← Product Code [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI) · K162053

# PhotonBlade (K162053)

_Invuity, Inc. · GEI · Sep 15, 2016 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K162053

## Device Facts

- **Applicant:** Invuity, Inc.
- **Product Code:** [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI.md)
- **Decision Date:** Sep 15, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The PhotonBlade is a monopolar RF device coupled with illumination that is indicated for cutting and coagulation of soft tissue during general surgical procedures.

## Device Story

PhotonBlade is a single-use, sterile, monopolar electrosurgical device with integrated LED illumination. It features a rotatable, extendable shaft with a distal electrode tip. The device connects to standard, FDA-cleared electrosurgical units (ESUs) via a universal cable to deliver RF energy for soft tissue dissection and coagulation. Integrated handle controls allow the surgeon to manage cut, coagulation, and light functions. The device provides localized illumination to the surgical field, powered by a replaceable 3V battery. Used in general surgery, it assists clinicians by providing both surgical cutting/coagulation and site visibility. The device benefits patients by combining tissue management and illumination in a single instrument, potentially improving surgical site visualization during procedures.

## Clinical Evidence

Bench testing only. Performance data included mechanical testing, electrical safety/EMC (IEC 60601-1, IEC 60601-2-2), operational testing, ex-vivo thermal effects on tissue, biocompatibility (ISO 10993-1), and packaging evaluation. No clinical testing was performed.

## Technological Characteristics

Monopolar RF electrosurgical device with integrated LED illumination. Materials: stainless steel blade electrode, enamel insulator. Features: rotatable/extendable shaft, replaceable 3V battery. Connectivity: universal cable for standard RF generators. Sterilization: Ethylene Oxide (EO). Compliance: IEC 60601-1 3rd Ed, IEC 60601-2-2, ISO 10993-1.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- PEAK Surgical PlasmaBlade 3.0S ([K093695](/device/K093695.md))
- PARE NOVA ES Pencil with Surgical Light Source ([K944363](/device/K944363.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features the department's name encircling a symbol. The symbol is a stylized representation of a human figure, with three profiles overlapping to suggest community and connection. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 15, 2016

Invuity, Inc. Mr. John Kang Senior Director, Quality Assurance and Regulatory Affairs 444 De Haro Street San Francisco, California 94107

Re: K162053

Trade/Device Name: PhotonBlade Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI Dated: July 20, 2016 Received: July 25, 2016

Dear Mr. Kang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR Part 803); good manufacturing practice requirements as set forth 

{1}------------------------------------------------

in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

# Christopher J. Ronk -S

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K162053

Device Name PhotonBlade™

Indications for Use (Describe)

The PhotonBlade is a monopolar RF device coupled with illumination that is indicated for cutting and coagulation of tissue during general surgical procedures.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for "INVUITY intelligent photonics". The logo consists of a circular graphic to the left of the word "INVUITY" in bold, black letters. Below the word "INVUITY" is the phrase "intelligent photonics" in a smaller, thinner font. The circular graphic is made up of several smaller circles in different shades of blue and green.

#### 5 510(K) SUMMARY

Date Summary Prepared July 20, 2016.

#### 5.1 Regulatory authority

This 510(k) Summary is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990, 21 CFR 807.92

#### 5.2 Company name:

INVUITY, INC. 444 De Haro Street San Francisco, CA 94107

#### 5.3 Contact person:

John Kang Senior Director RA/QA INVUITY, INC. 444 De Haro Street San Francisco, CA 94107

#### 5.4 Name of device

| Trade Name:            | PhotonBlade™                                                                        |
|------------------------|-------------------------------------------------------------------------------------|
| Common Name:           | Electrosurgical Accessory                                                           |
| Device Product Code:   | GEI                                                                                 |
| Classification Name:   | Electrosurgical Cutting and Coagulation Device and<br>Accessories (21 CFR 878.4400) |
| Device Panel:          | General and Plastic Surgery                                                         |
| Device Classification: | Class II                                                                            |

#### 5.5 Predicate device(s):

- 0 PEAK Surgical PlasmaBlade 3.0S (K093695)
- 0 PARE NOVA ES Pencil with Surgical Light Source (K944363)

#### 5.6 Device description

:

The PhotonBlade is a single use, sterile, electrosurgical device with a light. The PhotonBlade has an electrode at the distal tip, which delivers RF energy for cutting and coagulation of soft tissue. The electrode tip is located at the distal end of a rotatable and extendable shaft. The device handle is integrated with controls for cut, coagulation, and illumination (light). A universal cable attaches the device to a 510(k) cleared electrosurgical unit.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Invuity, a company that specializes in intelligent photonics. The logo features a circular icon with a gradient of blue and green colors, followed by the company name in bold, black letters. Below the company name, the words "intelligent photonics" are written in a smaller, lighter font.

#### 5.7 Indications for Use statement

The PhotonBlade™ is a monopolar RF device coupled with illumination that is indicated for cutting and coagulation of soft tissue during general surgical procedures.

