Bovie 3.3mm Disposable Bipolar Ablator 90°, Aspirating and Non-Aspirating, Bovie 2.4mm Disposable Bipolar Ablator 55°, Non-Aspirating, Bovie 1.8mm Disposable Bipolar Ablator 60°, Non-Aspirating

K161558 · Bovie Medical Corporation · GEI · Jul 29, 2016 · General, Plastic Surgery

Device Facts

Record IDK161558
Device NameBovie 3.3mm Disposable Bipolar Ablator 90°, Aspirating and Non-Aspirating, Bovie 2.4mm Disposable Bipolar Ablator 55°, Non-Aspirating, Bovie 1.8mm Disposable Bipolar Ablator 60°, Non-Aspirating
ApplicantBovie Medical Corporation
Product CodeGEI · General, Plastic Surgery
Decision DateJul 29, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Indications for Use

This device is intended to be used for cutting, vaporization, and coagulation of soft tissue during arthroscopic surgical procedures. This device is intended to be used with a standard electrosurgical generator with footswitch control and a standard return electrode connection, and the electrode is to be activated only when immersed in a conductive media such as standard saline solution.

Device Story

Disposable bipolar electrosurgical ablator for arthroscopic soft tissue procedures; utilizes radiofrequency (RF) energy. Device components: insulated handle, stainless steel electrode tip, ceramic thermal insulator, aspiration tubing (aspirating models), RF connector, return electrode connector. Operates in bipolar mode; requires immersion in conductive media (saline) for activation. Used in clinical arthroscopic settings by surgeons; controlled via footswitch. Aspirating models remove saline to dissipate heat and improve visualization. Wedge-shaped tip facilitates access to difficult areas. Device provides cutting, vaporization, and coagulation; benefits include precise tissue management and improved surgical visibility.

Clinical Evidence

Bench testing only. No clinical data provided. Verification and validation included mechanical testing (IEC 60601-1, IEC 60601-2-2), electrical safety, EMC (IEC 60601-1-2), biocompatibility (ISO 10993-1), and functional bench-top testing (aspiration flow, thermal performance, clogging, ignition delay, and cable integrity).

Technological Characteristics

Bipolar electrosurgical device; stainless steel electrode with ceramic insulator. Shaft diameters: 1.8mm, 2.4mm, 3.3mm. Working length: 160mm. Energy: RF. Connectivity: standard electrosurgical generator interface. Sterilization: Ethylene Oxide (ANSI/AAMI/ISO 11135-1). Standards: IEC 60601-1 (Ed 3.1), IEC 60601-1-2 (2007), IEC 60601-2-2 (2009), ISO 10993-1, ISO 10993-7.

