K032207 · Boston Scientific Corp · GEI · Aug 20, 2003 · General, Plastic Surgery
Device Facts
Record ID
K032207
Device Name
COBRA COOLED SURGICAL PROBE, MODEL M004159670
Applicant
Boston Scientific Corp
Product Code
GEI · General, Plastic Surgery
Decision Date
Aug 20, 2003
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Cobra Cooled Surgical Probe (Probe) is intended for the coagulation of cardiac tissue using radiofrequency (RF) energy during cardiac surgery. The Probe can be used during general surgery to coagulate soft tissues. The Probe may also be used to coagulate blood and soft tissue to produce hemostasis.
Device Story
Cobra Cooled Surgical Probe is an electrosurgical probe with a malleable or flexible shaft; used with Cobra Electrosurgical Unit (ESU). Device delivers radiofrequency (RF) energy to coagulate cardiac and soft tissues; produces hemostasis. Used in open surgical procedures by surgeons. Device features 9F shaft size, polycarbonate tip, and 2 to 7 electrodes. Provides thermal coagulation via RF energy; cooling mechanism integral to probe function. Surgeon controls energy delivery to achieve desired tissue effect; output affects clinical decision-making by enabling precise tissue coagulation and hemostasis during surgery. Benefits include effective tissue management during cardiac and general surgical interventions.
Clinical Evidence
Bench testing only. Non-clinical tests conducted in accordance with 21 CFR Part 58 (GLP) confirmed the device met design-input criteria and is safe and effective for its intended use.
Technological Characteristics
Electrosurgical probe; 9F shaft; polycarbonate tip; 2-7 electrodes (6-12.5 mm). Energy source: RF energy via Cobra ESU. Sterilization: Ethylene Oxide (EO). Form factor: Malleable or flexible shaft (15-35 cm).
Indications for Use
Indicated for patients undergoing cardiac surgery requiring cardiac tissue coagulation, or general surgery requiring soft tissue coagulation and hemostasis.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
K040219 — COBRA SURGICAL PROBE · Boston Scientific Corp · Feb 24, 2004
K023291 — COBRA COOLED CARDIAC SURGICAL PROBE, MODEL 1596X · Boston Scientific Corp · Oct 17, 2002
K040104 — COBRA ADHERE SURGICAL SYSTEM, MODEL M00415977C0 · Boston Scientific Corp · Feb 18, 2004
K021046 — COBRA COOLED SURGICAL PROBE, MODEL 1596X · Boston Scientific Corp · Apr 16, 2002
K051749 — COBRA SURGICAL SYSTEM · Estech, Inc. · Sep 13, 2005
Submission Summary (Full Text)
{0}------------------------------------------------
-
### 510(k) Summary of Safety and Effectiveness 3.
#### General Information a.
### Modified Device Information
| Category: | Comments: |
|--------------------------|--------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor: | Boston Scientific Corporation<br>2710 Orchard Parkway<br>San Jose, Ca 95134 |
| Correspondent: | April I. Malmborg<br>Regulatory Affairs Specialist II<br>Boston Scientific Corporation<br>2710 Orchard Parkway<br>San Jose, Ca 95134 |
| Contact Information: | E-mail: malmbora@bsci.com<br>Phone: (408) 895-3637<br>Fax: (408) 895-2202 |
| Device Common Name: | Electrosurgical Probe |
| Device Proprietary Name: | Cobra Cooled Surgical Probe (K032207) |
| Device Classification: | 21 CFR §878.4400 |
## Predicate Device Information
| Predicate Device: | Cobra Cooled Surgical Probe (K023291) |
|-----------------------------------------|---------------------------------------|
| Predicate Device Manufacturer: | Boston Scientific Corporation |
| Predicate Device Common Name | Electrosurgical Probe |
| Predicate Device Classification: | 21 CFR §878.4400 |
| Predicate Device Classification Number: | Class II |
Boston Scientific Corporation Special 510(k) Submission
{1}------------------------------------------------
K032207
Page ② z Pac
## SPECIAL 510(K) SUBMISSION Cobra Cooled Surgical Probe
### b. Date Summary Prepared
August 18, 2003
#### Description of Device C.
The Cobra Cooled Surgical Probe is an Electrosurgical Probe, with either a malleable or flexible shaft, used in conjunction with the Cobra Electrosurgical Unit (ESU). The system is intended for use by surgeons for the coagulation of cardiac and soft tissues during open surgical procedure.
### d. Intended Use
The intended use for the Cobra Cooled Surgical Probes is as follows:
The Cobra Cooled Surgical Probe (Probe) is intended for the coagulation of cardiac tissue using radiofrequency (RF) energy during cardiac surgery. The Probe can be used during general surgery to coagulate soft tissues. The Probe may also be used to coagulate blood and soft tissue to produce hemostasis.
#### Comparison to Predicate Device e.
See Table I- Comparison of Device Characteristics to Predicate on the following page.
000006
{2}------------------------------------------------
| Device Manufacturer<br>& Name | Cobra® Cooled Surgical Probe | |
|-------------------------------|-----------------------------------------------------------------------------------------------------------------------|---------|
| Boston Scientific Corporation | | Same |
| Device Description | Electrosurgical Probe | Same |
| 510(k) Number | K023291 | K032207 |
| Regulatory Class | II | Same |
| Device Classification | 21 CFR §878.4400 | Same |
| Intended Use | Coagulation of Cardiac Tissue<br>during Cardiac surgery and Soft<br>Tissue during General Open<br>Surgical Procedures | Same |
| Single Use? | Yes | Same |
| EO Sterilized? | Yes | Same |
| Shaft Size | 9F | Same |
| Tip Material | Polycarbonate | Same |
| Length (cm) | 15-35 | Same |
| Electrode Size (mm) | 6 to 12.5 mm | Same |
| Electrode Number | 2 to 7 | Same |
Table 1 - Comparison of Device Characteristics to Predicate
#### f. Summary of the Non-clinical Data
Where appropriate, testing conformed to the requirements of 21 CFR Part 58 (Good Laboratory Practices (GLP)). Specifically, non-clinical tests conducted for the Device showed the device met its design-input criteria, and was safe and effective for its intended use.
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird-like figure with stylized wings and body.
FEB 2 1 2008
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Boston Scientific Corporation c/o Ms. April I. Malmborg Regulatory Affairs Specialist II 2710 Orchard Parkway San Jose, CA 95134
Re: K032207
Trade/Device Name: Cobra Cooled Surgical Probe Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II (two) Product Code: OCL, GEI Dated: July 17, 2003 Received: July 22, 2003
Dear Ms. Malmborg:
This letter corrects our substantially equivalent letter of August 20, 2003.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. April I. Malmborg
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
el. Mall
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## SPECIAL 510(K) SUBMISSION Cobra Cooled Surgical Probe
# K032207
## 4. Premarket Notification -Indication for Use Statement
Device Name: Cobra Cooled Surgical Probe
Indication for Use:
The intended use for Cobra Cooled Surgical Probe is as follows:
The Cobra Cooled Surgical Probe is intended for the coagulation of cardiac tissue using radiofrequency (RF) energy during cardiac surgery. The Probe can be used during general surgery to coagulate soft tissues. The Probe may also be used to coagulate blood and soft tissue to produce hemostasis.
### (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IS NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
OR (Per 21 CFR §801.109) Over-the-Counter Use_
Mark N. Mellema
f General, Restorative and Neurological Devic
510(k) Number.
Boston Scientific Corporation Special 510(k) Submission
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.