← Product Code [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI) · K092383

# BOVIE ICON VS GENERAL PURPOSE RF GENERATOR, MODEL: VS300 (K092383)

_Bovie Medical Corporation · GEI · Mar 10, 2010 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K092383

## Device Facts

- **Applicant:** Bovie Medical Corporation
- **Product Code:** [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI.md)
- **Decision Date:** Mar 10, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Intended Use

The ICON VS Electrosurgical Generator is intended to be used with Bovie Seal and Cut Instruments for sealing vessels and cutting tissues.

## Device Story

Bovie ICON VS Electrosurgical Generator provides RF energy for surgical cutting, coagulation, and vessel sealing; used with Bovie Seal and Cut Instruments. Device transforms electrical input into specific waveforms (frequency/duration) corresponding to user-selectable modes (e.g., Pure Cut, Blend, Coagulation, Bipolar, Vessel Seal). Operated by clinicians in surgical settings via an integrated user interface screen for setting selection. Output effects include tissue cutting, coagulation, and vessel sealing. Device benefits include precise tissue management and vessel occlusion. Differences from predicate include compatibility with specific Bovie Seal and Cut instruments, reduced vessel sealing power (90W vs 400W), and increased power cut capability (100W vs 80W).

## Clinical Evidence

Bench testing only. No clinical data provided.

## Technological Characteristics

Electrosurgical generator; RF energy source. Features user interface screen for mode selection. Waveform profiles identical to predicate. Compliance with medical electrical equipment safety and electromagnetic compatibility standards. Standalone device.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- Bovie ICON GP General Purpose RF Generator (K-082109)

## Reference Devices

- Bovie Seal and Cut Instruments (K-092208)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Page 1 of 2

## 510(k) SUMMARY (As Required per 21 CFR 807.92(c))

# 383

### GENERAL INFORMATION:

け

.

MAR 1 0 2010

. . . .

| 510k Owner's Name    | Bovie Medical                                                                                                                           |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Address              | 5115 Ulmerton Road<br>Clearwater, Florida 33760-4004                                                                                    |
| Contact Person       | Richard A. Kozloff<br>Vice-President; Quality Assurance/Regulatory Affairs<br>Telephone #: (727) 803-8513<br>FAX Number: (727) 322-4465 |
| Date Prepared:       | 07/31/2009                                                                                                                              |
| DEVICE DESCRIPTION:  |                                                                                                                                         |
| Trade Name:          | Bovie ICON VS Electrosurgical Generator                                                                                                 |
| Common Name:         | Electrosurgical Generator                                                                                                               |
| Classification Name: | Electrosurgical Cutting and Coagulation Devices and<br>Accessories (21CFR 878.4400; Class II; Product Code GEI)                         |
| Predicate Devices:   |                                                                                                                                         |

Bovie Medical

Bovie ICON GP General Purpose RF Generator (K-082109)

{1}------------------------------------------------

### 510(k) SUMMARY (As Required per 21 CFR 807.92(c))

#### DEVICE OPERATION:

The ICON VS Electrosurgical Generator is intended to be used with Bovie Seal and Cut Instruments for sealing vessels and cutting tissues.

There is a characteristic electrical wave form associated with each mode. The electrical properties of the waveform (frequency and duration) produce the clinical effect (i.e. cut, coagulation, and seal). The shape and duration of waveforms are comparable between the generator and the predicate device.

The Generator functions in any of the following user selectable modes:

- 1. Pure Cut 2. Blend Mode 1
- 3. Blend Mode 2 4. Blend Mode 3
- 5. Special Cut 6. Laparoscopic Cut
- 7. Pinpoint Coagulation 8. Gentle Coagulation
- Spray Coagulation 9. 10. Laparoscopic Coagulation
- 11. Macro Bipolar 12. Micro Bipolar
- 13. Gentle Bipolar 14. Auto Stop Bipolar
- 15. Vessel Seal with Optional Power Cut

The generator is designed to comply with applicable Medical Electrical Equipment safety standards, including electromagnetic compatibility and other safety standards.

Like the predicate device, the ICON VS incorporates an ergonomically designed user interface screen for the selection of device settings. The ICON VS uses identical technology to the predicate device and the waveform profiles for all cut, coagulation, and bipolar modes are identical. The ICON VS differs from the predicate device in that 1) it is designed to work with Bovie Seal and Cut Instruments (K-092208 pending review), 2) the vessel sealing mode on the ICON VS is less aggressive than the predicate device (90 watts of power versus 400 watts of power), and 3) the power cut mode on the ICON VS delivers 100 watts of power versus the predicate device which delivers 80 watts of power.

There are no new hazards presented with the use of the Bovie ICON VS generator as compared with the named predicate device.

#### INTENDED USE:

The ICON VS Electrosurgical Generator is intended to be used with Bovie Seal and Cut Instruments for sealing vessels and cutting tissues.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference. Inside the circle is a stylized image of a bird or abstract human figures, with flowing lines suggesting movement or connection.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room W-O66-0609 Silver Spring, MD 20993-0002

Bovie Medical Corporation % Mr. Richard A. Kozloff Vice President, Quality Assurance/ Regulatory Affairs 5115 Ulmerton Road Clearwater, Florida 33760-4004

### MAR 1 0 2010

Re: K092383

Trade/Device Name: Bovie ICON VS Electrosurgical Generator Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: February 05, 2010 Received: February 12, 2010

#### Dear Mr. Kozloff:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must

{3}------------------------------------------------

Page 2 - Mr. Richard A. Kozloff

comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Sincerely yours,

Mark N. Melkerson
Director

Division of Surgical, Orthopedic

DED DIRECTOR
3/10/0

And Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Device Name: Bovie ICON VS Electrosurgical Generator

Indications for Use:

The ICON VS Electrosurgical Generator is intended to be used with Bovie Seal and Cut Instruments for sealing vessels and cutting tissues.

イ Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Neil R. Osada Csr mxm

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K092383

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K092383](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K092383)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
