← Product Code [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI) · K031082

# EVEREST BIPOLAR L HOOK & PK BIPOLAR L HOOK (K031082)

_Gyrus Medical, Inc. · GEI · May 21, 2003 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K031082

## Device Facts

- **Applicant:** Gyrus Medical, Inc.
- **Product Code:** [GEI](/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI.md)
- **Decision Date:** May 21, 2003
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

Electrosurgical coagulation, electrosurgical cutting, dissection of tissue, irrigation and aspiration during the performance of laparoscopic and general surgical procedures.

## Device Story

Everest Bipolar L Hook and Gyrus Bipolar L Hook are electrosurgical instruments used for coagulation, cutting, dissection, irrigation, and aspiration during laparoscopic and general surgery. Devices utilize bipolar electrosurgical energy; forceps jaws are electrically isolated to act as return electrodes, eliminating need for external return pad. Modification from predicate involves shaft length adjustment; fundamental technology remains unchanged. Operated by surgeons in clinical/surgical settings. Output is thermal energy applied to tissue for surgical effect; assists in achieving hemostasis and tissue separation.

## Clinical Evidence

No clinical data; bench testing only.

## Technological Characteristics

Bipolar electrosurgical instrument; energy source: bipolar electrosurgical energy; features electrically isolated forceps jaws acting as return electrodes; dimensions modified via shaft length; intended for laparoscopic and general surgery.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- [K904993](/device/K904993.md)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

| F: 510(k) Summary |
|-------------------|
|-------------------|

|  | K 031082 |
|--|----------|
|--|----------|

March 31, 2003

| Company:             | Gyrus Medical, Inc.<br>6655Wedgwood Road<br>Maple Grove, MN<br>55311-3602<br>Tel. No. (763) 416-3000<br>FAX. No. (763) 416-3070 |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------|
| Contact:             | Mercedes Bayani<br>Director, Regulatory Affairs                                                                                 |
| Common/Usual Name:   | Electrosurgical Instruments                                                                                                     |
| Classification Name: | Electrosurgical Cutting and Coagulation Device<br>and Accessories (21 CFR 878.4400)                                             |
| Proprietary Name:    | Everest Bipolar L Hook and Gyrus Bipolar L Hook                                                                                 |

The device is a Class II medical device. The Bipolar L Hook is a modification to the predicate device cleared under K904993. The Bipolar L Hook is identical in construction (with the exception of shaft length) and in component materials when compared to the predicate device. The forceps jaws are electrically isolated from each other enabling one jaw to act as a refurn electrode, eliminating the need for a return pad. The modification has not altered the fundamental technology of the predicate device cleared under K904993. The intended use, electrosurgical coagulation, grasping and dissection of the tissue during surgical procedures is similar to the predicate devices cleared under K904993. The energy source, bipolar electrosurgical energy, is the same energy type as used for the predicate devices.

In conclusion, as the design, materials of construction, function and intended use of the modified Bipolar L Hook is similar to that of the predicate devices currently cleared, Gyrus Medical Inc. believes that no new issues of safety and effectiveness are raised and that the submitted device is substantially equivalent.

{1}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/1/Picture/12 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its body and wings. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.

MAY 21 2003

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Ms. Mercedes Bayani Director, Regulatory Affairs Gyrus Medical, Inc. 6655 Wedgwood Road Maple Grove, Minnesota 55311

Re: K031082

Trade/Device Name: Everest Bipolar L Hook and Gyrus Bipolar L Hook Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI Dated: March 31, 2003 Received: April 24, 2003

Dear Ms. Bayani:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

Page 2 - Ms. Mercedes Bayani

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely vours.

Sincerely yours,

Mark A. Millbern

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Page 1 of 1

510(k) Number (if known): 139082 Device Name: Everest Bipolar L Hook & Gyrus Bipolar L Hook Indications for Use:

Electrosurgical coagulation, electrosurgical cutting, dissection of tissue, irrigation and aspiration during the performance of laparoscopic and general surgical procedures.

## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurence of CDRH, Office of Device Evaulation (ODE) (Optional Format 3-10-98) (Posted July 1, 1998)

Mark N. Milliken

(Division Sign-Off) Division of General, Restorative and Neurological Devices

510(k) Number K031082

---

**Source:** [https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K031082](https://fda.innolitics.com/submissions/SU/subpart-e%E2%80%94surgical-devices/GEI/K031082)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
