K160539 · Covidien · GEI · Jun 15, 2016 · General, Plastic Surgery
Device Facts
Record ID
K160539
Device Name
BiZact Open Sealer/Divider
Applicant
Covidien
Product Code
GEI · General, Plastic Surgery
Decision Date
Jun 15, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The BiZact device is a bipolar instrument intended for use in open surgical procedures where ligation of vessels, tissue bundles, and lymphatics is desired. The tissue fusion function of the device can be used on veins) and lymphatics up to and including 3 mm diameter. The BiZact device is indicated for use in open general surgical procedures. It is also indicated for adult ENT procedures, including tonsillectomy, for the ligation of vessels, tissue bundles and lymphatics 2-3 mm away from unintended thermally sensitive structures. The BiZact device has not been shown to be effective for tubal coagulation for sterilization procedures. Do not use for these procedures.
Device Story
BiZact is a sterile, single-use, hand-held bipolar electrosurgical instrument; used in open general surgery and adult ENT procedures (e.g., tonsillectomy). Device connects to Valleylab LS10 Vessel Sealing Generator via 10-foot cord; generator identifies device via embedded RFID tag. During operation, device delivers RF energy to tissue; tissue-sensing circuitry measures electrical resistance; generator output adjusts instantaneously to maintain controlled tissue response for permanent fusion. Physician operates device to ligate/divide vessels, tissue bundles, and lymphatics. Benefits include precise, controlled tissue fusion; reduced thermal spread to sensitive structures. Device output is physical tissue ligation/division; healthcare provider uses visual confirmation of tissue effect to guide surgical progress.
Clinical Evidence
No clinical data. Evidence consists of bench testing (electrical safety per IEC 60601-1, EMC per IEC 60601-1-2, HF equipment safety per IEC 60601-2-2, biocompatibility per ISO 10993-1), performance testing (device functionality, tissue burst pressure), in-vivo studies (acute hemostasis, thermal spread, lymphatic burst pressure; chronic hemostasis), and summative usability validation in simulated environments.
Technological Characteristics
Bipolar electrosurgical instrument; RF tissue fusion technology. Materials biocompatible per ISO 10993-1. Connectivity via proprietary connector with RFID tag for generator identification. Single-use, hand-held form factor. Energy source: Valleylab LS10 Vessel Sealing Generator. Software-controlled power output based on real-time tissue impedance sensing.
Indications for Use
Indicated for adult patients undergoing open general surgical procedures and ENT procedures, including tonsillectomy, requiring ligation of vessels, tissue bundles, and lymphatics up to 3 mm in diameter. Contraindicated for tubal coagulation or sterilization procedures.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
LigaSure™ Small Jaw, Open Sealer/Divider (K152286)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 15, 2016
Covidien Ms. Sharon McDermott Senior Product Specialist Regulatory Affairs 5920 Longbow Drive Boulder, Colorado 80301
Re: K160539
Trade/Device Name: Bizact Open Sealer/divider Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: May 24, 2016 Received: May 25, 2016
Dear Ms. McDermott:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Jennifer R. Stevenson -A
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K160539
Device Name BiZact Open Sealer/Divider
#### Indications for Use (Describe)
The BiZact device is a bipolar instrument intended for use in open surgical procedures where ligation of vessels, tissue bundles, and lymphatics is desired.
The tissue fusion function of the device can be used on veins) and lymphatics up to and including 3 mm diameter. The BiZact device is indicated for use in open general surgical procedures.
It is also indicated for adult ENT procedures, including tonsillectomy, for the ligation of vessels, tissue bundles and lymphatics 2-3 mm away from unintended thermally sensitive structures.
The BiZact device has not been shown to be effective for tubal coagulation for sterilization procedures. Do not use for these procedures.
Type of Use (Select one or both, as applicable)
| <span style="font-size: 10pt;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|--------------------------------------------------------------------------------------|
| <span style="font-size: 10pt;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the Covidien logo. The logo consists of a blue square with a white square in the center and a blue vertical line running through the center. To the right of the square is the word "COVIDIEN" in blue, sans-serif font. The logo is simple and modern.
## 510(k) Summary
Date summary prepared: June 10, 2016
#### 510(k) Submitter/Holder
Covidien, llc 5920 Longbow Drive Boulder, CO 80301
### Contact
Sharon McDermott Senior Product Specialist, Regulatory Affairs Telephone: 303-581-6789 Fax: 303-516-6313 Email: sharon.1.mcdermott@medtronic.com
### Name of Device
Trade Name: BiZact™ Open Sealer/Divider Common Name: Bipolar electrosurgical instrument Classification Name: Electrosurgical cutting and coagulation device and accessories (21 CFR § 878.4400, class II. GEI).
#### Predicate Devices
The BiZact, Open Sealer/Divider was compared and found to be substantially equivalent to the following products of comparable type in commercial distribution:
| Device Common Name: | Bipolar electrosurgical instrument |
|---------------------|------------------------------------------|
| Trade Name: | LigaSure™ Small Jaw, Open Sealer/Divider |
| Catalog Number: | LF1212A |
| 510(k) Number: | K152286 |
| Manufacturer: | Covidien, llc |
### Device Description
The BiZact Open Sealer/Divider is sterile, single use, hand-held electrosurgical device that incorporates RF tissue fusion technology for a desired tissue effect when used with the Valleylab™ LS10 Vessel Sealing Generator for ligation of vessels, tissue bundles, and lymphatics during open general surgical procedures.
The BiZact attaches to the Valleylab LS10 Vessel Sealing Generator with a 10 foot cord containing a connector. The generator is able to identify the BiZact device via the RFID tag embedded in the connector. The generator delivers precise energy through the device to tissue for a controlled tissue response to achieve complete and permanent tissue fusion by incorporating tissue sensing circuitry to constantly measure the electrical resistance of the tissue and instantaneously adjust the generator output to maintain the desired power.
{5}------------------------------------------------
### Indications for Use
The BiZact device is a bipolar instrument intended for use in open surgical procedures where ligation and division of vessels, tissue bundles, and lymphatics is desired.
The tissue fusion function of the device can be used on vessels (arteries and veins) and lymphatics up to and including 3 mm diameter. The BiZact device is indicated for use in open general surgical procedures.
It is also indicated for adult ENT procedures, including tonsillectomy, for the ligation and division of vessels, tissue bundles and lymphatics 2-3 mm away from unintended thermally sensitive structures.
The BiZact device has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use for these procedures.
#### Testing Summary:
Verification and validation results demonstrate that the BiZact Open Sealer/Divider performs as intended and is substantially equivalent to its predicate, LF1212A. The following summarizes testing conducted to establish safety and substantial equivalence:
- IEC 60601-1 Basic Electrical Safety ●
- IEC 60601-1-2 Electromagnetic Compatibility ●
- IEC 60601-2-2 Basic Safety for HF Equipment and HF Accessories ●
- ISO 10993-1 Biocompatibility ●
- Performance Testing - Device Functionality
- Performance Testing Bench Tissue burst testing ●
- Performance Testing In-vivo acute (hemostasis, thermal spread, lymphatic burst pressure,) ● and chronic (hemostasis)
- Usability - Design Validation - Summative testing in simulated use environment
### Conclusion on Substantial Equivalence
The proposed BiZact Open Sealer/Divider is substantially equivalent to the predicate LigaSure Small Jaw Open Sealer/Divider. Both devices have the same intended use and fundamental technology. The proposed device has some intentional design and indications differences that enable it to be well suited for its indicated uses. Testing demonstrated that these differences do not raise any new questions of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.