LAPAROSCOPIC AND ELECTROSURGICAL INSTRUMENTS AND ACCESSORIES, LAPAROSCOPY RUDOLF SYSTEM
K073498 · Rudolf Medical GmbH + Co. KG · GEI · Sep 19, 2008 · General, Plastic Surgery
Device Facts
Record ID
K073498
Device Name
LAPAROSCOPIC AND ELECTROSURGICAL INSTRUMENTS AND ACCESSORIES, LAPAROSCOPY RUDOLF SYSTEM
Applicant
Rudolf Medical GmbH + Co. KG
Product Code
GEI · General, Plastic Surgery
Decision Date
Sep 19, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Intended Use
Rudolf's laparoscopic and electrosurgical instruments are designed for dissecting, cutting and optional monopolar or bipolar coagulation of tissue during general surgical procedures. These devices have not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures and should not be used for these procedures.
Device Story
Laparoscopic and electrosurgical instruments used for tissue dissection, cutting, and coagulation during general surgery. Devices connect to high-frequency electrosurgical generators via cables to deliver monopolar or bipolar energy. Operated by surgeons in clinical/OR settings. Instruments vary in size and build but maintain consistent operational principles. Benefit includes precise tissue management during minimally invasive procedures. Supplied non-sterile; requires steam sterilization (pre-vacuum) at 134°C for 5 minutes before use.
Clinical Evidence
Bench testing only. Conformance to international standards (ANSI/AAMI HF18, IEC 60601-1, IEC 60601-1-1, IEC 60601-2-2, IEC 60601-2-18) provided to demonstrate safety and performance.
Technological Characteristics
Electrosurgical cutting and coagulation instruments. Monopolar and bipolar configurations. Connectivity via standard high-frequency electrosurgical generator cables. Conforms to ANSI/AAMI HF18, IEC 60601-1, IEC 60601-1-1, IEC 60601-2-2, and IEC 60601-2-18. Sterilization: Steam (pre-vacuum) at 134°C (273°F) for 5 minutes.
Indications for Use
Indicated for tissue dissection, cutting, and monopolar or bipolar coagulation during general surgical procedures. Contraindicated for tubal sterilization or tubal coagulation procedures.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
K082505 — KLS MARTIN ELECTROSURGICAL INSTRUMENTS · Kl Martin L.P. · Jun 12, 2009
K210942 — MicroFrance Monopolar and Bipolar Electrosurgical Instruments · Integra Microfrance · May 14, 2021
Submission Summary (Full Text)
{0}------------------------------------------------
| 510(k) Premarket Notification |
|-------------------------------|
|-------------------------------|
Laparoscopic and Electrosurgical Instruments
Image /page/0/Picture/2 description: The image shows the logo for Rudolf Medical. The logo consists of a stylized letter "G" followed by the word "rudolf" in a slanted, sans-serif font. Below the word "rudolf" is the word "MEDICAL" in a smaller font. The logo is black and white.
K
SEP 1 9 2008
# 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS IN ACCORDANCE WITH SMDA OF 1990
DATE: December 5, 2007
RUDOLF MEDICAL GmbH + Co. KG Submitted by: Zollerstr. 1 D-78567 Fridingen Germany Phone: +49 (7463) 9956-0 Fax: +49 (7463) 9956-56 Email: info@rudolf-med.com
#### Device Name 1.
Laparoscopic and Electrosurgical Instruments and Accessories Trade Name: Common Name: Laparoscopic and Electrosurgical Instruments and Accessories
#### Classification 2.
Our laparoscopic and electrosurgical Instruments can be classified according following Device Names and Product Codes:
| Device: | Device, Electrosurgical, Cutting & Coagulation & Accessories | |
|---------------------|-------------------------------------------------------------------------------------|--|
| Device description: | Laparoscopic and electrosurgical cutting and coagulation device and<br>accessories. | |
| Medical Specialty: | Part 878, General & Plastic Surgery | |
| Product Code: | GEI | |
| Regulation Number: | 878.4400 | |
| Device Class: | 2 | |
{1}------------------------------------------------
273495
\$\gamma^2-e\$
## 510(k) Premarket Notification
# Laparoscopic and Electrosurgical instruments
Image /page/1/Picture/23 description: The image shows the logo for Rudolf Medical. The logo consists of a stylized letter 'G' on the left, followed by the word 'rudolf' in a bold, sans-serif font. Below the word 'rudolf' is the word 'MEDICAL' in a smaller font. The logo is black and white.
#### 3. Substantial Equivalence
Rudolf's Laparoscopic and Electrosurgical Instruments are substantial equivalent to
- PAJUNK GMBH MEDIZINTECHNOLOGIE, K062047 N
- . GUNTER BISSINGER MEDIZINTECHNIK GMBH, K051429
- GUNTER BISSINGER MEDIZINTECHNIK GMBH, K042077 ■
- 피 ENABLE MEDICAL CORP., K010112
- AESCULAP, INC., K001330 피
#### Description of the Device 4.
Laparoscopic and electrosurgical instruments are used in various surgical applications. Therefore they are varying in size and build-up, though the general principle remains unchanged. By using a suitable connecting cable, the monopolar and bipolar instruments can be connected to high-frequency electrosurgical generators.
#### 5. Intended Use
Rudolf's laparoscopic and electrosurgical instruments are designed for dissecting, cutting and optional monopolar or bipolar coagulation of tissue during general surgical procedures. These devices have not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures and should not be used for these procedures.
#### Performance Standards 6.
The devices are conforming to following international standards: ANSI/AAMI HF18, IEC 60601-1, IEC 60601-1-1, IEC 60601-2-2, IEC 60601-2-18
## 7. Sterilization
Rudolf's laparoscopic and electrosurgical instruments are supplied non-sterile and must be sterilized before every use by a steam sterilization (pre-vacuum) process at 134°C (273°F) and 5 minutes minimum holding time.
#### 8. Conclusion
Based on the available 510(k) summaries and 510(k) statements (21 CFR 807) and the information provided herein, we conclude that Rudolf's Laparoscopic and Electrosurgical Instruments are substantially equivalent to the existing legally marketed devices under Federal Food, Drug and Cosmetic Act.
{2}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle.
SEP 1 9 2008
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Rudolf Medical GmbH & Co., KG % Medagent GmbH & Co., KG Mr. Franz Menean Griesweg 47 Muhlheim, Baden-Wurttemberg Germany
Re: K073498
Trade/Device Name: Rudolf Laparoscopic and Electrosurgical Instruments Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI Dated: August 23, 2008 Received: August 27, 2008
Dear Mr. Menean:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 - Mr. Franz Menean
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark N. Millman
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
1073498 p. l -sf1
# Indications for Use
510(k) Number: KO73498
Device Name: Rudolf Laparoscopic and Electrosurgical Instruments
Indications for Use:
Rudolf's laparoscopic and electrosurgical instruments are designed for dissecting, cutting and optional monopolar or bipolar coagulation of tissue during general surgical procedures.
These devices have not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures and should not be used for these procedures.
Prescription Use YES (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use NO (21 SFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mark A. Milheim
(Division Sign-Off) (Division Sign-On)
Division of Ceneral, Restorative, rological Devices 093498 and I
510(k) Number_
Page 1 of 1
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.