CRYO-MONO CRYOGENIC SYSTEM, MODEL CRYO-MONO WITH TUF PROBE
K991792 · Influence, Inc. · GEH · Aug 12, 1999 · General, Plastic Surgery
Device Facts
Record ID
K991792
Device Name
CRYO-MONO CRYOGENIC SYSTEM, MODEL CRYO-MONO WITH TUF PROBE
Applicant
Influence, Inc.
Product Code
GEH · General, Plastic Surgery
Decision Date
Aug 12, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4350
Device Class
Class 2
Attributes
Therapeutic
Intended Use
The Cryo-Mono Cryogenic System is intended for cryogenic destruction of tissues during surgical procedures. It is indicated for use as a cryosurgical tool in the fields of general surgery, dermatology, neurosurgery, thoracic surgery, ENT, gynecology, oncology, proctology, and urology for the ablation of tissue, including liver metastases, skin lesions, warts, and prostate tissues. Cryo-Mono with TUF probe is specifically indicated for the ablation of prostate tissue in cases of benign prostatic hyperplasia ("BPH").
Device Story
Cryo-Mono Cryogenic System uses argon gas to perform cryogenic tissue destruction. Device consists of console and single-use TUF probe; probe delivered transurethrally via cystoscope. Physician manually applies probe to target tissue; system delivers preset argon amounts for cooling. Iceball size monitored via external transrectal ultrasound. Used in surgical settings for ablation of various tissues, specifically prostate tissue for BPH. Benefits include minimally invasive tissue destruction. System features optional foot pedal control. Output is physical cooling of tissue; physician observes iceball formation via ultrasound to guide ablation. Device is smaller than predicate and uses fewer gas cylinders.
Clinical Evidence
Bench testing only. Biocompatibility established via predicate device materials. Conformance certified to IEC 60601-1, IEC 60601-1-2, and ASTM F882-84.
Technological Characteristics
Cryogenic system using argon gas. Single-use TUF probe. Manual application; optional foot pedal. Preset argon delivery. Transurethral delivery via cystoscope. Materials consistent with predicate Cryo-Hit and Impra Venaflo vascular graft. Conforms to IEC 60601-1, IEC 60601-1-2, and ASTM F882-84. Software modified to remove 'stick' mode.
Indications for Use
Indicated for patients requiring tissue ablation in general surgery, dermatology, neurosurgery, thoracic surgery, ENT, gynecology, oncology, proctology, and urology, including liver metastases, skin lesions, warts, and prostate tissue. Specifically indicated for prostate tissue ablation in patients with benign prostatic hyperplasia (BPH).
Regulatory Classification
Identification
(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.
Predicate Devices
Galil Medical, Ltd.'s Cryo-Hit Cryogenic System
Endocare's CryoCare™ cryogenic system
Vidamed's TUNA system
Related Devices
K031117 — SEEDNET SYSTEM AND SEEDNETGOLD SYSTEM · Galil Medical , Ltd. · May 23, 2003
K993965 — CRYO-HIT · Galil Medical , Ltd. · Feb 18, 2000
K051052 — SEEDNET FAMILY (SEEDNET/SEEDGOLD SYSTEM, CRYOTHERA SYSTEM, CRYO-HIT SYSTEM) · Galil Medical , Ltd. · May 6, 2005
K260377 — XSense Cryoablation System with CryoProbes · IceCure Medical , Ltd. · Feb 24, 2026
K141110 — CRYOCARE CS SURGICAL SYSTEM · Healthtronics · Jun 25, 2015
Submission Summary (Full Text)
{0}------------------------------------------------
# K991792
# 510(k) Summary Influence, Inc.'s Cryo-Mono with TUF probe Cryogenic System 510(k) Number
## Submitter's Name:
Peter A. Bick, M.D., President and CEO, Influence, Inc. 71 Stevenson Street, Suite 1120 San Francisco, California 94105 Telephone: 415-546-7700 / Fax: 415-546-7744
#### Trade Name:
Crvo-Mono Cryogenic System with TUF probe.
## Classification Name:
Cryosurgical unit
#### Product Code:
GEH
#### Predicate Devices:
Galil Medical, Ltd.'s Cryo-Hit Cryogenic System, Endocare's CryoCare™ cryogenic system and Vidamed's TUNA system.
