← Product Code [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI) · K170107

# Surgi Max Ultra (K170107)

_Elliquence, LLC · GEI · May 3, 2017 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K170107

## Device Facts

- **Applicant:** Elliquence, LLC
- **Product Code:** [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI.md)
- **Decision Date:** May 3, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

Orthopedic, arthroscopic, spinal, and neurological For resection, ablation, and coagulation of soft tissues and hemostasis of blood vessels in orthopedic, arthroscopic, spinal and neurological procedures. For soft tissue resection and ablation during arthroscopic surgical procedures of knee, shoulder, ankle, elbow, hip and wrist. Cutting Snoring, Submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control, epistaxis treatment, and turbinate shrinkage, skin incisions, biopsy, cysts, abscesses, tumors, cosmetic repairs, development of skin flaps, skin tags, Blepharoplasty. Blended Cutting and Coagulation Snoring, Submucosal palatal shrinkage, traditional uvulopalatoplasty (RAP), myringotomy with effective hemorrhage control, epistaxis treatment, and turbinate shrinkage, skin tags, papilloma Keloids, Keratosis, Verrucae, Basal Cell Carcinoma, Nevi, Fistulas, Epithelioma, Cosmetic Repairs, Cysts, Abscesses, Development of skin flaps. Hemostasis and Nonablative Coagulation Control of bleeding, Epilation, Telangiectasia Bipolar Pinpoint, Precise Coagulation, Pinpoint Hemostasis, in any field (wet or dry), snoring, sub-mucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control, epistaxis treatment, and turbinate shrinkage.

## Device Story

Surgi-Max Ultra is a high-frequency RF electrosurgical generator; provides monopolar and bipolar energy for soft tissue cutting, ablation, and coagulation. Device is a modification of Surgi-Max/Surgi-Max Plus (K100390); features updated internal electrical components for improved efficiency; maximum power output increased from 120W to 170W. Used in OR/clinical settings by physicians; operated via footswitch or fingerswitch. Input is electrical power; output is high-frequency RF energy delivered through compatible electrosurgical accessories. Healthcare providers use device to achieve precise hemostasis or tissue resection; benefits include effective hemorrhage control and surgical versatility across orthopedic, spinal, and ENT procedures.

## Clinical Evidence

No clinical data. Substantial equivalence demonstrated via bench testing, including electrical safety (AAMI ANSI ES60601-1, IEC 60601-1-2, IEC 60601-2-2), usability assessment (IEC 62366), software verification/validation (IEC 62304), energy output verification, and ex-vivo thermal effect comparison on bovine tissue.

## Technological Characteristics

High-frequency RF electrosurgical generator; 4 MHz (monopolar) and 1.7 MHz (bipolar) output frequency. Power output: 170W. Applied part type: CF. Connectivity: Standalone. Power input: 100-240V AC, 50/60Hz. Duty cycle: 10s ON/30s OFF. Complies with AAMI ANSI ES60601-1, IEC 60601-1-2, and IEC 60601-2-2.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- Surgi-Max® / Surgi-Max® Plus ([K100390](/device/K100390.md))
- Bovie® IDS-310 Electrosurgical Generator ([K134054](/device/K134054.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image shows the address of the Food and Drug Administration. The address is 10903 New Hampshire Avenue, Document Control Center - WO66-G609, Silver Spring, MD 20993-0002. The text is left-aligned and in a simple font.

May 3, 2017

Elliquence, LLC Mr. Paul Buhrke IV QA/RA Manager 2455 Grand Avenue Baldwin, New York 11510

Re: K170107

Trade/Device Name: Surgi-Max® Ultra (IEC6-SP170, IEC6-SU170, & IEC6-SV170) Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: April 03, 2017 Received: April 04, 2017

Dear Mr. Buhrke:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Actinclude requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Jennifer R. Stevenson -S

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### Indications for Use

510(k) Number (if known) K170107

Device Name

Surgi-Max Ultra (IEC6-SU170, IEC6-SV170, IEC6-SP170)

Indications for Use (Describe)

Orthopedic, arthroscopic, spinal, and neurological

For resection, ablation, and coagulation of soft tissues and hemostasis in orthopedic, arthroscopic, spinal and neurological procedures. For soft tissue resection and ablation during arthroscopic surgical procedures of knee, shoulder, ankle, elbow, hip and wrist.

