TargetCool-e

K253348 · Recensmedical, Inc. · GEH · Oct 15, 2025 · General, Plastic Surgery

Device Facts

Record IDK253348
Device NameTargetCool-e
ApplicantRecensmedical, Inc.
Product CodeGEH · General, Plastic Surgery
Decision DateOct 15, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4350
Device ClassClass 2
AttributesTherapeutic

Intended Use

TargetCool-e (Cooling mode / Boosting mode) is indicated for the temporary reduction of pain, swelling, inflammation, and hematoma from minor surgical procedures, minor sprains or other minor sports injuries, and as an adjunct to rehabilitative treatment (e.g., intermittent cold with stretch). TargetCool-e (Freezing mode) is indicated for the surgical destruction of target tissue by applying cryogenic gases at extreme low temperatures - Molluscum Contagiosum - Skin Tags - Actinic Keratosis - Lentigo - Verruca Plana - Verruca Vulgaris - Verruca Lesions - Genital Lesions - Seborrheic Keratosis

Device Story

Handheld cryosurgical device; delivers rapid, controlled cooling or freezing to skin tissue. Inputs: CO2 gas and cold-water (saline). Operation: Thermoelectric cooling; CO2 gas expansion for cryoablation. Outputs: Controlled temperature gas/water stream; real-time skin temperature, set temperature, and time displayed on LCD. Safety: Alarm and blue status light blink if skin temperature < -1°C for >1 second. Used in clinical settings by healthcare providers. Benefits: Pain/swelling reduction; surgical tissue destruction. Differences from predicate: Built-in puncture pin eliminates need for disposable filter; supports 120g CO2 cartridge in addition to 65g for extended duration.

Clinical Evidence

Bench testing only. Device complies with ISO 13485, IEC 60601-1 (electrical safety), and IEC 60601-1-2 (EMC). Performance testing verified temperature specifications and device function with both 65g and 120g CO2 cartridges. Biocompatibility of patient-contacting polycarbonate guard established via equivalence to predicate.

Technological Characteristics

Handheld cryosurgical unit. Materials: Polycarbonate guard (ISO 10993 compliant). Energy: Thermoelectric cooling; CO2 gas (cryogen). Dimensions/Form: Handheld with LCD interface. Connectivity: None. Sterilization: Not specified. Software: Embedded firmware for temperature monitoring and alarm logic.

Indications for Use

Indicated for temporary reduction of pain, swelling, inflammation, and hematoma from minor surgical procedures, minor sprains, or sports injuries; adjunct to rehabilitative treatment. Also indicated for surgical destruction of target tissue (Molluscum Contagiosum, Skin Tags, Actinic Keratosis, Lentigo, Verruca Plana, Verruca Vulgaris, Verruca Lesions, Genital Lesions, Seborrheic Keratosis) via cryoablation.

