K973686 · Endocare, Inc. · GEH · Dec 22, 1997 · General, Plastic Surgery
Device Facts
Record ID
K973686
Device Name
ENDOCARE CRYOCARE CRYOSURGICAL SYSTEM
Applicant
Endocare, Inc.
Product Code
GEH · General, Plastic Surgery
Decision Date
Dec 22, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4350
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ENDOcare CRYOcare™ Cryosurgical System is intended for use in general surgery, urology, gynecology, oncology, neurology, thoracic surgery, dermatology, ENT, and proctology, The system is designed to destroy tissue by the application of extreme cold temperatures including prostate, kidney, and breast tissue, liver metastases, tumors, skin lesions, and warts.
Device Story
ENDOcare CRYOcare™ Cryosurgical System is a console-based device for tissue destruction via extreme cold. System controls 1-8 single-use, disposable CRYOprobes™ and monitors temperatures using T-type needle thermocouples. Operates on 110 VAC power. Available in 1, 4, and 8-probe configurations. Used by clinicians in various surgical specialties (urology, oncology, dermatology, etc.) to treat tumors, lesions, and specific organ tissues. Console manages cooling delivery to probes; thermocouples provide real-time temperature feedback to the operator. Clinical benefit is the targeted destruction of diseased tissue through cryoablation.
Clinical Evidence
Bench testing only. Sterilization validation performed per AAMI guidelines for Ethylene Oxide (SAL 10^-6). Shelf-life testing conducted via accelerated aging (1-year equivalent) followed by environmental challenge (shipping/handling simulation), microbial challenge, and functional testing. All results met acceptance criteria.
Technological Characteristics
Console-based cryosurgical system; 1-8 disposable CRYOprobes; T-type needle thermocouples for temperature monitoring; 110 VAC power source. Sterilization of probes via Ethylene Oxide (EtO) per AAMI guidelines. Packaging consists of plastic film and coated paper.
Indications for Use
Indicated for patients requiring tissue destruction via cryosurgery in general surgery, urology, gynecology, oncology, neurology, thoracic surgery, dermatology, ENT, and proctology. Applicable to prostate, kidney, breast tissue, liver metastases, tumors, skin lesions, and warts.
Regulatory Classification
Identification
(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.
{0}------------------------------------------------
DEC ≤ 1997
# Section II - SMDA 1990 Requirements
K.973686
## Safety and Effectiveness Summary
#### Device Description
The ENDOcare CRYOcare™ Cryosurgical System consists of a console which can control one to eight single-use, disposable CRYOprobes™ and one to eight independent inouts to monitor temperatures from standard T-type needle thermocouples. The system is compact and operates off of standard 110 VAC wall power.
Several models will be provided depending on the needs of the customer. Currently, the system is available in a 1, 4 and 8-CRYOprobe™ configuration. The performance characteristics and internal design of each model are equivalent. The only differences are the number of valves to control the CRYOprobes™ (e.g., 1-8), number of thermocouple inputs (e.g., 1-8) and the size of the outer case.
#### Indications for Use
The ENDOcare CRYOcare™ Cryosurgical System is intended for use in general surgery, urology, gynecology, oncology, neurology, thoracic surgery, dermatology, ENT, and proctology. The system is designed to destroy tissue by the application of extreme cold temperatures including prostate, kidney, and breast tissue, liver metastases, tumors, skin lesions, and warts.
#### Substantial Equivalence
The ENDOcare CRYOcare™ CRYOsurgical System is substantially equivalent to the ENDOcare 8-probe system which was cleared on January 19, 1995 (reference K942299) and the 2 and 5-probe system which was cleared on December 18, 1996 (reference K963826). The ENDOcare system is also substantially equivalent to the Urosprobe System Model 1000 which was cleared on April 1, 1996 (reference K960387).
#### Sterilization Methodology
The following sterilization information is applicable to the CRYOprobes™ only. Sterilization validation was based on the recommendations in the current AAMI (American Association for the Advancement of Medical Instrumentation) Guideline for Industrial Ethylene Oxide Sterilization of Medical Devices. A minimum Sterility Assurance Level (SAL) of 10 was achieved. Sterile barrier packaging consists of standard disposable medical device packaging containing a plastic film and a coated paper side. Maximum levels of EtO residuals did not exceed:
{1}------------------------------------------------
25 ppm for ethylene oxide
25 ppm for ethylene chlorohydrin
250 ppm for ethylene glycol
### Shelf Life Determination
Ten (10) simulated products and one (1) control were packaged, sealed and processed through a full sterilization cycle to replicate sterilization parameter effects on package seal integrity. Following sterilization, the packaged CRYOprobes™ were submitted to an outside testing laboratory for accelerated aging equivalent to one year using the following parameters:
2.5 weeks at 55℃ and 75 + 5% relative humidity 24 hours at -15°C + 5°C 2.5 weeks at 55℃ and <20% relative humidity
Following accelerated aging, the CRYOprobes™ were shipped to another facility via UPS ground service. The packages were then returned via UPS Second Day Air Service. The purpose of this test was to simulate conditions that the product might experience in shipping and handling.
Following the environmental challenge, the samples were subjected to a microbial challenge and tested to determine if microbial penetration has occurred. Functional testing was conducted pre- and post-accelerated aging. All test results met the acceptance criteria equivalent to a one year shelf life.
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows the address of the Food and Drug Administration. The address is 9200 Corporate Boulevard, Rockville MD 20850. The text is black and the background is white.
Image /page/2/Picture/3 description: The image shows a date, "DEC 22 1997", in a simple, sans-serif font. The letters and numbers are bold and black against a white background. The date is formatted with the month abbreviated, followed by the day and then the year.
Mr. Vin Cutarelli Vice President, Regulatory Affairs and Quality Assurance Endocare, Incorporated 7 Studebaker Irvine, California 92618
Re: K973686
> Trade Name: ENDOcare CRYOcare™ Cryosurgical System Regulatory Class: II Product Code: GEH Dated: September 25, 1997 Received: September 26, 1997
Dear Mr. Cutarelli:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements , as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for
{3}------------------------------------------------
Page 2 - Mr. Cutarelli
devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indication For Use Statement
510(k) Number: K973686
Device Name: ENDOcare CRYOcare™ Cryosurgical System
The ENDOcare CR YOcare™ Cryosurgical System is intended for use in general surgery, urology, gynecology, oncology, neurology, thoracic surgery, dermatology, ENT, and proctology, The system is designed to destroy tissue by the application of extreme cold temperatures including prostate, kidney, and breast tissue, liver metastases, tumors, skin lesions, and warts.
Concurrence of CDRH, Office of Device Evaluation (ODE):
tecouto
(Division Sign-Off) Division of Dental, Infection Control, and General Hospital Devices 510(k) Number _
Prescription Use: X (Per 21 CFR 801.109)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.