Stryker Integrated Bipolar Cord and Tubing Sets

K172367 · Stryker Instruments · GEI · Aug 22, 2017 · General, Plastic Surgery

Device Facts

Record IDK172367
Device NameStryker Integrated Bipolar Cord and Tubing Sets
ApplicantStryker Instruments
Product CodeGEI · General, Plastic Surgery
Decision DateAug 22, 2017
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Stryker Integrated Bipolar Cord and Tubing Sets are intended to provide irrigation and energy simultaneously to bipolar forceps specifically designed for irrigation.

Device Story

Integrated tubing and bipolar cord sets; interface between irrigation unit/electrosurgical generator and irrigating bipolar forceps. Device delivers irrigation fluid and electrical energy simultaneously to surgical site. Available in gravity-type or rotary-type configurations; includes options for flying leads or unitized leads. Used in clinical settings by surgical staff. Provides irrigation and coagulation capabilities; supports surgical workflow by consolidating connections. Benefits patient by enabling precise electrosurgical coagulation with simultaneous irrigation to manage thermal effects.

Clinical Evidence

Bench testing only. No clinical data. Testing included biocompatibility (ISO 10993-5, -10, -11), mechanical/functional testing (durometer, PSI levels), electrical safety (ANSI/AAMI ES60601-1, IEC 60601-1-2, IEC 60601-2-2), and packaging/sterilization validation (ASTM D4169, F1886, F2096, F88; ISO 11135).

Technological Characteristics

Integrated irrigation tubing (silicone/PVC) and bipolar electrical cord. Sterilized via EtO (10-6 SAL). Electrical safety per ANSI/AAMI ES60601-1, IEC 60601-1-2, and IEC 60601-2-2. Packaging per ASTM D4169, F1886, F2096, F88. Biocompatibility per ISO 10993 series.

