← Product Code [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI) · K172610

# THUNDERBEAT 5 mm, 20 cm, Front-actuated Grip Type S, THUNDERBEAT 5 mm, 35 cm, Front-actuated Grip Type S,  THUNDERBEAT 5 mm, 45 cm, Front-actuated Grip Type S (K172610)

_Olympus Medical Systems Corp. · GEI · Apr 17, 2018 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K172610

## Device Facts

- **Applicant:** Olympus Medical Systems Corp.
- **Product Code:** [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI.md)
- **Decision Date:** Apr 17, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The THUNDERBEAT Type S hand instruments are intended to be used with the Ultrasonic Generator (USG-400), the Electrosurgical Generator (ESG-400), and the THUNDERBEAT Transducer (TD-TB400). Seal & Cut mode: The THUNDERBEAT Type S hand instruments when used in combination with the Seal & Cut mode are indicated for open, laparoscopic (including single-site surgery) general surgery and gynecological surgery (including urologic, thoracic, plastic and reconstructive, bowel resections, cholecystectomies, Nissen fundoplication, adhesiolysis, oophorectomy, hysterectomies (both vaginal assisted and abdominal) etc.), and endoscopic surgery or in any procedure in which cutting, vessel ligation (sealing and cutting), coagulation, grasping, and dissection is performed. These devices have been designed to seal and cut vessels (up to and including 7 mm in diameter), tissue bundles, and lymphatics. Seal mode: The THUNDERBEAT Type S hand instruments when used in combination with the Seal mode are indicated for open, laparoscopic (including single-site surgery) general surgery and gynecological surgery (including urologic, thoracic, plastic and reconstructive, bowel resections, cholecystectomies, Nissen fundoplication, adhesiolysis, oophorectomy, hysterectomies (both vaginal assisted and abdominal) etc.), and endoscopic surgery or in any procedure in which vessel sealing, coagulation, grasping is performed. These devices have been designed to seal vessels (up to and including 7 mm in diameter), tissue bundles, and lymphatics. The THUNDERBEAT Type S hand instruments have not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures, and should not be used for these procedures.

## Device Story

THUNDERBEAT Type S hand instruments are sterile, single-use surgical tools used with USG-400/ESG-400 generators and TD-TB400 transducer. Device features front-actuated grip, modified grasper jaw, and probe tip coating. Operates in 'Seal & Cut' (simultaneous ultrasonic and HF bipolar energy) or 'Seal' (HF bipolar only) modes. Surgeon uses device to grasp, seal, cut, coagulate, and dissect tissue bundles, lymphatics, and vessels up to 7mm. Used in open, laparoscopic, and endoscopic procedures. Output provides hemostasis and tissue separation, aiding surgical efficiency and patient recovery. Bench and animal testing confirm performance equivalence to predicate.

## Clinical Evidence

No clinical data. Evidence consists of bench testing (ex-vivo porcine vessel burst pressure, cutting performance, max temperature, grasping force, durability) and animal testing (acute and chronic porcine studies confirming seal performance, thermal spread, and healing).

## Technological Characteristics

Electrosurgical/ultrasonic instrument; 5mm diameter; 20/35/45cm lengths. Materials biocompatible per ISO-10993. Energy: HF Bipolar and Ultrasonic. Standards: IEC60601-1, IEC60601-2-2, IEC60601-2-18, ISO11135 (sterilization), ASTM F1980 (aging). Single-use, sterile.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- Olympus Medical Systems Corp. THUNDERBEAT TB-0520IC/0535PC/0545PC/0535IC/0545IC Hand Instruments ([K132703](/device/K132703.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. Next to that is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

April 17, 2018

Olympus Medical Systems Corp. % Christina Flores Senior Specialist, Regulatory Affairs Gyrus ACMI Inc. 136 Turnpike Road Southborough, Massachusetts 01772