#### Cutting

Snoring, Submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control, epistaxis treatment, and turbinate shrinkage, skin incisions, biopsy, cysts, abscesses, tumors, cosmetic repairs, development of skin flaps, skin tags, Blepharoplasty.

Blended Cutting and Coagulation

Snoring, Submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control, epistaxis treatment, and turbinage, skin tags, papilloma Keloids, Keratosis, Verrucae, Basal Cell Carcinoma. Nevi, Fistulas, Epithelioma, Cosmetic Repairs, Cysts, Abscesses, Development of skin flaps.

Hemostasis and Nonablative Coagulation Control of bleeding, Epilation, Telangiectasia

Bipolar

Pinpoint, Precise Coagulation, Pinpoint Hemostasis, in any field (wet or dry), snoring, sub-mucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control, epistaxis treatment, and turbinate shrinkage.

Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image is a logo of a person. The person is drawn in white lines on a dark blue background. The person is facing away from the viewer and has their right arm raised. The logo is circular in shape and has a glossy appearance.

# 510 (k) Summary

(As required by 21 CFR 807.92(a))

Date Prepared

May 1, 2017

# Submitter's Information (807.92(a)(1))

Company Name and Address:

Elliquence, LLC 2455 Grand Avenue Baldwin, NY 11510 Phone: (516) 277-9000 Fax: (516) 277-9001 www.elliquence.com

Establishment Registration for Elliquence, LLC is 3007024186.

Contact Information:

Daniel Connell - Requlatory Affairs Associate Phone: (516) 277-9036 Fax: (516) 277-9011 Email: dconnell@elliquence.com

Paul D. Buhrke IV- QA/RA Manager Phone: (516) 277-9010 Fax: (516) 277-9011 Email: pbuhrke@elliquence.com

# Device Information (807.92(a)(2))

Trade Name Surgi-Max® Ultra (IEC6-SP170, IEC6-SU170, & IEC6-SV170)

Common/Usual Name Electrosurgical unit and accessories, electrosurgical cutting and coagulation device and accessories

> elliquence, LLC. 2455 Grand Avenue · Baldwin, New York 11510 (516) 277.9000 · Fax: (516) 277.9001 · www.elliquence.com

{4}------------------------------------------------

Classification Name and Regulation Electrosurgical Cutting and Coagulation Device and Accessories; 21 CFR 878.4400

K170107 Page 2 of 5

Class FDA Classification: Class II FDA Product Code: GEI

#### Predicate Devices (807.92(a)(3))

- Surgi-Max® / Surgi-Max® Plus (K100390) ●
- . Bovie® IDS-310 Electrosurgical Generator (K134054)

#### Device Description (807.92(a)(4))

The elliquence Surgi-Max® Ultra (IEC6-SP170, IEC6-SU170, & IEC6-SV170) is a modification to the existing Surgi-Max® / Surgi-Max® Plus (K100390) line of radiofrequency (RF) electrosurgical generators. It is a compact source of high-frequency RF energy employed for cutting and coagulation of soft tissue in a variety of surgical procedures. It is designed to enable the performance of monopolar and bipolar functions in conjunction with separatelymarketed electrosurgical accessories and ancillary equipment which have been developed by elliquence for mutually-exclusive use with the Surgi-Max® line of generators.

The Surgi-Max® Ultra (IEC6-SP170, IEC6-SU170, & IEC6-SV170) is the result of an update to internal electrical components of the Surgi-Max® / Surgi-Max® Plus. As a result of this equipment update, device design as it relates to internal electrical system framework was simplified in many respects and overall efficiency was greatly improved upon. This increase in efficiency enabled an increase in the maximum power output of the device from 120W to 170W. Aside from this increase in maximum power output, all other operational aspects of the device were conserved. The only other modifications made to the device were aesthetic in nature.