Indications for Use

Indicated for cutting, vaporization, and coagulation of soft tissue during arthroscopic surgical procedures. No specific patient population (age/gender) or contraindications provided.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of three human profiles facing to the right, stacked on top of each other. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 July 29, 2016 Bovie Medical Corporation Ms. Rubiela Maldonado Regulatory Affairs Manager 5115 Ulmerton Road Clearwater, Florida 33760 Re: K161558 Trade/Device Name: Bovie 3.3mm Disposable Bipolar Ablator 90°, Aspirating and Nonaspirating: Bovie 2.4mm Disposable Bipolar Ablator 55° , Non-aspirating; Bovie 1.8mm Disposable Bipolar Ablator 60° . Non-aspirating. Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: June 3, 2016 Received: June 6, 2016 Dear Ms. Maldonado: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR {1}------------------------------------------------ Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. # Christopher J. Ronk -S - For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K161558 Device Name Bovie Disposable Bipolar Ablator #### Indications for Use (Describe) This device is intended to be used for cutting, vaporization of soft tissue during arthroscopic surgical procedures. This device is intended to be used with a standard electrosurgical generator with footswitch control and a standard return electrode connection, and the electrode is to be activated only when immersed in a conductive media such as standard saline solution. | Type of Use (Select one or both, as applicable) | | |-----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------| | <span style="text-decoration: underline;"><b> Prescription Use (Part 21 CFR 801 Subpart D) </b></span> | <span style="text-decoration: overline;">Over-The-Counter Use (21 CFR 801 Subpart C)</span> | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the word "Bovie" in a dark blue font. Above the word is a light blue curved line with a dot at the end. The word is written in all capital letters except for the "ie" at the end. # 510(k) SUMMARY ### (Submitted As Required per 21 CFR 807.92) # GENERAL INFORMATION: | Submitter Name: | Bovie Medical Corporation | |---------------------------------------|----------------------------------------------------------------------------------| | Establishment Registration<br>Number: | 3007593903 | | Submitter Address: | 5115 Ulmerton Road<br>Clearwater, Florida 33760-4004<br>United States of America | | Submitter Telephone Number: | 727) 803-8508 | | Submitter FAX Number: | (727) 322-4465 | | Contact Person: | Rubiela Maldonado<br>Regulatory Affairs Manager | | Date Prepared: | July 26, 2016 | | DEVICE IDENTIFICATION: | | | Proprietary Name: | Bovie Disposable Bipolar Ablator | |----------------------|-------------------------------------------------------------------| | Common Name: | Electrosurgical Generator Accessory | | Classification Name: | Electrosurgical Cutting and Coagulation Device and<br>Accessories | # Model Numbers: | Catalog # | Description | |-----------|------------------------------------------------------| | BA3390A | 3.3mm Disposable Bipolar Ablator 90°, Aspirating | | BA3390NA | 3.3mm Disposable Bipolar Ablator 90°, Non-Aspirating | | BA2455NA | 2.4mm Disposable Bipolar Ablator 55°, Non-Aspirating | | BA1860NA | 1.8mm Disposable Bipolar Ablator 60°, Non-Aspirating | {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the word "Bovie" in a dark blue font. Above the word is a light blue curved line with a dot at the end. The font is sans-serif and the word is slightly italicized. There is a registered trademark symbol in the bottom right corner of the image. 21CFR 878.4400; Class II; Product Code GEI Legally Marketed Predicate Device(s): Classification: - K120791, K083306 ArthroCare SuperTurbo Vac 90, TurboVac 90, Arthrocare ArthroWand® 2.3mm 35° Short Bevel and Microblator 30 With Integrated Cable - . K152777 Bovie Disposable Bipolar Ablator ### INTENDED USE/INDICATIONS This device is intended to be used for cutting, vaporization, and coagulation of soft tissue during arthroscopic surgical procedures. This device is intended to be used with a standard electrosurgical generator with footswitch control and a standard return electrode connection, and the electrode is to be activated only when immersed in a conductive media such as standard saline solution. #### DEVICE DESCRIPTION The device is comprised of the following major components: an insulated handle, an active electrode, a length of aspiration tubing, a standard foot-switch RF connector, and a standard return connector. The hand piece is designed to connect to electrosurgical generators that feature a standard foot switch controlled monopolar connector and a standard return connector. A stainless steel electrode tip is surrounded by a ceramic thermal insulator. The product is available in shaft diameters of 3.3mm, 2.4mm, and 1.8mm. The larger size is available in aspirating and non-aspirating version. The aspirating versions use