## Re: K172610

Trade/Device Name: THUNDERBEAT 5 mm, 20 cm, Front-actuated Grip Type S, THUNDERBEAT 5 mm, 35 cm, Front-actuated Grip Type S, THUNDERBEAT 5 mm, 45 cm, Front-actuated Grip Type S Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI, LFL Dated: February 23, 2018 Received: February 26, 2018

Dear Christina Flores:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good

{1}------------------------------------------------

manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

> Sincerely, Jennifer R. Stevenson -S For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K172610

#### Device Name

THUNDERBEAT 5 mm, 20 cm, Front-actuated Grip Type S THUNDERBEAT 5 mm, 35 cm, Front-actuated Grip Type S THUNDERBEAT 5 mm, 45 cm, Front-actuated Grip Type S

#### Indications for Use (Describe)

The THUNDERBEAT Type S hand instruments are intended to be used with the Ultrasonic Generator (USC-400), the Electrosurgical Generator (ESG-400), and the THUNDERBEAT Transducer (TD-TB400).

#### Seal & Cut mode:

The THUNDERBEAT Type S hand instruments when used in combination with the Seal & Cut mode are indicated for open, laparoscopic (including single-site surgery and gynecological surgery (including urologic, thoracic, plastic and reconstructive, bowel resections, cholecystectomies, Nissen fundoplication, adhesiolysis, oophorectomy, hysterectomies (both vaginal assisted and abdominal) etc.), and endoscopic surgery or in any procedure in which cutting, vessel ligation (sealing and cutting), coagulation, grasping, and dissection is performed. These devices have been designed to seal and cut vessels (up to and including 7 mm in diameter), tissue bundles, and lymphatics.

#### Seal mode:

The THUNDERBEAT Type S hand instruments when used in combination with the Seal mode are indicated for open, laparoscopic (including single-site surgery) general surgery (including urologic, thoracic, plastic and reconstructive, bowel resections, cholecystectomies, Nissen fundoplication, adhesiolysis, oophorectomy, hysterectomies (both vaginal assisted and abdominal) etc.), and endoscopic surgery or in any procedure in which vessel sealing, coagulation, grasping is performed. These devices have been designed to seal vessels (up to and including 7 mm in diameter), tissue bundles, and lymphatics.

The THUNDERBEAT Type S hand instruments have not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures, and should not be used for these procedures.

Type of Use (*Select one or both, as applicable*)

| <span style="font-size:10px"><b></b></span>                                                                                | <span style="font-size:10px"><b></b></span>                                                                               |
|----------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|
| <div style="display:inline-block"><span style="font-size:10px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div style="display:inline-block"><span style="font-size:10px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)</div> |

CONTINUE ON A SEPARATE PAGE IF NEEDED.#### This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary Gyrus ACMI, Inc. THUNDERBEAT Type S Hand Instruments

## General Information

Regulation Number:

Regulatory Class:

Product Codes:

Review Panel:

| Applicant:                         | OLYMPUS MEDICAL SYSTEMS<br>CORP.<br>2951 Ishikawa-cho, Hachioji-shi, Tokyo,<br>Japan 192-8507<br>Phone: (+81) 42-642-2694<br>Fax: (+81) 42-642-2307 |
|------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|
| Establishment Registration Number: | 8010047                                                                                                                                             |
| Manufacturer:                      | Aomori Olympus<br>2-248-1 okkonoki<br>kuroishi-shi aomori, Japan 036-0357<br>Phone: (+81) 172-52-8543<br>Fax: (+81) 172-52-8515                     |
| Establishment Registration Number: | 9614641                                                                                                                                             |
| 510(k) Submitter:                  | Gyrus ACMI, Inc.<br>136 Turnpike Rd.<br>Southborough, MA 01772-2104                                                                                 |
| Establishment Registration Number: | 3003790304                                                                                                                                          |
| Contact Person:                    | Christina Flores, RAC<br>Senior Specialist, Regulatory Affairs                                                                                      |
| Date Prepared:                     | April 16, 2018                                                                                                                                      |
| Device Description                 |                                                                                                                                                     |
| Classification Name:               | Electrosurgical cutting and coagulation<br>device and accessories                                                                                   |