Regulatory Classification

Identification

(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD &amp; DRUG ADMINISTRATION October 15, 2025 RecensMedical, Inc. Minju Choi Regulatory Affairs Manager #507, #908, SK V1 Center, 830 Dongtansunhwan-daero Hwaseong-si, Gyeonggi-do 18468 Republic Of Korea Re: K253348 Trade/Device Name: TargetCool-e Regulation Number: 21 CFR 878.4350 Regulation Name: Cryosurgical Unit And Accessories Regulatory Class: Class II Product Code: GEH, MLY Dated: September 29, 2025 Received: September 30, 2025 Dear Minju Choi: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" U.S. Food &amp; Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K253348 - Minju Choi Page 2 (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). {2} K253348 - Minju Choi Page 3 Sincerely, Colin K. Chen -S Digitally signed by Colin K. Chen -S Date: 2025.10.15 10:19:31 -04'00' Colin Kejing Chen Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} FORM FDA 3881 (6/20) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K253348 | | | Device Name | | | TargetCool-e | | | Indications for Use (Describe) TargetCool-e (Cooling mode / Boosting mode) is indicated for the temporary reduction of pain, swelling, inflammation, and hematoma from minor surgical procedures, minor sprains or other minor sports injuries, and as an adjunct to rehabilitative treatment (e.g., intermittent cold with stretch). | | | TargetCool-e (Freezing mode) is indicated for the surgical destruction of target tissue by applying cryogenic gases at extreme low temperatures - Molluscum Contagiosum - Skin Tags - Actinic Keratosis - Lentigo - Verruca Plana - Verruca Vulgaris - Verruca Lesions - Genital Lesions - Seborrheic Keratosis | | | Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {4} TargetCool-e K253348 # 510(k) Summary ## 1. ADMINISTRATIVE INFORMATION Manufacturer Name RecensMedical, Inc. #507, #908, SK V1 center, 830 Dongtansunhwan-daero, Hwaseong-si, Gyeonggi-do, Republic of Korea Phone: +82 31 8043 3064 Fax: +82 31 630 2092 Date of the summary preparation September 29, 2025 Contact person: Minju Choi/ Regulatory Affairs Manager mj.choi@recensmedical.com ## 2. DEVICE NAME AND CLASSIFICATION Trade name: TargetCool-e Common name: Cryosurgical Device Classification name: Cryosurgical Unit and Accessories Classification Regulations: 21 CFR 878.4350 / 21 CFR 878.4810 Class: Class II Classification Panel: General &amp; Plastic Surgery Product code: GEH, MLY ## 3. PRIMARY PREDICATE DEVICE 510(k) Number: K230599 Trade name: TargetCool™ Classification name: Cryosurgical Unit and Accessories Classification Regulations: 21 CFR 878.4350 Class: Class II Classification Panel: General &amp; Plastic Surgery Product code: GEH, MLY Page 1 of 5 {5} TargetCool-e K253348 # 4. INDICATIONS FOR USE TargetCool-e (Cooling mode, Boosting) is indicated for the temporary reduction of pain, swelling, inflammation, and hematoma from minor surgical procedures, minor sprains or other minor sports injuries, and as an adjunct to rehabilitative treatment (e.g., intermittent cold with stretch). TargetCool-e (Freezing mode) is indicated for the surgical destruction of target tissue by applying cryogenic gases at extreme low temperatures - Molluscum Contagiosum - Skin Tags - Actinic Keratosis - Lentigo - Verruca Plana - Verruca Vulgaris - Verruca Lesions - Genital Lesions - Seborrheic Keratosis # 5. DEVICE DESCRIPTION TargetCool-e is a handheld device that can deliver rapid, precise, and controlled cooling to the skin tissue. The TargetCool-e device consists of a main device, nozzle, a guard and a cartridge. The main device produces controlled cooling based on thermoelectric cooling, which controls the temperature of the targeted area. TargetCool-e displays the skin temperature measured in real time, the set cooling temperature and time, and the device status through the LCD display. Also, if the measured temperature is below -1°C and lasts for more than 1 second, the status light blinks in blue with a beep sound. # 6. PERFORMANCE DATA The Company's Performance Data for TargetCool-e is as follows: ## Bench Testing TargetCool-e complies with all applicable standards, including ISO 13485:2016, IEC 60601-1 for electrical safety and IEC 60601-1-2 for electromagnetic compatibility. TargetCool-e is substantially equivalent to the predicate device, TargetCool (K230599), with the only difference being that the subject device does not require a disposable filter for cartridge installation. TargetCool-e features a built-in puncture pin, and the CO₂ cartridge includes a seal cap on the mounting part, allowing direct attachment without a separate disposable filter. All other components used with TargetCool (K230599)—including the Guard, Cooling Nozzle, Boosting Nozzle, Freezing Nozzle, Boosting Container, and Boosting Guard—are identical for TargetCool-e. Performance testing was conducted with both the 65 g and 120 g CO₂ cartridges. Testing confirmed that all specifications met the acceptance criteria for each module and the overall device function. Page 2 of 5 {6} TargetCool-e K253348 TargetCool-e passed all testing and substantiates substantial equivalence as well as safe and effective operation. Biocompatibility evaluation was considered in accordance with ISO 10993. The only patient-contacting component of TargetCool-e is the guard, which is made of polycarbonate. This material is identical to that used in the predicate device (K230599). Therefore, additional biocompatibility testing was not required. The device hazard analysis was completed and risk-control implemented to mitigate identified hazards. The Performance testing results support that all the specifications have met the acceptance criteria of each module and interaction of processes. # 7. SUBSTANTIAL EQUIVALENCE The comparison chart below provides evidence to support the equivalence determination between TargetCool-e and the predicate device (K230599) with respect to intended use, technological characteristics and principles of operation. TargetCool-e shares the same indications for use, device operation, technical and functional capabilities, and therefore is substantially equivalent to the predicate device. [TargetCool-e-Cooling mode, Boosting mode] | Product Name | TargetCool-e (Subject Device) | | TargetCool™ (K230599) | | Comparison | | --- | --- | --- | --- | --- | --- | | Indications for Use / Intended