suction to remove saline irrigant during the surgical procedure for the dissipation of heat and improved vision. The device is available in electrode face angles of 55 degrees, and 90 degrees. There is a feature on the handle that lets the operator easily determine the orientation of the active electrode face. The wedge-shaped tapered tip allows access to difficult to reach locations. The working length is 160mm. The cable from the handle is 126" +/- 6" in length. The device operates in bipolar mode. The return electrode is located on the device and does not require the use of an external grounding pad or return electrode. The device has a simple user interface that uses a footswitch to activate the device. The Bovie Disposable Bipolar Ablator has the same intended use and energy type as the predicate device. There are no technological differences, no changes to the principle of operation or the method of application. #### PERFORMANCE TESTING {5}------------------------------------------------ Performance testing was completed to demonstrate substantial equivalence of the subject device to the predicates. The devices were subjected to the following verification and validation tests. as applicable: #### Mechanical testing Mechanical verification testing was conducted for the proposed device to ensure compliance with mechanical requirements of IEC-60601-1, Edition 3.1, IEC 60601-2-2: 2009, and Bovie self-enforced requirements. #### Electrical testing Electrical verification testing was conducted for the relevant components of the proposed device to ensure compliance with current electrical standard requirements. ### Electromagnetic compatibility Electromagnetic compatibility (EMC) testing has been completed for the applicable components of the proposed device. The results demonstrated compliance of the proposed device to current IEC 60601-1-2 standard requirements. #### Biocompatibility Biocompatibility verification was performed in accordance with requirements of ISO 10993-1 and FDA's modified ISO guidelines in accordance with FDA's blue book memorandum #G95-1 on biocompatibility. #### INE Testing Testing was conducted to demonstrate that the Bovie Disposable Bipolar RF Arthroscopy Ablator is able to withstand the insertion / navigation / extraction actions during a procedure. ### Bench-top validation testing - Visual Inspection & Electrical Continuity Testing: . Testing was conducted to: Meet general workmanship requirement. Pass electrical continuity test. - . Mold Stress Relief Testing: Testing was conducted to demonstrate: Any shrinkage or distortion of the molded thermoplastic components due to release of internal stresses caused by the molding operation does not result in an unacceptable risk. - Cable Strain (steady pull) Relief Testing: Testing was conducted to demonstrate: The ablator cable and its connections are capable of withstanding mechanical stress from a steady pull - . Cable Impulse Testing and Post-impulse continuity testing: Testing was conducted to demonstrate: The ablator cable and its connections are capable of withstanding mechanical stress from an impulse of force. - . Cable Flexure / Tension and Post-Flexure-testing Electrical Continuity Testing: Testing was conducted to: Evaluate the integrity of the anchorage (fastening, termination) of electrical conductors and insulation by exercising the cable / cords by flexing and applying specified tension. - Drop Testing: Testing was conducted to: Verify the ablator will not present a safety hazard as a result of a free fall onto a hard surface. - Front Panel Connection Verification Testing: {6}------------------------------------------------ Testing was conducted to: Verify that Proximal End (brass sockets) of the ablator cable connects to a compatible generator. - Connector Insertion / Extraction Force Testing: . Testing was conducted to: Verify that the insertion / Extraction force for inserting / extracting ablator cable connector into or from a compatible generator is within specified range. - . Measurement and Maximum Temperature During Use (Aspirating Model): Testing was conducted to: Assess the maximum temperature that ablator handle can achieve during a worst case use scenario - . Measurement and Maximum Temperature During Use (Non-Aspirating Model): Testing was conducted to: Assess the maximum temperature that ablator handle can achieve during a worst case use scenario - . Rigidity Testing of Handle and Shaft: Testing was conducted to: Verify that handle of an ablator is able to support a static weight located at the distal position on the shaft without permanently deforming - Compatibility Verification Testing: Testing was conducted to: Verify that the ablator is able to function as intended when connected to a compatible electrosurgical generators. - . Aspiration Flow Testing: Testing was conducted to: Verify aspiration flow rate under specific vacuum conditions. - . Thermal Performance & clogging Testing: Testing was conducted to: Verify the performance of the Bovie Disposable Bipolar RF Arthroscopic Ablator in terms of 