THUNDERBEAT

K132703 · Olympus Medical Systems Corp. · GEI · May 19, 2014 · General, Plastic Surgery

Device Facts

Record IDK132703
Device NameTHUNDERBEAT
ApplicantOlympus Medical Systems Corp.
Product CodeGEI · General, Plastic Surgery
Decision DateMay 19, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The THUNDERBEAT hand instruments are intended to be used with the Ultrasonic Generator (USG-400), the Electrosurgical Generator (ESG-400), and the THUNDERBEAT Transducer, (TD-TB400). Seal & Cut mode: The THUNDERBEAT hand instruments when used in combination with the Seal & Cut mode are indicated for open, laparoscopic (including single-site surgery) general surgery and gynecological surgery (including urologic, thoracic, plastic and reconstructive, bowel resections, cholecystectomies, Nissen fundoplication, adhesiolysis, oophorectomy, hysterectomies (both vaginal assisted and abdominal) etc) and endoscopic surgery or in any procedure in which cutting, vessel ligation (sealing and cutting), coagulation, grasping, and dissection is performed. These devices have been designed to seal and cut vessels (up to and including 7 mm in diameter), tissue bundles, and lymphatics. Seal mode: The THUNDERBEAT hand instruments when used in combination with the Seal mode are indicated for open, laparoscopic (including single-site surgery) general surgery and gynecological surgery (including urologic, thoracic, plastic and reconstructive, bowel resections, cholecystectomies, Nissen fundoplication, adhesiolysis, oophorectomy, hysterectomies (both vaginal assisted and abdominal) etc), and endoscopic surgery or in any procedure in which vessel sealing, coagulation, grasping is performed. These devices have been designed to seal vessels (up to and including 7 mm in diameter), tissue bundles, and lymphatics. The THUNDERBEAT hand instruments have not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures, and should not be used for these procedures. (Note: Additional indications for TB-0510IC/0520IC include open ENT procedures in adults for ligation/sealing of vessels, lymphatics, and tissue bundles away from thermally sensitive structures).

Device Story

THUNDERBEAT hand instruments are sterile, single-use surgical tools used with USG-400/ESG-400 generators and TD-TB400 transducer. Devices perform vessel sealing, cutting, coagulation, grasping, and dissection. Operation involves simultaneous HF Bipolar (FineCoag) and Ultrasonic output (Seal & Cut mode) or HF Bipolar (HardCoag) output (Seal mode) while grasping tissue between probe and grasping section. Operated by surgeons in OR settings (open, laparoscopic, endoscopic). Hand switches on grip handle control output. Output provides hemostasis and tissue division, reducing need for multiple instruments. Benefits include efficient vessel ligation up to 7mm and precise tissue management.

Clinical Evidence

No clinical data. Evidence consists of bench testing (ex-vivo sealing/cutting performance on blood vessels/mesentery, histopathological confirmation, heating comparisons) and animal testing (in-vivo performance on porcine vessels, lymphatics, and tissue bundles).

Technological Characteristics

Electrosurgical and ultrasonic device. Materials: Biocompatible, compliant with ISO-10993. Energy: Combined HF Bipolar and Ultrasonic. Form factor: Handheld instruments with pistol or inline grips, various shaft lengths (10cm-45cm). Connectivity: Connects to USG-400/ESG-400 generators. Sterilization: ISO 11135-1 compliant. Standards: AAMI/ANSI 60601-1, IEC 60601-1-1, IEC 60601-2-2, IEC 60601-1-2, IEC 60601-2-18.

