K172529 · Aerin Medical, Inc. · GEI · Dec 5, 2017 · General, Plastic Surgery
Device Facts
Record ID
K172529
Device Name
Vivaer ARC Stylus
Applicant
Aerin Medical, Inc.
Product Code
GEI · General, Plastic Surgery
Decision Date
Dec 5, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Vivaer ARC Stylus is indicated for use in otorhinolaryngology (ENT) surgery for the coagulation of soft tissue in the nasal airway, to treat nasal airway obstruction by shrinking submucosal tissue, including cartilage in the internal nasal valve area.
Device Story
Handheld bipolar radiofrequency (RF) probe; used in ENT surgery to treat chronic nasal airway obstruction. Clinician inserts tip into nostril; delivers low-power RF energy to target submucosal tissue/cartilage. Energy generates heat; tissue repositioned via lateral pressure; creates coagulation lesion. Healing process causes tissue retraction and stiffening; shrinks/reshapes tissue to reduce obstruction. Device connects to Aerin Console via flexible cable; includes temperature sensor for real-time monitoring (50-70 °C). Features authentication and use-limit protocols. Benefits include improved nasal breathing and symptom relief.
Clinical Evidence
Prospective, multi-center, non-randomized study (n=50). Primary endpoint: >15-point improvement in NOSE score at 26 weeks. Results: Mean NOSE score decreased 55 points (SD 21) from baseline (p<0.0001). Responder rate (≥15-point improvement) was 94% (95% CI [84, 98]). No device-related adverse events reported. In vivo canine study (up to 24 weeks) confirmed long-term safety and tissue healing without significant fibrosis or cartilage degeneration.
Technological Characteristics
Bipolar RF probe; handle, shaft, and electrode array. Temperature sensor for feedback control (50-70 °C). Biocompatibility per ISO 10993-1. Sterilization per ISO 11135. Electrical safety per IEC 60601-1, IEC 60601-2-2. EMC per IEC 60601-1-2. Packaging per ASTM D4169, D4332, F1980, F2096, F88.
Indications for Use
Indicated for patients with chronic nasal airway obstruction or poor nasal breathing requiring ENT surgery for coagulation of soft tissue and cartilage in the internal nasal valve area.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters FDA in a blue box, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font below.
December 5, 2017
Aerin Medical, Inc. Shannon Scott Director, Regulatory Affairs 232 E. Caribbean Drive Sunnyvale, CA 94089
Re: K172529
Trade/Device Name: Vivaer ARC Stylus Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI Dated: November 2, 2017 Received: November 6, 2017
Dear Shannon Scott:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely yours,
Eric A. Mann -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K172529
# Device Name
Vivaer ARC Stylus
#### Indications for Use (Describe)
The Vivaer ARC Stylus is indicated for use in otorhinolaryngology (ENT) surgery for the coagulation of soft tissue in the nasal airway, to treat nasal airway obstruction by shrinking submucosal tissue, in the internal nasal valve area.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Aerin Medical. The word "Aerin" is in a large, blue, sans-serif font, with the "A" stylized to look like a triangle. Below "Aerin" is the word "MEDICAL" in a smaller, red, sans-serif font. To the right of "Aerin" is a cluster of green circles that fade from dark green to light green.
# 510(k) Summary
This 510(k) Summary is being submitted in accordance with the requirements of 21 CFR 807.92.