1) thermal performance (the temperature of irrigation fluid near the active electrode during activation of a device); 2) tissue removal, 3) clogging, and 4) ignition delay - . Thermal Performance Testing (Non-Aspirating ablator): Testing was conducted to: Assess the fluid temperature near the active electrode during activation of an ablator. The electrosurgical generator is to be set at the maximum allowable power for ablator model in order to simulate the worst case scenario. All test requirements were met as specified by applicable standards and the test protocols. The Bovie Disposable Bipolar Ablator was designed in accordance with the following standards: | International Standard | Description | |--------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | IEC-60601-1, Edition 3.1 | Medical Electrical Equipment - Part 1: General Requirements<br>For Safety | | IEC 60601-1-2:2007 | Medical electrical equipment - Part 1-2: General requirements<br>for basic safety and essential performance - Collateral<br>Standard: Electromagnetic disturbances - Requirements and | {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the word "Bovie" in a bold, dark blue font. Above the word is a light blue curved line with a dot at the end. The word appears to be a logo for a company or product. | | tests | |--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | IEC-60601-2-2 : 2009 | Particular requirements for the safety of high frequency<br>surgical equipment | | ANSI / AAMI / ISO 10993-1:2009 | Biological Evaluation of Medical Devices – Part 1: Evaluation<br>and Testing | | ANSI/AAMI/ISO 11135-1: 2007 | Sterilization of Healthcare Products – Ethylene Oxide – Part 1:<br>Requirements for the development, validation, and routine<br>control of a sterilization process for medical devices. | | ANSI / AAMI / ISO 10993-7:2008 | Biological Evaluation of Medical Devices – Part 7: Ethylene<br>Oxide Residuals | {8}------------------------------------------------ Image /page/8/Picture/0 description: The image shows the word "Bovie" in a bold, dark blue font. Above the word is a curved, light blue line with a dot at the end. The registered trademark symbol is located at the bottom right of the word. # SUBSTANTIAL EQUIVALENCE SUMMARY | Feature/<br>Characteristic | Bovie Disposable Bipolar<br>Ablator | ArthroCare<br>SuperTurbo Vac, Turbovac 90<br>Arthrocare ArthroWand®<br>2.3mm 35° Short Bevel<br>Arthrocare ArthroWand®<br>Microblator 30 With<br>Integrated Cable | Bovie Disposable Bipolar<br>Ablator BA3350NA,<br>BA3350A | |--------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------| | Intended Use | Intended to be used for<br>cutting, vaporization, and<br>coagulation of soft tissue<br>during arthroscopic surgical<br>procedures. | Intended to be used for<br>resection, ablation, and<br>coagulation of soft tissue and<br>hemostasis of blood vessels<br>in arthroscopic and<br>orthopedic procedures. | Intended to be used for<br>cutting, vaporization,<br>and coagulation of soft<br>tissue during<br>arthroscopic surgical<br>procedures. | | Energy Type | Radiofrequency | Radiofrequency | Radiofrequency | | Generator<br>Compatibility | To be used with a standard<br>electrosurgical generator<br>with footswitch control and<br>a standard return electrode<br>connection | Dedicated generator | To be used with a<br>standard electrosurgical<br>generator with<br>footswitch control and a<br>standard return<br>electrode connection | | Use only in<br>Conductive<br>Media | The electrode is to be<br>activated only when<br>immersed in a conductive<br>media such as standard<br>saline solution | Same | The electrode is to be<br>activated only when<br>immersed in a<br>conductive media such<br>as standard saline<br>solution | | Mode | Bipolar | Bipolar | Bipolar | | User Interface | Footswitch | Footswitch and Hand | Footswitch | | Aspirating and<br>Non-Aspirating<br>Models | Yes | Yes | Yes | | Shaft Diameter | 3.3mm, 2.4mm, 1.8mm | 3.75mm<br>2.3mm and 1.4mm | 3.3mm | | Electrode Face<br>Angle | 90, 55, 60 degrees | 90,35, 30 degrees | 50 degrees | | Working<br>Length | 160mm | 160mm | 160mm | {9}------------------------------------------------ ### CONCLUSION In summary, the only differences between the subject and the predicate devices are: - Generator compability . - user interface - . Shaft diameter - . Electrode Face angle These differences between the Bovie Disposable Bipolar Ablator and the predicate devices do not raise new or different questions of safety and efficacy. The Bovie Disposable Bipolar Ablator was subjected to verification testing to confirm device performance. There is no new technology and no difference that would raise new or different questions of safety or efficacy. Comparative performance testing demonstrated the device performed as well as, or better than, the predicate device. There is no difference between the Bovie Disposable Bipolar Ablator and the predicate devices in terms of intended use, principle of operation, and the technology used for device performance.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...