← Product Code [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI) · K150959

# Sutter Bipolar Forceps-Calvian (K150959)

_Sutter Medizintechnik GmbH · GEI · Sep 10, 2015 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K150959

## Device Facts

- **Applicant:** Sutter Medizintechnik GmbH
- **Product Code:** [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI.md)
- **Decision Date:** Sep 10, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

Sutter Bipolar Forceps- Calvian are intended to remove tissue and control bleeding.

## Device Story

Sutter Bipolar Forceps - Calvian are reusable electrosurgical instruments used for tissue removal and hemostasis. The device consists of stainless steel forceps with varying handle styles and tip sizes (0.7-2.5 mm), partially coated with Polyamide PA 11 for electrical insulation. The forceps connect to an electrosurgical generator via a bipolar cable. Operated by surgeons in clinical settings, the device delivers bipolar electrical energy to the target tissue to achieve coagulation. The device is provided non-sterile and requires steam sterilization by the user before initial and subsequent use. Maximum peak voltage is 500 Vp. The device functions as an accessory to an electrosurgical generator, allowing the surgeon to precisely control bleeding and remove tissue during surgical procedures.

## Clinical Evidence

Bench testing only. The device was evaluated for compliance with IEC 60601-2-2 standards for electrosurgical accessories. No clinical data was required or provided.

## Technological Characteristics

Materials: Medical grade stainless steel, Polyamide PA 11 coating. Energy: Bipolar electrosurgical energy (max 500 Vp). Form factor: Reusable forceps with various handle styles and tip sizes (0.7-2.5 mm). Connectivity: Bipolar cable connection to electrosurgical generator. Sterilization: Steam-sterilization by user. Standards: Complies with IEC 60601-2-2.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- Microfrance Electrosurgical Instruments ([K993655](/device/K993655.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features the department's name encircling a symbol. The symbol is a stylized representation of human profiles facing right, suggesting a sense of community and care. The text is arranged in a circular pattern around the symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 10, 2015

Sutter Medizintechnik GmbH Ulrike Zeissler Manager, Regulatory Affairs Tullastrasse 87 79108 Freiburg Germany

Re: K150959

Trade/Device Name: Sutter Bipolar Forceps - Calvian Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: April 1, 2015 Received: April 1, 2015

Dear Zeissler:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 

{1}------------------------------------------------

807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

for

Sincerely yours,

## David Krause -S

- Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation
Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K150959

Device Name Sutter Bipolar Forceps - Calvian

| Indications for Use (Describe) |
|--------------------------------|
|--------------------------------|

| Intended Use: | Sutter Bipolar Forceps- Calvian are intended to remove tissue and control bleeding. |
|---------------|-------------------------------------------------------------------------------------|
|---------------|-------------------------------------------------------------------------------------|

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

APPLICANT: SUTTER MEDIZINTECHNIK GMBH DEVICES: SUTTER BIPOLAR FORCEPS - Calvian

# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

Section 807.92(c)

| Date:<br>807.92(a)(1)                  | March 31, 2015                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                             |                             |
|----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|-----------------------------|
| Submitter:<br>807.92(a)(1)             | Name:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | SUTTER MEDIZINTECHNIK GmbH                                                                  |                             |
|                                        | Address:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Tullastrasse 87<br>79108 Freiburg<br>Germany                                                |                             |
|                                        | Managing Director:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Bert Sutter                                                                                 |                             |
|                                        | Telephone:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | +49 (0) 761 51551-0                                                                         |                             |
|                                        | Fax:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | +49 (0) 761 51551-30                                                                        |                             |
|                                        | Contact person:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Ulrike Zeissler                                                                             |                             |
|                                        | Telephone:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | +49 (0) 761 51551-14                                                                        |                             |
|                                        | E-mail:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | zeissler@sutter-med.de                                                                      |                             |
| Product:<br>807.92(a)(2)               | Trade Name:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Sutter Bipolar Forceps - Calvian                                                            |                             |
|                                        | Classification:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Class II; CFR 21 § 878.4400                                                                 |                             |
|                                        | Common Name:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Electrosurgical Instruments and Accessories                                                 |                             |
|                                        | Product Code and<br>Classification Name:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | GEI - Electrosurgical Cutting & Coagulation<br>Device & Accessories                         |                             |
| Predicate<br>Device:<br>807.92(a)(3)   | Predicate device to which Sutter Bipolar Forceps - Calvian are claimed<br>to be substantially equivalent is manufactured by<br>• Xomed Surgical Products, Microfrance Electrosurgical Instruments,<br>various, (K993655)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                             |                             |
| Device<br>Description:<br>807.92(a)(4) | Sutter Bipolar Forceps - Calvian is an electrosurgical tool available in<br>different handle styles and tip sizes. It is constructed with medical grade<br>stainless steel, coated with Polyamide PA 11 as electrical insulator and<br>possesses tips made of stainless steel that are partly not insulated. The<br>forceps can be connected through an appropriate bipolar cable with the<br>bipolar output of an electrosurgical generator. As an electrosurgical<br>accessory it is intended to remove tissue and control bleeding. The<br>maximum peak voltage to use the forceps is 500 Vp. The forceps are<br>provided non-sterile, are reusable and must be sterilized prior initial and<br>subsequent use. |                                                                                             |                             |
| Intended<br>Use:<br>807.92(a)(5)       | Intended Use: Sutter Bipolar Forceps - Calvian are intended to remove<br>tissue and control bleeding.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                             |                             |
| Substantial<br>Equivalence:            | Sutter Bipolar Forceps - Calvian are substantial equivalent to the<br>predicate devices since the basic features, design and intended uses<br>are the same.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                             |                             |
|                                        | Feature                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Sutter Calvian                                                                              | Predicate device<br>K993655 |
|                                        | Intended Use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | As shown above under<br>Intended Use                                                        | Same                        |
|                                        | Branches Style                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Angled                                                                                      | Straight, angled            |
|                                        | Dimensions<br>Tip size [mm]:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | 0.7-2.5                                                                                     | 1.3 - 2.5                   |
|                                        | Material<br>Tips:<br>Coating:<br>Connector:<br>Base material:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Stainless steel<br>Polyamide (PA) 11<br>Stainless steel/Plastic material<br>Stainless steel | Same                        |
|                                        | Bipolar                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | yes                                                                                         | Same                        |
|                                        | Meets IEC 60601-2-2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | yes                                                                                         | Same                        |
|                                        | Maximum peak voltage                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | 500 Vp                                                                                      | 833 Vp                      |
|                                        | Sterility                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Non-sterile, reusable                                                                       | Same                        |
|                                        | Sterilization method                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Steam-sterilisation by user                                                                 | Same                        |

{4}------------------------------------------------

## 510(k) Summary for K150959

APPLICANT: SUTTER MEDIZINTECHNIK GMBH DEVICES: SUTTER BIPOLAR FORCEPS - Calvian

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K150959](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K150959)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
