← Product Code [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI) · K170296

# CoreCath 2.7S (K170296)

_Medtronic Advanced Energy · GEI · Mar 20, 2017 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K170296

## Device Facts

- **Applicant:** Medtronic Advanced Energy
- **Product Code:** [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI.md)
- **Decision Date:** Mar 20, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

The CoreCath 2.7S is a single use electrosurgical instrument designed to be used with flexible bronchoscopes and qualified electrosurgical units. It is indicated for cutting of soft tissue obstructions in upper airways and tracheobronchial tree, providing electrosurgical hemostasis, and evacuation of surgical smoke during such procedures.

## Device Story

CoreCath 2.7S is a disposable, single-use monopolar RF electrosurgical instrument; designed for use with flexible bronchoscopes and Medtronic AEX Generator. Device features a flexible multi-lumen shaft with a distal electrode; connects to generator output port; activated via footswitch. During bronchoscopic procedures, device delivers RF energy to ablate/cut soft tissue obstructions and provide hemostasis; integrated suction channel evacuates surgical smoke. Used by physicians in clinical settings; provides real-time tissue management. Benefits include improved visualization and procedural efficiency through simultaneous cutting, coagulation, and smoke evacuation.

## Clinical Evidence

No clinical data. Evidence consists of bench testing (sterilization, transit, shelf life, biocompatibility, electrical safety, EMC, endoscope compatibility, mechanical, activation, and suction testing) and comparative performance testing in in-vivo animal and ex-vivo tissue models.

## Technological Characteristics

Monopolar RF electrosurgical instrument; polymeric flexible multi-lumen shaft; 750 mm active length; 2.68 mm max diameter. Compatible with 2.8 mm bronchoscope working channels. Energy source: Medtronic AEX Generator (K143175). Sterilization: ETO. Features integrated suction for smoke evacuation.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- EC 2.7 Endoscopic Cutter ([K120909](/device/K120909.md))

## Reference Devices

- Medtronic Advanced Energy AEX Generator ([K143175](/device/K143175.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 20, 2017

Medtronic Advanced Energy Ms. Lydia Sakakeeny Principal Regulatory Affairs Specialist 180 International Drive Portsmouth, New Hampshire 03801

Re: K170296

Trade/Device Name: Corecath 2.7s Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: January 27, 2017 Received: January 31, 2017

Dear Ms. Sakakeeny:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

Jennifer R. Stevenson -S

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K170296

Device Name

CoreCath 2.7S

### Indications for Use (Describe)

The CoreCath 2.7S is a single use electrosurgical instrument designed to be used with flexible bronchoscopes and qualified electrosurgical units. It is indicated for cutting of soft tissue obstructions in upper airways and tracheobronchial tree, providing electrosurgical hemostasis, and evacuation of surgical smoke during such procedures.

| Type of Use (Select one or both, as applicable)                                                              |                                                                                                             |
|--------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|
| <div> <span style="text-decoration: overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)  </div> | <div> <span style="text-decoration: overline;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)  </div> |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# CoreCath2.7S

## Submitter: Medtronic Advanced Energy 180 International Drive Portsmouth, NH 03801 Contact Person: Lydia Sakakeeny, PhD Principal Requlatory Affairs Specialist Phone: (603) 294-5482 Fax: (603) 742-1488 E-mail: lydia.sakakeeny@medtronic.com Date Summary Prepared: January 27, 2017 Device Trade Name: CoreCath 2.7S (K170296) Common Name: Electrosurgical Instrument Classification Name: Electrosurgical cutting and coagulation device and accessories (21 CFR 878.4400) Product Code: GEI Predicate Device: EC 2.7 Endoscopic Cutter (K120909) DeviceDescription: The CoreCath 2.7S is a disposable, single-use monopolar RF (radiofrequency) surgical device intended to be use for ablation and hemostasis of soft tissue obstructions in upper airways and tracheobronchialtree. The CoreCath 2.7S is terminally sterilized via ETO. It is comprised of a cylindrically shaped, disposable monopolar cutting and coagulation electrode affixed to the distal end of a polymeric. flexible multi lumen shaft and cabling, intended for connection to the monopolar output port of the Medtronic Advanced Energy AEX Generator (K143175). The CoreCath 2.7S is intended to be energized for use by the footswitch accessory supplied with the AEX Generator. It has an active length of approximately 750 mm and a maximum diameter crossing profile of 2.68 mm. It is sized such that may be delivered through the working channel of a flexible bronchoscope 2.8 mm diameter working channel and working length of 600 mm. Indications for use: The CoreCath 2.7S is a single use electrosurgical instrument designed to be used with flexible bronchoscopes and qualified electrosurgical units. It is indicated for cutting of soft tissue obstructions in upper airways and tracheobronchial tree, providing electrosurgical hemostasis, and evacuation of surgical smoke during such procedures.

510(k) Summary (K170296)

### Technoloqical Characteristics:

The CoreCath 2.7S is similar to the predicate device in that they are single use, sterile, electrosurgical instruments used to cut and coagulate soft tissue obstructions in upper airways and tracheobronchial tree, while providing electrosurgical hemostasis. In addition, the subject device includes a suction feature enabling the suction of smoke during the

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the word "Medtronic" in a bold, sans-serif font. The color of the text is a dark blue. The letters are closely spaced together, giving the word a solid and unified appearance. The overall impression is one of a professional and established brand.

CoreCath 2.7S

indicated procedures. This additional feature does not impact the safety or effectiveness of the device.

### Summary of Non-Clinical Testing:

The design and performance of the modified CoreCath 2.7S were verified and validated through bench testing.

Bench testing included the following:

- Sterilization ●
- Transit
- Shelf Life
- Biocompatibility ●
- Electrical Safety ●
- EMC .
- Endoscope Compatibility ●
- Mechanical Testing ●
- Activation Testing .
- Suction Testing ●

Comparative performance testing was conducted in an in-vivo animal model. Ex-vivo tissue testing was also completed. Thermal effect of the CoreCath 2.7S in both the in-vivo and ex-vivo model was substantially equivalent to that of the predicate device.

Summary of Clinical Tests: Clinical testing was not required for this product.

Conclusion: The indications for use, technology and performance characteristics of the CoreCath 2.7S device are equivalent to the predicate device and therefore Medtronic Advanced Energy claims substantial equivalence to the predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K170296](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K170296)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
