← Product Code [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI) · K142350

# PK J-HOOK (K142350)

_Gyrus Acmi, Inc. · GEI · Jan 8, 2015 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K142350

## Device Facts

- **Applicant:** Gyrus Acmi, Inc.
- **Product Code:** [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI.md)
- **Decision Date:** Jan 8, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The PK J-Hook is indicated for resection and coagulation of soft tissue and blood vessels in laparoscopic and general surgical procedures when used with the Olympus ESG-400 Generator.

## Device Story

PK J-Hook is a bipolar electrosurgical instrument for laparoscopic and general surgery; used for soft tissue and blood vessel resection and coagulation. Device features active 'J' shaped tip; activated via handle buttons or foot pedal. Physician operates device through 5mm cannula or laparoscope working channel. Device connects to Olympus ESG-400 generator; generator automatically presets output settings upon connection. Nosecone allows electrode tip rotation independent of handle. Benefits include precise tissue management and coagulation without removing eyes from surgical site.

## Clinical Evidence

Bench testing only. Performance testing included dimensional measurements, cutting and coagulation equivalency, force testing, design feature testing (rotation/activation), shelf life, sterilization validation, and biocompatibility testing per ISO 10993 standards.

## Technological Characteristics

Bipolar electrosurgical instrument. Materials: fluoropolymer sheath, metal electrode tip. Dimensions: compatible with 5mm cannula/working channel. Energy: bipolar RF via Olympus ESG-400 generator. Connectivity: proprietary connection to generator for automatic preset configuration. Sterilization: ethylene oxide (ANSI/AAMI/ISO 11135-1). Compliance: IEC 60601-1, IEC 60601-2-2, ISO 10993 series.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- Gyrus Plasmacision and Plasmablend Electrodes ([K050460](/device/K050460.md))

## Reference Devices

- Olympus ESG-400 generator ([K141225](/device/K141225.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol depicts a stylized caduceus, with a double helix intertwined with a staff, representing health and medicine.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 8, 2015

Gyrus ACMI Incorporated Mr. Neil Kelly Senior Regulatory Affairs Specialist 136 Turnpike Road Southborough, Massachusetts 01772

Re: K142350

Trade/Device Name: PK J-Hook Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: December 2, 2014 Received: December 3, 2014

Dear Mr. Kelly:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{1}------------------------------------------------

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

for

Sincerely yours,

# David Krause -S

Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K142350

Device Name

PK J-Hook

Indications for Use (Describe)

The PK J-Hook is indicated for resection and coagulation of soft tissue and blood vessels in laparoscopic and general surgical procedures when used with the Olympus ESG-400 Generator.

Type of Use (Select one or both, as applicable)

| <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)     </span> | <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)     </span> |
|---------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|
|---------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

### 510(k) Summary of Safety and Effectiveness Gyrus ACMI, Inc.

## Olympus PK J-Hook

### General Information

| Manufacturer:                                                                                 | Gyrus ACMI, Inc.<br>6655 Wedgwood Road<br>Maple Grove, MN 55311<br>Phone: 508-804-2739                                                 |
|-----------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|
| Establishment Registration Number:                                                            | 2183680                                                                                                                                |
| 510(k) Submitter:                                                                             | Gyrus ACMI, Inc.<br>136 Turnpike Rd.<br>Southborough, MA 01772-2104                                                                    |
| Establishment Registration Number:                                                            | 3003790304                                                                                                                             |
| Contact Person:                                                                               | Neil Kelly<br>Senior Regulatory Affairs Specialist<br>508-804-2690<br>Neil.kelly@olympus-osta.com                                      |
| Date Prepared:                                                                                | August 22, 2014                                                                                                                        |
| Device Description                                                                            |                                                                                                                                        |
| Classification Name:<br>Regulation number<br>Product code<br>Regulatory class<br>Review Panel | Electrosurgical cutting and coagulation<br>device and accessories<br>21 CFR 878.4400<br>GEI<br>Class II<br>General and Plastic Surgery |
| Trade Name:                                                                                   | Olympus - PK J-Hook                                                                                                                    |
| Generic/Common Name:                                                                          | Electrosurgical cutting and coagulation<br>device                                                                                      |
| Predicate Devices                                                                             |                                                                                                                                        |