## General Information
| Submitter Information | | |
|--------------------------------------|----------------------------------------------------------------------|--|
| Company: | Aerin Medical, Inc. | |
| Submitter's Address: | 232 E. Caribbean Drive<br>Sunnyvale, CA 94089 | |
| Contact Person: | Shannon Scott<br>Director, Regulatory Affairs<br>Phone: 512-221-9956 | |
| Establishment<br>Registration Number | 3011625895 | |
| Date Prepared: | November 2, 2017 | |
| Name of the Device | | |
| Proprietary Name: | Vivaer ARC Stylus, Model FG257 | |
| Common Name: | Radiofrequency probe | |
| Classification Name: | Electrosurgical cutting and coagulation device and<br>accessories | |
| Classification Panel: | General and Plastic Surgery | |
| Device Class: | Class II | |
| Product Code: | GEI | |
| CFR Section: | 21 CFR 878.4400 | |
| Predicate Device: | InSeca ARC Stylus (K162810) | |
## Device Description
The Vivaer ARC Stylus is a handheld bipolar radiofrequency (RF) probe designed for use in otorhinolaryngology (ENT) surgery. The Stylus comprises a handle, shaft and treatment tip. The treatment tip consists of an array of bipolar electrodes and a temperature sensor that allows for monitoring of tissue temperature during RF energy delivery. The Stylus is designed for use with the Aerin Console. It includes features to allow compatibility with and authentication by the Aerin Console. The Stylus connects to the Aerin Console via a flexible cable.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Aerin Medical. The word "Aerin" is in a large, blue, sans-serif font, with the "A" stylized to have a pointed top. To the right of the "n" are several green circles of varying sizes, arranged in a gradient pattern. Below "Aerin" is the word "MEDICAL" in a smaller, sans-serif font, also in blue.
The Vivaer ARC Stylus is used to treat patients experiencing chronic nasal airway obstruction. During a treatment procedure, the clinician inserts the tip of the Vivaer ARC Stylus into a patient's nostril to deliver low power RF energy to the target tissue of the nasal airway. The low-power radiofrequency energy generates heat within the tissue, allowing the tissue to be repositioned by applying lateral pressure, and creates a coagulation lesion. As the lesion heals, the tissue retracts and stiffens. The treatment shrinks and reshapes the tissue to lessen the degree of obstruction.
#### Indications for Use
The Vivaer ARC Stylus is indicated for use in otorhinolaryngology (ENT) surgery for the coagulation of soft tissue in the nasal airway, to treat nasal airway obstruction by shrinking submucosal tissue, including cartilage in the internal nasal valve area.
#### Summary of the technological characteristics of the device compared to the predicate device
The Vivaer ARC Stylus was found to be equivalent to the predicate device in design and intended use to generate and deliver bipolar RF energy to treat tissue in ENT procedures.
| Characteristic | Vivaer ARC Stylus<br>(Model FG257)<br>Subject Device | InSeca ARC Stylus<br>(Model FG256)<br>Predicate Device K162810 |
|-------------------------------|------------------------------------------------------|----------------------------------------------------------------|
| Design configuration | Same | Integrated cable, handle and<br>electrode |
| Energy type | Same | Bipolar radiofrequency |
| Tissue temperature | Same | 50 - 70 °C<br>(temperature controlled) |
| RF generator<br>compatibility | Same | Aerin Console,<br>Model FG226 |
| Use limit feature | Yes | Yes |
| Stylus validation feature | Yes | Yes |
## Summary of non-clinical tests
Device performance testing included system responsiveness and effectiveness for use in a clinical setting. Force load testing was conducted to verify adequate shaft strength. The efficacy of the subject device is supported by thermocouple accuracy and response time testing via water bath immersion and tissue heating time testing. Tissue heating time testing was performed using tissues representative of the soft tissues of the nasal airway. The subject device met all the performance testing requirements. Additional performance testing was performed to verify substantial equivalence to the predicate devices in terms of thermal effects. The testing showed that the Vivaer ARC Stylus creates thermal
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Aerin Medical. The word "Aerin" is in a large, blue, sans-serif font, with the "A" stylized to look like a triangle. Below "Aerin" is the word "MEDICAL" in a smaller, red, sans-serif font. To the right of "Aerin" are several green circles of varying sizes, arranged in a cluster.
lesion volumes that are comparable to volumes created with the predicate device.
