K172757 · Covidien, LLC · GEI · Nov 2, 2017 · General, Plastic Surgery
Device Facts
Record ID
K172757
Device Name
Valleylab FX8 Electrosurgical Platform
Applicant
Covidien, LLC
Product Code
GEI · General, Plastic Surgery
Decision Date
Nov 2, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Valleylab FX8 FX Series Energy Platform is a high frequency electrosurgical generator intended for use with monopolar and bipolar accessories for cutting and coagulating tissue.
Device Story
Radio-frequency (RF) electrosurgical generator; delivers energy to compatible monopolar and bipolar surgical instruments. Energy concentration at instrument tip creates heat; produces surgical effects (cutting, coagulation, desiccation, fulguration) based on waveform variations. Used in hospitals/healthcare facilities; operated by surgeons/clinical staff. Provides power for tissue management; output controlled via footswitches and instruments. Benefits patient by enabling precise tissue manipulation and hemostasis during surgery.
Clinical Evidence
No clinical data. Bench testing only. Verification and validation included electrical safety (IEC 60601-1, IEC 60601-2-2), EMC (IEC 60601-1-2), and ex vivo porcine tissue testing to compare thermal effects.
Technological Characteristics
RF electrosurgical generator; isolated monopolar and bipolar outputs. Complies with IEC 60601-1:2005/A1:2012, IEC 60601-1-2:2014, and IEC 60601-2-2:2009. Operates on electrical mains (47-63 Hz).
Indications for Use
Indicated for use in surgical procedures requiring tissue cutting and coagulation via monopolar and bipolar electrosurgical accessories. Applicable to general surgery patient populations.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
K050579 — MEGA POWER ELECTROSURGICAL GENERATOR · Megadyne Medical Products, Inc. · Mar 24, 2005
K241085 — Valleylab FT10 Energy Platform Software Version 5.0 (VLFT10GEN VLFT10FXGEN VLFT10LSGEN) · Covidien, LLC · Jul 12, 2024
K100390 — SURGIMAX/ SURGIMAX PLUS · Elliquence, LLC · Aug 25, 2010
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
November 2, 2017
Covidien LLC Ms. Sharon McDermott Sr. Specialist, Regulatory Affairs 5920 Longbow Drive Boulder, Colorado 80301
Re: K172757
Trade/Device Name: Valleylab FX8 FX Series Energy Platform Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: September 11, 2017 Received: September 13, 2017
Dear Sharon McDermott:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
> Sincerely, Jennifer R. Stevenson -S3
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K172757
Device Name Valleylab FX8 FX Series Energy Platform
Indications for Use (Describe)
The Valleylab FX8 FX Series Energy Platform is a high frequency electrosurgical generator intended for use with monopolar and bipolar accessories for cutting and coagulating tissue.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------|--|
| <div style="display:inline-block; width:20px; height:20px; border: 1px solid black;"></div> | |
| <div style="display:inline-block; width:20px; height:20px; border: 1px solid black;"></div> | |
لله | Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 6 510(k) Summary
### 510(k) Summary
Date summary prepared: November 1, 2017
### 510(k) Submitter/Holder
Covidien 5920 Longbow Drive Boulder, CO 80301
#### Contact
Sharon McDermott Sr. Regulatory Affairs Product Specialist Telephone: 303-581-6789 Fax: 303-530-6313 Email: sharon.mcdermott@medtronic.com
### Identification of Device
| Trade Name: | Valleylab™ FX8 FX Series Energy Platform |
|----------------------|------------------------------------------------------------------------|
| Catalog Number: | VLFX8GEN |
| Common Name: | Electrosurgical Generator |
| Classification Name: | Electrosurgical cutting and coagulation device and accessories (21 CFR |
| | 878.4400, Class II, GEI) |
### ldentification of Predicate Device
| Trade/Proprietary Name: | Valleylab FT10 Energy Platform |
|----------------------------|----------------------------------------------------------------|
| Common/Usual Name: | Electrosurgical Generator |
| Classification Name: | Electrosurgical Cutting and Coagulation Device and Accessories |
| Class/Panel: | Class II, General and Plastic Surgery |
| Regulation: | 21 CFR 878.4400 |
| Product Code: | GEI |
| 510(k) Submitter / Holder: | Covidien |
| | 5920 Longbow Drive |
| | Boulder, CO, 80301 |
| Predicate 510(k) #: | K151649, K170170 |
### Identification of Reference Predicate
| Trade/Proprietary Name: | Force FX Generators |
|-------------------------|----------------------------------------------------------------|
| Common/Usual Name: | Electrosurgical Generator |
| Classification Name: | Electrosurgical Cutting and Coagulation Device and Accessories |
| Class/Panel: | Class II, GEI, 21 CFR 878.4400 |
{4}------------------------------------------------
| 006_510(k) Summary | K172757 | Page 2 of 3 |
|----------------------------|--------------------|-------------|
| 510(k) Submitter / Holder: | Covidien | |
| | 5920 Longbow Drive | |
| | Boulder, CO, 80301 | |
| 510(k) #: | K944602, K143161 | |
A reference predicate, Force FX Electrosurgical Generators, (K944602, K143161) was used for comparison testing of the "FX Bipola" modes included on the subject Valleylab FX8 FX Series Energy Platform.
