← Product Code [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI) · K171593

# InMode RF System (K171593)

_Inmode MD , Ltd. · GEI · Oct 10, 2017 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K171593

## Device Facts

- **Applicant:** Inmode MD , Ltd.
- **Product Code:** [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI.md)
- **Decision Date:** Oct 10, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The InMode RF System is indicated for use in dermatological and general surgical procedures for electro-coagulation and hemostasis.

## Device Story

InMode RF System is a computerized electrosurgical device for electro-coagulation and hemostasis. It consists of an AC/DC power supply, RF generator, controller, touch screen interface, and handpieces with internal/external electrodes. The device delivers monopolar RF energy to target tissue. An integral feedback mechanism continuously monitors tissue temperature and impedance, automatically adjusting energy delivery to maintain safe, effective heating. Operated by clinicians in surgical or dermatological settings, the system uses a foot switch for activation. The clinician selects and adjusts treatment settings via the control panel. By providing controlled thermal effects, the device enables precise coagulation and hemostasis, potentially improving procedural outcomes in dermatological and general surgery.

## Clinical Evidence

No clinical data. Evidence consists of bench performance testing (output waveform, power density, and intensity comparisons) and an ex-vivo study on porcine tissue (muscle, liver, fat) using TTC staining to evaluate thermal effects.

## Technological Characteristics

Electrosurgical unit; 1 MHz RF output frequency; 40W max power. Features integral temperature and impedance feedback. Components: AC/DC power supply, RF generator, touch screen interface, foot switch, and disposable electrodes. Sterilization via radiation (ISO 11137). Complies with AAMI/ANSI 60601-1, IEC 60601-1-2, and IEC 60601-2-2.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- InMode RF System ([K160193](/device/K160193.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference. Inside the circle is a stylized emblem featuring three human profiles facing to the right, with a design element above them that resembles a bird in flight or a stylized wing.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 10, 2017

InMode MD Ltd c/o Amit Goren, Ph.D. A. Stein - Regulatory Affairs Consulting, Ltd. 20 Hata'as Str.,Suite 102 Kfar Saba, Israel 4442520

Re: K171593

Trade/Device Name: InMode RF System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: September 14, 2017 Received: Sept. 18, 2017

Dear Dr. Goren:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice

{1}------------------------------------------------

requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Jennifer R. Stevenson -S3

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

510(k) Number (if known)

K171593

Device Name

InMode RF system

Indications for Use (Describe)

The InMode RF System is indicated for use in dermatological and general surgical procedures for electro-coagulation and hemostasis.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{3}------------------------------------------------

## 510(K) SUMMARY

# INMODE RF SYSTEM

# 510(k) Number K171593

## Applicant Name:

| Company Name: | InMode MD Ltd                                                    |
|---------------|------------------------------------------------------------------|
| Address:      | Tabor Building, Shaar Yokneam<br>POB 44, Yokneam 20692<br>Israel |
| Tel:          | +972-4-9097470                                                   |
| Fax:          | +972-4-9097471                                                   |
| E-mail:       | amit@asteinrac.com; ahava@asteinrac.com                          |

## Contact Person:

| Official Correspondent: | Amit Goren                                              |
|-------------------------|---------------------------------------------------------|
| Company Name:           | A. Stein – Regulatory Affairs Consulting Ltd.           |
| Address:                | 20 Hata'as Str., Suite 102<br>Kfar Saba 4442520 Israel  |
|                         | Tel: +972-9-7670002                                     |
|                         | Fax: +972-9-7668534                                     |
|                         | E-mail: amit@asteinrac.com                              |
| Date Prepared:          | October 08, 2017                                        |
| Trade Name:             | InMode RF System                                        |
| Classification Name:    | CFR Classification section 878.4400; (Product code GEI) |
| Classification:         | Class II Medical Device                                 |

## Predicate Device:

InMode RF System is substantially equivalent to the previously cleared, InMode RF System, also manufactured by InMode MD Ltd.

| Device           | Manufacturer   | 510(k) No. |
|------------------|----------------|------------|
| InMode RF System | InMode MD Ltd. | K160193    |

{4}------------------------------------------------

## Device Description:

The InMode RF System (InMode MD Ltd.) is a computerized system generating RF energy with integral temperature and impedance feedback mechanism for procedures requiring electrocoagulation and hemostasis. The InMode RF System constantly monitors the temperature and impedance of the target treatment tissue, automatically adjusting energy delivery to maintain effective and safe tissue heating.

The InMode RF System consists of an AC/DC power supply unit, RF generator, controller and user interface including touch screen. The RF hand piece is connected to the console via a cable and a foot switch activates the energy delivery to the hand piece. The hand piece is comprised of a disposable, single use; internal and external electrodes.