Device safety is supported through biocompatibility testing, sterilization and packaging validation activities, electrical safety testing and electromagnetic Biocompatibility of the Vivaer ARC Stylus was compatibility testing. demonstrated through testing performed in accordance with AAMI/ANSI/ISO 10993-1 (2009/(R) 2013). The sterilization was performed in accordance with ANSI/AAMI/ISO 11135:2014 and ISO 10993-7:2008. Electrical safety was demonstrated through testing to IEC 60601-1:2005/A1:2012 and IEC 60601-2-2:2009 and electromagnetic compatibility testing was performed according to IEC 60601-1-2:2007/AC:2010. The Vivaer ARC Stylus packaging system was subjected to transit testing, visual inspection, accelerated aging, gross leak detection and seal strength testing in accordance with the applicable ASTM standards (ASTM D4169-14, ASTM D4332-14, ASTM F1980-07(2011), ASTM F2096-11, and ASTM F88/F88M-15), and met all acceptance criteria. In vivo testing was conducted in canines to evaluate the healing and long-term effects of temperature-controlled bipolar RF energy delivered to nasal tissue. The treatment sites were evaluated endoscopically post-procedure and at several timepoints out to 24 weeks (6 months). The treated tissue was evaluated histologically by a veterinary pathologist after 1. 4 and 24 weeks of healing. The study results at the end of the 6-month follow-up period support the long-term safety of the Vivaer ARC Stylus to treat the submucosal tissue including cartilage. Treatment with the Stylus resulted in a durable change of the submucosal gland population and did not result in significant fibrosis, tissue contraction, or cartilage degeneration after 6 months of healing. At 6 months, the healing process was complete or near-complete.
## Summary of clinical tests
A prospective, multi-center, non-randomized, non-significant risk clinical study was conducted in the United States to confirm safety and effectiveness in a population presenting with a history of poor nasal breathing or chronic nasal obstruction. The primary efficacy endpoint was defined as a statistically significantly greater than 15-point improvement in the mean Nasal Obstruction Symptom Evaluation (NOSE) score from baseline to 26 weeks post-procedure. The secondary efficacy endpoint was defined as a treatment responder rate of 55% or greater, where a responder is defined as achieving a 15 point or greater improvement (decrease) in the NOSE score from baseline to 26 weeks. Safety was assessed by characterizing the type and frequency of adverse events reported throughout the study. Additional evaluations included physical and endoscopic nasal assessment and Patient Satisfaction Survey.
Fifty (50) subjects from 8 study sites were treated in the study. The subjects received bilateral treatment of the internal nasal valve with the Vivaer Stylus. The primary and secondary efficacv endpoints of the study were met and demonstrated significant and clinically meaningful improvement in the nasal
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Aerin Medical. The word "Aerin" is in a large, blue, sans-serif font, with the "A" being significantly larger than the other letters. Below "Aerin" is the word "MEDICAL" in a smaller, gray, sans-serif font. To the right of "Aerin" are several green circles of varying sizes, arranged in a cluster.
obstructions symptoms reported by the subjects.
Significant clinical improvement in symptoms was demonstrated by marked improvement in the NOSE score of nearly all subjects and across the multiple study centers. The mean NOSE score at the 26-week endpoint decreased 55 points (SD 21) from baseline (paired t-test. p<0.0001. lower 95% confidence bound=50). The responder rate, defined as at least a 15-point improvement in the NOSE Score, was 94% (95% CI [84, 98]). Three subjects were nonresponders at the 26-week evaluation. Each of these non-responders had external circumstances that likely contributed to their status at the final evaluation. One subject was lost-to-follow-up for the evaluation at 26 weeks; at previous evaluations up through 18 weeks the subject had significant improvement and was a responder.
Patient-reported opinion and satisfaction with the tolerability, recovery process, and effect of the procedure was high. Physical and endoscopic nasal status assessments showed improvement from baseline condition and a favorable recovery process following the procedure.
Two serious adverse events in 2 subjects were reported. Neither of the events was related to the device or procedure. Both subjects continued to participate in the study and were treatment responders. There were no device-related adverse events and no unanticipated adverse events. The adverse events associated with the procedure were relatively mild, transient, and not unexpected for this type of procedure, demonstrating that the risks associated with the Vivaer ARC Stylus and procedure are minimal.
The results from this study support the safety and efficacy of the proposed indication expansion to include treatment the internal nasal valve area.
## Conclusion
Testing demonstrates that the Vivaer ARC Stylus is substantially equivalent to the predicate in terms of both intended use and delivered RF treatment and is as safe and effective for its intended use.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.