Venclose Radiofrequency System (digiRF Generator, EVSRF Catheter)
K160754 · Venclose, Inc. · GEI · Sep 9, 2016 · General, Plastic Surgery
Device Facts
Record ID
K160754
Device Name
Venclose Radiofrequency System (digiRF Generator, EVSRF Catheter)
Applicant
Venclose, Inc.
Product Code
GEI · General, Plastic Surgery
Decision Date
Sep 9, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Venclose™ Radiofrequency System (Generator and Catheters) is intended for use in endovascular of blood vessels in patients with superficial vein reflux.
Device Story
Venclose RF System consists of digiRF generator and EVSRF catheter; used for endovascular ablation of incompetent superficial veins (e.g., great saphenous vein). Procedure involves percutaneous catheter insertion under ultrasound guidance; generator delivers temperature-controlled bipolar RF energy (460 kHz) to resistive element in catheter tip. Thermal energy heats vein wall to cause irreversible luminal occlusion, fibrosis, and resorption; blood reroutes to healthy veins. Performed in outpatient setting by physician; no general anesthesia required. Output includes thermal energy delivery; healthcare provider monitors procedure via generator touchscreen. Benefits include minimally invasive treatment of venous reflux with reduced recovery time.
Clinical Evidence
No human clinical data provided. Evidence consists of non-clinical bench testing (catheter simulation, temperature, pressure, tensile, and fluid tests) and an animal study. Animal model (veins sized similarly to human) demonstrated successful catheter navigation, ability to reach and maintain target treatment temperatures, and effective vein diameter reduction comparable to predicate devices.
Technological Characteristics
System comprises digiRF generator and EVSRF catheter. Energy: Bipolar RF, 460 kHz, max 24V. Catheter: 6F (2.0mm) diameter, adjustable heating element (10cm or 2.5cm). Materials: Biocompatible, non-pyrogenic (ISO 10993 compliant). Software: Controlled. Sterilization: Catheters provided sterile. Standards: IEC 60601-1, IEC 62304, IEC 62366, ISO 10555-1, EN ISO 13485, EN ISO 14971.
Indications for Use
Indicated for endovascular coagulation of blood vessels in patients with superficial vein reflux.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of three human profiles facing to the right, resembling a stylized bird.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 9, 2016
Venclose, Inc. % Ms. Pamela Buckman Pamela M. Buckman, MSN 2800 Pleasant Hill Road, Suite 175 Pleasant Hill, California 94523
Re: K160754
Trade/Device Name: Venclose™ Radiofrequency System (digiRF Generator, EVSRF Catheter) Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: August 8, 2016 Received: August 11, 2016
Dear Ms. Buckman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR Part 803); good manufacturing practice requirements as set forth
{1}------------------------------------------------
in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Christopher J. Ronk-S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K160754
Device Name Venclose Radiofrequency System
Indications for Use (Describe)
The Venclose™ Radiofrequency System (Generator and Catheters) is intended for use in endovascular of blood vessels in patients with superficial vein reflux.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
### K Number: K160754
# General Information
| Date Prepared: | September 8, 2016 |
|------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|
| Proprietary Name: | Venclose™ RF Generator and Catheter System |
| Common Name: | Electrosurgical, Cutting & Coagulation & Accessories |
| Classification Name and Reference: | Electrosurgical cutting and coagulation device and accessories<br>21 CFR §878.4400 |
| Regulatory Class: | Class II |
| FDA Panel: | General and Plastic Surgery |
| Product Code: | GEI |
| Applicant: | Venclose™, Inc.<br>2570 N. First St.<br>2nd Floor #221<br>Tel: 844 834 6292<br>Fax: 844 834 6293<br>Brady Esch, CEO<br>besch@venclose.com |
| Contact: | Pamela M. Buckman, MSN<br>2800 Pleasant Hill Rd., Suite 175<br>Pleasant Hill, CA 94523<br>Tel: 925 980 7007<br>Fax: 925 705 7381<br>pmbuckman@gmail.com |
#### Indication for Use:
The Venclose™ Radiofrequency System (Generator and Catheters) is intended for use in endovascular coagulation of blood vessels in patients with superficial vein reflux.
#### Predicate Device(s):
The Venclose™ Radiofrequency System is substantially equivalent to the following predicate devices.
| K040638 | VNUS RF Generator |
|---------|---------------------------|
| K061373 | VNUS ClosureFast Catheter |
{4}------------------------------------------------
# Device Description:
The Venclose RF System is composed of two basic components: an energy console or generator (digiRF) and catheters (EVSRF) designed specifically to be used with the generator. The Venclose RF System uses resistive radiofrequency ablation via energy delivery to heat the wall of an incompetent vein with temperature-controlled RF energy to cause irreversible luminal occlusion, followed by fibrosis and ultimately resorption of the vein. The blood then naturally reroutes to healthy veins.
The technique involves percutaneous access and insertion of the EVSRF Catheter into the great saphenous vein (GSV) or other superficial vein under ultrasound guidance, injection of local anesthesia, and thermal energy from a radiofrequency generator (digiRF) applied into the target vein. As the EVSRF Catheter is withdrawn stepwise down the treated length of the vein, thermal damage is inflicted upon the venous endothelium and through the vein wall, resulting in contraction and ultimately destruction of the vessel. The procedure can be performed in an outpatient setting, without the need for general anesthesia, allowing for a walkin/walk-out procedure with minimal postoperative recovery time.
