          |
| Electrode Rod                            | 303 SST                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Same                                                                                                                                                                                                                                                                                                                                                                                          |
| Electrode Rod<br>Diameter                | 2.36mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Same                                                                                                                                                                                                                                                                                                                                                                                          |
| Electrode Connector<br>Shape             | Pentagon                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Heat shrink sheath                                                                                                                                                                                                                                                                                                                                                                            |
| Electrode Connector<br>Material          | ABS Polylac PA-757                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | PTFE Heat shrink                                                                                                                                                                                                                                                                                                                                                                              |
| Electrode shaft<br>insulation (overmold) | ABS Polylac PA-757                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | None                                                                                                                                                                                                                                                                                                                                                                                          |
| Electrode Tip<br>Material                | 303 SST                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Same                                                                                                                                                                                                                                                                                                                                                                                          |
| Electrode Tip Coating                    | None                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Same                                                                                                                                                                                                                                                                                                                                                                                          |
| Electrode Tip<br>Working length          | 70-165mm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | 70mm                                                                                                                                                                                                                                                                                                                                                                                          |
| Electrode Tip<br>Insulation              | None                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | None                                                                                                                                                                                                                                                                                                                                                                                          |
| Substantial Equivalence Table            |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                                                                                                                               |
| Characteristic                           | Predicate Device<br>Stryker Neptune E-SEP Smoke Evacuation<br>Pencil K160693                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Subject Device<br>SafeAir Telescopic Smoke Pencil                                                                                                                                                                                                                                                                                                                                             |
| Adjustable Suction<br>Sleeve Material    | Acrylonitrile Butadiene Styrene with barium<br>sulfate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Trianmx731 transparent                                                                                                                                                                                                                                                                                                                                                                        |
| Evacuation Tubing<br>Dimension           | 10 mm dia x 3 m length                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Same                                                                                                                                                                                                                                                                                                                                                                                          |
| Smoke Evacuation<br>System connector     | 8mm, 22m                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Same                                                                                                                                                                                                                                                                                                                                                                                          |
| Handpiece Housing<br>Material            | Acrylonitrile Butadiene Styrene with<br>Thermoplastic Elastomer                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Same                                                                                                                                                                                                                                                                                                                                                                                          |
| Handpiece<br>Dimensions                  | 15mm dia x 190mm length                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | 18 mm x 200 mm length                                                                                                                                                                                                                                                                                                                                                                         |
| Operation Function<br>Switches           | CUT/COAG buttons available in both Rocker-<br>switch and Pushbutton-switch configuration                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | CUT/COAG buttons available in<br>Pushbutton configuration                                                                                                                                                                                                                                                                                                                                     |
| Accessories                              | Holster-Clip provided                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | None                                                                                                                                                                                                                                                                                                                                                                                          |
| Tissue contacting<br>Materials           | Compliant with ISO 10993-1                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Same                                                                                                                                                                                                                                                                                                                                                                                          |
| Range of Electrodes                      | Neptune® E-SEP TM 165mm Blade<br>Electrode, Coated and Insulated<br>Neptune® E-SEP TM 125mm Blade<br>Electrode, Coated and Insulated<br>Neptune® E-SEP TM 70mm Blade<br>Electrode, Coated and Insulated<br>Neptune E-SEP 165mm blade electrode<br>Neptune E-SEP loop t-bar electrode W20<br>D15 L60<br>Neptune E-SEP conization electrode W20<br>D20 L120<br>Neptune E-SEP 125mm blade electrode<br>Neptune E-SEP conization electrode W13<br>D20 L120<br>Neptune E-SEP 70mm Blade electrode<br>Neptune E-SEP 70mm needle electrode<br>Neptune E-SEP 5mm ball electrode<br>Neptune E-SEP loop t-bar electrode W20<br>D20 L120<br>Neptune E-SEP loop t-bar electrode W20 | Same                                                                                                                                                                                                                                                                                                                                                                                          |
| Substantial Equivalence Table            |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                                                                                                                               |
| Characteristic                           | Predicate Device<br>Stryker Neptune E-SEP Smoke Evacuation<br>Pencil K160693                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Subject Device<br>SafeAir Telescopic Smoke Pencil                                                                                                                                                                                                                                                                                                                                             |
|                                          | Neptune E-SEP 165mm blade electrode<br>coated                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                               |
|                                          | Neptune E-SEP conization electrode W13<br>D15 L120                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                                                               |
|                                          | Neptune E-SEP loop u-bar electrode W20<br>D20 L120                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                                                               |
|                                          | Neptune E-SEP conization electrode W16<br>D18 L 120                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                               |
|                                          | Neptune E-SEP loop t-bar electrode W10<br>D10 L120                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                                                               |
|                                          | Neptune E-SEP 70mm blade electrode<br>coated                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                                               |
|                                          | Neptune E-SEP loop t-bar electrode W20<br>D15 L120                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                                                               |
|                                          | Neptune E-SEP 3mm ball electrode                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                               |
|                                          | Neptune E-SEP conization electrode W16<br>D8 L120                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                                               |
|                                          | Neptune E-SEP 125mm blade electrode<br>coated                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                               |
|                                          | Neptune E-SEP loop t-bar electrode W15<br>D12 L120                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                                                               |