Indications for Use

Indicated for patients undergoing open, laparoscopic, or endoscopic general, gynecological, urologic, thoracic, plastic, reconstructive, or ENT (adults only) surgeries. Used for vessel ligation (up to 7mm), tissue bundle/lymphatic sealing, coagulation, grasping, and dissection. Contraindicated for tubal sterilization or coagulation for sterilization.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Traditional 510(k) Notification May 5, 2014 ## 510(k) Summary Gyrus ACMI, Inc. THUNDERBEAT TB-0535PC/0545PC/0510IC/0520IC/0535IC/0545IC ## MAY 1 9 2014 #### General Information Applicant: Establishment Registration Number: Manufacturer: Establishment Registration Number: 510(k) Submitter: Establishment Registration Number: Contact Person: Date Prepared: ## Device Identification Classification Name: Regulation Number: Product Code(s): Regulatory Class: Review Panel: Generic/Common Name: Trade Name: OLYMPUS MEDICAL SYSTEMS CORP. 2951 Ishikawa-cho, Hachioji-shi, Tokyo, Japan 192-8507 Phone: (+81) 42-642-2694 Fax: (+81) 42-642-2307 8010047 Aomori Olympus 248-1 Okkonoki 2-chome Kuroishi-shi. Aomori, Japan 036-0367 Phone: (+81) 172-52-8543 Fax: (+81) 172-52-8515 9614641 Gyrus ACMI, Inc. 136 Turnpike Rd. Southborough, MA 01772-2104 3003790304 Graham Baillie Manager, Regulatory Affairs May 5, 2014 Electrosurgical cutting and coagulation device and accessories 21 CFR 878.4400 GEI, LFL II. Unclassified General & Plastic Surgery Ultrasonic and electrosurgical devices THUNDERBEAT Hand Instruments 9 {1}------------------------------------------------ Traditional 510(k) Notification May 5, 2014 | Model Name | Device Name | |------------|------------------------------------| | TB-0545PC | THUNDERBEAT 5mm, 45cm, Pistol Grip | | TB-0535PC | THUNDERBEAT 5mm, 35cm, Pistol Grip | | TB-0545IC | THUNDERBEAT 5mm, 45cm, Inline Grip | | TB-0535IC | THUNDERBEAT 5mm, 35cm, Inline Grip | | TB-0520IC | THUNDERBEAT 5mm, 20cm, Inline Grip | | TB-0510IC | THUNDERBEAT 5mm, 10cm, Inline Grip | ## Predicate Devices | Olympus Medical Systems Corp. THUNDERBEAT TB-0535PC, | | |------------------------------------------------------|---------| | 0545PC, 0510IC, 0520IC, 0535IC and 0545IC | K111202 | | Covidien LigaSure Small Jaw | K113572 | ## Product Description The THUNDERBEAT hand instruments are intended to be used with the Ultrasonic Generator (USG-400), the Electrosurgical Generator (ESG-400), and the THUNDERBEAT Transducer, (TD-TB400). The THUNDERBEAT hand instruments are provided as sterile, single use devices. These are functional devices capable of vessel sealing & cutting, tissue coagulating & cutting, grasping, dissecting. These instruments have been designed to seal and cut vessels up to and.including 7 mm in diameter, tissue bundles, and lymphatics for open, laparoscopic, open ENT and endoscopic procedures in surgery. ## Technological Characteristics The THUNDERBEAT activates combined HF Bipolar (FineCoag) output and Ultrasonic output [Sea] and Cut model simultaneously while grasping a tissue or a vessel between the Probe and the Grasping section. The THUNDERBEAT also performs the HF Bipolar (HardCoag) output [Seal mode]. Those outputs lead to sealing and cutting/sealing of vessels or coagulating and cutting soft tissue. The hand switches provided on the grip handle enable those output operations. However, the technological features of the subject devices are identical to the predicate (K111202) such as the shaft length, shaft rotation, and shaft diameter. ## Material The patient contacting materials are well established and identical to the predicate devices cleared under K111202. Full biocompatibility testing on all patient contacting surfaces has been performed in compliance to the relevant requirements of ISO-10993. {2}------------------------------------------------ Traditional 510(k) Notification May 5, 2014 ## Intended Uses The purpose of this submission is to demonstrate equivalence to predicate devices and to expand the Indication for Use for the existing (subject) TB-0535PC/0545PC/0535IC/05451C/0510IC/05201C. The additional indications appear as bolded text under the Intended Use section below. ## THUNDERBEAT TB-0535PC/0545PC/0535IC/0545IC The THUNDERBEAT hand instruments are intended to be used with the Ultrasonic Generator (USG-400), the Electrosurgical Generator (ESG-400), and the THUNDERBEAT Transducer, (TD-TB400). #### Seal & Cut mode: The THUNDERBEAT hand instruments when used in combination with the Seal & Cut mode are indicated for open, laparoscopic (including single-site surgery) general surgery and gynecological surgery (including urologic, thoracic, plastic and reconstructive, bowel resections, cholecystectomies, Nissen fundoplication, adhesiolysis, oophorectomy, hysterectomies (both vaginal assisted and abdominal) etc) and endoscopic surgery or in any procedure in which cutting, vessel ligation (sealing and cutting), coagulation, grasping, and dissection is performed. These devices have been designed to seal and cut vessels (up to and including 7 mm in diameter), tissue bundles, and lymphatics. #### Seal mode: The THUNDERBEAT hand instruments when used in combination with the Seal mode are indicated for open, laparoscopic (including single-site surgery) general surgery and gynecological surgery (including urologic, thoracic, plastic and reconstructive, bowel