← Product Code [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI) · K143161

# Force FX-8(S), Force FX-8C(S), Force FX-8CA(S) (K143161)

_Covidien · GEI · Dec 19, 2014 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K143161

## Device Facts

- **Applicant:** Covidien
- **Product Code:** [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI.md)
- **Decision Date:** Dec 19, 2014
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Force FX™ Electrosurgical Generator is an electrosurgical generator containing monopolar and bipolar technology. It is intended for use with accessories during surgical procedures where the surgeon requires electrosurgical cutting (resecting, dividing, or separating) and coagulating (hemostasis).

## Device Story

Microcontroller-based isolated output generator; provides RF power for cutting, desiccating, and fulgurating tissue. Inputs: tissue impedance (sensed via electrodes), user settings (mode/power), activation signals (handswitch/footswitch). Processing: Instant Response™ technology adjusts output voltage based on tissue resistance to maintain consistent effect; Autobipolar function (on specific models) uses impedance sensing to auto-start/stop RF activation. Outputs: RF energy for surgical application; audible activation tones. Used in OR/surgical settings by surgeons. Benefits: consistent tissue effect across varying densities; REM™ system monitors return electrode contact to prevent pad site burns; supports simultaneous activation.

## Clinical Evidence

No clinical data. Substantial equivalence demonstrated via design verification testing, software regression/validation, and conformance to IEC 60601-1:2005, IEC 60601-2-2:2009, and IEC 60601-1-2:2007 standards.

## Technological Characteristics

Microcontroller-based isolated output generator. Features: Instant Response™ technology, REM™ contact quality monitoring, monopolar/bipolar modes. Connectivity: RF activation port, RS-232 serial port, expansion port. Standards: IEC 60601-1:2005, IEC 60601-2-2:2009, IEC 60601-1-2:2007. Modifications include five PCBAs and firmware/software updates.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- Force FX ([K944602](/device/K944602.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of a human profile in triplicate, suggesting a sense of community and care. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the central image, emphasizing the organization's name and national scope. The logo is presented in black and white, giving it a clean and official appearance.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 19, 2014

Covidien Ms. Sharon McDermott Senior Product Regulatory Specialist 5920 Longbow Drive Boulder, Colorado 80301

Re: K143161 Trade/Device Name: Force FX Generators Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: December 4, 2014 Received: December 5, 2014

Dear Ms. McDermott:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{1}------------------------------------------------

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Binita SDAshar -S

Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K143161

Device Name Force FX Generators

#### Indications for Use (Describe)

The Force FX™ Electrosurgical Generator is an electrosurgical generator containing monopolar and bipolar technology. It is intended for use with accessories during surgical procedures where the surgeon requires electrosurgical cutting (resecting, dividing, or separating) and coagulating (hemostasis).

| Type of Use (Select one or both, as applicable)                          |                                                                         |
|--------------------------------------------------------------------------|-------------------------------------------------------------------------|
| <div> <span>☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | <div> <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains the word "COVIDIEN" in blue font. To the left of the word is a blue square with a white rectangle in the center and a light blue vertical line through the center. The word "COVIDIEN" is in a sans-serif font and is slightly offset to the right of the square. The image is likely a logo for a medical device company.

# 510(k) Summary

Date summary prepared: December 18, 2014

#### 510(k) Submitter/Holder

Covidien (formerly Valleylab, Inc.) 5920 Longbow Drive Boulder, CO 80301

#### Contact

Sharon McDermott Sr. Regulatory Affairs Product Specialist Telephone: 303-581-6789 Fax: 303-530-6313 Email: sharon.mcdermott@covidien.com

#### Name of Device

| Trade Name:          | Force FX TM electrosurgical generators                                                            |
|----------------------|---------------------------------------------------------------------------------------------------|
| Common Name:         | Electrosurgical Generator                                                                         |
| Classification Name: | Electrosurgical cutting and coagulation device and accessories (21 CFR § 878.4400, class II, GEI) |

# Predicate Device

| Trade Name:     | Force FX                   |
|-----------------|----------------------------|
| Common Name:    | Electrosurgical Generator  |
| Catalog Number: | Force FX                   |
| 510(k) Number:  | K944602 (cleared 6/5/1995) |
| Manufacturer:   | Covidien                   |

#### Device Description

The Force FX™ Electrosurgical Generators are microcontroller-based isolated output generators that provide power for cutting, desiccating, and fulgurating tissue during bipolar and monopolar surgery. Features included are:

- . Instant Response™ Technology - Automatically senses resistance and adjusts the output voltage to maintain a consistent effect across different tissue density.
- Three bipolar modes: precise, standard, and macro Low voltage, continuous current provides faster . desiccation without sparking.
- Three monopolar Cut modes: low, pure, blend Allows a wide range of power settings necessary to ● perform diverse surgical procedures.
- . Three monopolar Coag modes: desiccate, fulgurate, and spray - Helps control the size of the area and the depth of penetration during tissue coagulation.
- Support for Simultaneous Activation ●
- Covidien Energy-based Devices REM™ Contact Quality Monitoring System Monitors the quality . of electrical contact between the return electrode and the patient to eliminate risk of pad site burns.
- Handswitch or footswitch activation
- . Recall of most recently used mode and power settings

{4}------------------------------------------------

- Adjustable activation tone volume
- An RF activation port, RS-232 serial port, and expansion port .

The precise and standard bipolar modes of the Force FX™-8CA(S) models are equipped with an Autobipolar function. The autobipolar feature senses tissue impedance between the two bipolar electrodes and then uses the impedance information to automatically start or stop bipolar RF activation. Optionally, the user may choose between footswitch start and auto start, or program a delay between auto start and RF activation.

#### Indications for Use

The Force FX™ Electrosurgical Generator is an electrosurgical generator containing monopolar and bipolar technology. It is intended for use with accessories during surgical procedures where the surgeon requires electrosurgical cutting (resecting, dividing, or separating) and coagulating (hemostasis).

#### Technological Characteristics

The Force FX generators have the same technological and performance characteristics as the predicate, K944602. This Special 510(k) proposes changes to five printed circuit board assemblies (PCBAs), firmware/software, and a minor modification to the simultaneous (coag) activation tone. The function of the devices has not changed.

# Performance Characteristics

The modifications made to the Force FX generator were found to not affect safety or performance through design verification testing, which continued conformance to applicable technical design specifications and performance requirements, including requirements associated with industry safety and performance standards, as follows:

- Basic safety and essential performance in accordance with IEC 60601-1:2005 and IEC 60601-2-. 2:2009
- Electromagnetic compatibility in accordance with IEC 60601-1-2:2007
- . Design verification testing including application of multiple AC mains voltages and temperatures to ensure the proposed modifications perform within design parameters under varying environmental conditions.
- . Software regression/validation

# Clinical Studies

This premarket submission did not rely on the assessment of clinical performance data to demonstrate substantial equivalence.

# Conclusions

The intended use of the Force FX generator was not altered. In addition, the results of testing demonstrate that the modifications to the Force FX generator do not affect the safety or performance of the generator which is substantially equivalent to the predicate Force FX.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K143161](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K143161)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
