← Product Code [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI) · K062082

# CONMED ABC DISSECTING BLADE ELECTRODE, MODEL 139330; BLUNT NEEDLE ELECTRODE, MODEL 139331 (K062082)

_Conmed Corporation · GEI · Jan 25, 2007 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K062082

## Device Facts

- **Applicant:** Conmed Corporation
- **Product Code:** [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI.md)
- **Decision Date:** Jan 25, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

Sterile, singe-use electrosurgical accessory used in conjunction with Argon Beam Coagulation generator for delivery of argon gas and electrosurgical current to achieve cutting and coagulation of tissue.

## Device Story

ABC Dissecting Electrodes are sterile, single-use electrosurgical accessories designed for use with existing Argon Beam Coagulation (ABC) generators and handpieces. The device delivers argon gas and RF electrosurgical current to tissue. During cutting, the electrode contacts tissue within an argon gas envelope; during coagulation, the electrode is held at a distance for non-contact fulguration. The argon gas flow clears fluids from the surgical site, allowing direct coagulation on the stroma and reducing carbonization. Used in open and laparoscopic procedures by clinicians to achieve hemostasis, create homogenous eschar, and minimize tissue damage. The system allows clinicians to exchange electrodes on existing handpieces to switch between cutting and coagulation modes.

## Clinical Evidence

Bench testing only. The device was evaluated for compliance with recognized consensus standards including ANSI/AAMI HF-18 (electrosurgical devices), IEC 60601-2-2 (safety of electrosurgical equipment), ISO 14971 (risk management), and ISO 10993 (biocompatibility). No clinical data was required or provided.

## Technological Characteristics

Sterile, single-use electrosurgical accessory. Materials comply with ISO 10993 biocompatibility standards. Operates via RF energy and argon gas delivery. Designed to interface with existing cleared ABC handpieces and generators. Complies with ANSI/AAMI HF-18 and IEC 60601-2-2. No software or electronic components are contained within the electrode itself; it relies on the host generator for energy and control.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- Valleylab Force Argon™ II Argon Enhanced Electrosurgical System ([K964636](/device/K964636.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Page 1 2
k) Premarket Notification

510/k) Pre CONMED ABC Dissecting Electrodes B

K062082

### Summary of Safety and Effectiveness

| Submitted by:        | CONMED Electrosurgery Division<br>14603 East Fremont Avenue<br>Centennial, CO 80112 USA<br>Telephone: 303-269-8224<br>Facsimile: 303-699-9854 |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:      | Shawn Riedel                                                                                                                                  |
| Date Prepared:       | July 19, 2006                                                                                                                                 |
| Proprietary Name:    | CONMED ABC Dissecting Electrodes                                                                                                              |
| Common Name:         | Electrosurgical Electrode                                                                                                                     |
| Classification Name: | Electrosurgical Cutting and Coagulation Device and accessories<br>21 CFR 878.4400<br>79 GEI                                                   |
| Predicate Device:    | Valleylab Force Argon™ II Argon Enhanced Electrosurgical System<br>K964636, cleared March 11, 1997                                            |

Device Description: The ABC Dissecting Electrodes are electrodes used in coniunction with an Argon Beam Coagulation generator. The system is designed to provide a controlled flow of argon to electrosurgical handsets. The handsets allow the clinicians to perform argon enhanced Electrosurgery in open and laproscopic procedures. The electrodes attach to several previously cleared ConMed Electrosurgery ABC® hand pieces to convert from Argon Beam Coagulation only to Argon Beam Dissect.

When cutting, the edge of the electrode shall be in contact with the tissue while being enclosed in an envelope of argon gas. The electrosurgical energy is passed through the electrode as it is drawn across the tissue. When coagulating, the accessory electrode is not in contact with the tissue and is separated from the tissue by distance for fulguration to achieve the desired result. The electrodes utilizes previously cleared/marketed generators and hand pieces.

### Argon Beam Coagulation (ABC)

Argon Beam Coagulation was developed to satisfy a need for a more even spray coagulation effect. With normal electrosurgery spray coagulation may over coagulate and carbonize some areas and under coagulate others due to the randomness of the arcing process. With ABC technology the RF current is focused in a directional, non-contacting, room temperature beam of argon gas. The RF current follows the argon beam from the ABC unit to the patient's tissue and forms a reticulum of arc tunnels which are smaller, more numerous, more uniform size and depth, and more evenly distributed on the tissue. The advantages are faster hemostasis, more homogenous eschar, less tissue damage, and enhanced healing ability. The flow of argon gas also serves to clear the surgical site of fluids allowing direct coagulation on the stroma of the target tissue and reduces carbonization of the tissue to prevent floating eschar.

The sterile disposable electrodes allow the clinician the capability of using their existing hand pieces while being able to exchange the electrodes based on the clinician's desire and need clinical coagulation or cutting.

Intended Use of Device: Sterile, singe-use electrosurgical accessory used in conjunction with Argon Beam Coagulation generator for delivery of argon gas and electrosurgical current to achieve cutting and coagulation of tissue.

{1}------------------------------------------------

K062082

Page 2 of (2)

Technological Characteristics: The proposed device is equivalent to the identified predicate device with respect to technological characteristics and function. The device has been designed to comply with the applicable sections of ANSI/AAMI American National Standard for Electrosurgical Devices HF-18, the International Electrotechnical Commission Standard for Electrosurgical Devices, 60601-2-2, Risk Management ISO 14971 and Biocompatibility ISO 10993.

- There are no changes to the generator and therefore no changes to the accompanying . software.
- The operator controls the use of the device; .
- The generator provides alarms for conditions that could pose a risk to the patient; .
- The operator sets the appropriate mode and output settings for the device. .

{2}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Conmed Corporation % Mr. Shawn Riedel Vice President, Quality Assurance and Regulatory Affairs 14603 E. Fremont Avenue Centennial, Colorado 80112

JAN 2 5 2007

Re: K062082

Trade/Device Name: CONMED ABC® Hip Stem Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: BEI Dated: November 27, 2006 Received: November 29, 2006

Dear Mr. Riedel:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2 - Mr. Shawn Riedel

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115 Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

for Nom

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

K062082

# Indications for Use

510(k) Number (if known): K062082

JAN 2 5 2007

Device Name: CONMED ABC® Dissect Electrodes ™

Indications for Use:

Sterile, singe-use electrosurgical accessory used in conjunction with Argon Beam Coagulation generator for delivery of argon gas and electrosurgical current to achieve cutting and coagulation of tissue.

Prescription Use
(Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (Part 21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)

| Concurrence of CDRH, Office of Device Evaluation (ODE)                                          |             |
|-------------------------------------------------------------------------------------------------|-------------|
| (Division Sign-Off)                                                                             |             |
| Division of General, Restorative, and Neurological Devices                                      | Page 1 of 1 |
| 510(k) Number                                                                                   | 2062082     |
| Additional information for ConMed ABC <sup>  TM  </sup> Dissecting Electrodes <sup>  TM  </sup> | 5 of 11     |

Add K062082 】【

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K062082](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K062082)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
