K182354 · Lina Medical Aps · GEI · Dec 14, 2018 · General, Plastic Surgery
Device Facts
Record ID
K182354
Device Name
SafeAir Telescopic Smoke Pencil
Applicant
Lina Medical Aps
Product Code
GEI · General, Plastic Surgery
Decision Date
Dec 14, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
SafeAir Telescopic Smoke Pencil is designed for general electrosurgical applications, including cutting and coagulation, and for removing smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system. The pencil enables the operator to remotely conduct an electrosurgical current from the output connector of an electrosurgical unit to the operative site for the desired surgical effect.
Device Story
SafeAir Telescopic Smoke Pencil is a sterile, single-use electrosurgical handpiece integrating smoke evacuation. It connects to an electrosurgical unit (generator) and an external smoke evacuation system. The device features a telescopic shaft extendable by up to 120 mm via a locking ring, allowing operators to adjust reach during surgery. It conducts electrosurgical current to the operative site while simultaneously evacuating surgical smoke. Used in clinical surgical environments by physicians/surgeons. The device benefits patients by reducing exposure to surgical smoke and providing adjustable reach for precise electrosurgical cutting and coagulation.
Clinical Evidence
No clinical data. Substantial equivalence established through non-clinical bench testing, including biocompatibility (ISO 10993), electrical safety (IEC 60601-1/2-2), sterilization (ISO 11135), packaging/aging (ASTM F88, F1929, F1980, D4169), thermal spread, flammability, and smoke evacuation performance testing.
Technological Characteristics
Monopolar electrosurgical pencil with integrated smoke evacuation. Materials: 303 SST (electrode), Trianmx731 (suction sleeve), ABS/TPE (housing). Dimensions: 18mm x 200mm. Telescopic mechanism extends shaft up to 120mm. Electrical: 5.5 kV peak voltage, US-3-Pin connector. Sterilization: EtO (SAL 10^-6). Standards: ISO 10993-1, AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-2-2, ISO 11135.
Indications for Use
Indicated for general electrosurgical applications (cutting and coagulation) and smoke removal during electrosurgery. For use by clinicians in surgical settings requiring electrosurgical current delivery and smoke evacuation.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 14, 2018
LiNA Medical ApS % Kevin MacDonald Clinical/Regulatory Consultant MacDonald Regulatory Consulting 4297 D Street Sacramento, California 95819
Re: K182354
Trade/Device Name: SafeAir Telescopic Smoke Pencil Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: November 16, 2018 Received: November 19, 2018
Dear Kevin MacDonald:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{2}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Image /page/1/Picture/5 description: The image shows a digital signature. The signature is from Long H. Chen -S, and the date of the signature is 2018.12.14. The time of the signature is 14:56:52 -05'00'. The signature is for something.
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K182354
Device Name SafeAir Telescopic Smoke Pencil
#### Indications for Use (Describe)
SafeAir Telescopic Smoke Pencil is designed for general electrosurgical applications, including cutting and coagulation, and for removing smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system. The pencil enables the operator to remotely conduct an electrosurgical current from the output connector of an electrosurgical unit to the operative site for the desired surgical effect.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|
| <div><span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div><span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# K182354 510(k) Summary SafeAir Telescopic Smoke Pencil
#### 1. Submission Sponsor:
LiNA Medical ApS Formervangen 5 DK-2600 Glostrup, Denmark Natalia Szychulska Regulatory Affairs Officer Email: nhe@lina-medical.com Office number: +48 61 222 21 43
#### 2. Submission Correspondent:
Kevin MacDonald U.S. Regulatory Consultant Email: kma@lina-medical.com Office number: 1-415-609-9875
#### 3. Date prepared:
December 11, 2018
#### 4. Device Identification
| Type of 510(k) Submission: | Traditional |
|----------------------------|---------------------------------------------------------------------|
| Trade or Proprietary Name: | SafeAir Telescopic Smoke Pencil |
| Regulation Number: | 21 CFR 878.4400 |
| Product Code: | GEI, Electrosurgical Cutting And Coagulation Device And Accessories |
| Class of Device: Class | II |
| Panel: | General and Plastic Surgery |
| Reason for Submission: | New device |
| Prior Related Submissions: | No prior submissions for the device |
| Multiple Devices: | n/a |
{5}------------------------------------------------
#### K182354
### 5. Legally Marketed Predicate Device(s)
Stryker Neptune E-SEP Smoke Evacuation Pencil (K160693)
#### 6. Indication for Use Statement
SafeAir Telescopic Smoke Pencil is designed for general electrosurgical applications, including cutting and coagulation, and for removing smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system. The pencil enables the operator to remotely conduct an electrosurgical current from the output connector of an electrosurgical unit to the operative site for the desired surgical effect
#### 7. Device Description
SafeAir Telescopic Smoke Pencil is designed for general electrosurgical applications and for removing smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system. The pencil enables the operator to remotely conduct an electrosurgical current from the output connector of an electrosurgical unit to the operative site for the desired surgical effect.