## Intended Use and Indication for Use:
Influence, Inc.'s Cryo-Mono Cryogenic System and Galil Medical, Ltd.'s Cryo-Hit Cryogenic System are intended for cryogenic destruction of tissues during surgical procedures. They are indicated for use as a cryosurgical tool in the fields of general surgery, dermatology, neurosurgery, thoracic surgery, ENT, gynecology, oncology, proctology, and urology for the ablation of tissue, including liver metastases, skin lesions, warts, and prostate tissues. Cryo-Mono with TUF probe is specifically indicated for the ablation of prostate tissue in cases of benign prostatic hyperplasia ("BPH").
# Technological Characteristics:
The Cryo-Mono is a modified version of Cryo-Hit. They have the same technological characteristics with several exceptions: The primary technological differences between the Cryo-Mono with TUF probe and the Cryo-Hit are: (1) Cryo-Mono with TUF accepts only one cryoprobe, the TUF probe, which is one of the options for the Cryo-Hit; (2) The Cryo-Mono with TUF probe is smaller than Cryo-Hit; (3) The Cryo-Mono with TUF probe has fewer gas cylinders; (4) The physician manually applies the Cryo-Mono's
{1}------------------------------------------------
K991792
probe while with the Cryo-Hit, the physician has the option to use the "Stick" mode, which automatically adheres the probe to the tissue; (5) The Cryo-Mono with TUF probe has an optional foot pedal control; (6) The Cryo-Mono with TUF probe delivers preset amounts of argon for cooling while the Cryo-Hit adjusts the amount of argon delivered based on the tissue temperature measured by its thermocouples; (7) The Cryo-Mono with TUF probe's iceball size is monitored by transrectal ultrasound; the use of ultrasound, as well as external theromcouples, is optional with the Cryo-Hit; and (8) The Cryo-Mono's TUF probe is delivered transurethrally through a cystoscope rather than transperineally as with the Cryo-Hit.
The only modification to the device' s software was the removal of the "stick" option mentioned above.
The Cryo-Mono's single-use probes are within the range of probes that can be used with the Cryo-Hit, i.e. one to eight probes.
# Performance Data:
The only component of the Cryo-Mono that comes into contact with the patient's body is its cryoprobe. The biocompatibility of the Cryo-Mono is based on FDA premarket clearance of the Cryo-Hit. Its components are made of the same materials, as the Cryo-Hit's probes and Impra's Venaflo™ vascular graft. Influence has certified to the Cryo-Mono's conformance with IEC 60601-1. IEC 60601-1-2 and ASTM F882-84.
# Substantial Equivalence:
The Cryo-Mono has the same general intended use and virtually identical technological characteristics as the 510(k) cleared Cryo-Hit. The minor technological differences between these devices, namely, the Cryo-Mono smaller size, use of fewer probes, lack of a "stick" option, foot pedal, modified software, fewer gas cylinders and acceptance of a transurethral flexible probe do not raise any new questions of safety and effectiveness. Thus, the Cryo-Mono is substantially equivalent to a legally marketed cryosurgical device. The Cryo-Mono with TUF probe is substantially equivalent to EndoCare's CryoCare™ in terms of its technology and indication for use. The Cryo-Mono with TUF probe is substantially equivalent also to Vidamed's TUNA system in terms of its indication.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a symbol consisting of three stylized human profiles facing to the right, with flowing lines beneath them.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG 1.2 1999
Influence, Inc. c/o Jonathan S. Kahan Hogan & Hartson 555 Thirteenth Street, N.W. Washington, D.C. 20004
Re: K991792
> Trade Name: Cryo-Mono Cryogenic System with TUF Probe Regulatory Class: II Product Code: GEH Dated: May 23, 1999 Received: May 26, 1999
Dear Mr. Kahan:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition. FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
# Page 2 - Mr. Jonathan S. Kahan
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
# 510(k) Number (if known):
### Device Name:
Indications for Use:
Cryo-Mono Cryogenic System with TUF probe.
The Cryo-Mono Cryogenic System is intended for cryogenic destruction of tissues during surgical procedures. It is indicated for use as a cryosurgical tool in the fields of general surgery, dermatology, neurosurgery, thoracic surgery, ENT, gynecology, oncology, proctology, and urology for the ablation of tissue, including liver metastases, skin lesions, warts, and prostate tissues. Cryo-Mono with TUF probe is specifically indicated for the ablation of prostate tissue in cases of benign prostatic hyperplasia ("BPH").
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-off) Division of General and Restorative Devices
510(k) Number K991792
Prescription Use
(Per 21 CFR 801.109)
OR Over-The-Counter-Use
(Division Sign-Off) Division of General Re 510(k) Number
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.