← Product Code [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI) · K160041

# Cobbra RF Tissue Dissector (K160041)

_Elliquence, LLC · GEI · Mar 7, 2016 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K160041

## Device Facts

- **Applicant:** Elliquence, LLC
- **Product Code:** [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI.md)
- **Decision Date:** Mar 7, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Cobbra RF Tissue Dissector is intended for use by a physician familiar with resection, incision, and hemostasis in soft tissue surgical procedures. The types of surgery intended are: General surgery, Laparoscopic procedures, Endoscopic procedures, Open abdominal, Orthopedic coagulation, Thorascopic coagulation, Neurosurgical coagulation Gynecological coagulation, (except for use in female sterilization), Ear, Nose, Throat coagulation.

## Device Story

Single-use, sterile, monopolar electrosurgical instrument; used exclusively with elliquence Surgi-Max Plus RF generator. Device consists of plastic handpiece, partially-insulated electrode with active edge, cable, and plug. RF electrical power from generator transfers to tissue via active electrode edge; time-varying voltage induces tissue cutting or coagulation. Used in OR settings by physicians for soft tissue surgical procedures. Output provides hemostasis and tissue resection/incision. Benefits include precise tissue management during surgery.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing only, including mechanical testing, thermal damage measurements, simulated use, biocompatibility (ISO 10993-1), and electrical safety/EMC (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2).

## Technological Characteristics

Monopolar electrosurgical instrument. Materials: 17-4PH stainless steel, ABS plastic, PFA, PVC, copper. Biocompatibility per ISO 10993-1. Energy source: RF generator (Surgi-Max Plus). Five models varying by electrode diameter (0.25 to 1.5 inches). EtO-sterilized, single-use. Electrical safety per IEC 60601-1, IEC 60601-1-2, IEC 60601-2-2.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- Elliquence Electrodes ([K142410](/device/K142410.md))
- PEAK Surgery System ([K082786](/device/K082786.md))

## Reference Devices

- elliquence Surgi-Max Plus radiofrequency generator ([K100390](/device/K100390.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 7, 2016

Elliquence LLC. Mr. Paul Buhrke IV QA/RA Manager 2455 Grand Avenue Baldwin, New York 11510

Re: K160041

Trade/Device Name: Cobbra RF Tissue Dissector Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: January 6, 2016 Received: January 8, 2016

Dear Mr. Buhrke:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Jennifer R. Stevenson -A

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K160041

Device Name Cobbra RF Tissue Dissector

#### Indications for Use (Describe)

The Cobbra RF Tissue Dissector is intended for use by a physician familiar with resection, incision, and hemostasis in soft tissue surgical procedures. The types of surgery intended are: General surgery, Laparoscopic procedures, Endoscopic procedures, Open abdominal, Orthopedic coagulation, Thorascopic coagulation, Neurosurgical coagulation Gynecological coagulation, (except for use in female sterilization), Ear, Nose, Throat coagulation.

| Type of Use (Select one or both, as applicable)                                       |                                                          |
|---------------------------------------------------------------------------------------|----------------------------------------------------------|
| <div> <span>☑</span> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image is a logo that features a stylized white outline of a person's back and shoulder. The person is facing away from the viewer, and the outline emphasizes the spine and shoulder muscles. The background is a dark blue circle with a glossy effect, giving it a three-dimensional appearance. A white shadow is visible at the bottom right of the circle, adding depth to the image.

Page 1 of 3

510 (k) Summary

(As required by 21 CFR 807.92(a))

Date Prepared

January 6, 2016

# Submitter's Information (807.92(a)(1))

Company Name and Address:

Elliquence, LLC. 2455 Grand Avenue Baldwin, NY 11510 Phone: (516) 277-9000 Fax: (516) 277-9001 www.elliquence.com

Establishment Registration for Elliquence, LLC is 3007024186.

