← Product Code [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI) · K051214

# ALUMA SKIN RENEWAL SYSTEM (K051214)

_Lumenis, Inc. · GEI · Oct 24, 2005 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K051214

## Device Facts

- **Applicant:** Lumenis, Inc.
- **Product Code:** [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI.md)
- **Decision Date:** Oct 24, 2005
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Aluma Skin Renewal System is a noninvasive device intended for use in Dermatologic and General Surgical procedures non-invasive treatment of wrinkles and rhytids.

## Device Story

Aluma Skin Renewal System is a non-invasive, non-ablative electrosurgical device for treating wrinkles and rhytids. System components include a user interface, programmable logic controller (PLC), RF power module, vacuum pump, and a treatment handpiece with interchangeable tips. The device delivers RF energy at a 468 kHz sinusoidal frequency to the skin. Used in clinical settings by healthcare professionals, the system allows parameter selection via interface buttons with settings displayed on an LCD screen. The PLC manages operational and safety functions. The vacuum mechanism assists in tissue engagement during treatment. By applying controlled RF energy, the device aims to improve skin appearance. Clinical benefit is derived from the non-invasive reduction of wrinkles and rhytids.

## Clinical Evidence

No clinical data provided; substantial equivalence is based on technological characteristics and intended use compared to predicate devices.

## Technological Characteristics

Electrosurgical device; RF energy source (468 kHz sinusoidal signal); vacuum-assisted handpiece with interchangeable tips; programmable logic controller (PLC) for system management; non-invasive, non-ablative operation; user interface with LCD display.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- Thermage, Inc ThermaCool™ TC System ([K033942](/device/K033942.md) and [K040135](/device/K040135.md))
- Syneron Medical Ltd., Polaris™ ([K031671](/device/K031671.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

OCT 2 4 2005

K 051214

## 510(k) Summary of Safety and Effectiveness

This 510(k) Summary of Safety and Effectiveness for the Aluma™ Skin Renewal System is submitted in accordance with the requirements of the Safe Medical Device Act (SMDA) of 1990 and follows the Office of Device Evaluation (ODE) guidance concerning the organization and content of a 510(k) summary.

| Applicant:                          | Lumenis, Inc.                                                                                  |
|-------------------------------------|------------------------------------------------------------------------------------------------|
| Address:                            | 2400 Condensa Street<br>Santa Clara, CA 95051                                                  |
| Contact Person:                     | Connie Hoy                                                                                     |
| Telephone:<br>Fax:                  | (408) 764-3303<br>(408) 764-3500                                                               |
| Preparation Date:                   | May 1, 2005                                                                                    |
| Device Trade Name:                  | Aluma™ Skin Renewal System                                                                     |
| Common Name:                        | Electrosurgical cutting and coagulation and<br>device and accessories                          |
| Classification Name:                | Device, electrosurgical, cutting and coagulation<br>device and accessories (see CFR 878.4400). |
| Legally Marketed Predicate Devices: | Thermana Inc                                                                                   |

Thermage, Inc ThermaCool™ TC System (K033942 and K040135) Syneron Medical Ltd., Polaris™ (K031671)

{1}------------------------------------------------

The Aluma Skin Renewal System is a non-System Description: invasive, non-ablative unit consisting of a user interface, a programmable logic controller (PLC), an RF power module, internal electronics, a vacuum pump, and a treatment handpiece with interchangeable tips (small and large). The interface allows the selection of treatment parameters by pressing on the treatment buttons; an LCD screen displays the current treatment settings. The PLC is a specially configured computer that provides the operational and safety function of the system. The RF power module provides RF energy to the handpiece, producing a sinusoidal signal at a 468 kHz frequency Intended Use of the Device: The Aluma Skin Renewal System is a noninvasive device intended for use in Dermatologic and General Surgical procedures non-invasive treatment of wrinkles and rhytids. Technical characteristics; The technological characteristics of the Aluma Skin Renewal System are the same or very similar to those of the claimed predicate devices. Summary: Based on the foregoing, the Aluma Skin Renewal System is substantially equivalent to the legally marketed, claimed predicate devices for the purposes of this 510(k) submission.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health and Human Services (HHS) in the USA. The logo features a stylized eagle with its wings spread, symbolizing strength and protection. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" is arranged in a circular pattern around the eagle, indicating the department's name and country.

Public Health Service

OCT 2 4 2005

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Connie Hoy Global Director, RA/QS Lumenis, Incorporate 2400 Condensa Street Santa Clara, California 95051

Re: K051214

Trade/Device Name: Aluma Skin Renewal System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI Dated: August 22, 2005 Received: August 23, 2005

Dear Ms. Hoy:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html

Sincerely yours.

Barbara Buchner

Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## INDICATION FOR USE STATEMENT

510(k) Number:

K051214

Device Name:

Aluma Skin Renewal System

Indications for Use:

The Aluma Skin Renewal System is intended for use in Dermatologic and The Anama UKIN Keneral Systems of the non-invasive treatment of wrinkles and rhytids.

(Please do not write below this line – Continue on another page if needed)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use

OR (per 21 CFR 801.109)

Over-the-Counter Use _________________________________________________________________________________________________________________________________________________________

charlene mello
Division Sign Off

(Division Sign-Off) Division of General, Restorative, and Neurological Devices

510(k) Number K051214 11

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K051214](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K051214)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