{5}------------------------------------------------

{6}------------------------------------------------

{7}------------------------------------------------

The SafeAir Telescopic Smoke Pencil shares the same intended use, device operation, overall technical and functional capabilities as the predicate device. In addition, the SafeAir Telescopic Smoke Pencil is similar in design and function to the Stryker Neptune E-SEP Smoke Evacuation Pencil (K160693) in terms of mode of operation and use. In reference to manufacturing, the SafeAir Telescopic Smoke Pencil and the Stryker Neptune E-SEP Smoke Evacuator Pencil are manufactured in the LiNA Medical's Poland manufacturing facility under the same environmental conditions.

### 9. Non-Clinical Performance Data

As part of demonstrating safety and effectiveness of the SafeAir Telescopic Smoke Pencil and in showing substantial equivalence to the predicate devices that are subject to this 510(k) submission, LiNA completed a number of non-clinical performance tests. The SafeAir Telescopic Smoke Pencil meets all the requirements for overall design, sterilization, biocompatibility, and electrical safety results confirming that the design output meets the design inputs and specifications for the device.

The SafeAir Telescopic Smoke Pencil was subjected to safety performance testing in accordance with the FDA's Premarket Notification [510(k)] Submissions for Electrosurgical Devices for General Surgery -Guidance for Industry and Food and Drug Administration Staff. Issued: August 15, 2016.

The SafeAir Telescopic Smoke Pencil passed all applicable testing in accordance with internal requirements, national standards, and international standards shown below to support substantial equivalence of the subject device:

{8}------------------------------------------------

- Biocompatibility testing per ISO 10993-1, ISO 10993-10, ISO 10993-10, ISO 10993-12
- . Electrical safety testing per AAMI / ANSI ES60601-1, IEC 60601-1-2, IEC 60601-2-2
- . Sterilization per ISO 11135, ISO 11737-2
- Packaging, Aging and Transport testing per ASTM F88-06, ASTM F1929-98, ASTM F1980-16, ASTM D4169-16
- Thermal Spread Testing
- Flammability Testing
- Smoke Evacuation Testing

#### 10. Clinical Performance Data

Based on the similarities in design, intended use and comparability in design verification results to the predicate device, human clinical testing was not required to establish substantial equivalence. These types of devices, including the predicate devices, have been a long market history with demonstrated safety and effectiveness for the use of the device. The non-clinical testing detailed in this submission supports the substantial equivalence determination.

#### 11. Statement of Substantial Equivalence

Based on the similarities in design, intended use statement and comparative verification test results, the SafeAir Telescopic Smoke Pencil raises no new issues of safety and effectiveness when compared to the predicate device, Stryker Neptune E-SEP Smoke Evacuation Pencil.

The SafeAir Telescopic Smoke Pencil function differs only in the telescopic feature which does not impact the intended use or significantly impact the technological features and therefore is determined to be substantially equivalent to the referenced predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K182354](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K182354)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