Use | TargetCool-e is indicated for the temporary reduction of pain, swelling, inflammation, and hematoma from minor surgical procedures, minor sprains or other minor sports injuries, and as an adjunct to rehabilitative treatment (e.g., intermittent cold with stretch). | | TargetCool is indicated for the temporary reduction of pain, swelling, inflammation, and hematoma from minor surgical procedures, minor sprains or other minor sports injuries, and as an adjunct to rehabilitative treatment (e.g., intermittent cold with stretch). | | Same | | Component | Main system, Control button, LCD, Cooling nozzle, Freezing nozzle, Guard, CO2 cartridge and Boosting accessories (Boosting nozzle, a Boosting guard, and a Boosting Container). | | Main system, Control button, LCD, Cooling nozzle, Freezing nozzle, Guard, filter, CO2 cartridge and Boosting accessories (Boosting nozzle, a Boosting guard, and a Boosting Container). | | Difference | | Mechanism of Action | The unit blows very low-temperature gas and cold-water at temperature and time settings, onto the desired treatment area | | The unit blows very low-temperature gas and cold-water at temperature and time settings, onto the desired treatment area | | Same | | Mode type | Cooling Mode, Boosting Mode | | Cooling Mode, Boosting Mode | | Same | | Cryogen Type | CO2, Cold-water (Saline solution) | | CO2, Cold-water (Saline solution) | | Same | | Temperature | 2-4 °C within 5 sec | | 2-4 °C within 5 sec | | Same | | Treatment Duration | Cooling Mode | Continuous (0 sec) PL (2 sec pre-set) 1~60sec | Cooling Mode | Continuous (0 sec) PL (2 sec pre-set) 1~60sec | Same | | | Boosting Mode | 3~4 min (65g) 5~6 min (120g) | Boosting Mode | 3~4 min | Difference | {7} TargetCool-e K253348 | Gas Volume | 65g Cartridge, 120g Cartridge | 65g Cartridge | Difference | | --- | --- | --- | --- | | Safety feature | Alarm and status light blinking if the temperature of the skin is determined to be less than -1 °C for 1 second. | Alarm and status light blinking if the temperature of the skin is determined to be less than -1 °C for 1 second. | Same | The Cooling mode &amp; Boosting mode of TargetCool-e is equivalent in intended use, principles of operation, and performance temperature to TargetCool™ (K230599) and raises no new issues of safety or effectiveness. The subject device differs from the predicate device in the following aspects: First, the subject device allows cartridges to be attached directly without a filter. The CO₂ cartridge includes a seal cap on the mounting part, which simplifies installation and eliminates the need for a separate filter. Second, a 120 g CO₂ cartridge is available in addition to the existing 65 g option. This larger cartridge increases the volume of CO₂ gas available, thereby extending the treatment duration while maintaining the same temperature performance as the predicate device. These differences do not raise new issues of safety or effectiveness. [TargetCool-e-Freezing mode] | Product Name | TargetCool-e (Subject Device) | TargetCool™ (K230599) | Comparison | | --- | --- | --- | --- | | Indications for Use / Intended Use | TargetCool-e are intended for the surgical destruction of target tissue by applying cryogenic gases at extreme low temperatures - Molluscum Contagiosum - Skin Tags - Actinic Keratosis - Lentigo - Verruca Plana - Verruca Vulgaris - Verruca Lesions - Genital Lesions - Seborrheic Keratosis | The TargetCool is intended for the surgical destruction of target tissue by applying cryogenic gases at extreme low temperatures - Molluscum Contagiosum - Skin Tags - Actinic Keratosis - Lentigo - Verruca Plana - Verruca Vulgaris - Verruca Lesions - Genital Lesions - Seborrheic Keratosis | Same | | Component | Main system, Control button, LCD, Cooling nozzle, Freezing nozzle, Guard, CO₂ cartridge and Boosting accessories (Boosting Nozzle, a Boosting guard, and a Boosting Container). | Main system, Control button, LCD, Cooling nozzle, Freezing nozzle, Guard, filter, CO₂ cartridge and Boosting accessories (Boosting Nozzle, a Boosting guard, and a Boosting Container). | Difference | | Mechanism of Action | Cryogen, CO₂ is delivered to the treatment site to effect cellular destruction | Cryogen, CO₂ is delivered to the treatment site to effect cellular destruction | Same | | Temperature | CO₂ (-79°C) | CO₂ (-79°C) | Same | | Gas dispensing rate | 0.578 g/sec | 0.578 g/sec | Same | | Gas Volume | 65g cartridge, 120g cartridge | 65g cartridge | Difference | | Tissue damage | Cell necrosis occurs only inside the ice ball | Cell necrosis occurs only inside the ice ball | Same | Cryoablation is the fundamental technological principle for the freezing mode of the subject device, Page 4 of 5 {8} TargetCool-e K253348 consistent with the predicate device (TargetCool™). The freezing mode of TargetCool-e is equivalent in intended use, principles of operation, and performance temperature to TargetCool™ (K230599), and raises no new issues of safety or effectiveness. The subject device incorporates two design differences from the predicate device. First, the subject device allows the $\mathrm{CO}_{2}$ cartridge to be attached directly without a filter. The cartridge is equipped with a seal cap at the mounting part, simplifying installation and eliminating the need for a separate filter. Second, a $120\mathrm{gCO_2}$ cartridge option has been added in addition to the existing $65\mathrm{g}$ cartridge. This larger cartridge increases the available $\mathrm{CO}_{2}$ volume, thereby extending the treatment duration while maintaining the same temperature performance as the predicate device. These modifications do not raise new questions regarding safety or effectiveness. ## 8. CONCLUSION TargetCool-e and the legally marketed predicate devices have the same intended use, Indications for Use statement and the technological characteristics. While the technological characteristics differ between the two systems, the differences are minor. Performance testing data established that the TargetCool-e is safe and effective as the legally marked predicate devices and that the TargetCool-e does not raise any different questions of safety and effectiveness than the predicate. On this basis and in accordance with 21 CFR § 807.100(b), TargetCool-e is substantially equivalent to the predicate device. Page 5 of 5
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...