← Product Code [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI) · K143662

# Bissinger Monopolar Forceps and Cables (K143662)

_Medagent, Inc. · GEI · Mar 12, 2015 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K143662

## Device Facts

- **Applicant:** Medagent, Inc.
- **Product Code:** [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI.md)
- **Decision Date:** Mar 12, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The monopolar forceps and clamps are designed to grasp, dissect and coagulate tissue. They must be connected to the monopolar output of an electrosurgical generator and must only be used with parameters for monopolar coagulation. Do not exceed a maximum output of 2000 Vp of your generator.

## Device Story

Bissinger Monopolar Forceps and Cables are surgical instruments used to grasp, dissect, and coagulate tissue. The forceps feature straight or bayonet-shaped handles and are connected to high-frequency electrosurgical generators via attached monopolar cables. The cables conduct electrical power from the generator to the instrument. The system requires a neutral electrode for operation. Used by surgeons in clinical settings, the device facilitates tissue coagulation through monopolar current. The output is controlled by the electrosurgical generator, with a maximum voltage limit of 2000 Vp for forceps and 6250 Vp for cables. The device benefits patients by providing precise tissue management during surgery.

## Clinical Evidence

Bench testing only. Testing included electrical safety per EN 60601-1:2006 and EN 60601-2-2:2009, validation of steam sterilization, manual cleaning validation, and biocompatibility assessment per ISO 10993-1.

## Technological Characteristics

Reusable monopolar forceps (straight/bayonet) and cables. Materials meet ISO 10993-1 biocompatibility requirements. Electrical safety per EN 60601-1 and EN 60601-2-2. Steam sterilizable. Cables feature dual spring connectors compatible with Erbe, Martin, Berchtold, Bovie, Valleylab, and Codmann units.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- EGON FAULHABER Bipolar Forceps, Non Stick Bipolar Forceps and Monopolar Forceps ([K101080](/device/K101080.md))
- Black and Black Electrosurgical Cutting and Coagulation Forceps and Electrodes ([K072124](/device/K072124.md))
- Olsen Medical Electrosurgical Cable/Adapters ([K111262](/device/K111262.md))
- Suffer Electrosurgical Cables ([K073450](/device/K073450.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, arranged in a row. Above the profiles is a curved line that resembles a wing or a stylized wave. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the image.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 12, 2015

Medagent Incorporated Mr. Frank Menean Managing Director 112 Corporate Drive, Unit 1 Portsmouth, New Hampshire 03801

Re: K143662

Trade/Device Name: Bissinger Monopolar Forceps and Cables Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: November 25, 2014 Received: December 23, 2014

Dear Mr. Menean:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

## Jennifer R. Stevenson -S

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the word "MEDAGENT" in a stylized font, with the letters "ME" in black and the letters "AGENT" in gray. The word is enclosed in parentheses, with the left parenthesis in gray and the right parenthesis in gray. Below the word "MEDAGENT" are the words "Individual · Cooperative · Efficient" in a smaller font.

### Indications for Use Statement

510(k) Number: K143622

Device Name: Monopolar Forceps

Indications for Use:

The monopolar forceps and clamps are designed to grasp, dissect and coagulate tissue. They must be connected to the monopolar output of an electrosurgical generator and must only be used with parameters for monopolar coagulation.

Do not exceed a maximum output of 2000 Vp of your generator.

Prescription Use X (Part 21 CFR 801 Subpart D) Over-The-Counter Use _____________ (21 SFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

AND/OR

Concurrence of CDRH, Office of Device Evaluation (ODE)

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for MEDAGENT. The word "MEDAGENT" is written in a combination of black and gray letters, with the "MED" in black and the "AGENT" in gray. Below the logo, the words "Individual · Cooperative · Efficient" are written in a smaller font.

... Kok ' ** &

## VOLUME 006

# 510(k) Summary

DATE OF APPLICATION: 2014-01-07

### Type of 510(K) submission: Traditional 510(K) submission

APPLICANT: MEDAGENT Inc. 112 Corporate Drive, Unit 1 Portsmouth, NH 03801 USA Tel.: +1.603.570.2039 Fax: +1.603.570.2055 E-Mail: mail@medagent-usa.com Internet: www.medagent-usa.com Establishment Registration Number: 3007219401

- CONTACT PERSON: Franz Menean Tel.: +49.7463.9954.0 E-Mail: franz.menean@medagent.de

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for MEDAGENT. The word "MEDAGENT" is written in bold, gray letters, and it is surrounded by parentheses. Below the word, the words "Individual," "Cooperative," and "Efficient" are written in a smaller font, separated by dots.

#### Device Name 1.

| Trade Names:         | Bissinger Monopolar Forceps and Cables                          |
|----------------------|-----------------------------------------------------------------|
| Common Name:         | Monopolar Instruments and Accessoires                           |
| Classification Name: | Electrosurgical cutting and coagulation device and accessories. |

#### Classification Product Code / Subsequent Code 2.

| Device                                   | Medical<br>Specialty | Review Panel                 | Product Code | Device Class | Regulation Number |
|------------------------------------------|----------------------|------------------------------|--------------|--------------|-------------------|
| Monopolar Instruments<br>and Accessoires | Part 878             | General & Plastic<br>Surgery | GEI          | 2            | 878.4400          |

#### 3. Prior Submissions

There have been no prior submissions of the subject devices so far.

