ACCUPROBE 800 SERIES MODEL NUMBERS : 810, 820, 830, 840, 850, 860, 870 AND 880
K982055 · Cryomedical Sciences, Inc. · GEH · Sep 8, 1998 · General, Plastic Surgery
Device Facts
Record ID
K982055
Device Name
ACCUPROBE 800 SERIES MODEL NUMBERS : 810, 820, 830, 840, 850, 860, 870 AND 880
Applicant
Cryomedical Sciences, Inc.
Product Code
GEH · General, Plastic Surgery
Decision Date
Sep 8, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4350
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The CMS AccuProbe® 800 Series (all models depicted as A) has the same indications for use as its predicate device, the AccuProbe® 600 Series [510(k) number K973190 cleared November 21, 1997]. Urology - The (A) system may be used to ablate prostatic tissue. - . The (A) system may be used for the ablation of prostate tissue in cases of prostate cancer and benign prostatic hyperplasia. Oncology - . system may be used for ablation of cancerous or The (A) malignant tissue. - system may be used for ablation of benign tumors. . The (A) - The . (A) system may be used for palliative intervention. Dermatology - system may be used for the ablation or freezing of The (A) . skin cancers and other cutaneous disorders. Gynecology - system may be used for the ablation of malignant . The (A) neoplasia or benign dysplasia of the female genitalia. General Surgery - (A) system may be used for the ablation of The . leukoplakia of mouth, anaiomas, sebaceous hyperpalsia, basal cell tumors of the eyelid or canthus area, ulcerated basal cell tumors, dermatofibromas, small hemangiomas, mucocele cysts, multiple warts, plantar warts, hemorrhoids, anal fissures, perianal condylomata, pilonidal cysts, actinic and seborrheic keratoses , cavernous hemanaiomas recurrent cancerous lesions. - . The (A) system may be used for the destruction of warts or lesions. - . The (A) system may be used for the palliation of tumors of the oral cavity, rectum, and skin. Thoracic Surgery - The . (A) system may be used for the ablation of arrhythmic cardiac tissue. - . The system may be used for the ablation of cancerous (A) lesions. Proctology - The (A) system may be used for the ablation of benign or malignant growths of the anus and rectum - . systems may be used for the ablation of The (A) hemorrhoids.
Device Story
Cryosurgical unit; uses liquid nitrogen to destroy unwanted tissue via extreme cold. System comprises console with pumping mechanism and cryoprobes with interchangeable tips. Physician positions cryoprobe at target site using ultrasound or other imaging; liquid nitrogen forced through probe to freeze tissue. Real-time monitoring via imaging; procedure duration ~20 minutes. Post-freeze, warming gas facilitates thawing for probe removal. Optional on-board computer provides redundant display and procedure archival; does not control cryoprobe operation. Used in clinical settings by physicians.
Clinical Evidence
No clinical data provided. Substantial equivalence based on technological comparison and intended use.
Technological Characteristics
Cryosurgical unit; liquid nitrogen cryogen; 65psi operating pressure; nitrogen gas for warming; one-dewar system; console with pumping mechanism; interchangeable cryoprobe tips of various materials/shapes. Optional on-board computer for display/archival. Class II device per 21 CFR 878.4350 and 882.4250.
Indications for Use
Indicated for ablation of benign/malignant tissue in urology (prostate cancer, BPH), oncology, dermatology (skin cancers, cutaneous disorders), gynecology (neoplasia, dysplasia), general surgery (warts, lesions, tumors, hemorrhoids, cysts), thoracic surgery (arrhythmic cardiac tissue, cancerous lesions), and proctology. Prescription use only.
Regulatory Classification
Identification
(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.
{0}------------------------------------------------
#### 8 1998 SEb
K982055
### 510(k) SUMMARY
June 11, 1998 Submitted by:
ﻤﺼﻨﻌﺔ
Contact:
Richard J. Reinhart, Ph.D. President and CEO
Cryomedical Sciences, Inc. 1300 Piccard Drive Suite L105 Rockville, Maryland 20850 (301) 417-7070 Fax (301) 417-7077
Proprietary name: CMS AccuProbe® Models 810, 820, 830, 840, 850, 860, 870, and 880
Common name: Cryosurgical units and accessories
Cryosuraical units with Liquid Nitrogen, Nitrous Oxide, Carbon Classification: Dioxide, Class II, under CFR § 878.4350. Under CFR § 882.4250 a cryogenic surgical device also is classified in Class II.
The CMS AccuProbe® 800 Series consists of eight models, each of which has two components: i) a console with a pumping mechanism to compress and pressurize the cryogen and ii) one to eight (models 810 through 880) cryoprobes with tips extending from the console. The may include an optional on-board computer which is as redundant display system and archival system to store information about the procedure. The computer does not control or affect the operation of the cryoprobes in any fashion. The cryoprobes consist of interchangeable tips of various materials, sizes and shapes which are used a necessary.
The AccuProbe® 800 Series is used to destroy unwanted tissue by the application of extreme cold to the selected site. Liquid nitrogen is forced through the cryoprobes under pressure. The cryoprobe, having been placed in the appropriate position through the used of utlrasound or other imaging device, then becomes cold and freezes the unwanted tissue. The procedure may be
{1}------------------------------------------------
monitored in real-time through the use of imaging modes such as ultrasound, ﻤﺴﺴﺴﺴﺴﺴﺴﺴ Once the freezing is completed (approximately 20 minutes), the cryoprobe will be heated using warming gas in order to facilitate thawing and removal of the cryoprobe from the frozen tissue.
