The DiLumen Endolumenal Interventional Scissors ("DiLumen Is™") is a disposable monopolar electrosurgical device intended to be used for cutting, and cauterizing tissue within the digestive tract during endoscopic procedures.
Device Story
Disposable monopolar electrosurgical device for endoscopic digestive tract procedures; operated by clinician via pistol-style handle. Input: manual clinician control of shaft rotation, blade articulation, and blade actuation. Device transforms manual inputs into mechanical movement of distal scissor blades; optional monopolar electrical current applied via external generator for tissue cutting/cauterization. Used in clinical settings; inserted through endoscopic tool channels (≥6mm). Output: mechanical cutting/dissection and electrosurgical cauterization. Benefits: enables precise endolumenal tissue management; flexible shaft facilitates navigation through gastrointestinal tract.
Clinical Evidence
Bench testing only. Includes biocompatibility, dissection force, tip/shaft articulation and rotation accuracy, cutting durability, electrical safety, electromagnetic compatibility, usability evaluation, packaging/transit testing, and EO residuals. No clinical data presented.
Technological Characteristics
Materials: Stainless steel (jaws), Pebax with Aramid braid (shaft), Fluoropolymer (sleeve). Principle: Manual mechanical actuation with optional monopolar electrosurgical energy. Form factor: Pistol-handle, flexible shaft (95cm/140cm), 5mm diameter. Connectivity: Banana plug for electrosurgical generator. Sterilization: Ethylene Oxide (EO).
Indications for Use
Indicated for cutting and cauterizing tissue within the digestive tract during endoscopic procedures in patients requiring such interventions.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
Cambridge Endoscopic Devices, Inc. pureWrist™ electrocautery laparoscopic instruments (K061425)
K151474 — Fujifilm Diathermic Slitter (FlushKnife), Diathermic Slitter (ClutchCutter) · Fujifilm Medical System U.S.A., Inc. · Feb 18, 2016
K090583 — ENDOSCOPIC MONOPOLAR SCISSORS, MODELS EMS-235, EMS-165 · Apollo Endosurgery, Inc. · Jun 26, 2009
K212999 — Multi-Functional Electrosurgical Knife · Anrei Medical (Hangzhou) Co., Ltd. · Mar 21, 2022
K241855 — SB Knife Jr2 (MD-47702 and MD-47702L) · Sb-Kawasumi Laboratories, Inc. · Mar 26, 2025
K091824 — ETHICON ENDO-SURGERY ROTATING ENDOSCOPIC SCISSORS, MODEL IN2507 · Ethicon Endo-Surgery, Inc. · Nov 27, 2009
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
May 22, 2018
Lumendi, LLC % John J. Smith, MD, JD Partner Hogan Lovells U.S. LLP 555 13th Street NW Washington, DC 20004
Re: K173405
Trade/Device Name: DiLumen Endolumenal Interventional Scissors ("DiLumen Is™") Regulation Number: 21 CFR§ 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: II Product Code: GEI Dated: April 6, 2018 Received: April 6, 2018
Dear John J. Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Joyce M. Whang -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K173405
#### Device Name
DiLumen Endolumenal Interventional Scissors ("DiLumen IsTM")
Indications for Use (Describe)
The DiLumen Endolumenal Interventional Scissors ("DiLumen Is™") is a disposable monopolar electrosurgical device intended to be used for cutting, and cauterizing tissue within the digestive tract during endoscopic procedures.