21 CFR 878.4400

General & Plastic Surgery

Class II,

GEI, LFL

{4}------------------------------------------------

Trade Name:

## THUNDERBEAT Type S Hand Instruments

| Model Name | Device Name                                         |
|------------|-----------------------------------------------------|
| TB-0520FCS | THUNDERBEAT 5 mm, 20 cm, Front-actuated Grip Type S |
| TB-0535FCS | THUNDERBEAT 5 mm, 35 cm, Front-actuated Grip Type S |
| TB-0545FCS | THUNDERBEAT 5 mm, 45 cm, Front-actuated Grip Type S |

Generic/Common Name:

Ultrasonic and electrosurgical devices

# Predicate Devices

Olympus Medical Systems Corp. THUNDERBEAT TB-0520IC/0535PC/0545PC/0535IC/0545IC Hand Instruments K132703

# Product Description

The THUNDERBEAT Type S Hand Instruments is intended to be used with the Ultrasonic Generator (USG-400), the Electrosurgical Generator (ESG-400), and the THUNDERBEAT Transducer, (TD-TB400).

The THUNDERBEAT Type S Hand Instruments is provided as a sterile, single use device. It is a functional device capable of vessel sealing & cutting, tissue coagulating & cutting, grasping, and dissecting. This device has been designed to seal and cut vessels up to and including 7 mm in diameter, tissue bundles, and lymphatics.

# Comparison of Technological Characteristics

The THUNDERBEAT Type S Hand Instruments have the same intended use and fundamental scientific technological characteristics as the predicate THUNDERBEAT Hand Instruments cleared under K132703. The subject and predicate devices activate combined HF Bipolar (FineCoag) output and Ultrasonic output [Seal & Cut mode] simultaneously while grasping vessels, tissue bundles and lymphatics between the Probe and the Grasping section.

The THUNDERBEAT Type S Hand Instruments, like the predicate THUNDERBEAT, also activates the HF Bipolar (HardCoag) output [Seal mode]. Those outputs can be used for sealing and cutting/sealing of vessels, tissue bundles and lymphatics and or coagulating and cutting/coagulating tissues. The hand switches provided on the grip handle enable those output operations.

{5}------------------------------------------------

The differences between the THUNDERBEAT Type S and the predicate THUNDERBEAT include the following:

- Grasper shape difference in upper jaw .
- Coating on underside of probe tip
- . Front actuation grip on handle

## Material

Full biocompatibility testing on all patient contacting surfaces has been performed in compliance to the relevant requirements of ISO-10993 series.

## Indications for use

The THUNDERBEAT Type S hand instruments are intended to be used with the Ultrasonic Generator (USG-400), the Electrosurgical Generator (ESG-400), and the THUNDERBEAT Transducer (TD-TB400).

## Seal & Cut mode:

The THUNDERBEAT Type S hand instruments when used in combination with the Seal & Cut mode are indicated for open, laparoscopic (including single-site surgery) general surgery and gynecological surgery (including urologic, thoracic, plastic and reconstructive, bowel resections, cholecystectomies, Nissen fundoplication, adhesiolysis, oophorectomy, hysterectomies (both vaginal assisted and abdominal) etc.), and endoscopic surgery or in any procedure in which cutting, vessel ligation (sealing and cutting), coagulation, grasping, and dissection is performed. These devices have been designed to seal and cut vessels (up to and including 7 mm in diameter), tissue bundles, and lymphatics.

Seal mode:

The THUNDERBEAT Type S hand instruments when used in combination with the Seal mode are indicated for open, laparoscopic (including single-site surgery) general surgery and gynecological surgery (including urologic, thoracic, plastic and reconstructive, bowel resections, cholecystectomies, Nissen fundoplication, adhesiolysis, oophorectomy, hysterectomies (both vaginal assisted and abdominal) etc.), and endoscopic surgery or in any procedure in which vessel sealing, coagulation, grasping is performed. These devices have been designed to seal vessels (up to and including 7 mm in diameter), tissue bundles, and lymphatics.

