ENDOCARE CRYOCARE CRYOSURGICAL SYSTEM URETHRAL WARMING SYSTEM

K963970 · Endocare, Inc. · GEH · Feb 25, 1997 · General, Plastic Surgery

Device Facts

Record IDK963970
Device NameENDOCARE CRYOCARE CRYOSURGICAL SYSTEM URETHRAL WARMING SYSTEM
ApplicantEndocare, Inc.
Product CodeGEH · General, Plastic Surgery
Decision DateFeb 25, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4350
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The ENDOcare Urethral Warming System is an accessory to the cryosurgical system and is indicated to transfer heat to the urethral tissue during urological cryosurgery procedures using the ENDOcare CRYOcare™ Cryosurgical System.

Device Story

System consists of balloon catheter, PVC tubing, heater/stand, and roller pump. Operates as closed-circuit heat transfer system; circulates warm water/saline from I.V. bag to catheter balloon to warm urethral tissue during cryosurgery. Heater maintains fluid at 37.5°C; integrated shut-off prevents exceeding 42°C. Used in clinical urological settings by physicians. Provides thermal protection to urethra during tissue ablation. Benefits patient by preventing unintended cold injury to urethral structures during cryosurgical procedures.

Clinical Evidence

Bench testing only. Functional performance compared to predicate device at 37.5°C set point demonstrated equivalence.

Technological Characteristics

Components: polyurethane shaft catheter with polyester balloon, PVC tubing, heater, roller pump. Energy: thermal (heated fluid). Sterilization: Ethylene oxide (catheter/tubing) per AAMI guidelines; Gamma radiation (heat exchanger cassette) per AAMI guidelines. Biocompatibility: ISO-10993 compliant. SAL 10^-6. EtO residuals: <25 ppm EtO, <25 ppm ethylene chlorohydrin, <250 ppm ethylene glycol.

Indications for Use

Indicated for patients undergoing urological cryosurgery procedures requiring urethral tissue warming.

Regulatory Classification

Identification

(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.

Predicate Devices

Submission Summary (Full Text)

{0} K963970 FEB 25 1997 # Section 2.0 - SMDA 1990 Requirements ## 2.1 510(k) Summary ### Device Description The ENDOcare Urethral Warming System consists of a balloon catheter, PVC tubing set, heater/heater stand and roller pump. The urethral catheter consists of a polyurethane shaft with a polyester balloon. The system provides a steady flow of warm water or saline from an I.V. bag to the catheter in order to transfer heat to the urethral tissue during urological cryosurgery procedures. The catheter has an inlet port which allows the warm water or saline to flow into the balloon, warming the surrounding urethral tissue. An outlet port allows the water or saline to exit the catheter and return to the I.V. bag. The water or saline in the I.V. bag is maintained at a set temperature of 37.5°C by the heater. A shut-off feature prevents the temperature from exceeding 42°C. The water or saline is circulated by a roller pump from the catheter to the I.V. bag in a closed circuit. ### Biocompatibility The biocompatibility requirements were determined through use of the International Standard ISO-10993, “Biological Evaluation of Medical Devices Part 1: Evaluation and Testing.” The urethral warming catheter is the only component of the system which has patient contact. The catheter has the same material, manufacturing process, chemical composition, body contact and sterilization method as the marketed catheter. Therefore, per the FDA matrix the biocompatibility requirements were met and no additional testing was performed. ### Substantial Equivalence Support The ENDOcare CRYOcare™ Cryosurgical System is designed for use in general surgery, dermatology, neurology, thoracic surgery, ENT, gynecology, oncology, protocology and urology for the ablation and destruction of tissue by the application of extreme cold. The ENDOcare Urethral Warming System is an accessory to the cryosurgical system and is indicated to transfer heat to the urethral tissue during urological cryosurgery procedures using the ENDOcare CRYOcare™ Cryosurgical System. The Urethral Warming System is substantially equivalent to the Cryomedical Sciences, Inc. Urethral Warmer which was determined to be substantially equivalent on October 24, 1995 (reference K952895). The ENDOcare Urethral Warming system consists of a balloon catheter, tubing, heater and roller pump which are substantially equivalent to the marketed device. The catheter is the only component with direct patient contact and it is manufactured with the same materials as the marketed device. {1} # In Vitro Testing Bench testing was conducted to compare the functional performance of the ENDOcare Urethral Warming System to the Cryomedical Sciences Urethral Warmer. At a set point temperature of $37.5^{\circ}\mathrm{C}$, the performance of the two systems was determined to be equivalent. # Sterilization Methodology Ethylene oxide will be utilized to sterilize the catheter and tubing set. Sterilization validation will be based on the recommendations in the current AAMI (American Association for the Advancement of Medical Instrumentation) Guideline for Industrial Ethylene Oxide Sterilization of Medical Devices. A minimum Sterility Assurance Level (SAL) of $10^{-6}$ will be achieved. Sterile barrier packaging will consist of standard disposable medical device packaging containing a plastic film and a coated paper side. This is the same type of sterile barrier packaging that is used with the marketed device. Maximum levels of EtO residuals will not exceed: - 25 ppm for ethylene oxide - 25 ppm for ethylene chlorohydrin - 250 ppm for ethylene glycol The heat exchanger cassette will be received sterile from the supplier. This component is gamma sterilized. Sterilization validation was based on the recommendations of the current AAMI Guideline for Gamma Radiation to achieve a minimum Sterility Assurance Level (SAL) of $10^{-6}$.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...