← Product Code [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI) · K113597

# MINI-LAP BIPOLAR ELECTROCAUTERY DEVICES (K113597)

_Minilap Technologies · GEI · Jun 3, 2012 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K113597

## Device Facts

- **Applicant:** Minilap Technologies
- **Product Code:** [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI.md)
- **Decision Date:** Jun 3, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The MiniLap Technologies family of BiPolar instruments with bipolar cautery have applications in a variety of general, thoracic, gynecologic (except for use in female sterilization), urologic, laparoscopic and endoscopic procedures for manipulation and coagulation of tissue.

## Device Story

Disposable bipolar electrosurgical grasper; used in general, thoracic, gynecologic, urologic, laparoscopic, and endoscopic procedures. Device features 2.4mm insulated needle/cannula shaft with retractable grasper; percutaneously inserted to surgical site without traditional conduit. Connects to standard bipolar cables and generators for tissue cauterization. Operated by surgeons in clinical settings. Provides precise tissue manipulation and coagulation; benefits include reduced invasiveness via percutaneous access.

## Clinical Evidence

Bench testing only. Mechanical and electrical performance evaluations conducted. Biocompatibility of materials confirmed via history of use in medical industry.

## Technological Characteristics

Disposable bipolar electrosurgical grasper. Materials: Medical grade stainless steel (17-7 SS, 300 SS). Dimensions: 2.4mm (+/- 0.1mm) shaft diameter; 150-300mm length. Power: Bipolar electrosurgical generator (70W max). Features: Retractable grasper, 175-degree shaft rotation, safety interlock, insulated shaft/jaws. Compliant with IEC electrical safety standards. Sterilization: Sterile, single-use.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- Wolf Medical Instruments Bipolar Forceps ([K023813](/device/K023813.md))
- MINI LAP Instruments ([K070686](/device/K070686.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# 510(k) Summary

# JUN - 3 2012

### Date Prepared [21 CFR 807.92(a)(1)] November 1, 2011, Updated May 31, 2012

### Submitter's Information [21 CFR 807.92(a)(1)]

Regulatory Contact Allan Alward 88 Ashford Avenue Dobbs Ferry, NY 10522

Sponsor/Manufacturer Mini Lap Technologies, Inc. 88 Ashford Avenue Dobbs Ferry, NY 10522 Contact: Dr. S. Ravikumar Tel: 914 591 8400

FDA Establishment Registration Number is 3007123990

### Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)|

Trade Names

- MINI LAP Bipolar Electrocautery Devices ◆
Device Common, Usual or Classification Names

Laparoscopic Instruments, Graspers, Cannula, Trocar, Manual Surgical Instruments, Bipolar Instruments

#### Classification Panel

Classification of this device would fall under the responsibility of the General & Plastic Surgery panel.

#### Class

Class 2 device under the following product codes/regulations:

- KOG, 21 CFR 876.1500, Endoscope Accessories
- 트 FBQ, 21 CFR 878.5090, Trocar
- . GEI, 21 CFR 878.4400, Electrosurgical cutting and coagulation accessories

{1}------------------------------------------------

#### Predicate Device [21 CFR 807.92(a)(3)]

The following devices have been identified as predicate devices:

- . Wolf Medical Instruments Bipolar Forceps - K023813
- MINI LAP Instruments K070686 .

### Description of the Device [21 CFR 807.92(a)(4)]

The MiniLap Bipolar Electrocautery devices are a family of disposable bipolar electrosurgical graspers that directly penetrate soft tissue to access certain areas of the human anatomy without the need for a traditional insertion conduit. These disposable instruments are used for electrosurgical cauterization during general laparoscopic procedures.

Fabricated from a stainless steel alloy, MiniLap Bipolar instruments consist of an integrated 2.4mm (+/- .1mm) insulated needle/cannula shaft that houses a retractable grasper instrument. The shaft can be introduced percutaneously to the surgical site, after which the working portion of the instrument can be deployed to approximate and cauterize soft tissue. Each BiPolar instrument has a male connection provided as an integral part of the instrument that may be utilized for the bipolar connection when attached to standard two plug bipolar cables and their generators. The bipolar connection is compatible with manufacturers of standard dual female bipolar plug cables.

The instruments come in multiple lengths from 150mm to 300mm.

#### Intended Use [21 CFR 807.92(a)(5)]

The MiniLap Technologies family of BiPolar instruments with bipolar cautery have applications in a variety of general, thoracic, gynecologic (except for use in female sterilization), urologic, laparoscopic and endoscopic procedures for manipulation and coagulation of tissue.