#### Predicate Device 4.

Bissinger's Monopolar Forceps and Cables are substantially equivalent to the following predicate devices, already cleared by the FDA:

| Predicate Device                                                                        | 510(k) Number | 510(k) Holder                                 |
|-----------------------------------------------------------------------------------------|---------------|-----------------------------------------------|
| EGON FAULHABER Bipolar Forceps, Non<br>Stick Bipolar Forceps and Monopolar For-<br>ceps | K101080       | EGON FAULHABER Surgical Instruments Pinzetten |
| Black and Black Electrosurgical Cutting and<br>Coagulation Forceps and Electrodes       | K072124       | Black & Black surgical Inc.                   |
| Olsen Medical Electrosurqical Ca-<br>ble/Adapters                                       | K111262       | Olsen Medical                                 |
| Suffer Electrosurqical Cables                                                           | K073450       | Suffer Medizintechnik GmbH                    |

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the word "MEDAGENT" in a stylized font, enclosed in parentheses. The first three letters, "MED", are in black, while the last four letters, "AGENT", are in gray. Below the word, there are three words separated by dots: "Individual", "Cooperative", and "Efficient". The words are in a smaller font size than "MEDAGENT". The overall design appears to be a logo or branding element.

#### Description of the Device 5.

Bissinger's Monopolar Forceps are used in various surgical applications. Therefore they are varying in size and build-up, though the general principle remains unchanged.

By using the firmly attached connecting cable, the monopolar forceps can be connected to high-frequency electrosurgical generators.

#### Monopolar Forceps 5.1.

Monopolar Forceps are used for grasping, dressing and coagulating of tissue. They must be connected to a monopolar output of a high frequency generator by using the attached monopolar cable. The application of monopolar current and connection with a neutral electrode is mandatory.

Bissinger's monopolar forceps either have a straight or bayonet-shaped handle. They are equipped with a firmly attached connecting cable with dual spring connector.

#### Monopolar Cables 5.2.

Bissinger Cables are a line of non-sterile, reusable Monopolar Cables fitting Erbe, Martin, Berchtold, Bovie, Valleylab, Codmann Electrosurgical units. They are designed to conduct electrical power from the output of a high frequency generator to the instrument.

#### Indications for Use 6.

#### 6.1. Monopolar Forceps

The monopolar forceps and clamps are designed to grasp, dissect and coagulate tissue. They must be connected to the monopolar output of an electrosurgical generator and must only be used with parameters for monopolar coagulation.

Do not exceed a maximum output of 2000 Vp of your generator

#### 6.2. .Monopolar Cables

Cables for electrosurgery are designed to conduct electrical power from the output of a high-frequency electrosurgical generator to the electrosurgical instrument.

Do not exceed a maximum output of 6250 Vp of your generator.

#### Technological Characteristics 7.

The technological characteristics of Bissinger's Forceps and Cables are identical or substantial equivalent as the previously cleared predicate devices stated in section 4 in terms of design, dimensions, intended use and materials.

#### 8. Testing

Testing in order to proof safety and effectiveness of Bissinger's Monopolar Forceps and Cables has been performed according to recognized consensus Standards and results are conforming to the respective requirements.

#### 8.1. Electrical Safety

The devices subject to this submission have been tested according to the requirements of EN 6061-1:2006 and EN 60601-2-2:2009.

<sup>001 510</sup>K Summary v.4

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for MEDAGENT. The word "MEDAGENT" is written in a sans-serif font, with the "MED" portion in black and the "AGENT" portion in gray. The word is enclosed in parentheses. Below the word, the words "Individual · Cooperative · Efficient" are written in a smaller font.

#### 8.2. Sterilization

The sterilization process has been validated under consideration of recognized standards. Testing shows that the Products can be steam sterilized with a sufficient sterlity assurance level by use of standard sterilization parameters.

#### 8.3. Reprocessing

Reprocessability was tested and validated under consideration of recognized standards.

#### Manual cleaning 8.4.

Manual cleaning validation was performed under consideration of recognized standards.

#### Biocompatibility 9.

All requirements of biocompatibility are met through the composition of the used materials which demonstrate the appropriate levels of biocompatibility for its clinical use. The used materials are also used in many other medical devices and have an established history of safe use and biocompatibility outlined in ISO 10993-1.

### 10. Substantial Equivalence Summary / Conclusion

Based on available 510(k) information provided herein, Bissinger's Monopolar Forceps and Cables are considered substantial equivalent to the predicate devices in terms of indications for use, material, technoloqy, design and performance specifications.

There are no differences between the devices which would raise new issues of safety or effectiveness.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K143662](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K143662)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