Type of Use (*Select one or both, as applicable*)
| <span style="font-size:12pt">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|------------------------------------------------------------------------------------|
| <span style="font-size:12pt">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) SUMMARY
### Lumendi, LLC's DiLumen Endolumenal Interventional Scissors (DiLumen I、™)
#### Submitter's Information:
LUMENDI, LLC 253 Post Road West Westport, CT 06880 Phone: (203) 463-2669 Fax: (203) 557-0459
#### Contact Person:
Dennis J. Daniels Senior Director, Requlatory Affairs and Quality Assurance Telephone: (203) 557-6335 Fax: (203) 557-0459 Email: Dennis.Daniels@Lumendi.com
Date Prepared: October 31, 2017
#### Device Identification:
Trade Name: Dilumen Endolumenal Interventional Scissors (DiLumen III) Common Name: Electrosurgical cutting & coagulation device Classification Regulation: 21 C.F.R. § 878.4400, Electrosurgical cutting & coagulation device and accessories Product Code: GEI Device Class: II
Predicate Device: Cambridge Endoscopic Devices, Inc. pureWrist™ electrocautery laparoscopic instruments (K061425)
Reference Device: Sumitomo Bakelite Co., Ltd. SB Knife (K152771)
### Device Description:
The DiLumen I , is a sterile, single patient use, disposable instrument that consists of a pistol style handle, a flexible shaft with an articulating section at the distal end, and scissor blades. The handle incorporates controls that enable one-handed operation of the device, including rotating the shaft; opening, closing and rotating the blades independently from the shaft; articulating the distal end of the shaft in a specific plane; and locking the articulation in a fixed position. This allows the clinician maximum flexibility to cut, dissect and cauterize tissue during endoscopic interventions in the digestive tract. The clinician can insert the DiLumen Is into any endoscopic tool channel with a diameter of at least 6mm. With the DiLumen Is under direct visualization, the clinician can position the scissor blades as appropriate for the particular procedure using blade rotation, blade articulation, and shaft rotation as necessary. The clinician can then cut gastrointestinal tissue either with or without energizing the blades, depending on the specific need during the procedure.
In addition, the DiLumen Is has a plug at the bottom of the handle. If desired by the clinician, this plug can be used to connect the device to an electrosurgical generator via a dedicated cord in order
{4}------------------------------------------------
to apply monopolar electrical current through the blades to the target site in the gastrointestinal tract. When the device is connected to the generator, monopolar current can be applied through the blades to cut or cauterize tissue electrosurgically.
The DiLumen I , can also be used in conjunction with other endoscopic devices such as graspers, knives, etc. to perform endolumenal interventions.
## Intended Use / Indications for Use:
The DiLumen Endolumenal Interventional Scissors ("DiLumen I,") is a disposable monopolar electrosurgical device intended to be used for cutting, and cauterizing tissue within the digestive tract during endoscopic procedures.
## Summary of Technological Characteristics:
The technological principle underlying both the subject and predicate devices is manual and/or monopolar electrosurgical cutting, dissection, and cauterization of tissue. Both devices are based on the following same technological elements:
- . Both are sterile, single use, monopolar scissors that allow for one-handed clinician operation.
- . Both share key functional characteristics, including the control handle, an insulated working shaft with the same diameter, an articulating tip, and two curved rotating stainless-steel blades.
- . Each device can be connected to an electrosurgical generator using a dedicated cable plugged into the handle, in order to make use of monopolar electrical current to facilitate or enhance the scissors' functioning.
- . Both are prescription-use devices employed by medical professionals in a healthcare facility.
The primary technological difference between the subject and predicate devices is that the predicate's working shaft is rigid and shorter than that of the DiLumen Is. This does not raise different questions of safety or effectiveness, because the greater flexibility and length of the DiLumen Is working shaft simply makes it easier for the device to follow the path inside a flexible tool channel within the gastrointestinal tract; it does not alter the device's principles of operation or the purposes for which it is used. Furthermore, FDA has cleared other endoscopic cutting devices with longer flexible shafts, such as the reference device, Sumitomo Bakelite's SB Knife (K152771). The DiLumen Is also has one feature that is not found in the predicate: a tethered Blades Lock attached to the bottom of the handle that can be inserted between the body of the handle to keep the trigger squeezed and the blades closed. However, this addition is for user convenience; the predicate's blades could also be kept closed by the clinician.