#### 5.8 Substantial equivalence comparison

Two predicates have been identified for the PhotonBlade, PlasmaBlade 3.0S and NOVA ES Pencil. In terms of the indications for use, the PlasmaBlade 3.0S and the NOVA ES Pencil are predicates for soft tissue cutting and coagulation and the NOVA ES Pencil is a predicate for surgical light (illumination).

| Attribute                          | Invuity PhotonBlade<br>Subject Device                                                                                                                   | PEAK Surgical PlasmaBlade<br>(Medtronic)<br>K093695                                                                                                         |
|------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use                       | The PhotonBlade is<br>intended for soft tissue<br>dissection and coagulation<br>coupled with illumination<br>for use in General Surgical<br>procedures. | The Peak PlasmaBlade is<br>intended for soft tissue<br>dissection and coagulation<br>for use in General, Plastic,<br>and Orthopedic surgical<br>procedures. |
| Principle of<br>Operation          | Electrosurgical cutting and<br>coagulation                                                                                                              | Electrosurgical cutting and<br>coagulation                                                                                                                  |
| Anatomical Site(s)                 | Soft Tissue                                                                                                                                             | Soft Tissue                                                                                                                                                 |
| Use                                | Single Use                                                                                                                                              | Single Use                                                                                                                                                  |
| Materials                          | Metal electrode/enamel<br>insulator                                                                                                                     | Metal electrode/enamel<br>insulator                                                                                                                         |
| Specification                      |                                                                                                                                                         |                                                                                                                                                             |
| Energy Type                        | Radiofrequency (RF)                                                                                                                                     | Radiofrequency (RF)                                                                                                                                         |
| Power Modes                        | Monopolar                                                                                                                                               | Monopolar                                                                                                                                                   |
| Electrode                          | Stainless steel blade<br>insulated with enamel                                                                                                          | Stainless steel blade<br>insulated with enamel                                                                                                              |
| Compliance                         |                                                                                                                                                         |                                                                                                                                                             |
| Sterilization                      | Sterile-Ethylene Oxide                                                                                                                                  | Sterile-Ethylene Oxide                                                                                                                                      |
| EMC & Electrical<br>Safety Testing | IEC 60601-1 3rd Edition<br>IEC 60601-2-2                                                                                                                | IEC 60601-1 2nd Edition                                                                                                                                     |
| Biocompatibility                   | ISO 10993-1                                                                                                                                             | ISO 10993-1                                                                                                                                                 |

Comparison to Predicate Device(s)

{5}------------------------------------------------

| Attribute                    | Invuity PhotonBlade<br>Subject Device | PARE Surgical NOVA ES<br>(Electrosurgical) Pencil with<br>Light K944363 |
|------------------------------|---------------------------------------|-------------------------------------------------------------------------|
| Illuminate Surgical<br>Field | Yes                                   | Yes                                                                     |
| Illumination<br>source       | LED                                   | LED                                                                     |
| Illumination<br>location     | Shaft                                 | Handle                                                                  |
| Nominal Light<br>Output      | 28 lumens (n=29)                      | 30 lumens (n=1)                                                         |
| Illumination<br>power source | 3V Battery                            | 3V Battery                                                              |
| Battery<br>characteristic    | Replaceable                           | Non-replaceable                                                         |
| Light Color                  | White                                 | White                                                                   |

# Comparison to Predicate Device(s)-Illumination (Lighted ES Pencil)

#### Technological Characteristics 5.9

The PhotonBlade device is similar to both predicate devices in that they are electrosurgical instruments used to cut and coagulate soft tissue utilizing RF energy.

PhotonBlade and PlasmaBlade both have insulated electrodes that extend and rotate. Both devices utilize blade electrodes where only the edge of the blade is exposed or active. As such, the principles of operation involved in cutting and coagulation of tissue are the same on either device. The differences between the PhotonBlade and PlasmaBlade are:

- 0 The PlasmaBlade device has aspiration and the PhotonBlade does not.
- . The PlasmaBlade is compatible with a dedicated electrosurgical unit (ESU) or generator and the PhotonBlade is intended for use with general purpose FDA-cleared RF generators with the standard or traditional (Bovie) three plug receptacle, such as the Valleylab (Medtronic) Force FX c and Conmed S5000 for example.
- The PhotonBlade has a light and the PlasmaBlade does not. 0

The PhotonBlade and NOVA ES Pencil lights both provide illumination using LED technology. In both devices, the illumination is powered by a battery; however, the PhotonBlade battery is replaceable, while the NOVA ES Pencil battery is not. The NOVA ES Pencil electrode tip is not insulated, and the device is provided with two tip configurations (one long and one short). The

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Invuity, a company that specializes in intelligent photonics. The logo consists of a circular icon with a blue and green color scheme, followed by the company name in bold, black letters. Below the company name, the words "intelligent photonics" are written in a smaller, lighter font.

PhotonBlade has an insulated tip with only one configuration that can be adjusted via an extendable shaft. The NOVA ES Pencil has three LEDs that are located in the handle of the device; the distance from the electrode tip to the LEDs varies depending on the tip configuration. The PhotonBlade has one LED that is located in the shaft of the device; the distance from the electrode tip to the LED is constant (approx. 2cm).

The differences identified between the PhotonBlade and its two predicates (PlasmaBlade and NOVA ES Pencil Surgical Light Source) do not present any new concerns of safety and effectiveness.

## 5.10 Summary of Performance Data

Performance testing was conducted on PhotonBlade devices that were exposed to EO sterilization and simulated transportation to demonstrate compliance with the product requirements and to demonstrate safety, efficacy, and substantial equivalence to the predicate. The following product performance tests are included: mechanical, electrical safety/electromagnetic compatibility, operational, ex-vivo thermal effects on tissue comparisons, biocompatibility, and packaging evaluation. Product Shelf Life studies for 7 months and 37 months will be conducted.

Like the predicate device(s), no clinical testing was necessary to support substantial equivalence. The data submitted support the substantial equivalence claim for the proposed indications for use; the PhotonBlade is as safe and effective as the predicate device(s).

## 5.11 Conclusion

Invuity Inc. considers the PhotonBlade device to be substantially equivalent to the legally marketed predicate device(s) with respect to the device function, intended use, and patient population. Any differences in technological characteristics between the PhotonBlade device and the predicate device(s) do not raise any new issues of safety and effectiveness. The Invuity PhotonBlade is substantially equivalent to the proposed predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K162053](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K162053)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