The modifications enacted do not affect the device's intended use or alter its fundamental scientific technology.

{5}------------------------------------------------

#### Intended Use (807.92(a)(5))

#### Orthopedic, arthroscopic, spinal, and neurological

For resection, ablation, and coagulation of soft tissues and hemostasis of blood vessels in orthopedic, arthroscopic, spinal and neurological procedures. For soft tissue resection and ablation during arthroscopic surgical procedures of knee, shoulder, ankle, elbow, hip and wrist.

#### Cutting

Snoring, Submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control, epistaxis treatment, and turbinate shrinkage, skin incisions, biopsy, cysts, abscesses, tumors, cosmetic repairs, development of skin flaps, skin tags, Blepharoplasty.

#### Blended Cutting and Coagulation

Snoring, Submucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control, epistaxis treatment, and turbinate shrinkage, skin tags, papilloma Keloids, Keratosis, Verrucae, Basal Cell Carcinoma, Nevi, Fistulas, Epithelioma, Cosmetic Repairs, Cysts, Abscesses, Development of skin flaps.

### Hemostasis and Nonablative Coagulation

Control of bleeding, Epilation, Telangiectasia

### Bipolar

Pinpoint, Precise Coagulation, Pinpoint Hemostasis, in any field (wet or dry), snoring, sub-mucosal palatal shrinkage, traditional uvulopalatoplasty (RAUP), myringotomy with effective hemorrhage control, epistaxis treatment, and turbinate shrinkage.

### Substantial Equivalence Comparison (807.92(a)(6))

The Surgi-Max® Ultra is substantially equivalent in both intended use and technological characteristics to the Surgi-Max® / Surgi-Max® Plus (K100390) as well as the Bovie® IDS-310 Electrosurgical Generator (K134054). An extensive comparison chart concerning key characteristics is provided on the following pages (Table 5.1).

{6}------------------------------------------------

# Table 5.1 Substantial Equivalence Comparison Chart

|                                               | Surgi-Max® Ultra<br>(K170107) - SUBJECT<br>DEVICE                                            | Surgi-Max® / Surgi-Max<br>Plus (K100390) –<br>PREDICATE                                      | Bovie® IDS-310<br>Electrosurgical Generator<br>(K134054) - PREDICATE                                                                                                                      |
|-----------------------------------------------|----------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer                                  | elliquence, LLC                                                                              | elliquence, LLC                                                                              | Bovie Medical Corp.                                                                                                                                                                       |
| Product code                                  | GEI                                                                                          | GEI                                                                                          | GEI                                                                                                                                                                                       |
| Regulation<br>number                          | 21 CFR 878.4400                                                                              | 21 CFR 878.4400                                                                              | 21 CFR 878.4400                                                                                                                                                                           |
| Regulation<br>name                            | Electrosurgical Cutting and<br>Coagulation Device and<br>accessories                         | Electrosurgical Cutting and<br>Coagulation Device and<br>accessories                         | Electrosurgical Cutting and<br>Coagulation Device and<br>accessories                                                                                                                      |
| Indications for<br>use                        | See previous page                                                                            | See previous page                                                                            | The IDS-310 High<br>Frequency Electrosurgical<br>Generator is used to deliver<br>RF energy via an assortment<br>of surgical devices to cut and<br>coagulate different kinds of<br>tissue. |
| Prescription<br>device<br>(21 CFR<br>801.109) | Yes                                                                                          | Yes                                                                                          | Yes                                                                                                                                                                                       |
| Principle of<br>Operation                     | Conversion of electrical<br>energy to high-frequency RF<br>energy<br>(monopolar and bipolar) | Conversion of electrical<br>energy to high-frequency RF<br>energy<br>(monopolar and bipolar) | Conversion of electrical<br>energy to high-frequency RF<br>energy<br>(monopolar and bipolar)                                                                                              |
| Modes of<br>operation                         | Cut<br>Blend<br>Hemo<br>Bipolar<br>Bipolar Turbo                                             | Cut<br>Blend<br>Hemo<br>Bipolar<br>Bipolar Turbo                                             | 9 Monopolar and 4 bipolar<br>modes                                                                                                                                                        |
| Output<br>frequency                           | 4 MHz (monopolar)<br>1.7 MHz (bipolar)                                                       | 4 MHz (monopolar)<br>1.7 MHz (bipolar)                                                       | 350 - 490 kHz (monopolar)<br>490 kHz (bipolar)                                                                                                                                            |
| Maximum power<br>output                       | 170W                                                                                         | 120W                                                                                         | 300W                                                                                                                                                                                      |
| Voltage<br>(peak-to-peak)                     | 1,200V                                                                                       | 1,600V                                                                                       | 8,000V                                                                                                                                                                                    |
| Duty Cycle                                    | 10 sec On / 30 sec OFF                                                                       | 10 sec On / 30 sec OFF                                                                       | 10 sec On / 30 sec OFF                                                                                                                                                                    |
| Supply input<br>power                         | 100 V AC – 240 V AC, 50/60<br>Hz                                                             | 100 V AC – 240 V AC, 50/60<br>Hz                                                             | 100 V AC – 240 V AC, 50/60<br>Hz                                                                                                                                                          |