SafeAir Telescopic Smoke Pencil has got telescopic function which allows to extend the length of pencil by up to 120 mm. Extension is possible after twisting the green lock-ring.
| Model number | |
|--------------|---------------------------------|
| SHK-TSP-US | SafeAir Telescopic Smoke Pencil |
#### 8. Substantial Equivalence Discussion
Both the SafeAir Telescopic Smoke Pencil and the Stryker Neptune E-SEP Smoke Evacuation Pencil (K160693) are manufactured by LiNA Medical in Poland. The subject and predicate devices are substantially equivalent in terms of technological characteristics and intended use. The primary difference is the telescopic feature present on the SafeAir Telescopic Smoke Pencil (Subject device).
| Substantial Equivalence Table | | |
|------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Characteristic | Predicate Device<br>Stryker Neptune E-SEP Smoke Evacuation<br>Pencil K160693 | Subject Device<br>SafeAir Telescopic Smoke Pencil |
| Product Code | GEI | Same |
| Regulatory Class | II | Same |
| Regulation Name | Electrosurgical Cutting And Coagulation Device<br>And Accessories | Same |
| Regulation Number | 21 CFR 878.4400 | Same |
| Prescription | Yes | Same |
| Intended Use | The Neptune E-SEP is an Integrated Smoke<br>Evacuation Pencil (pencil) designed for general | SafeAir Telescopic Smoke Pencil is<br>designed for general electrosurgical |
| Substantial Equivalence Table | | |
| Characteristic | Predicate Device<br>Stryker Neptune E-SEP Smoke Evacuation<br>Pencil K160693 | Subject Device<br>SafeAir Telescopic Smoke Pencil |
| | electrosurgical applications including cutting<br>and coagulation, and for removing smoke<br>generated by electrosurgery when used in<br>conjunction with an effective smoke<br>evacuation system. The pencil enables the<br>operator to remotely conduct an<br>electrosurgical current from the output<br>connector of an electrosurgical unit (generator)<br>to the operative site for the desired surgical<br>effect | applications, including cutting and<br>coagulation, and for removing smoke<br>generated by electrosurgery when<br>used in conjunction with an effective<br>smoke evacuation system. The pencil<br>enables the operator to remotely<br>conduct an electrosurgical current<br>from the output connector of an<br>electrosurgical unit to the operative<br>site for the desired surgical effect. |
| Overall Design | Designed to integrate smoke evacuation into<br>electrosurgery by combining both features into<br>a single handpiece | Designed to integrate smoke<br>evacuation into electrosurgery by<br>combining both features into a single<br>handpiece. Incorporates a<br>telescoping feature allowing the<br>length top extended up to 120mm |
| Power Supply | Monopolar Generator | Same |
| Voltage Rating | 5.5 kV peak | Same |
| Electrical Connector | US-3-Pin | Same |
| Electrical Safety<br>Testing | ISO 60601-1<br>ISO 60601-1-2<br>ISO 60601-2-2 | Same |
| Sterility | Sterile Single Use, EtO, SAL 10-6 | Same |
| Packaging | Individually packaged pencil and electrode in<br>Tyvek pouch, sold 10 per box | Same |
| Electrode Rod | 303 SST | Same |
| Electrode Rod<br>Diameter | 2.36mm | Same |
| Electrode Connector<br>Shape | Pentagon | Heat shrink sheath |
| Electrode Connector<br>Material | ABS Polylac PA-757 | PTFE Heat shrink |
| Electrode shaft<br>insulation (overmold) | ABS Polylac PA-757 | None |
| Electrode Tip<br>Material | 303 SST | Same |
| Electrode Tip Coating | None | Same |
| Electrode Tip<br>Working length | 70-165mm | 70mm |
| Electrode Tip<br>Insulation | None | None |
| Substantial Equivalence Table | | |
| Characteristic | Predicate Device<br>Stryker Neptune E-SEP Smoke Evacuation<br>Pencil K160693 | Subject Device<br>SafeAir Telescopic Smoke Pencil |
| Adjustable Suction<br>Sleeve Material | Acrylonitrile Butadiene Styrene with barium<br>sulfate | Trianmx731 transparent |
| Evacuation Tubing<br>Dimension | 10 mm dia x 3 m length | Same |
| Smoke Evacuation<br>System connector | 8mm, 22m | Same |
| Handpiece Housing<br>Material | Acrylonitrile Butadiene Styrene with<br>Thermoplastic Elastomer | Same |
| Handpiece<br>Dimensions | 15mm dia x 190mm length | 18 mm x 200 mm length |
| Operation Function<br>Switches | CUT/COAG buttons available in both Rocker-<br>switch and Pushbutton-switch configuration | CUT/COAG buttons available in<br>Pushbutton configuration |
| Accessories | Holster-Clip provided | None |
| Tissue contacting<br>Materials | Compliant with ISO 10993-1 | Same |