### Substantial Equivalence Overview
The proposed Valleylab FX8 FX Series Energy Platform was compared and found to be substantially equivalent to the predicate Valleylab FT10 Energy Platform. Characteristics identified as important for substantial equivalence comparison purposes are shown in the tables below, and similarities and differences of aspects found to be relevant to the potential impact on safety and performance are discussed in this section. None of the differences raise new questions of safety or effectiveness when compared to the predicate device.
Valleylab FX8 FX Series Energy Platform is used in hospitals and other health care facilities where surgical procedures are carried out.
Valleylab FX8 FX Series Energy Platform can be used with a variety of legally marketed accessories including monopolar and bipolar instruments, footswitches, and return electrode pads. The Valleylab FX8 FX Series Energy Platform connects to electrical mains and operates at an input line frequency of 47-63 Hz.
### Principle of Operation
Valleylab FX8 FX Series Energy Platform is a radio-frequency (RF) electrosurgical generator that delivers energy to compatible surgical instruments. The concentration of energy at the tip of the instrument in conjunction with tissue characteristics produces heat. The heating of tissue provides the desired surgical effect (cutting, coagulation, sealing). Variations in the waveform result in the different surgical effects achieved by different modes.
### Indications for Use
The Valleylab FX8 FX Series Energy Platform is a high frequency electrosurgical generator intended for use with monopolar and bipolar accessories for cutting and coagulating tissue.
# Technological Characteristics
The technological characteristics of the proposed Valleylab FX8 FX Series Energy Platform are the same as the predicate, VLFT10. Both are full-featured general-surgery energy platform. The monopolar and bipolar sections of the generator are isolated outputs that provide the appropriate power for cutting, desiccating, and fulgurating tissue during monopolar and bipolar surgery.
During monopolar electrosurgery, radio frequency (RF) energy exits/enters a monopolar pencil and electrode in high concentration and disperses as it enters/exits the return pad. The concentration of energy at the monopolar pencil in conjunction with tissue characteristics produces heat that provides the desired surgical effect.
{5}------------------------------------------------
#### Performance Characteristics
Extensive verification and validation activities were successfully completed that the Valleylab FX8 FX Series Energy Platform performs as intended and is substantially equivalent to its predicate, VLFT10. Testing included the following:
- Compliance with Electrical Safety and EMC standards:
| IEC 60601-1:2005/A1:2012 | Medical electrical equipment – Part 1: General requirements for<br>basic safety and essential performance – Edition 3.1 |
|--------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-1-2:2014 | Medical electrical equipment, Part 1-2: General requirements<br>for basic safety and essential performance – Collateral<br>standard: Electromagnetic compatibility – Requirements and<br>tests |
| IEC 60601-2-2:2009 | Medical electrical equipment – Part 2-2: Particular requirements<br>for the basic safety and essential performance of high frequency<br>surgical equipment and high frequency surgical accessories -<br>Edition 5.0 |
- Ex vivo testing using porcine tissue showed comparable performance with regard to thermal ● effects.
- System verification showed that the Valleylab FX8 FX Series Energy Platform has all required functionality and that it meets system specifications.
- Software verification and validation testing was conducted and documentation provided in accordance with FDA's, Guidance or the Content of Premarket Submissions for Software Contained in Medical Devices
The minor differences that were seen during testing are consistent with design changes and do not raise safety or effectiveness concerns.
#### Clinical Studies
This premarket submission did not rely on the assessment of clinical performance data to demonstrate substantial equivalence.
#### Conclusion
Verification and validation testing demonstrates the proposed Valleylab FX8 FX Series Energy Platform is substantially equivalent to the predicate, Valleylab VLFT10 Energy Platform. The two electrosurgical generators have the same intended use and fundamental technology with the exception that the subject platform does not include vessel sealing functionality. The new generator has similar performance when compared to the predicate device. The differences do not raise any new questions of safety and efficacy when compared with the predicate.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.