K052686 · Arthrocare Corp. · GEI · Oct 5, 2005 · General, Plastic Surgery
Device Facts
Record ID
K052686
Device Name
ARTHROCARE ARTHROWANDS
Applicant
Arthrocare Corp.
Product Code
GEI · General, Plastic Surgery
Decision Date
Oct 5, 2005
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ArthroCare ArthroWands are indicated for resection, ablation, and coagulation of soft tissue and hemostasis of blood vessels in arthroscopic and orthopedic procedures: [Table of procedures including ACL/PCL, acromioplasty, articular cartilage, bursectomy, chondroplasty, facia, ligament, notchplasty, scar tissue, soft tissue, subacromial decompression, synovectomy, tendon, acetabular labrum, articular labrum, capsule, capsular release, cartilage flaps, cysts, discoid meniscus, frozen shoulder release, glenoidale labrum, lateral release, loose bodies, meniscal cystectomy, meniscectomy, plica removal, synovial membrane, triangular fibrocartilage (TFCC), villusectomy, carpal ligaments, glenohumeral capsule, medial retinaculum, rotator cuff, and wrist tendons across various joints including ankle, elbow, hip, knee, shoulder, and wrist].
Device Story
ArthroCare ArthroWands are bipolar, single-use, high-frequency electrosurgical devices. Used in arthroscopic and orthopedic procedures; operated by surgeons. Device delivers high-frequency energy to soft tissue for resection, ablation, and coagulation; also provides hemostasis of blood vessels. Benefits include precise tissue management during minimally invasive joint surgery. No software or complex algorithms involved.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
Bipolar, high-frequency electrosurgical device. Single-use. Designed for arthroscopic/orthopedic soft tissue management. No software or electronic processing components.
Indications for Use
Indicated for patients undergoing arthroscopic and orthopedic procedures requiring soft tissue resection, ablation, coagulation, or hemostasis of blood vessels in the ankle, elbow, hip, knee, shoulder, or wrist.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
K083306 — MODIFICATION TO ARTHROCARE ARTHROWANDS · Arthrocare Corp. · Dec 10, 2008
K020557 — ARTHROCARE CONTROLLER; ARTHROCARE CABLE; FOOT CONTROL; POWER CORD; WANDS · Arthrocare Corp. · Mar 21, 2002
Submission Summary (Full Text)
{0}------------------------------------------------
OCT 5 - 2005
K052686
## 510(k) Summary
ArthroCare Corporation ArthroCare ArthroWands
Page 1 of Page (3)
### General Information
Submitter Name/Address:
ArthroCare Corporation 680 Vaqueros Avenue Sunnyvale, CA 94085-3523
Establishment Registration Number:
Contact Person:
Director, Regulatory Affairs September 27, 2005
Valerie Defiesta-Ng
2951580
Date Prepared:
Device Description
Trade Name:
Generic/Common Name:
Classification Name:
Predicate Devices
ArthroCare® ArthroWands®
Electrosurgical Device and Accessories
Electrosurgical Cutting and Coagulation Device and Accessories (21 CFR 878.4400)
ArthroCare® ArthroWands®
K020557, K033584
#### Product Description
The ArthroCare ArthroWands are bipolar, single use, high frequency electrosurgical devices designed for specific indications in arthroscopic and orthopedic procedures.
{1}------------------------------------------------
# K052686
## ( 3) 2 6
## Intended Uses
The ArthroCare ArthroWands are indicated for resection, ablation, and coagulation of soft tissue and hemostasis of blood vessels in arthroscopic and orthopedic procedures:
| Arthroscopic and Orthopedic Procedures | Joint Specific or All<br>Joints (ankle, elbow,<br>hip, knee, shoulder, and<br>wrist) |
|----------------------------------------|--------------------------------------------------------------------------------------|
| Ablation and Debridement | |
| • ACL/PCL | Knee |
| • Acromioplasty | Shoulder |
| • Articular Cartilage | All Joints |
| • Bursectomy | All Joints |
| • Chondroplasty | All Joints |
| • Facia | All Joints |
| • Ligament | All Joints |
| • Notchplasty | Knee |
| • Scar Tissue | All Joints |
| • Soft Tissue | All Joints |
| • Subacromial Decompression | Shoulder |
| • Synovectomy | All Joints |
| • Tendon | All Joints |
| Excision and Resection | |
| • Acetabular Labrum | Hip |
| • Articular Labrum | All Joints |
| • Capsule | All Joints |
| • Capsular Release | Knee |
| • Cartilage Flaps | Knee |
| • Cysts | All Joints |
| • Discoid Meniscus | Knee |
| • Frozen Shoulder Release | Shoulder |
| • Glenoidale Labrum | Shoulder |
| • Lateral Release | Knee |
| • Ligament | All Joints |
| • Loose Bodies | All Joints |
| • Meniscal Cystectomy | Knee |
| • Meniscectomy | Knee |
| Continued | |
| Arthroscopic and Orthopedic Procedures | Joint Specific or All Joints (ankle, elbow, hip, knee, shoulder, and wrist) |
| • Plica Removal | All Joints |
| • Scar Tissue | All Joints |
| • Soft Tissue | All Joints |
| • Synovial Membrane | All Joints |
| • Tendon | All Joints |
| • Triangular Fibrocartilage (TFCC) | Wrist |
| • Villusectomy | Knee |
| Coagulation | |
| • ACL/PCL | Knee |
| • Articular Cartilage | All Joints |
| • Carpal Ligaments | Wrist |
| • Glenohumeral Capsule | Shoulder |
| • Ligament | All Joints |
| • Medial Retinaculum | Knee |
| • Rotator Cuff | Shoulder |
| • Tendon | All Joints |
| • Wrist Tendons | Wrist |
{2}------------------------------------------------
K052686
Page 3 of (3)
#### Substantial Equivalence
This Special 510(k) proposes modifications in performance specifications, labeling, and packaging configuration for the ArthroCare ArthroWands, which were previously cleared in K020557 (March 21, 2002) and K033584 (November 28, 2003). The indications for use, technology, principle of operation, and sterilization parameters of the ArthroCare ArthroWands remain the same as in the predicate cleared 510(k)s.