| | Lumendi DiLumen Endolumenal<br>Interventional Scissors (Iₛ) | Cambridge Endoscopic Devices, Inc.<br>pureWrist™ electrocautery<br>laparoscopic Instruments |
|---------------------------------------|-------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
| 510(k) | TBD | K061425 |
| Intended Use /<br>Indications for Use | The DiLumen Endolumenal<br>Interventional Scissors ("DiLumen Iₛ") | The pureWrist™ electrocautery<br>laparoscopic Instruments have |
| | is a disposable monopolar | application in a variety of minimally |
| | electrosurgical device intended to be | invasive procedures to facilitate grasping, |
| | used for cutting, dissecting, and | mobilization, dissection and transection of |
The table below compares the key features of the subject and predicate devices.
{5}------------------------------------------------
| | Lumendi DiLumen Endolumenal<br>Interventional Scissors (Is) | Cambridge Endoscopic Devices, Inc.<br>pureWrist™ electrocautery<br>laparoscopic Instruments |
|---------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | cauterizing tissue within the digestive<br>tract during endoscopic procedures. | tissue. |
| Sterility | Sterile - EO | Sterile - Radiation |
| Shelf Life | One year | Two years |
| Single use/reusable | Single Use, fully disposable | Single Use, fully disposable |
| Basic Design | Proximal handle with controls, flexible<br>insulated shaft, articulating section<br>and rotating scissor tip | Proximal handle with controls, rigid<br>insulated shaft, articulating section and<br>rotating scissor tip |
| Patient-Contacting<br>Materials | Stainless steel (jaws), medical-grade<br>polymers: Pebax with embedded<br>Aramid braid (flexible shaft),<br>Fluoropolymer (Articulation sleeve) | Stainless steel (jaws); FEP, PEEK,<br>Pebax, Tecoflex, Cyanoacrylate, PET |
| Electrical connector | Banana plug at bottom of handle | Banana plug at bottom of handle |
| Cautery Type | Monopolar cautery | Monopolar cautery |
| Shaft Outer diameter | 5 mm | 5 mm |
| Shaft length | 95 cm, 140 cm | 15 - 45 cm |
| Operating principle | The DiLumen Is are operated<br>manually using one hand to control<br>blade position and blade actuation.<br>The end effector can be rotated,<br>articulated and positioned, allowing<br>the scissor blades to cut, dissect, and<br>cauterize tissue either with or without<br>monopolar energy | The pureWrist™ electrocautery<br>laparoscopic Instruments are operated<br>manually using one hand to control blade<br>position and blade actuation. The end<br>effector can be rotated, articulated and<br>positioned, allowing the scissor blades to<br>cut, dissect, and cauterize tissue with or<br>without monopolar energy. |
### Performance Data:
Performance testing has demonstrated that the DiLumen Is meets specifications and is as safe and effective as the predicate. The following performance data were provided in support of this Premarket Notification:
- 1. Biocompatibility (cytotoxicity, sensitization, irritation, systemic toxicity, material mediated pyrogenicity)
- 2. Dissection Force Test
- 3. Tip Articulation Accuracy Test
- 4. Tip Rotation Accuracy Test
- 5. Shaft Rotation Accuracy Test
- 6. Cutting Durability Test
- 7. Monopolar Cable Insertion and Removal Force Test
- 8. Electrical Safety Test
- 9. Electromagnetic Compatibility Test
- 10. Usability Evaluation
- 11. Packaging and Transit Test
- 12. User Validation
- 13. EO Residuals
In all instances, the device functioned as intended and the results observed were as expected.
### Conclusion:
The DiLumen Is device has the same intended use and similar indications, technological
{6}------------------------------------------------
characteristics, and principles of operation as the identified predicate. The minor technological differences between the DiLumen I, and its predicate device do not raise any new questions of safety of effectiveness. Furthermore, the performance data presented in the 510(k) notice further support that the DiLumen I functions as safely and performs as well as the identified predicate. Thus, the DiLumen I , is substantially equivalent to the predicate pureWrist™ electrocautery laparoscopic instruments (K061425).
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.