#### Technological Characteristics [21 CFR 807.92(a)(6)]

We believe the MINI LAP Bipolar Electrocautery Devices are substantially equivalent to the predicate devices. see chart

{2}------------------------------------------------

| Characteristic       | Mini Lap Technologies                                                                                                                                                                                                                                                                                       | Richard Wolf Medical<br>Instruments Corp.                                                                           | Mini Lap Technologies                                                                                                                                                                                   |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                      | BiPolar Electrocautery Devices                                                                                                                                                                                                                                                                              |                                                                                                                     | MINI LAP Instruments                                                                                                                                                                                    |
|                      |                                                                                                                                                                                                                                                                                                             | Bipolar Forceps                                                                                                     |                                                                                                                                                                                                         |
|                      |                                                                                                                                                                                                                                                                                                             |                                                                                                                     |                                                                                                                                                                                                         |
| 510(k)               | K113597                                                                                                                                                                                                                                                                                                     | K023813                                                                                                             | K070686                                                                                                                                                                                                 |
| Indications for Use  | The MiniLap Technologies family<br>of BiPolar instruments with<br>bipolar cautery have applications<br>in a variety of general, thoracic,<br>gynecologic (except for use in<br>female sterilization), urologic,<br>laparoscopic and endoscopic<br>procedures for manipulation and<br>coagulation of tissue. | For use in open and laparoscopic<br>surgery where grasping,<br>coagulating, and transecting<br>tissue is indicated. | Penetrate soft tissue to access<br>certain areas of the human<br>anatomy. The devices are used to<br>grasp, hold and manipulate other<br>soft internal tissues as well as<br>items such as hernia mesh. |
| Needle diameter      | 2.4mm ± .1mm                                                                                                                                                                                                                                                                                                | 3.5 mm, 5 mm                                                                                                        | 2.4 mm                                                                                                                                                                                                  |
| Device length        | 150-300 mm                                                                                                                                                                                                                                                                                                  | 320 mm, 450 mm                                                                                                      | 150-300 mm                                                                                                                                                                                              |
| Material Composition | Medical Grade Stainless Steel<br>17-7 SS, 300 SS                                                                                                                                                                                                                                                            | Medical Grade Stainless Steel                                                                                       | Medical Grade Stainless Steel<br>470 SS, 300 SS                                                                                                                                                         |
| Sterilization        | Sterile, single use                                                                                                                                                                                                                                                                                         | Non sterile, reusable                                                                                               | Sterile, single use                                                                                                                                                                                     |
| Power Source         | Bipolar Electrosurgical Generator                                                                                                                                                                                                                                                                           | Bipolar Electrosurgical Generator                                                                                   | N/A                                                                                                                                                                                                     |
| Power Ranges         | 70 watts maximum, 20-35 watts<br>normal operating range                                                                                                                                                                                                                                                     | 50 watts maximum, 20-35 watts<br>normal operating range                                                             |                                                                                                                                                                                                         |
| Safety               | Safety Interlock                                                                                                                                                                                                                                                                                            | Safety Interlock is not applicable                                                                                  | Safety Interlock                                                                                                                                                                                        |
|                      | Insulated shaft and jaws                                                                                                                                                                                                                                                                                    | Insulated jaws                                                                                                      |                                                                                                                                                                                                         |
|                      | Compliant to applicable IEC<br>electrical safety standards                                                                                                                                                                                                                                                  | Compliant to applicable IEC<br>electrical safety standards                                                          |                                                                                                                                                                                                         |
| Rotation             | 175 degree shaft rotation, each<br>direction                                                                                                                                                                                                                                                                |                                                                                                                     | 360 degree shaft rotation                                                                                                                                                                               |

# Technological Characteristics Comparison Chart

### Performance Data [21 CFR 807.92(b)(1)]

The subject device has been subjected to and passed a variety of mechanical and electrical tests and evaluations. Additionally, the device is composed of biocompatible materials with a history of usage in the medical device industry.

# Conclusion [21 CFR 807.92(b)(3)]

We believe the changes are minimal and conclude that the subject device is as safe and effective as the predicate device.

{3}------------------------------------------------

### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three wing-like shapes, symbolizing health and human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

#### Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

MiniLap Technologies % Mr. Allan Alward Vice President Research and Development 88 Ashford Ave. Dobbs Ferry. NY 10522

JUN. - 3 2012

Re: K113597

Trade/Device Name: Mini-Lan Bipolar Electrocautery Devices Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI Dated: May 15, 2012 Received: May 16, 2012

Dear Mr. Alward:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{4}------------------------------------------------

# Page 2 - Mr. Allan Alward

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Sincerely yours,

Mark N. Melkerson

Director Division of Surgical, Orthopedic & Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

# Indications for Use

510(k) Number (if known): K113597

Device Name: Mini-Lap Bipolar Electrocautery Devices

Indications for Use:

The MiniLap Technologies family of BiPolar instruments with bipolar cautery have applications in a variety of general, thoracic, gynecologic (except for use in female sterilization), urologic, laparoscopic and endoscopic procedures for manipulation and coagulation of tissue.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Sign-Off

(Division Sign-Off) (Division Sign-Off)
Division of Surgical, Orthopedic.
Division of Surgical, Orthopedic. Division of Surgices and Restorative Devices

K113597

510(k) Number

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K113597](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K113597)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