Contact Information:

Daniel Connell - Regulatory Affairs Associate Phone: (516) 277-9036 Fax: (516) 277-9011 Email: dconnell@eliquence.com

Paul Buhrke – QA/RA Manager Phone: (516) 277-9010 Fax: (516) 277-9011 Email: pbuhrke@elliquence.com

## Device Information (807.92(a)(2))

Trade Name Cobbra™ RF Tissue Dissector

Common/Usual Name Monopolar electrosurgical device

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for "elliquence Innovative Medical Solutions". The word "elliquence" is written in a bold, sans-serif font, with the "e" at the beginning stylized as a circle with a dot inside. Below "elliquence" is the phrase "Innovative Medical Solutions" in a smaller, sans-serif font. The logo is in black and white.

Page 2 of 3

Classification Name and Regulation Electrosurgical Cutting and Coagulation Device and Accessories; 21 CFR 878.4400.

Class FDA Classification: Class II FDA Product Code: GEI

## Predicate Devices (807.92(a)(3))

- Elliquence Electrodes (510(k) K142410)
- PEAK Surgery System (510(k) K082786) .

#### Device Description (807.92(a)(4))

The Cobbra™ RF Tissue Dissector is a single-use, sterile, monopolar electrosurgical instrument intended for use exclusively with the elliquence Surgi-Max® Plus radiofrequency generator (510(k) K100390). The basic design of the device is a plastic handpiece, a partiallyinsulated electrode with an uninsulated, active edge, and a cable and plug. The electrical power operating at radio frequency (RF) is transferred to tissue at the surgical site via the active edge of the Cobbra™ RF Tissue Dissector's electrode. The time-varying voltage produced by RF electrical power source vields a predetermined electrosurgical effect, such as tissue cutting or coagulation.

The Cobbra™ RF Tissue Dissector is classified as a sterile, surgically-invasive, active device with an intended patient contact period of <1hr (transient). It is composed of 17-4PH medical-grade stainless steel, ABS plastic, PFA, PVC, and copper. The biocompatibility of patient contacting materials has been assured through appropriate testing in accordance with ISO 10993-1.

The Cobbra™ RF Tissue Dissector is individually-packaged and supplied EtO-sterilized for single use. It is available in five models which differ only by the diameter of the electrode: RF-COBB/XS (0.25 in.), RF-COBB/S (0.375 in.), RF-COBB/M (0.5 in.), RF-COBB/L (1.0 in.), and RF-COBB/XL (1.5 in.).

#### Intended Use (807.92(a)(5))

The Cobbra™ RF Tissue Dissector is intended for use by a physician familiar with resection, dissection, incision, and hemostasis in soft tissue surgical procedures. The types of surgery intended are: General surgery, Laparoscopic procedures, Endoscopic procedures, Open

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image is a company logo for "elliquence Innovative Medical Solutions". The logo is in black font and the words "Innovative Medical Solutions" are smaller than the company name. Below the logo is the text "510(k) K160041 Page 3 of 3".

abdominal, Orthopedic coagulation, Thorascopic coagulation, Neurosurgical coagulation, Gynecological coagulation, (except for use in female sterilization), and Ear, Nose, Throat coaqulation.

## Substantial Equivalence Comparison (807.92(a)(6))

The Cobbra™ RF Tissue Dissector is substantially equivalent in its intended use, technology/principle of operation, materials, and performance to the two chosen predicate devices:

- (1) elliquence Electrodes (510(k) K142410)
- (2) PEAK Surgery System (510(k) K082786)

An extensive comparison chart and discussion is provided within the 510(k) submission.

## Non-Clinical Testing (807.92(b)(1))

Bench testing performed to support substantial equivalence included:

- Various performance tests including mechanical testing, thermal damage measurements, . and simulated use tests
- Applicable biocompatibility tests conducted in accordance with ISO 10993-1 ●
- . Electrical safety and electromagnetic compatibility testing in accordance with IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-2
- . Testing to support the shelf life determination and sterility of the product

# Clinical Testing (807.92(b)(2))

This premarket submission did not rely on the assessment of clinical performance data to demonstrate substantial equivalence.

## Conclusion (807.92(b)(3))

Based upon a comparison of the intended uses, technological characteristics, and performance testing, we have concluded that the subject device is as safe and effective as the predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K160041](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K160041)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