Indications for Use

Indicated for use with irrigating bipolar forceps to provide simultaneous irrigation and electrosurgical energy during surgical procedures.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, featuring three human profiles in a row. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 August 22, 2017 Stryker Instruments Ms. Susanne Galin Principal Regulatory Affairs Specialist 4100 E. Milham Avenue Kalamazoo, Michigan 49001 Re: K172367 Trade/Device Name: Stryker Integrated Bipolar Cord and Tubing Sets Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: August 2, 2017 Received: August 4, 2017 Dear Ms. Galin: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Actinclude requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device {1}------------------------------------------------ related adverse events) (21 CFR 803): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, Jennifer R. Stevenson -S3 For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K172367 Device Name Stryker Integrated Bipolar Cord and Tubing Sets Indications for Use (Describe) The Stryker Integrated Bipolar Cord and Tubing Sets are intended to provide irrigation and energy simultaneously to bipolar forceps specifically designed for irrigation. Type of Use (*Select one or both, as applicable*) | <span></span> Residential Real Estate (CRE) Sales or Financing | <span></span> Other Transaction Not Involving CRE | |----------------------------------------------------------------|---------------------------------------------------| |----------------------------------------------------------------|---------------------------------------------------| X | Prescription Use (Part 21 CFR 801 Subpart D) | | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(K) SUMMARY In accordance with 21 CFR 807.87(h) and (21 CFR 807.92) the 510(k) Summary for the Stryker Integrated Bipolar Cord and Tubing Sets is provided below. | Device Common Name: | Electrosurgical, Cutting & Coagulation & Accessories | |----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Trade Name: | Stryker Integrated Bipolar Cord and Tubing Sets | | Applicant: | Stryker Instruments<br>4100 E. Milham Avenue<br>Kalamazoo, MI 49001 | | Contact: | Susanne Galin<br>Principal Regulatory Affairs Specialist<br>Stryker Instruments<br>4100 E. Milham Avenue<br>Kalamazoo, MI 49001<br>Desk Phone +1 269 389 4132<br>Cell Phone +1 812 345 7917<br>Email: susanne.galin@stryker.com | | Date Prepared: | August 21, 2017 | | Classification Regulation: | 21 CFR §878.4400- Electrosurgical, Cutting &<br>Coagulation & Accessories, Class II | | Panel: | General & Plastic Surgery | | Product Code: | GEI - Electrosurgical, Cutting & Coagulation &<br>Accessories | | Predicate Device: | K890648, Malis Bipolar Cord/Irrigation Tubing Set | ### Indication for Use: The Stryker Integrated Bipolar Cord and Tubing Sets are intended to provide irrigation and energy simultaneously to bipolar forceps specifically designed for irrigation. ### Device Description: The subject devices are the Integrated Tubing and Bipolar Cord Sets. They are integrated irrigation tubing and electrical cord sets to be used with irrigating bipolar forceps, and interface with both the irrigation unit and the electrosurgical generator. There are two types of sets available from Stryker: one type that is for use with a rotary irrigator unit and one type that is for use with a gravity irrigation unit. Additionally, each type is available with either "flying leads" or a "unitized lead". {4}------------------------------------------------ | Part Number | Item Name | Description | |--------------|-----------------------------------------------------------------------|------------------------------| | 6790-100-001 | Integrated Tubing and Bipolar Cord Set,<br>Disposable | Gravity-type, flying lead. | | 6790-100-002 | Integrated Tubing and Bipolar Cord Set,<br>Disposable, Unitized Lead | Gravity-type, unitized lead. | | 6790-100-003 | Bipolar Irrigator Integrated Tubing Set,<br>Disposable | Rotary-type, flying lead. | | 6790-100-004 | Bipolar Irrigator Integrated Tubing Set,<br>Disposable, Unitized Plug | Rotary-type, unitized lead. | ## Listing of Subject Devices # Device Comparison Table: | Technological<br>Characteristic | Integrated Tubing &<br>Bipolar Cord Set (as<br>cleared in K890648) | Currently Marketed Gravity Tubing & Bipolar<br>Cord Set | | |-----------------------------------|--------------------------------------------------------------------|---------------------------------------------------------|------------------------------------------------| | | | Gravity-type | Rotary-type | | Electrical Cord<br>Length | 12 ft bipolar electrical cord | 10.958 ft ± 0.375 ft | 10.958 ft ± 0.375 ft | | Lead type | flying leads | flying leads and unitized<br>leads | flying leads and unitized<br>leads | | Electrical Cord<br>Connector | Female connector which<br>connects to forceps | Female connector which<br>connects to forceps | Female connector which<br>connects to forceps | | Irrigation<br>Tubing Length | Total Length: 15 ft<br>irrigation tubing | Total length: 15.833 ft<br>(nominal) | Total Length: 15.583 ft<br>(nominal) | | | 8.5 inch section serves as<br>pump chamber | 8.5 inch serves as pump<br>chamber | 5.5 inch serves as pump<br>chamber | | Irrigation<br>Tubing<br>Connector | Spike with drip chamber | Spike with