The THUNDERBEAT Type S hand instruments have not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures, and should not be used for these procedures.

{6}------------------------------------------------

## Compliance to Voluntary Standards

The design of the THUNDERBEAT Type S Hand Instruments comply with the following standards:

IEC60601-1: 2005+A1:2012 IEC60601-1-2: 2001+A1, 2007 IEC60601-2-2:2009 IEC60601-2-18:2009 ISO10993-1:2009 ISO10993-5: 2009 ISO10993-7: 2008 ISO10993-10: 2010 ISO10993-11: 2006 ISO11135:2014 ISO14971:2007 ASTM F1980-16

## Device-specific guidance

- Premarket Notification (510(k)) Submissions for Bipolar Electrosurgical Vessel Sealers for General Surgery - Guidance for Industry and Food and Drug Administration Staff, 08/15/2016
- Premarket Notification (510(k)) Submissions for Electrosurgical Devices for General Surgery - Guidance for Industry and Food and Drug Administration Staff, 08/15/2016

## Summary of Sterilization and Shelf Life Discussion

The sterilization has been tested to comply with ISO11135:2014. Accelerated Aging testing confirmed that the subject device has a three-year shelf life.

## Summary of Performance Testing

The following performance testing was conducted in support of the substantial equivalence determination.

| Item                                    | Contents                                                                                                                                                                                                |
|-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <i>Ex-vivo</i> Vessel Burst<br>Pressure | <i>Ex-vivo</i> burst pressure testing of porcine blood vessels (arteries, veins, and tissue bundles) was conducted on both the subject and predicate devices to demonstrate vessel sealing performance. |
| <i>Ex-vivo</i> Cutting<br>Performance   | <i>Ex-vivo</i> Cutting performance (Cutting time) testing of porcine mesentery was conducted on both the subject and predicate devices to demonstrate cutting performance.                              |

#### 1. Bench Testing

{7}------------------------------------------------

| Item                                                | Contents                                                                                                                                                                     |
|-----------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Max Temperature                                     | Max Temperature testing was conducted on both the subject and predicate devices to demonstrate for the inside/outside temperature at the grasping section.                   |
| Grasping force                                      | Grasping force Testing was conducted on both the subject and predicate devices to demonstrate homogeneousness of grasping force across grasping section.                     |
| Durability of the coating on underside of probe tip | Durability testing was conducted to confirm durability of coating on underside of probe tip after device output.                                                             |
| Durability against twisting                         | Durability testing against twisting was conducted on both the subject and predicate devices to demonstrate torque when spark occurs between probe and jaw during activation. |

## 2. Animal Testing

| Item                 | Contents                                                                                                                                                                                                                                                          |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Chronic Animal Study | Chronic animal study of porcine was conducted on both the subject and predicate devices to confirm seal performance, thermal spread and healing degree including vessels up to 7.0mm in diameter, lymphatics and tissue bundles.                                  |
| Acute Animal Study   | Acute animal study of porcine was conducted on both the subject and predicate devices to confirm seal performance, thermal spread and degeneration degree including vessels (artery and artery with vein) up to 7.0mm in diameter, lymphatics and tissue bundles. |

3. Clinical Testing

Clinical testing was not performed.

{8}------------------------------------------------

## Substantial Equivalence

The subject and predicate devices have the same fundamental technology and indications for use. The performance, dimensions and materials of the subject device are similar to those of the predicate. Compared to the predicate device (K132703), the subject device has a different shape in the Grasping section, a coating of insulating material has been added to the underside of the probe, and the handle is now a frontactuated grip. To support the proposed design modifications, the performance tests summarized above were conducted.

## Conclusion:

In summary, the THUNDERBEAT Type S Hand Instruments are substantially equivalent to the predicate devices and present no new questions of safety or efficacy.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K172610](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K172610)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
