← Product Code [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI) · K111670

# VENUS FREEZE (MP)2 SYSTEM (K111670)

_Venus Concept , Ltd. · GEI · Mar 2, 2012 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K111670

## Device Facts

- **Applicant:** Venus Concept , Ltd.
- **Product Code:** [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI.md)
- **Decision Date:** Mar 2, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Venus Freeze (MP)2 system is a non-invasive device intended for use in dermatologic and general surgical procedures for the non-invasive treatment of moderate to severe facial wrinkles and rhytides in Fitzpatrick skin types I- IV.

## Device Story

Venus Freeze (MP)2 is a non-invasive, non-ablative dermatologic device. It consists of a console with a touch screen interface, an RF power module, and two applicators: Octipolar (for large areas) and Diamondpolar (for small areas). The device emits 1MHz RF energy through electrodes and generates Pulsed Magnetic Fields (PMF) via integrated coils. Operated by clinicians in a clinic setting, the user selects the applicator and adjusts RF power and treatment time based on the patient's skin condition and anatomical site. The device provides non-invasive treatment for facial wrinkles and rhytides. By combining RF and PMF technologies, the device aims to reduce facial wrinkles. Clinical performance data indicated no unexpected adverse side effects.

## Clinical Evidence

Clinical performance data involved patients treated for face wrinkle reduction. No unexpected adverse side effects were reported. The study indicates that the simultaneous use of RF and PMF technology is safe and effective.

## Technological Characteristics

Non-invasive, non-ablative electrosurgical device (21 CFR 878.4400). Emits 1MHz RF energy and PMF. Features Octipolar and Diamondpolar applicators with RF electrodes and magnetic coils. Complies with ANSI AAMI 60601-2-2, EN 60601-1, and IEC 60601-1-2 standards.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- Venus Freeze System ([K100586](/device/K100586.md))

## Reference Devices

- NeuroStar® TMS Therapy System ([K083538](/device/K083538.md))
- CIRRUS OPEN Magnetic Resonance Diagnostic Device ([K090080](/device/K090080.md))
- OPER Series Open Type Permanent-Magnet MRI ([K092899](/device/K092899.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

VENUSCONCEPT

+

Page 1 of 3

#### 510(K) SUMMARY

MAR - 2 2012

.

### Venus Freeze (MP)2 System 510(k) Number K111670

Applicant's Name: Venus Concept Ltd. 62 HaHermesh Street Karmiel, 21652, Israel Tel: (972)3-644-6656 Fax: (972)3-644-6319

- Contact Person: Yoram Levy, Osite 31 Haavoda St. Binyamina, Israel 30500 Tel (972)4-638-8837; Fax (972)4-638-0510 Yoram@gsitemed.com
Venus Freeze (MP)2 Trade Name:

Summary May 12, 2011 Preparation Date:

Classification: Name: Electrosurgical, cutting & coagulation device & accessories Product Code: GEI Regulation No: 21 CFR 878.4400 Class: II Panel: General and Plastic Surgery

#### Device Description:

The Venus Freeze (MP)2 is a noninvasive, non-ablative device consisting of:

- Main Unit (console) 1
- -Touch Screen user interface
- RF Power module -
- Controller unit ﮯ
- Two treatment applicators:

(1) Octipolar™ applicator - for large area treatments, composed of 8 RF electrodes, 8 electrode coils assembled over electrodes and 1 central coil (the RF is emitted through the electrodes. The PMF is generated by the coils).

{1}------------------------------------------------

## 670

## VENUSCONCEPT

ﻴﺴﺎﺳ

Page 2 of (3)

(2) Diamondpolar™ applicator - for small area treatments, composed of 4 RF electrodes, 4 electrode coils (the RF is emitted through the electrodes. The PMF is generated by the coils). .

Touch screen user interface provides:

- Applicator selection (Octipolar™ / Diamondpolar™) -
- RF Power Output and Treatment Time parameter adjustments to fit . individual patient's skin condition and anatomical site treated.
- Current treatment parameters display. -

The RF power module provides RF energy to the selected applicator, producing a 1MHz signal.

#### Intended Use Statement:

The Venus Freeze (MP)2 system is a non-invasive device intended for use in dermatologic and general surgical procedures for the non-invasive treatment of moderate to severe facial wrinkles and rhytides in Fitzpatrick skin types I- IV.

Predicate Devices: Substantial equivalence to the following predicate device is claimed:

| Device Name                | 510k No | Date of Clearance |
|----------------------------|---------|-------------------|
| <i>Venus Freeze System</i> | K100586 | Nov 29, 2010      |

The Venus Freeze (MP)2 magnetic flux safety claim is based on the following devices:

| Device Name                                         | 510k No | Date of Clearance |
|-----------------------------------------------------|---------|-------------------|
| NeuroStar® TMS Therapy System                       | K083538 | Dec 16, 2008      |
| CIRRUS OPEN Magnetic<br>Resonance Diagnostic Device | K090080 | March 23, 2009    |
| OPER Series Open Type<br>Permanent-Magnet MRI       | K092899 | Nov 10, 2009      |

These devices are not considered predicate devices for the Venus Freeze (MP)2. They are presented as devices that apply much higher magnetic flux densities than the Venus Freeze (MP)2 onto the same parts of the body, to support the claim that there are no questions of safety with this energy in the proposed device.

Venus Freeze – 510k Notification

{2}------------------------------------------------

KIII 670

# VENUSCORCEPT

. * *

Page 3 of (3)

#### Performance Standards

Venus Freeze (MP)2 complies with ANSI AAMI 60601-2-2 for safety of high frequency surgical equipment.

In addition, the device complies with the European Medical Directive 93/42/EEC concerning medical devices (Annex II) and with the following voluntary standards:

EN 60601-1 (Medical Electrical Equipment-Part 1: General . Requirements for Safety-1. Collateral Standard: Safety Requirements for Medical Electrical Systems).

- IEC 60601-1-2 (Medical Electrical Equipment Part 1-2: ● Standard: Electromagnetic Compatibility -Collateral Requirements and Tests)
A detailed description appears in Section 14.

#### Summary of Clinical performance data

Patients were treated with the Venus Freeze (MP)2 System for face wrinkle reduction.

No unexpected adverse side effects were detected or reported.

The data reported in this study clearly indicate that the Venus Freeze (MP) System that simultaneously use RF and PMF technology, combines the well established safe and effective Freeze RF technology with the PMF technology.

Venus Freeze - 510k Notification

{3}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three bars representing its wings, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged in a circular pattern around the eagle. The eagle is facing to the right. The logo is black and white.

#### Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

MAR - 2 2012

Venus Concept Ltd % Osite Yoram Levy 31 Haavoda St. Binyamina, Israel 30500

Re: K111670

Trade/Device Name: Venus Freeze (MP)2 Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI Dated: February 3, 2012 Received: February 7, 2012

Dear Yoram Levy:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{4}------------------------------------------------

#### Page 2 - Yoram Levy

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

## VENUSCONCEPT

### INDICATIONS FOR USE STATEMENT

#### 510(k) Number (if known): K11670

Device Name:

Venus Freeze (MP)"

Indications for Use:

The Venus Freeze (MP)2 system is a non-invasive device intended for use in dermatologic and general surgical procedures for females for the non-invasive treatment of moderate to severe facial wrinkles and rhytides in Fitzpatrick skin types I- IV.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-off) Division of General, Restorative and Neurological Devices 510(k) Number

Nil Re Ogde for mkm
(Division Sign-Off)

Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number_K111670

Venus Freeze - 510k Notification

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K111670](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K111670)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