no drip<br>chamber | Spike with no drip<br>chamber | | | Male Luer connects to<br>forceps (tubing) | Male Luer connects to<br>forceps (tubing) | Male Luer connects to<br>forceps (tubing) | | Materials | Silicone tubing – Pump<br>PVC tubing - bag | Silicone tubing – Pump<br>PVC tubing - Bag | Silicone tubing – Pump<br>PVC tubing - Bag | | Silicone tubing<br>wall thickness | Not provided in K890648 | OD: 0.167" – 0.173"<br>ID: 0.117" - 0.123" | OD: 0.117 (+0.008/-0.0)"<br>ID: 0.028"- 0.035" | {5}------------------------------------------------ | Technological<br>Characteristic | Integrated Tubing &<br>Bipolar Cord Set (as<br>cleared in K890648) | Currently Marketed Gravity Tubing & Bipolar<br>Cord Set | | |---------------------------------|-------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------| | | | Gravity-type | Rotary-type | | Packaging | Peel pouch, Mylar to Tyvek | Poly bag inside a peel-<br>pouch (60GA BIAX<br>Nylon/2.0 MIL LDPE,<br>Uncoated Tyvek 1073B) | Poly bag inside a peel-<br>pouch (60GA BIAX<br>Nylon/2.0 MIL LDPE,<br>Uncoated Tyvek 1073B) | | Sterilization<br>Method | EtO | EtO | EtO | | Sterilization<br>Parameters | 120° F Temperature<br>50° F Relative Humidity<br>750 Mgs. Per liter of Pure<br>EtO<br>3 hr Gas Exposure<br>2 Air washes | 130° F Temperature<br>Humidity Dwell: 30 min at<br>4.0"HgA<br>790 Mg/L 100% EtO<br>6 hrs 1 min gas exposure<br>2 Air/nitrogen/steam<br>washes | 130° F Temperature<br>Humidity Dwell: 30 min<br>at 4.0"HgA<br>790 Mg/L 100% EtO<br>6 hrs 1 min gas exposure<br>2 Air/nitrogen/steam<br>washes | | SAL | 10-6 SAL | 10-6 SAL | 10-6 SAL | | Shelf Life | Not provided in K890648 | 3 years | 3 years | ## Performance Data: The following summary of V&V Testing is provided to establish substantial equivalence. | V&V Testing Summary | | |---------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------| | Biocompatibility | | | Cytotoxicity (MEM<br>Elution) | ISO 10993-5:2009 Biological Evaluation of Medical Devices, Part 5:<br>Tests for In Vitro Cytotoxicity | | Systemic Toxicity | ISO 10993-11: 2006 Biological Evaluation of Medical Devices Part 11:<br>Tests for Systemic Toxicity | | Material Mediated<br>Pyrogenicity | ISO 10993-11:2006, Biological Evaluation of Medical Devices, Part 11:<br>Tests for Systemic Toxicity. | | Irritation/Intracutaneous<br>Reactivity | ISO 10993-10: 2010 Standard, Biological Evaluation of<br>Medical Devices, Part 10: Tests for Irritation and Skin Sensitization. | | Sensitization (Guinea Pig<br>Maximization) | ISO 10993-10:2010. Biological Evaluation of Medical Devices, Part 10:<br>Tests for Irritation and Skin Sensitization. | | Cytotoxicity (MEM<br>Elution) after 2X<br>sterilization | ISO 10993-5:2009 Biological Evaluation of Medical Devices, Part 5:<br>Tests for In Vitro Cytotoxicity | {6}------------------------------------------------ | Mechanical and Performance Testing | | |------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Functional Testing of<br>Cord & Tube Set | Functional Testing of Cord & Tube Set after Packaging Testing<br>Testing conducted on: | | | • Unaged Samples | | | • Samples Accelerated Aged to 1.5 years | | | • Samples Accelerated Aged to 3 years | | | Cord Set Durometer Testing | | | Verification PSI Level Testing for Disposable Cable with Tubing, Rotary | | Electrical Safety Testing | | | Electrical Safety Test<br>Report | ANSI/AAMI ES60601-1:2005, Medical Electrical Equipment – Part 1:<br>General Requirements for Basic Safety and Essential Performance;<br>Consolidated Reprint (2009); Amendment 2 (2010) | | | ANSI/AAMI IEC 60601-1-2:2007(R)2012, Medical Electrical Equipment –<br>Part 1-2: General Requirements for Safety and Essential Performance –<br>Collateral Standard : Electromagnetic Compatibility – Requirements and<br>Tests (Edition 3) | | | ANSI/AAMI IEC 60601-2-2:2009, Medical Electrical Equipment – Part<br>2-2: Particular Requirements for the Basic Safety and Essential<br>Performance of High Frequency Surgery Equipment and High Frequency<br>Surgical Accessories | | Sterilization Validation | | | Sterilization Validation | ANSI / AAMI / ISO 11135: 2014 Sterilization of health care products–<br>Ethylene Oxide-Requirements for the development, validation and routine<br>control of a sterilization process for medical devices. | | Packaging Design Verification Testing | | | T=0 Design Verification | ASTM D4169, Standard Test Method for<br>Performance Testing of Shipping Containers and Systems | | T=1.5 Year Accelerated<br>Aging 60C | ASTM F1886/F1886M, Standard Test Method for Determining Integrity<br>of Seals for Flexible Packaging by Visual Inspection | | T=3 Year Accelerated<br>Aging 60C | ASTM F2096, Standard Test Method for Detecting<br>Gross Leaks in Porous Medical Packaging by Internal Pressurization | | | ASTM F88-15, Standard Test Method for Seal Strength of Flexible<br>Barrier Materials | {7}------------------------------------------------ ### Substantial Equivalence Conclusion: As shown in the table above and evidenced by the Verification and Validation Activities described in the submission, the differences in technological characteristics between the currently marketed Integrated Tubing and Cord Sets and the versions cleared in K890648 do not raise new questions of safety and effectiveness, and can be found substantially equivalent.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...