Image /page/1/Figure/1 description: The image shows the text "Page 2 of 3". This indicates that the image is a page from a document, specifically page 2 of a 3-page document. The text is written in a simple, clear font and is likely a page number indicator.
{2}------------------------------------------------
### COMPARISON OF FEATURES BETWEEN THE AccuProbe® 600 Series [CLEARED 510(k) (K964336)] AND THE AccuProbe® 800 SERIES
| FEATURES | AccuProbe® 600 Series | AccuProbe® 800 SERIES |
|---------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Liquid Nitrogen<br>Capacity<br>(Dewar size) | Model 610 15 liters<br>Model 620 25 liters<br>Model 630 40 liters<br>Model 640 55 liters<br>Model 650 70 liters<br>Model 660 85 liters<br>Model 670 100 liters<br>Model 680 115 liters | Model 810 25 liters<br>Model 820 25 liters<br>Model 830 25 liters<br>Model 840 35 liters<br>Model 850 35 liters<br>Model 860 45 liters<br>Model 870 45 liters<br>Model 880 45 liters |
| Normal Operating<br>Pressure | 65psi | 65psi |
| Regen/recirculating<br>time | Two minutes | Less than 10 seconds |
| Regen mode | Depressurizes the supply<br>dewar and pressurizes the<br>return dewar (this allows<br>capture of LN2 to flow from<br>the return dewar through the<br>cryoprobe and back to the<br>supply dewar) | The LN2 return line from the<br>cryoprobe feeds directly into<br>the LN2 supply |
| Dimensions | Model 610 45"x24"x29"<br>Model 620 49"x24"x29"<br>Model 630 52"x24"x29"<br>Model 640 52"x24"x29"<br>Model 650 54"x28"x32"<br>Model 660 54"x30"x34"<br>Model 670 54"x32"x34"<br>Model 680 54"x32"x34" | Model 810 38"x24"x24"<br>Model 820 38"x24"x24"<br>Model 830 38"x24"x24"<br>Model 840 38"x24"x24"<br>Model 850 38"x26"x26"<br>Model 860 38"x26"x26"<br>Model 870 38"x26"x26"<br>Model 880 38"x26"x26" |
| Operation | Uses a three dewar system | Uses a one dewar system |
| Warming Gas | Nitrogen Gas | Nitrogen Gas |
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with flowing lines, representing a sense of movement or progress. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.
FEB 2 1 2008
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Cryomedical Sciences, Inc. c/o Mr. Richard J. Reinhart, Ph.D President and CEO 1300 Piccard Drive, Suite L105 Rockville, MD 20850
Re: K982055
Trade Name: Accuprobe 800 Series Model Numbers: 810, 820, 830, 840, 850, 860, 870 and 880 Regulatory Class: II (two) Product Code: OCL. GEH Dated: June 11, 1998 Received: June 11, 1998
Dear Dr. Reinhart:
This letter corrects our substantially equivalent letter of September 8, 1998.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Richard J. Reinhart, Ph.D.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours.
eempe.
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications For Use
The CMS AccuProbe® 800 Series (all models depicted as A) has the same indications for use as its predicate device, the AccuProbe® 600 Series [510(k) number K973190 cleared November 21, 1997].
#### Urology
- The (A) system may be used to ablate prostatic tissue.
- . The (A) system may be used for the ablation of prostate tissue in cases of prostate cancer and benign prostatic hyperplasia.
#### Oncology
- . system may be used for ablation of cancerous or The (A) malignant tissue.
- system may be used for ablation of benign tumors. . The (A)
- The . (A) system may be used for palliative intervention.
#### Dermatology
- system may be used for the ablation or freezing of The (A) . skin cancers and other cutaneous disorders.
#### Gynecology
- system may be used for the ablation of malignant . The (A) neoplasia or benign dysplasia of the female genitalia.
ccer
**Prescription Use**
**(Per 21 CFR 801.109)**
(Division Sign-Off
Division of General Restorative Devices K982055
510(k) Number
{6}------------------------------------------------
# Indications for Use
(continued for CMS AccuProbe® device family)
#### General Surgery
- (A) system may be used for the ablation of The . leukoplakia of mouth, anaiomas, sebaceous hyperpalsia, basal cell tumors of the eyelid or canthus area, ulcerated basal cell tumors, dermatofibromas, small hemangiomas, mucocele cysts, multiple warts, plantar warts, hemorrhoids, anal fissures, perianal condylomata, pilonidal cysts, actinic and seborrheic keratoses , cavernous hemanaiomas recurrent cancerous lesions.
- . The (A) system may be used for the destruction of warts or lesions.
- . The (A) system may be used for the palliation of tumors of the oral cavity, rectum, and skin.
#### Thoracic Surgery
- The . (A) system may be used for the ablation of arrhythmic cardiac tissue.
- . The system may be used for the ablation of cancerous (A) lesions.
#### Proctology
- The (A) system may be used for the ablation of benign or malignant growths of the anus and rectum
- . systems may be used for the ablation of The (A) hemorrhoids.

(Division Sign-Off)
Division of General Restorative Devices
510(k) Number K982655
**Prescription Use**
(Per 21 CFR 801.109)
6
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.