The InMode RF System is compatible with the following hand pieces:

- HP101306A (optional sterile/non-sterile)
- HP172206A (optional sterile/non-sterile ●
- HP172246A (non-sterile) ●

## Device Specifications:

| Main Line Frequency (nominal)          | 50 - 60 Hz             |
|----------------------------------------|------------------------|
| Input Voltage (nominal)                | 100 - 240 VAC          |
| Electrosurgical Unit dimensions (inch) | 14.2"W x 18.2"D x 40"H |
| Platform weight (lb.)                  | 33                     |
| RF Max Output Power (Watt)             | 40                     |
| RF Output Frequency (MHz)              | $1 \pm 2%$             |

### Intended Use/Indication for Use:

The InMode RF System is indicated for use in dermatological and general surgical procedures for electrocoagulation and hemostasis.

### Performance Standards:

InMode RF System complies with the following FDA recognized consensus standards:

- AAMI/ANSI 60601-1 (2012), Medical Electrical Equipment Part 1: General . Requirements For Basic Safety And Essential Performance (IEC 60601-1:2005, Mod).
- . IEC 60601-1-2 (Edition 3.0, 2007), Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements And Tests.

{5}------------------------------------------------

- . IEC 60601-2-2 Edition 5.0 2009-02. Medical Electrical Equipment - Part 2-2: Particular Requirements For The Basic Safety And Essential Performance Of High Frequency Surgical Equipment And High Frequency Surgical Accessories.
- AAMI / ANSI / ISO 11137-1:2006/(R)2010, Sterilization Of Health Care ● Products - Radiation - Part 1: Requirements For Development, Validation, And Routine Control Of A Sterilization Process For Medical Devices
- . AAMI / ANSI / ISO 11137-2:2013, Sterilization Of Health Care Products -Radiation - Part 2: Establishing The Sterilization Dose.
- AAMI / ANSI / ISO 11737-1:2006 (R) 2011, Sterilization Of Health Care ● Products - Microbiological Methods - Part 1: Determination Of The Population Of Microorganisms On Product.
- AAMI / ANSI / ISO 11737-2:2009/(R) 2014, Sterilization Of Medical Devices -. Microbiological Methods - Part 2: Tests Of Sterility Performed In The Definition, Validation And Maintenance Of A Sterilization Process.
- AAMI / ANSI / ISO 11607-1:2006/(R)2010, Packaging For Terminally . Sterilized Medical Devices - Part 1: Requirements For Materials, Sterile Barrier Systems And Packaging Systems
- ASTM F1980-07 (Reapproved 2011). Standard Guide For Accelerated Aging Of . Sterile Barrier Systems For Medical Devices.

## Non-Clinical (Bench) Performance Data:

The InMode RF System with the HP172246A hand piece was utilized in bench performance tests to evaluate the system performance specifications. A side by side comparison tests were performed utilizing the HP172246A and the HP172206A hand pieces to evaluate the system output measurements, such as; output waveform at the rated load, identifying the associated mode, amplitude, frequency, duty cycle, load used, and crest factor, power output at maximum and half-of-maximum intensity over the range of expected loads, and compared measured output power density. The results of the bench tests demonstrate that the InMode RF System with the HP172246A hand piece operates within the predefined system performance requirements, exerting similar output power density as to the predicate device (K160193).

## Animal Performance Data / Histology Data:

The thermal effects of the InMode RF System with the HP172246A hand piece on the target tissue were evaluated in an ex-vivo study. The study was conducted on three different porcine tissue models (muscle, liver & fat) and included a single RF treatment followed by TTC staining analysis. The ex-vivo study results show that the InMode RF System with the HP172246A hand piece is safe for use and effective in achieving the specified indications of dermatological and general electrocoagulation and hemostasis.

{6}------------------------------------------------

## Clinical Performance Data:

Not Applicable

## Substantial Equivalence:

The subject of this 510(k) pre-market notification is the InMode RF System. It is composed of the same main device components as its predicate device the InMode RF System FDA cleared under 510(k) file No.K160193 accept for the addition of modified hand piece version HP172246A. The modified HP is emits a maximal RF output power of 40W. The components of modified InMode RF System, similarly to cleared, predicate device, generate its mechanism of operation using the same underlying technology for the same intended use. Delivery of monopolar RF energy through each specific hand piece is monitored and controlled by the device ESU. The user interface control panel provides the user with the optimal treatment settings. Using both modified and cleared devices the device user can decide on the optimal treatment settings and adjust these treatment settings through the control panel. Furthermore, the modified device, as the cleared device, introduces similar safety features and comply with same relevant consensus standards.

The device modifications were evaluated under design control activities and in the frame of conformity with relevant consensus standards and all potential hazards were mitigated in a set of performance activities. Performance bench and ex-vivo tests results show that the modified device was able to produce and deliver the desired RF energy according to the design requirements, which are comparable with the predicate device. Labeling material was revised to support the device modifications. All performance activates show that the modifications made to the cleared device do not pose any new safety and effectiveness concerns.

## Conclusions:

Based on the comparison to predicate device and on the performance testing, the InMode RF System is substantially equivalent to the predicate InMode RF System for the mentioned indication for use.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K171593](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K171593)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
