← Product Code [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI) · K160128

# Endopath Electrosurgery Probe Plus II (K160128)

_Ethicon Endo-Surgery, LLC · GEI · Apr 28, 2016 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K160128

## Device Facts

- **Applicant:** Ethicon Endo-Surgery, LLC
- **Product Code:** [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI.md)
- **Decision Date:** Apr 28, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

The ENDOPATH Electrosurgery Probe Plus II system has application in minimally invasive procedures to facilitate tissue dissection, coagulation, irrigation, and fluid evacuation through a common trocar sleeve.

## Device Story

Monopolar electrosurgical probe used in minimally invasive surgery; facilitates tissue dissection, coagulation, irrigation, and fluid evacuation via trocar sleeve. Device consists of handle and interchangeable shafts; connects to standard monopolar electrosurgery unit (ESU). Surgeon operates device manually; activation via hand buttons or ESU foot pedal. Provides suction/irrigation and hemostasis; improves surgical ergonomics and activation force compared to predicate. Benefits patient by enabling precise tissue management and fluid control during laparoscopic procedures.

## Clinical Evidence

No human clinical trial data. Evidence based on bench testing (flow rates, impedance, mechanical forces) and an acute animal study comparing tissue effects (dissection, hemostasis) against the predicate device. Results confirmed subject device performs as well as the predicate.

## Technological Characteristics

Monopolar electrosurgical device; suction/irrigation modalities. Materials evaluated per ISO 10993:2009. Sterilization via gamma radiation. Complies with IEC 60601-1:2005 (electrical safety) and IEC 60601-1-2:2007 (EMC). Features include improved button force, ergonomic design, and dual hand/foot activation capability.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- Ethicon Evacuation/Irrigation/Electrosurgical Device ([K912492](/device/K912492.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Ethicon Endo-Surgery, LLC David Locke Senior Regulatory Affairs Specialist 4545 Creek Road Cincinnati, Ohio 45242

Re: K160128

Trade/Device Name: Endopath Electrosurgery Probe Plus II Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: January 19, 2016 Received: January 20, 2016

Dear David Locke:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Jennifer R. Stevenson -A

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

K160128

Device Name ENDOPATH Electrosurgery Probe Plus II

#### Indications for Use (Describe)

Indications

The ENDOPATH Electrosurgery Probe Plus II system has application in minimally invasive procedures to facilitate tissue dissection, coagulation, irrigation, and fluid evacuation through a common trocar sleeve.

Contraindications

· These instruments are not intended for contraceptive coagulation of fallopian tissue but may be used to achieve

hemostasis following transection of the fallopian tube.

· These instruments are not intended for use when minimally invasive techniques are contraindicated.

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### 510(k) Summary

# Company

Ethicon Endo-Surgery 475 Calle C Guaynabo, PR 00969

# Contact

David Locke, DRSc, MS, RAC Senior Regulatory Affairs Program Lead Ethicon Endo-Surgery, Inc. 4545 Creek Road Cincinnati, OH 45242 Telephone: (513) 337-1281 Fax: (513) 337- 4366 Email: dlocke1 @its.jnj.com

# Date Prepared 19 January 2016

#### Device Name

Trade Name: Evacuation/Irrigation/Electrosurgical Device Classification Name: Laparoscope, General & Plastic Surgery Common Name: Endopath® Electrosurgery Probe Plus II Catalog Number: EPH02, EPS01, EPS02, EPS03, EPS04, EPS05, EPS06, EPS07

Device Class Class II

Device Panel General & Plastic Surgery

Product Code GEI

# Predicate Device

Ethicon Evacuation/Irrigation/Electrosurgical Device cleared under K912492

#### Device Description

The Endopath® Electrosurgery Probe Plus II device (product codes beginning with EPH and EPS) work with a standard monopolar unit and has applications for minimally invasive procedures to facilitate tissue dissection, coagulation, irrigation, and fluid evacuation through a common trocar sleeve. The subject device system offers one handle (EPH02) and seven shafts (EPS01-EPS07) which allow for a variety of configurations to best meet the surgeon's needs. Shafts EPS01 through EPS07 are for use through a 5 mm diameter surgical trocar or a larger trocar with a 5 mm reducer.

{4}------------------------------------------------

#### Indications for Use

The ENDOPATH Electrosurgery Probe Plus II system has application in minimally invasive procedures to facilitate tissue dissection, coagulation, irrigation and fluid evacuation through a common trocar sleeve.

#### Contraindications

- These instruments are not intended for contraceptive coagulation of fallopian tissue . but may be used to achieve hemostasis following transection of the fallopian tube.
- These instruments are not intended for use when minimally invasive techniques are . contraindicated.

# Technological Characteristics

The subject ENDOPATH Electrosurgery Probe Plus II device and predicate Ethicon Evacuation/Irrigation/Electrosurgical device use the same monopolar technology to perform their intended use. The subject and predicate devices also use the same standard suction and irrigation modalities to perform their intended use. The subject ENDOPATH Electrosurgery Probe Plus II device is similar to the predicate Ethicon Evacuation/Electrosurgical device with respect to the primary modes of action which include suction, irrigation and hemostasis. A few technological differences were identified between the subject device and the currently marketed predicate devices, attributable to the respective design of each. As compared with the predicate devices, the subject device does however incorporates several design enhancements which includes suction and irrigation button force improvement, monopolar activation force improvement, elimination of the button pin, improved ergonomics, and combined hand and foot activation capability (hand activation through the ENDOPATH Electrosurgery Probe Plus II device itself and foot activation through a monopolar electrosurgery unit (ESU) generator foot pedal).

# Performance Data

Bench testing and laboratory evaluations were conducted to demonstrate that the ENDOPATH Electrosurgery Probe Plus II performed as intended.

#### Sterilization

The subject device and the predicate device are sterilized via gamma radiation; both devices are sterilized to the same sterility assurance level.

#### Biocompatibility

The biocompatibility evaluation for the ENDOPATH Electrosurgery Probe Plus II device was conducted in accordance with the following standards: ISO 10993:2009. Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process and FDA Blue Book Memorandum #G95-1: Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing". The device passed all biocompatibility ISO testing which included cytotoxicity, sensitization, intracutaneous reactivity and acute system toxicity testing.

{5}------------------------------------------------

# EMC

Electrical safety and EMC testing were conducted on the ENDOPATH Electrosurgery Probe Plus II; the system complies with IEC 60601-1-2:2007 for electromagnetic compatibility and IEC 60601-1:2005 for electrical safety.

# Bench

Flow rate of suction and irrigation, impedance, minimum distal retraction force and sheath extension force were evaluated for the ENDOPATH Electrosurgery Probe Plus II to support substantial equivalent to the predicate device. Data generated from these tests met the predetermined acceptance criteria.

# Acute

Testing was performed in an acute study with the ENDOPATH Electrosurgery Probe Plus II vs. the currently marketed predicate device to demonstrate that the tissue effects were not different than the predicate device. The results of the study demonstrated the ability of the subject device to create tissue planes, dissect tissue, and provide hemostasis in different tissue types.

# Clinical

This premarket notification does not rely on human clinical trial data to demonstrate substantial equivalence.

# Conclusion

The results of the bench testing and laboratory evaluations in an animal model demonstrate that the Endopath® Electrosurgery Proble Plus II device is as safe and effective and performs as well as the identified legally marketed predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K160128](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K160128)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