resections, cholecystectomies, Nissen fundoplication, adhesiolysis, oophorectomy, hysterectomies (both vaginal assisted and abdominal) etc), and endoscopic surgery or in any procedure in which vessel sealing, coagulation. grasping is performed. These devices have been designed to seal vessels (up to and including 7 mm in diameter), tissue bundles, and lymphatics. The THUNDERBEAT hand instruments have not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures, and should not be used for these procedures. {3}------------------------------------------------ K132703 Page 4 of 6 ## THUNDERBEAT TB-0510IC/0520IC The THUNDERBEAT hand instruments are intended to be used with the Ultrasonic Generator (USG-400), the Electrosurgical Generator (ESG-400), and the THUNDERBEAT Transducer, (TD-TB400). ## Seal & Cut mode: The THUNDERBEAT hand instruments when used in combination with the Seal & Cut mode are indicated for open, laparoscopic (including single-site surgery) general surgery and gynecological surgery (including urologic, thoracic, plastic and reconstructive, bowel resections, cholecystectomies, Nissen fundoplication, adhesiolysis, oophorectomy, hysterectomies (both vaginal assisted and abdominal) etc) and endoscopic surgery or in any procedure in which cutting, vessel ligation (sealing and cutting), coagulation, grasping, and dissection is performed. These devices have been designed to seal and cut vessels (up to and including 7 mm in diameter), tissue bundles, and lymphatics. This mode is also indicated for open ENT procedures in adults for ligation (sealing and cutting) of vessels, lymphatics and tissue bundles away from unintended thermally sensitive structures. #### Seal mode: The THUNDERBEAT hand instruments when used in combination with the Seal mode are indicated for open, laparoscopic (including single-site surgery) general surgery and gynecological surgery (including urologic, thoracic, plastic and reconstructive, bowel resections, cholecystectomies, Nissen fundoplication, adhesiolysis, oophorectomy, hysterectomies (both vaginal assisted and abdominal) etc), and endoscopic surgery or in any procedure in which vessel sealing, coagulation, grasping is performed. These devices have been designed to seal vessels (up to and including 7 mm in diameter), tissue bundles, and lymphatics. This mode is also indicated for open ENT procedures in adults for sealing of vessels, lymphatics and tissue bundles away from unintended thermally sensitive structures. The THUNDERBEAT hand instruments have not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures, and should not be used for these procedures. {4}------------------------------------------------ ditional 510(k) Notification lav 5. 2014 ## Compliance to Voluntary Standards The design of the THUNDERBEAT TB-0535PC. 0545PC. 0510IC. 0520IC. 0535IC and 0545IC complies with the following standards: AAMI/ ANSI60601-1: 2005/(R)2012 and CI:2009/(R)2012 IEC60601-1-1:2000 IEC60601-2-2:2009 IEC60601-1-2:2007 IEC60601-2-18: 2009 ISO10993-1:2009 ISO11135-1:2007 ISO14971:2007 ## Summary of Sterilization and Shelf Life Discussion The sterilization had been tested to comply with ISO11135-1: 2007. Real Time Aging testing confirmed that the subject devices have a three year shelf life. ## Summary of Performance Testing The following performance testing was conducted. 1. Bench Testing Test #A: THUNDERBEAT Ex-vivo Sealing & Cutting Performance Confirmation Test on Blood Vessels Test #B: THUNDERBEAT Ex-vivo Cutting Performance Confirmation Test on Blood Vessels and Mesentery. Test #Fa: Sealing performance of the THUNDERBEAT SEAL & CUT mode Test #Fb: Sealing performance and characteristic of the THUNDERBEAT SEAL & CUT mode during activation Test #Fc: Sealing performance of the THUNDERBEAT SEAL & CUT mode Test #H: Comparison of dissecting performance between the THUNDERBEAT vs. Control devices Test #I/J: A comparison of heating at the grasping section of THUNDERBEAT Hand Instruments vs. Control devices. 13 {5}------------------------------------------------ Traditional 510(k) Notification May 5, 2014 Test #P: THUNDERBEAT Ex-vivo Sealing & Cutting and Sealing Performance -Histopathological Confirmation Test on Blood Vessels. 2. Animal Test Test Ea: THUNDERBEAT In-vivo Sealing & Cutting and Sealing Performance Confirmation Testing on Porcine Blood Vessels, Lymphatics and Tissue Bundles. #### Substantial Equivalence The subject devices have identical technology, performance, dimensions and materials. The differences between the predicate and the subject devices are expanded indications for use and additional marketing claims. To support the proposed indications for use and marketing claims, the performance tests summarized above were conducted. #### Conclusion: In summary, the Gyrus ACMI THUNDERBEAT TB-0535PC, 0545PC, 0510IC, 05201C, 0535IC and 0545IC handpieces are substantially equivalent to the predicate devices and present no new questions of safety or efficacy. {6}------------------------------------------------ Image /page/6/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines representing its body and wings. The eagle is positioned within a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is written around the circumference of the circle. #### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WObb-Golle Silver Spring, MD 20993-0002 May 19, 2014 OLYMPUS