#### Summary of Safety and Effectiveness
The modified ArthroCare ArthroWands, as described in this Special 510(k), are substantially equivalent to the predicate device. The proposed modifications in the performance specifications, labeling, and packaging configuration are not substantial changes or modifications, and do not significantly affect the safety or efficacy of the devices.
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its body and wings. The logo is encircled by the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" in a circular arrangement.
OCT 5 - 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Valerie DeFiesta-Ng Director, Regulatory Affairs ArthroCare Corporation 680 Vaqueros Avenue Sunnyvale, California 94085-3523
Re: K052686 Trade/Device Name: ArthroCare® Arthro Wands® Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI Dated: September 27, 2005 Received: September 28, 2005
Dear Ms. DeFiesta-Ng:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complics with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
#### Page 2- Ms. Valerie DeFiesta-Ng
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
farbara Buchelt
Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use Statement
KOS2686 510(k) Number:
Page 1 of 2
ArthroCarc® ArthroWands® Device Name
Indications for Use:
The ArthroCare ArthroWands are indicated for resection, ablation, and coagulation of soft tissue and hemostasis of blood vessels in arthroscopic and orthopedic procedures:
| Arthroscopic and Orthopedic Procedures | Joint Specific or All<br>Joints (ankle, elbow,<br>hip, knee, shoulder, and<br>wrist) |
|----------------------------------------|--------------------------------------------------------------------------------------|
| <b>Ablation and Debridement</b> | |
| • ACL/PCL | Knee |
| • Acromioplasty | Shoulder |
| • Articular Cartilage | All Joints |
| • Bursectomy | All Joints |
| • Chondroplasty | All Joints |
| • Facia | All Joints |
| • Ligament | All Joints |
| • Notchplasty | Knee |
| • Scar Tissue | All Joints |
| • Soft Tissue | All Joints |
| • Subacromial Decompression | Shoulder |
| • Synovectomy | All Joints |
| • Tendon | All Joints |
| <b>Excision and Resection</b> | |
| • Acetabular Labrum | Hip |
| • Articular Labrum | All Joints |
| • Capsule | All Joints |
| • Capsular Release | Knee |
| • Cartilage Flaps | Knee |
| • Cysts | All Joints |
| • Discoid Meniscus | Knee |
| • Frozen Shoulder Release | Shoulder |
| • Glenoidale Labrum | Shoulder |
| • Lateral Release | Knee |
| • Ligament | All Joints |
| • Loose Bodies | All Joints |
| • Meniscal Cystectomy | Knee |
| Meniscectomy | Knee |
(Division Sign-Off)
410/1) Numbe
Division of General, Restorative,
**and Neurological Devices**
Barbara Buchner
{6}------------------------------------------------
K052686
Page. 2 of (2)
| Arthroscopic and Orthopedic Procedures | | Joint Specific or All Joints (ankle, elbow, hip, knee, shoulder, and wrist) |
|----------------------------------------|----------------------------------|-----------------------------------------------------------------------------|
| • | Plica Removal | All Joints |
| • | Scar Tissue | All Joints |
| • | Soft Tissue | All Joints |
| • | Synovial Membrane | All Joints |
| • | Tendon | All Joints |
| • | Triangular Fibrocartilage (TFCC) | Wrist |
| • | Villusectomy | Knee |
| Coagulation | | |
| • | ACL/PCL | Knee |
| • | Articular Cartilage | All Joints |
| • | Carpal Ligaments | Wrist |
| • | Glenohumeral Capsule | Shoulder |
| • | Ligament | All Joints |
| • | Medial Retinaculum | Knee |
| • | Rotator Cuff | Shoulder |
| • | Tendon | All Joints |
| • | Wrist Tendons | Wrist |
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
X
Over-The-Counter Use (21 CFR 801 Subpart C)
(PI.EASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Barbara Buchus for MXY
(Division Sign-Off)
Division of General, Restorative, and Neurological Devices
510(k) Number_KUS 2605 ﺻﺎﻛ
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.