← Product Code [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI) · K142154

# PK NEEDLE (K142154)

_Gyrus Acmi, Inc. · GEI · Dec 19, 2014 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K142154

## Device Facts

- **Applicant:** Gyrus Acmi, Inc.
- **Product Code:** [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI.md)
- **Decision Date:** Dec 19, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

The PK Needle is indicated for resection of soft tissue in laparoscopic and general surgical procedures when used with the Olympus ESG-400 Generator.

## Device Story

PK Needle is a bipolar electrosurgical instrument for soft tissue resection in laparoscopic and general surgery. Device features blunt needle-shaped tip; passes through 5mm cannula or laparoscope working channel. Operated by surgeon via handle button or foot pedal. Plugs into Olympus ESG-400 generator, which automatically recognizes device and presets output settings. Device delivers electrosurgical energy to tissue for cutting and coagulation. Benefits include precise soft tissue resection during minimally invasive procedures.

## Clinical Evidence

Bench testing only. Testing included dimensional measurements, cutting equivalency, mechanical force testing, design feature verification, shelf life, sterilization validation, and biocompatibility testing per ISO 10993 standards.

## Technological Characteristics

Bipolar electrosurgical instrument. Materials: fluoropolymer sheath, polyimide-equivalent electrode tip/insulation. Dimensions: compatible with 5mm cannula/working channel. Energy: electrosurgical (bipolar) via ESG-400 generator. Connectivity: proprietary connection for automatic generator preset recognition. Sterilization: ethylene oxide (ISO 11135-1). Standards: ISO 10993, ISO 14971, IEC 60601-1, IEC 60601-2-2.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- Everest Bipolar Needle Electrode ([K031079](/device/K031079.md))

## Reference Devices

- Olympus ESG-400 Generator ([K141225](/device/K141225.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized image of an eagle with three lines representing its wings and head. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 19, 2014

Gyrus ACMI Incorporated Mr. Neil Kelly, MBA, RAC Senior Regulatory Affairs Specialist 136 Turnpike Road Southborough, Massachusetts 01772

Re: K142154

Trade/Device Name: PK Needle Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: November 25, 2014 Received: November 26, 2014

Dear Mr. Kelly:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Binita S. Ashar -S

Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### Indications for Use

510(k) Number (if known) K142154

Device Name PK Needle

Indications for Use (Describe)

The PK Needle is indicated for resection of soft tissue in laparoscopic and general surgical procedures when used with the Olympus ESG-400 Generator.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

__ Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary of Safety and Effectiveness Gyrus ACMI, Inc.

# Olympus PK Needle

## General Information

| Manufacturer:                                                                                 | Gyrus ACMI, Inc.<br>6655 Wedgwood Road<br>Maple Grove, MN 55311<br>Phone: 508-804-2739                                                 |
|-----------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|
| Establishment Registration Number:                                                            | 2183680                                                                                                                                |
| 510(k) Submitter:                                                                             | Gyrus ACMI, Inc.<br>136 Turnpike Rd.<br>Southborough, MA 01772-2104                                                                    |
| Establishment Registration Number:                                                            | 3003790304                                                                                                                             |
| Contact Person:                                                                               | Neil Kelly<br>Senior Regulatory Affairs Specialist<br>508-804-2690<br>Neil.kelly@olympus-osta.com                                      |
| Date Prepared:                                                                                | August 1, 2014                                                                                                                         |
| Device Description                                                                            |                                                                                                                                        |
| Classification Name:<br>Regulation number<br>Product code<br>Regulatory class<br>Review Panel | Electrosurgical cutting and coagulation<br>device and accessories<br>21 CFR 878.4400<br>GEI<br>Class II<br>General and Plastic Surgery |
| Trade Name:                                                                                   | Olympus - PK Needle                                                                                                                    |
| Generic/Common Name:                                                                          | Electrosurgical cutting and coagulation<br>device                                                                                      |
| Predicate Devices                                                                             |                                                                                                                                        |

Everest Bipolar Needle Electrode Gyrus Bipolar Needle

K031079

{4}------------------------------------------------

# Comparison to Predicate Devices:

The Olympus - PK Needle has been compared to our own legally marketed Everest Bipolar Needle Electrode (K031079) with respect to intended use and technological characteristics. The comparison and testing results presented in this 510(k) notification show that the device is substantially equivalent to predicate devices and raises no new concerns or safety or effectiveness.

## Product Description

The PK Needle is a bipolar electrosurgical instrument with the capability to resect soft tissue and blood vessels in laparoscopic and general surgical procedures. The instrument will pass through a 5mm cannula or through an operating laparoscope working channel of 5mm or larger diameter. The device has an active blunt needle shaped tip and is activated via a button on the handle, or by a foot pedal. The proposed device plugs into the Olympus ESG-400 generator (K141225). The generator and device make up a medical electrical system. The instrument is to be used only with the Gyrus ESG-400 Generator.

#### Technological Characteristics

For safety and convenience the Olympus ESG-400 generator recognizes and automatically presets the default output settings once the proposed PK Needle is connected.

The PK Needle is activated using a button located on the device handle, or via a foot pedal.

# Material

The predicate and proposed devices share many common materials. The two patient contact material differences are the sheath, which is now flouropolymer rather than Polyimide tubing, and a new ink was added on the device shaft as well. Biocompatibility testing has been carried out with passing results. As for the electrode tip and insulation all materials remain the same as the predicate.

#### Intended Uses

The PK Needle is indicated for resection of soft tissue in laparoscopic and general surgical procedures when used with the Olympus ESG-400 Generator.

{5}------------------------------------------------

# Compliance to Voluntary Standards

The design of the proposed device complies with the following standards:

ISO 10993-1. 2009 ISO 10993-5, 2009 ISO 10993-7 2008 ISO 10993-10, 2010 ANSI/AAMI/ISO 11607-1, 2006 ANSI/AAMI/ISO 11135-1, 2007 ISO 14971, 2007 ISO 15223-1; 2012 IEC 60601-1: 2005 IEC 60601-2-2: 2009

## Summary of Sterilization and Shelf Life Discussion

The Olympus - PK Needle is delivered in a sterile state and is intended for single patient use only. The sterilization method used is ethylene oxide and has a shelf life of three (3) years.

## Summary of Performance Testing

The following performance tests were conducted:

- Dimensional Measurements
- Cutting equivalency to predicate ●
- Expected forces on devices ●
- Design feature testing (button activation) ●
- Shelf Life .
- Sterilization ●
- Biocompatibility ●

#### Substantial Equivalence

The proposed Olympus - PK Needle has the same intended use, design, and scientific technology as the Predicate Everest Bipolar Needle Electrode (K031079). Both devices are of similar design and technology and have been shown to perform substantially equivalent in bench testing. There were no new issues of safety or effectiveness with the proposed device.

#### Conclusion:

In summary, the Olympus - PK Needle is substantially equivalent to the predicate devices and presents no new questions of safety or efficacy.

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K142154](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K142154)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