elliquence, LLC. 2455 Grand Avenue • Baldwin, New York 11510 (516) 277.9000 • Fax: (516) 277.9001 • www.elliquence.com

{7}------------------------------------------------

|                                                   | Surgi-Max® Ultra<br>(K170107) – SUBJECT<br>DEVICE     | Surgi-Max® / Surgi-Max<br>Plus (K100390) –<br>PREDICATE | Bovie® IDS-310<br>Electrosurgical Generator<br>(K134054) - PREDICATE |
|---------------------------------------------------|-------------------------------------------------------|---------------------------------------------------------|----------------------------------------------------------------------|
| Power<br>activation<br>control                    | Footswitch and/or<br>fingerswitch                     | Footswitch and/or<br>fingerswitch                       | Footswitch and/or<br>fingerswitch                                    |
| Electrical safety<br>and EMC testing<br>standards | AAMI ANSI ES60601-1<br>IEC 60601-1-2<br>IEC 60601-2-2 | AAMI ANSI ES60601-1<br>IEC 60601-1-2<br>IEC 60601-2-2   | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-2                        |
| Applied part<br>type                              | CF                                                    | BF                                                      | CF                                                                   |

# Non-Clinical Testing (807.92(b)(1))

The modifications enacted for the Surgi-Max® Ultra were made in accordance with the design control requirements of 21 CFR 820.30. Appropriate non-clinical verification and validation activities were planned and conducted in order to address identified risks and ensure the safety and effectiveness of the device with the modifications enacted. This testing included:

- . Electrical safety and EMC testing in accordance with AAMI ANSI ES60601-1, IEC 60601-1-2. and IEC 60601-2-2
- Usability assessment in accordance with IEC 62366
- Software verification and validation in accordance with IEC 62304 and the FDA Final . Guidance document titled "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices"
- Energy output verification of each of the five modes and comparison to predicate ●
- . Thermal effect comparison for representative modes of the subject device and predicate on ex-vivo bovine tissue

### Clinical Testing (807.92(b)(2))

This premarket submission did not rely on the assessment of clinical performance data to demonstrate substantial equivalence.

### Conclusion (807.92(b)(3))

Based upon a comparison of the intended uses, technological characteristics, and performance testing, we have concluded that the subject device is substantially equivalent to the selected predicate devices.

> elliquence, LLC. 2455 Grand Avenue · Baldwin, New York 11510 (516) 277.9000 · Fax: (516) 277.9001 · www.elliquence.com

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K170107](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K170107)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