Gyrus Plasmacision and Plasmablend Electrodes

K050460

{4}------------------------------------------------

## Comparison to Predicate Devices:

The Olympus - PK J-Hook has been compared to our own legally marketed Gyrus PlasmaCision and Plasmablend Electrodes (K050460) with respect to intended use and technological characteristics. The comparison and testing results presented in this 510(k) notification to the FDA show that the device is substantially equivalent to predicate devices and raises no new concerns or safety or effectiveness.

### Product Description

The PK J-Hook is a bipolar electrosurgical instrument with the capability to cut and coagulate soft tissue and blood vessels in laparoscopic and general surgery. The instrument will pass through a 5mm cannula or through an operating laparoscope working channel of 5mm or larger diameter. The device has an active "J" shaped tip and is activated via buttons on the handle, or by a foot pedal. The proposed device plugs into the Olympus ESG-400 generator (K141225). The generator and device make up a medical electrical system. The instrument is to be used only with the Gyrus ESG-400 Generator.

#### Technological Characteristics

The PK J-Hook uses bipolar energy in order to cut and coagulate soft tissue and blood vessels in laparoscopic and general surgical procedures. For safety and convenience the ESG-400 generator recognizes and automatically presets the default output settings once the proposed PK J-Hook is connected.

The PK J-Hook is activated using buttons located on the device handle, or via a foot pedal. This allows the physician to activate either cut or coagulation (coag) mode without taking their eyes off the surgical site. A nosecone located at the distal end of the handle and at the proximal end of the device shaft allows the physician to alter the orientation of the electrode tip without altering the orientation of the handle.

#### Material

The predicate and proposed devices share many common materials. The two patient contact material differences are the sheath, which is now flouropolymer rather than Polyimide tubing, and a new ink was added on the device shaft as well. Biocompatibility testing has been carried out with passing results. As for the electrode tip all materials remain the same as the predicate.

#### Intended Uses

The PK J-Hook is indicated for resection and coagulation of soft tissue and blood vessels in laparoscopic and general surgical procedures when used with the Olympus ESG-400 Generator.

{5}------------------------------------------------

### Compliance to Voluntary Standards

The design of the proposed device complies with the following standards:

ISO 10993-1. 2009 ISO 10993-5, 2009 ISO 10993-7 2008 ISO 10993-10, 2010 ANSI/AAMI/ISO 11607-1, 2006 ANSI/AAMI/ISO 11135-1, 2007 ISO 14971, 2007 ISO 15223-1; 2012 IEC 60601-1: 2005 IEC 60601-2-2: 2009

#### Summary of Sterilization and Shelf Life Discussion

The Olympus - PK J-Hook is delivered in a sterile state and is intended for single patient use only. The sterilization method used is ethylene oxide and has a shelf life of three (3) years.

#### Summary of Performance Testing

The following performance tests were conducted:

- Dimensional Measurements ●
- Cutting and Coagulation equivalency to predicate ●
- Expected forces on devices ●
- Design feature testing (rotation and button activation) ●
- Shelf Life .
- Sterilization ●
- Biocompatibility ●

#### Substantial Equivalence

The proposed Olympus - PK J-Hook has the same intended use, design, and scientific technology as the Predicate Gyrus ACMI Plasmacision and Plasmablend electrodes (K050460). Both devices are of similar design and technology and have been shown to perform substantially equivalent in bench testing. There were no new issues of safety or effectiveness with the proposed device.

#### Conclusion:

In summary, the Olympus - PK J-Hook is substantially equivalent to the predicate devices and presents no new questions of safety or efficacy.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K142350](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K142350)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