# Materials:
Materials for use in the console are consistent with industry practice for durable cases containing electronic elements. The console is supplied in a non-sterile condition and is not intended to be in contact with the surgical field or the patient. The catheters are made of industry-typical, biocompatible, nonpyroqenic materials and are provided packaged in a sterile condition. They are intended for use in a sterile surgical environment and will make contact with the patient. Biocompatibility data demonstrates that the catheters are in compliance with ISO 10993.
# Comparative Summary of Technological Characteristics:
The materials, design, energy source, software integration, sterility methods (EVSRF Catheter component only), environment of use and clinical applications for the Venclose™ Radiofrequency System are comparable to those identified for the predicate devices.
{5}------------------------------------------------
| COMPARISON<br>ITEMS | VNUS RF<br>Generator<br>K040638 | VNUS Closurefast™<br>Catheter<br>K061373 | SUBJECT DEVICE<br>K160754 |
|-------------------------------------|---------------------------------------------------------------------------|------------------------------------------------------------------------------------|---------------------------------------------------------------------------|
| Instrument<br>Cable | Attached to<br>catheter, connects<br>to generator, eight<br>feet long | | Attached to catheter,<br>connects to<br>generator, eight feet<br>long |
| Energy/<br>Frequency | Bipolar 460.8 kHz,<br>max 90 Volts | | Bipolar 460 kHz, max<br>24 Volts |
| Software | Software<br>Controlled | | Software Controlled |
| Maximum<br>Output Power | 6 W/cm (40 W into<br>catheter 7 cm<br>heating element) | | 6 W/cm (60 W into<br>catheter when<br>heating 10 cm<br>heating element) |
| Temperature<br>functional limits | 10°C to 130°C | | 10°C to 130°C |
| Input Voltage | 100/240 VAC 50-60<br>Hz | | 100/240 VAC 50-60<br>Hz |
| Touch Screen | None | | Included 7in<br>touchscreen |
| Weight | 20 lbs max | | 10 lbs max |
| Dimensions | 38 cm x 38 cm x 19<br>cm (15 in x 15 in x<br>7.5 in) | | 23 cm x 25 cm x 17<br>cm<br>(9 in x 10 in x 7 in) |
| Mode of Action | Delivers heat to<br>resistive element<br>within the catheter<br>component | | Delivers heat to<br>resistive element<br>within the catheter<br>component |
| Foot Pedal | Not available | | Optional |
| Length of Use | | Single<br>Use/Disposable | Single<br>Use/Disposable |
| Catheters | | | |
| Diameters | | 7F (2.3mm) | 6F (2.0 mm) |
| Insertable<br>Lengths | | 60 cm or 100 cm | 60 cm or 100 cm |
| Heating<br>element<br>Lengths | | 7cm (predicate<br>K061373) or separate<br>catheter with 3cm<br>(reference K111887) | Single, adjustable, 10<br>cm (default) or 2.5<br>cm |
| Handle, Cable<br>& Connector | | Integral to Catheter | Integral to Catheter |
| Heating Set<br>Temperature<br>Range | | 95-120°C — User<br>Select, default at<br>120°C | 120°C, non-<br>adjustable |
{6}------------------------------------------------
### Non-Clinical Testing:
Device testing was conducted to evaluate conformance to product specifications and applicable standards.
Bench testing included but not limited to the following tests: Catheter Simulation Test, Temperature Test, Pressure Test, Tensile Test, Fluid Test.
Applicable Standards and Guidances are as follows:
- . Guidance on Premarket Notification [510(k)] Submissions for Short-Term and Long-Term Intravascular Catheters, March 16, 1995.
- ISO 10555-1:2013 Sterile, Single-Use Intravascular Catheters, Part 1: . General Requirements
- EN ISO 13485:2012 Medical devices. Quality management systems. Requirements for regulatory purposes
- EN ISO 14971:2012 Medical devices. Application of risk management to medical devices
- IEC 62304:2006 Medical device software. Software life cycle processes
- . IEC 62366:2008 Medical devices. Application of usability engineering to medical devices
- . IEC 60601-1:2005 Medical electrical equipment Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-6:2010 Medical electrical equipment Part 1-6: General . requirements for safety - Collateral Standard: Usability
# Pre-clinical testing:
Animal testing was conducted in a model previously used for research in endovenous ablation for similar devices. The chosen model provided great saphenous and common jugular veins that are sized similarly to human veins. The purpose of the study was to demonstrate proper functionality of the Venclose System in a pre-clinical environment. Objectives included insertion of the catheter though a sheath into the vein, navigation to the intended treatment site, ability to reach a predetermined treatment temperature within a specified time frame and then to hold that temperature for the remainder of the treatment cycle, and capability of reducing the diameter of the vein by an amount similar to that done with the predicate device. All of the objectives were met resulting in a satisfactory study conclusion.
# Rationale for Substantial Equivalence
Conclusions drawn from the non-clinical bench and animal testing demonstrate that the Venclose System is as safe, as effective and performs at least as safely and as effectively as the legally marketed devices (K040638 and K061373) for the requested intended use as identified in this Summary.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.