MEDICAL SYSTEMS CORPORATION % Mr. Graham Baillie Gyrus ACMI Incorporated 136 Turnpike Road Southborough, Massachusetts 01772 Rc: K132703 Trade/Device Name: Thunderbeat Regulation Number: 21 CFR 878.4400 Regulation Name; Electrosurgical, Cutting & Coagulation and Accessories Regulatory Class: Class II Product Code: GEI, LFL Dated: April 16, 2014 Received: April 18, 2014 · Dear Mr. Baillie: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {7}------------------------------------------------ Page 2 - Mr. Graham Baillie forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # David Krause -S - for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {8}------------------------------------------------ #### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ## Indications for Use 510{k} Number (if known) K132703 #### Device Name ## THUNDERBEAT TB-0535PC/0545PC/0510IC/0520IC/05351C/05451C Indications for Use (Describe) THUNDERBEAT TB-0535PC/0545PC/0535IC/05451C The THUNDERBEAT hand instruments are intended to be used with the Ultrasonic Generator (USG-400), the Electrosurgical Generator (ESG-400), and the THUNDERBEAT Transducer, (TD-TB400). #### Seal & Cut mode: The THUNDERBEAT hand instruments when used in combination with the Seal & Cut mode are indicated for open, laparoscopic (including single-site surgery) general surgery and gynecological surgery (including urologic, thoracic, plastic and reconstructive, bowel resections, cholecystectomies, Nissen fundoplication, adhesiolysis, oophorectomy, hysterectomies (both vaginal assisted and abdominal) etc) and endoscopic surgery or in any procedure in which cutting, vessel ligation (sealing and cutting), coagulation, grasping, and dissection is performed. These devices have been designed to seal and cut vessels (up to and including 7 mm in diameter), tissue bundles, and lymphatics. #### Seal mode: The THUNDERBEAT hand instruments when used in combination with the Seal mode are indicated for open, laparoscopic (including single-site surgery) general surgery and gynecological surgery (including urologic, thoracic, plastic and reconstructive, bowel resections, cholecystectomies, Nissen fundoplication, adhesiolysis, oophorectorny, hysterectomies (both vaginal assisted and abdominal) etc), and endoscopic surgery or in any procedure in which vessel sealing, coagulation, grasping is performed. These devices have been designed to seal vessels (up to and including 7 mm in diameter), tissue bundles, and lymphatics. The THUNDERBEAT hand instruments have not been shown to be effective for tubal coagulation for sterilization procedures, and should not be used for these procedures. ## THUNDERBEAT TB-0510IC/0520IC The THUNDERBEAT hand instruments are intended to be used with the Ultrasonic Generator (USG-400), the Electrosurgical Generator (ESG-400), and the THUNDERBEAT Transducer, (TD-TB400). Seal & Cut mode: The THUNDERBEAT hand instruments when used in combination with the Seal & Cut mode are indicated for open, laparoscopic (including single-site surgery) general surgery and gynecological surgery (including wologic, thoracic, laboropio (including mage on resections, cholecystectomies, Nissen fundoplication, adhesiolysis, oophorectomy, hysterectomies (both vaginal assisted and abdominal) etc) and endoscopic surgery or in any procedure in which cutting, vessel ligation (sealing and cuting), coagulation, grasping, and dissection is performed. These devices have been designed to seal and cut vessels (up to and including 7 mm in diameter), tissue bundles, and lymphatics. This mode is also indicated for open ENT procedures in adults for ligation (sealing and cutting) of vessels, lymphatics and tissue bundles away from unintended thermally sensitive structures. Seal mode: The THUNDERBEAT hand instruments when used in combination with the Seal mode are indicated for open, FORM FDA 3881 (1/14) Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below. {9}------------------------------------------------ laparoscopic (including single-site surgery) general surgery and gynecological surgery (including urologic, thoracic, plastic and reconstructive, bowel resections, cholecystectomies, Nissen fundoplication, adhesiolysis, oophorectorny, hysterectomies (both vaginal assisted and abdominal) etc), and endoscopic surgery or in any procedure in which vessel sealing, coasing is performed. These devices have been designed to seal vessels (up to and including 7 mm in diameter), tissue bundles, and lymphatics. This mode is also indicated for open ENT procedures in adults for sealing of vessels, lymphatics and tissue bundles away from unintended thermally sensitive structures. The THUNDERBEAT hand instruments have not been shown to be effective for tubal coagulation for sterilization procedures, and should not be used for these procedures. Type of Use (Select one or both, as applicable) 2 Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. FOR FOR FOR FOR FOR FOR FOR FOR FOR Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) Image /page/9/Picture/10 description: The image shows the name "Joshua C. Nipper -S" in a bold, sans-serif font. The text is black against a white background, creating a high contrast. The name appears to be a signature or a title, with the "-S" possibly indicating a designation or initial. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...