| Range of Electrodes | Neptune® E-SEP TM 165mm Blade<br>Electrode, Coated and Insulated<br>Neptune® E-SEP TM 125mm Blade<br>Electrode, Coated and Insulated<br>Neptune® E-SEP TM 70mm Blade<br>Electrode, Coated and Insulated<br>Neptune E-SEP 165mm blade electrode<br>Neptune E-SEP loop t-bar electrode W20<br>D15 L60<br>Neptune E-SEP conization electrode W20<br>D20 L120<br>Neptune E-SEP 125mm blade electrode<br>Neptune E-SEP conization electrode W13<br>D20 L120<br>Neptune E-SEP 70mm Blade electrode<br>Neptune E-SEP 70mm needle electrode<br>Neptune E-SEP 5mm ball electrode<br>Neptune E-SEP loop t-bar electrode W20<br>D20 L120<br>Neptune E-SEP loop t-bar electrode W20 | Same |
| Substantial Equivalence Table | | |
| Characteristic | Predicate Device<br>Stryker Neptune E-SEP Smoke Evacuation<br>Pencil K160693 | Subject Device<br>SafeAir Telescopic Smoke Pencil |
| | Neptune E-SEP 165mm blade electrode<br>coated | |
| | Neptune E-SEP conization electrode W13<br>D15 L120 | |
| | Neptune E-SEP loop u-bar electrode W20<br>D20 L120 | |
| | Neptune E-SEP conization electrode W16<br>D18 L 120 | |
| | Neptune E-SEP loop t-bar electrode W10<br>D10 L120 | |
| | Neptune E-SEP 70mm blade electrode<br>coated | |
| | Neptune E-SEP loop t-bar electrode W20<br>D15 L120 | |
| | Neptune E-SEP 3mm ball electrode | |
| | Neptune E-SEP conization electrode W16<br>D8 L120 | |
| | Neptune E-SEP 125mm blade electrode<br>coated | |
| | Neptune E-SEP loop t-bar electrode W15<br>D12 L120 | |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
The SafeAir Telescopic Smoke Pencil shares the same intended use, device operation, overall technical and functional capabilities as the predicate device. In addition, the SafeAir Telescopic Smoke Pencil is similar in design and function to the Stryker Neptune E-SEP Smoke Evacuation Pencil (K160693) in terms of mode of operation and use. In reference to manufacturing, the SafeAir Telescopic Smoke Pencil and the Stryker Neptune E-SEP Smoke Evacuator Pencil are manufactured in the LiNA Medical's Poland manufacturing facility under the same environmental conditions.
### 9. Non-Clinical Performance Data
As part of demonstrating safety and effectiveness of the SafeAir Telescopic Smoke Pencil and in showing substantial equivalence to the predicate devices that are subject to this 510(k) submission, LiNA completed a number of non-clinical performance tests. The SafeAir Telescopic Smoke Pencil meets all the requirements for overall design, sterilization, biocompatibility, and electrical safety results confirming that the design output meets the design inputs and specifications for the device.
The SafeAir Telescopic Smoke Pencil was subjected to safety performance testing in accordance with the FDA's Premarket Notification [510(k)] Submissions for Electrosurgical Devices for General Surgery -Guidance for Industry and Food and Drug Administration Staff. Issued: August 15, 2016.
The SafeAir Telescopic Smoke Pencil passed all applicable testing in accordance with internal requirements, national standards, and international standards shown below to support substantial equivalence of the subject device:
{9}------------------------------------------------
- Biocompatibility testing per ISO 10993-1, ISO 10993-10, ISO 10993-10, ISO 10993-12
- . Electrical safety testing per AAMI / ANSI ES60601-1, IEC 60601-1-2, IEC 60601-2-2
- . Sterilization per ISO 11135, ISO 11737-2
- Packaging, Aging and Transport testing per ASTM F88-06, ASTM F1929-98, ASTM F1980-16, ASTM D4169-16
- Thermal Spread Testing
- Flammability Testing
- Smoke Evacuation Testing
#### 10. Clinical Performance Data
Based on the similarities in design, intended use and comparability in design verification results to the predicate device, human clinical testing was not required to establish substantial equivalence. These types of devices, including the predicate devices, have been a long market history with demonstrated safety and effectiveness for the use of the device. The non-clinical testing detailed in this submission supports the substantial equivalence determination.
#### 11. Statement of Substantial Equivalence
Based on the similarities in design, intended use statement and comparative verification test results, the SafeAir Telescopic Smoke Pencil raises no new issues of safety and effectiveness when compared to the predicate device, Stryker Neptune E-SEP Smoke Evacuation Pencil.
The SafeAir Telescopic Smoke Pencil function differs only in the telescopic feature which does not impact the intended use or significantly impact the technological features and therefore is determined to be substantially equivalent to the referenced predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.